
WARNING INDICATIONS: Failure to observe audible and/or visual
warnings and take corrective action may result in injury or death. In
general, if the WARNING legend is flashing, some other digits should be
flashing to indicate the anomaly.
•If the diver is ascending faster than the selected maximum ascent
rate, then the top bar of the ascent bar graph will flash and the
“WARNING” legend will illuminate. The unique high to low audible
sweep alarm will continue to sound once per second until the
situation is corrected.
•If the diver descends below the user set Depth Alarm, the Depth
digits will flash. A single beep audible alarm will sound once per
second for five seconds and will repeat every two minutes. The
Depth alarm is not active in the Decompression Mode to avoid
confusion with the “Shallower Than Ceiling” alarm.
•If the battery voltage goes below 2.5 volts, the “BATT” legend will
illuminate, at 2.2 volts it will flash once per second.
•If the diver has less than two minutes of No-Decompression Time
(NDC) remaining, the “WARNING” legend will illuminate and flash
along with the Remaining NDC time digits. A single beep audible
alarm will sound once per second for five seconds and repeat every
two minutes.
•If the diver enters the Decompression Mode, a single beep audible
alarm will sound once per second for five seconds.
•During a Decompression dive, if the Depth is less than the CEILING,
the “WARNING” legend will illuminate and flash along with the Depth
and Ceiling digits. A unique high to low audible sweep alarm will
continue to sound once every two seconds until the situation is
corrected.
•If the dive computer determines that either the Depth or Temperature
sensor is malfunctioning. The “WARNING” legend will illuminate, the
computer will also issue a 5-beep two tone audible alarm once every
two minute to alert the diver to this condition. On the Appropriate
Screen the Maximum Depth will be replaced with “S-d“ or "S--" to
indicate the low-pressure transducer or “S-t” for the temperature
transducer.
•For High PO2, see “OXYGEN TOXICITY FACTORS” on page 7.
•For High CNS, see “OXYGEN TOXICITY FACTORS” on page 7.
•For High OTU, see “OXYGEN TOXICITY FACTORS” on page 7.
If an audible alarm is being issued, the computer will not switch to the
Alternate display nor will the TACLITE™ activate.
SENSOR WARNING: The EMC-16 has the capability of monitoring the
integrity of its' sensors, both the low-pressure (depth/altitude) and the
temperature. When the computer detects an error in one of the
transducers, the diver is alerted to this condition by the illumination of the
“WARNING” legend, and the computer will also issue a 5-beep two tone
audible alarm once every two minute to alert the diver to this condition.
The Maximum Depth will be replaced with “S-d“ or "S--" to indicate the low-
pressure transducer or “S-t” for the temperature transducer (see figure
18a). The “WARNING” legend, along with either the Temperature digits or
the Depth digits and the error code will flash once per second. In the
highly unlikely situation were both sensors are detected as having errors,
the display will alternate between “S-d“ or "S--" and “S-t”. This warning will
be issued whether the computer is in the Surface Interval, Dive Mode,
Decompression Mode or Post Dive Interval. In the unlikely case that your
computer issues one of these warnings the unit should be returned to the
factory for evaluation and/or repair.
Figure 18a shows a Sensor Warning, in this case a Temperature Sensor,
as it would be displayed in the Dive Mode. If the Sensor Warning was for
the depth sensor it would display “S-d“ or "S--".
OXYGEN TOXICITY FACTORS: The EMC-16 has the ability to track
Oxygen Toxicity levels for the Central Nervous System (CNS) as well as
the Mission Oxygen Tolerance Units Dose (OTU). In addition, a maximum
Partial Pressure of Oxygen (PO2) warning alarm can also be set. While
most other audible alarms of the dive computer consist of five long beeps,
the CNS, OTU, and PO2have a distinctive audible alarm that consists of
short double-beeps that sound once per second for five seconds.
As long as one or more of these three parameters is outside its limits, the
“WARNING” legend on the display will continue to flash and the audible
alarm will be repeated once every two minutes.
These three functions are not active if the NITROX capability is disabled
via the Analyst®PC Interface.
PARTIAL PRESSURE OF OXYGEN (PO2): High levels of PO2can cause
severe Oxygen poisoning. Individual divers can be affected by widely
different levels of PO2. The user via the Analyst®can set the PO2alarm to
any level between 0.50 ATA and 1.59 ATA. As shipped from the factory,
this is set to 1.40 ATA. Should the PO2be above the alarm set point, the
“WARNING” legend will illuminate and the audible alarm will sound and the
PO2value that is displayed will flash.
CENTRAL NERVOUS SYSTEM (CNS) TOXICITY: The user via the
Analyst®can set the CNS Toxicity alarm to any level between 40% and
80% of the maximum allowable limit. As shipped from the factory, this is
set to 50%. Should the CNS Toxicity reach 50% of the maximum
allowable, the “WARNING” legend will illuminate and the audible alarm will
sound and on the Alternate Screen the displayed CNS percentage will be
flashing along with the “WARNING” legend.
By the accepted definition of CNS toxicity, should a PO2value of greater
than 1.6 ATA be measured, the CNS Toxicity will be 100%. During the
Surface Interval, this percentage will decrease as the CNS declines toward
zero. Whatever the current CNS Toxicity level, it can also be viewed on
the Surface Interval Alternate Screen or on the InFormation screen in the
Programming Mode.
OXYGEN TOLERANCE UNITS (OTU): An issue with long term breathing
of higher partial pressures of Oxygen above 0.5 ATA is Pulmonary Oxygen
Toxicity or sometimes called WHOLE BODY which must be tracked
properly.
The EMC-16 will track the OTU based on Dr. Bill Hamilton’s ‘REPEX’
method of oxygen exposure management. The OTU Dose is an
exponential function of oxygen partial pressure and time.
The time-dependent limit varies with length of time (days) that the diver
continues to dive without full recovery to zero OTU. The Mission OTU
Clock tracks the OTU, which is a running clock that tracks long-term
Oxygen exposure. This clock may run for several weeks if frequent dives
7