Covidien Genius 2 User manual

GeniusTM 2
Tympanic Thermometer and Base
Operating Manual

Tympanic Thermometer and Base
en
Table of Contents
Page
Section I Introduction 1
Initial Set Up 1
Section II General Information 1
Features 1
Section III Important Precautions 2
Section IV Icon Identication 3
Section V Instructions for Use 4
Peak Select System™ 4
Equivalence Modes 4
Probe Covers 4
Temperature Measurement 5
Temperature Recall 6
Temperature Display - Toggle 6
Sleep Mode 6
Pulse Timer Mode 6
Thermometer Display Icons and Alarms 7
Biotech Mode 8
Section VI Preventative Maintenance 9
Section VII Cleaning 9
Section VIII Battery Replacement 9
Section IX Mounting Instructions 9
Section X Troubleshooting 10
Section XI Specications 10
Section XII Customer Service 12
Section XIII Warranty 13
Section XIV Electromagnetic
Conformity Declaration 13
This product contains software solely owned by Covidien llc. Covidien llc grants the user a non-exclusive, limited license to use the
software according to the operating instructions. A copy of the license can be obtained from Covidien llc.

en
Tympanic Thermometer and Base 1Table of Contents
Section I — Introduction
This operator’s manual was prepared for the operator of the Genius™ 2 Tympanic Thermometer and Base. This manual contains
instructions for use, precautions, and available maintenance and service information. To achieve satisfactory results, the operator
must read this manual thoroughly before attempting to use the thermometer.
Initial Set Up
• Unpack the Genius 2 Tympanic Thermometer and Base and examine it for structural integrity.
• If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the
respective system.
• When rst used, the Genius 2 Tympanic Thermometer and Base will show the factory preset settings: ear mode (EAR) and the
celsius scale (°C).
Section II — General Information
The Genius 2 Tympanic Thermometer and Base is a fast, accurate, and convenient clinical instrument for measuring patient
temperatures. The Genius 2 Tympanic Thermometer and Base is an ear canal thermometer with measurement site equivalence
modes including oral, core, and rectal equivalent temperatures. The equivalence modes are explained further in Section V,
Instructions forUse.
Features
• Very fast temperature acquisition that meets CEN and ASTM standards - see Section XI.
• Peak Select System
• Temperature measurement range of 33.0°C to 42.0°C (91.4°F to 107.6°F).
• After a temperature has been acquired, the °C/°F button alternates the displayed temperature between °C and °F.
• Audible and visual indication of completed temperature acquisition.
• Disposable single use probe covers aid in the prevention of a cross contamination of infectious diseases.
• Low and dead battery indicators.
• “Sleep” mode stores the last temperature and conserves battery life.
• 15, 30, 45, and 60 second pulse timer functions.
• The thermometer housing can be wiped clean with common cleaning agents. See Section VII, Cleaning for instructions.
• Base unit protects the thermometer tip and stores probe covers for easy access.
• Easy to read LCD display with icons.
• Unit is designed for ambidextrous use.
• Improved probe design to allow ease of use on the pediatric population.

en
Tympanic Thermometer and Base 2Table of Contents
Section III — Important Precautions
• Warning: No modication of this equipment is allowed.
• The Genius 2 Tympanic Thermometer and Base is a precision optical instrument. For reliable and trouble free operation, handle
carefully and do not drop.
• Read the operator’s manual thoroughly before using the Genius 2 Tympanic Thermometer and Base.
• Prior to use, ensure that the probe tip is clean and clear of any material. If the thermometer tip becomes soiled, gently clean with
a lens wipe or lint free swab. The tip should appear shiny and free of ngerprints and/or debris. For further cleaning instructions
refer to Section VII Cleaning.
• Always install a new probe cover prior to taking a temperature. The probe cover membrane should be smooth with no holes,
tears, or wrinkles.
• The thermometer was designed to be used only with Genius 2 Tympanic Thermometer and Base probe covers. Use of probe
covers manufactured by any company other than Covidien llc may result in erroneous readings.
• Ensure that the probe tip seals the ear canal prior to taking a temperature. Failing to seal the ear canal will result in a loss
ofaccuracy.
• When not in use, the Genius 2 Tympanic Thermometer and Base should be placed in the thermometer base unit.
• The Genius 2 Tympanic Thermometer and Base should not be used on patients with ear drainage, blood, cerebrospinal uid,
vernix, ear wax plugs, or foreign bodies in the ear canal.
• Patients with removable hearing aids should remove the device at least 10 minutes prior to ear temperature assessment.
Implanted devices generally do not aect ear temperature.
• Pressure equalization (PE) or tympanostomy tubes will not adversely aect accuracy. For patient comfort, wait one week after
surgery before using Genius 2 Tympanic Thermometer and Base.
• When assessing patient temperatures during cold weather conditions, allow the patient to equilibrate to room temperature
before use.
• Excessive eardrum scarring may cause lowered temperature readings.
• Under normal conditions, ear wax does not aect accuracy. However, cerumen plugs or impactions containing debris can lower
the temperature measurement by several tenths of a degree.
• Always wait at least two minutes before taking another measurement in the same ear.
• Used probe covers must be treated as infectious biological waste and disposed of in accordance with current medical practices
and local regulations.
• Do not use lithium batteries with this device.
• Do not mix alkaline and rechargeable batteries.
• Removal of the batteries is recommended if the unit is not going to be used for an extended period of time.
• Alkaline batteries must be disposed of in accordance with local environmental and institutional policies.
• Expired or old equipment must be disposed of in accordance with institutional policy.
• Fluid ingress may cause a reduction in battery life and interfere with unit functionality. Cleaning guidelines should be followed as
per Section VII to prevent uid entering the unit.
• Do not use this device near ammable anesthetics. Not suitable for use in the presence of ammable anesthetic mixture with air,
oxygen or nitrous oxide, or in an oxygen rich environment.

en
Tympanic Thermometer and Base 3Table of Contents
Section IV — Icon Identication
Humidity
limitations
0%
95%
Medical Electrical Equipment
Genius 2 Tympanic Thermometer and Base
(1) Classied with respect to electrical
shock, re and mechanical hazards in
accordance with UL60601-1
(2) Classied with respect to electrical
shock, re, mechanical and other
specied hazards in accordance with
CAN/CSA C22.2 No. 60601-1-08
Probe cover installed Probe cover not installed
Serial number Date of manufactureManufacturer
0123
CE mark Authorized
representative in the
European Community
Catalog numberType BF Protection
(Degree of protection
against electrical
shock -there is no
conductive connection
to thepatient)
Non-ionizing
electromagnetic
radiation
Dispose of as
Electrical and
Electronic waste
-13°F
-25°C
131°F
55°C
Temperature
limitations
Keep away from
sunlight
Keep dryDEHP-free
Latex-free
Consult instructions
for use
By prescription only
Scan Button Choking Hazard Non-Sterile°C/°F Button Timer ButtonEject Button

en
Tympanic Thermometer and Base 4Table of Contents
Section V — Instructions for Use
Peak Select System
The Genius 2 Tympanic Thermometer and Base uses the patented Peak Select System. This system is well known from our Genius
Thermometer. It reduces technique errors by taking multiple readings and automatically selecting the most accurate temperature
for display. Whereas the Genius only used 32 readings for Peak Select, the new and improved Genius 2 Tympanic Thermometer
and Base uses 100 readings.
Equivalence Modes
Genius 2 Tympanic Thermometer and Base is an ear canal thermometer for neonates, newborns, children, andadults.
Prior to the introduction of tympanic thermometry, patient temperatures were measured in the mouth (oral), in the rectum
(rectal), or in special cases with an indwelling Swan-Ganz™* catheter (core). If a patient’s temperature was measured
simultaneously with each of these methods, dierent absolute temperatures would be obtained. The Genius 2 Tympanic
Thermometer and Base compensates for the average dierence in temperature at each of these sites by adjusting the
displayedtemperature.
A checker / calibrator is available for this device. The device should be checked if it is dropped or if it is stored at less than -25°C or
above 55°C.
“ASTM laboratory accuracy requirements in the display range of 37 to 39°C (98 to 102°F) for Infrared thermometers is ± 0.2°C (±
0.4°F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is
± 0.1°C (± 0.2°F).”
The following equivalence modes are available on the Genius 2 Tympanic Thermometer and Base. Data is available from
Covidien llc on request.
Ear: In ear (EAR) mode, the display will indicate the absolute temperature without adjustment.
Oral: In oral (ORL) mode, the tympanic temperature is adjusted to display an oral temperature equivalent.
Oral Mode = Ear Mode + 0.60°C.
Core: In core (CORE) mode, the tympanic temperature is adjusted to display the core temperature equivalent.
Core Mode = Ear Mode + 1.04°C.
Rectal: In rectal (REC) mode, the tympanic temperature is adjusted to display the rectal temperature equivalent.
Rectal Mode = Ear Mode + 1.16°C.
Probe Covers
The Genius 2 Tympanic Thermometer and Base utilizes a single use disposable probe cover during temperature measurement.
The probe cover was designed especially for use with the Genius 2 Tympanic Thermometer and Base. The use of probe covers
produced by any company except Covidien llc will result in erroneous temperature measurement. In order to aid in infection
prevention and control, always install a new probe cover prior to taking a temperature. The probe covers are contained in a
cassette that is located in the thermometer base. To load a probe cover on the thermometer, rmly insert the probe tip into the
probe cover. When installed on the thermometer, the probe cover membrane should be smooth with no holes, tears, or wrinkles.
After the patient temperature has been taken, eject the probe cover by pressing the eject button and then dispose of properly.

en
Tympanic Thermometer and Base 5Table of Contents
Temperature Measurement
Training on the use of the Genius 2 Tympanic Thermometer and Base is important for user competence. Please follow these basic
steps below and for further information please visit www.Genius2Training.com
1. Visually inspect the patient’s ear canal. Remove the Genius 2 Tympanic Thermometer and Base from
thebase.
2. Inspect the probe lens. If any debris is present, clean the probe tip per the directions in Section VII,
Cleaning. If the probe tip is clean, proceed to step 3.
3. Press the scan button to verify functionality and mode selection on the LCD screen. Install a probe cover
by rmly inserting the probe tip into a probe cover. After the probe cover is installed, the thermometer
will perform a system reset. The thermometer will then display dashes, the equivalence mode, and the
thermometer icon.
4. Inspect the probe cover to make sure it is fully seated (no space between cover and tip base) and no
holes, tears, or wrinkles are present in the plastic lm.
5. Place the probe in the ear canal and seal the opening with the probe tip. For consistent results, ensure
that the probe shaft is aligned with the ear canal.
6. Once positioned lightly in the ear canal press and release the scan button. Wait for the triple beep
before removing thethermometer.
7. Remove the probe from the ear as soon as the triple beep isheard.

en
Tympanic Thermometer and Base 6Table of Contents
8. The patient temperature and the probe eject icons will bedisplayed.
9. Press the eject button to eject the probe cover into a suitable waste receptacle.
10. Always return the thermometer to base forstorage.
Temperature Recall
After a temperature has been acquired, the thermometer will enter “o“ mode after approximately 10 seconds. The temperature
can be recalled by pressing and releasing the scan button or by pressing and holding the °C/°F button.
Temperature Display - Toggle °C or °F
When a temperature is in the display, the user may press and hold the °C/°F button to toggle between degrees Celsius and
degrees Fahrenheit.
Sleep Mode
The Genius 2 Tympanic Thermometer and Base enters sleep mode after 30-40 seconds of non-use. To wake up the thermometer,
eject the attached probe cover and/or install a new probe cover. This energy conserving function optimizes battery life.
Pulse Timer Mode
1. Press and hold the timer button to enter Timer mode. Press again to start the timer. The timer will run from 0 to 60 seconds.
2. The thermometer will issue a single beep at 15 seconds, a dual beep at 30 seconds, a triple beep at 45 seconds, and four beeps
at 60 seconds.
3. Pressing the timer button at any point during the timer function will return the thermometer to “o“ mode.
4. At the end of the 60 seconds, the thermometer will wait two seconds and then enter sleep mode.
5. Return the thermometer to base for storage.

en
Tympanic Thermometer and Base 7Table of Contents
Thermometer Display Icons and Alarms
Thermometer operation is communicated to the user via the LCD display and an audible device contained inside the thermometer
case. After the probe cover is installed or the batteries have been changed, the thermometer will perform a system reset. The
thermometer will perform an internal test to verify that the system components are functioning properly.
Alarm Condition Display Mode
The LCD displays the low battery icon. The low battery icon will remain on until the
battery is replaced or until the dead battery icon is displayed. After the low battery
icon is displayed, the user may take approximately 100 temperatures until the dead
battery icon isdisplayed.
Dead Battery The LCD displays the dead battery condition. When any button is pressed, the
battery icon will ash several times and then the LCD will turn OFF. After the battery
icon is displayed, the batteries must be changed before the thermometer can
beused.
Low Battery
Ambient Temperature below specied range
Ambient Temperature above specied range
Patient temperature below specied range
Patient temperature above specied range
If the display shows a system error 4 and the system error icon, the ambient conditions are changing too rapidly for the device to
be used. Allow the device to equilibrate for 20 minutes before using.
If the display shows any other system error, then reset the thermometer by installing a probe cover. If the system error does not
clear, then contact the factory for available service options. The applicable manufacturer addresses are located in Section XII,
Customer Service.

en
Tympanic Thermometer and Base 8Table of Contents
Biotech Mode
The biotech function is used to select the operational mode of the thermometer and to verify the installed software version. All
operational mode settings in biotech mode are stored in nonvolatile memory and retained through system power cycles, such as
changing batteries. All factory calibration parameters are also stored in nonvolatile memory.
The factory default settings are shown below:
Temperature mode °C (unlocked)
Site mode Ear
Site text On
The biotech mode is accessed by pressing and holding the timer and °C/°F buttons for four seconds. All LCD segments will light for
one second, the thermometer will issue a single beep, and the display will show scrolling dashes. Pressing the timer button cycles
through the biotech modes. When options are available within a mode, the °C/°F button cycles through the options.
Pressing the timer button after the site text display will return the user to the installed software version. To exit biotech mode,
two options are available: (1) press and hold the °C/°F and timer buttons for one second, or (2) the device will automatically exit
biotech mode after about 30 seconds of inactivity. Any changes are saved.
The biotech mode sequence is shown below:
Software version
Displays the installed software version of device. Where “00” is the current software version.
Temperature mode
°C (unlocked) Locked °C Locked °F °F (unlocked)
Site mode
Oral Core Rectal Ear
Site text
Pressing the °C/°F button when in this mode turns the body site text labels on or o. The labels will remain on when an “X” appears
inside the box icon, and the text will remain o when the box is empty.

en
Tympanic Thermometer and Base 9Table of Contents
Section VI — Preventative Maintenance
Functional checks must be performed by technicians trained by Covidien llc and based on the procedure in the Genius2 Tympanic
Thermometer and Base Checker Operation Manual. Contact your Covidien llc representative for details. Covidien llc recommends
functional checks at 12 month intervals. Harsh use or environmental conditions may require reduced functional check intervals at
some facilities. If the unit is dropped or abused, check the unit prior to next use.
Section VII — Cleaning
• The Genius 2 Tympanic Thermometer and Base body may be wiped clean with a damp cloth. Water temperature should not
exceed 55° C (130°F). Do not soak, rinse, spray or submerge the Genius 2 Tympanic Thermometer and Base. When cleaning the
thermometer body a probe cover should be installed on the thermometer. This will prevent harsh chemicals from damaging the
tip and probe lens.
• A mild detergent or detergent wipe may be used for cleaning purposes. Do not use cleaners such as Spray-Nine™*, Phisohex™*,
Hibiclens™*, Vesta-Syde™* , Anios DDSH™*, Sufanios Citroen™*, Perasafe powder™*, Sekusept Pulver, Classic™*, Sekulyse™*, Diesin
HG™*, Incides N™*, Surfa`Safe™* disinfectants as they may result in damage to the thermometer case or electronics.
• We do not recommend using a cleaning agent that contains Chlorine, but if it is necessary, use a 10:1 water hypochlorite mixture.
A damp isopropyl alcohol wipe, Cidex™*, Manuklenz™*, VIROX™* or CaviWipes™* cleansing agents are acceptable. However,
prolonged or repeated use of these chemicals may result in damage to the thermometer case, display or electronics. When
cleaning the thermometer, make sure the wipe has all excess uid squeezed from it. If too wet, the chemical may penetrate the
handset and aect the thermometer functionality. Never spray the thermometer directly with cleaning chemicals. Spray a cloth
or sponge lightly dampening it with the cleaning agent and then apply it to the thermometer.
• The probe tip can be cleaned using an Isopropyl alcohol wipe such as Webcol™* or Curity™*. Remove all foreign matter from the
thermometer tip. After the foreign matter has been removed, dry the window at the end of the probe tip using a lint free swab,
cotton ball or lens wipe. The thermometer lens must be free from ngerprints and / or smudges for proper operation. After
cleaning the tip allow the thermometer to air dry for 30 minutes prior to use.
• Use of a clean cloth or sponge is recommended for cleaning. Never use an abrasive pad or an abrasive cleaner on the Genius 2
Tympanic Thermometer and Base.
• The thermometer is a non-sterile device. Do not use ethylene oxide gas, heat, autoclave, or any other harsh method to sterilize
this thermometer.
• Allow sucient time for air drying (30 minutes) before attempting to use the thermometer after cleaning.
Section VIII — Battery Replacement
The Genius 2 Tympanic Thermometer and Base batteries should be replaced when the low battery icon is shown on the LCD
display. After the low battery icon is displayed, the user will be able to take approximately 100 temperatures before the dead
battery icon is displayed and the unit will not allow a temperature to be taken. To replace the batteries, access the battery
compartment by unscrewing the battery door that retains the battery cover. Note the polarity of the installed batteries. Remove
the old batteries and install fresh batteries, ensuring the correct polarity. Reinstall the battery cover and secure it with the screw.
Section IX — Mounting Instructions
If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the
respective system.

en
Tympanic Thermometer and Base 10 Table of Contents
Section X — Troubleshooting
If the Genius 2 Tympanic Thermometer and Base is not functioning properly, check the following items:
Symptom Action
Temperature reading unusually high Check the probe cover for tears or gaps.
Temperature reading unusually low Check the probe cover and thermometer tip for obstructions.
Check the patient ear canal for obstructions.
Low battery indicator lit Replace battery
Dead battery indicator lit Replace battery
Display blank Replace battery
System error displayed If system error “4”, then let the thermometer equilibrate in the
room for 20 minutes before using. For all other system errors, reset
the thermometer by installing a probe cover. If the system error
does not clear, send the thermometer to the facility Biomedical
Engineering department or contact the factory for available
service options. The service information is located in Section XII,
Customer Service.
The Genius 2 Tympanic Thermometer and Base advisory and alarm conditions are described in the Thermometer Display Icons
and Alarms subheading of Section V, Instructions for Use.
Section XI — Specications
Clinical accuracy characteristics and procedures are available from the manufacturer on request. To verify accuracy, use a certied
blackbody as specied in EN 12470-5-2003, Annex C or use a Genius 2 Checker/Calibrator – order part number 303097.
Calibrated Accuracy Limits:
Ambient Temperature Target Temperature Accuracy
25°C 36.7°C to 38.9°C ± 0.1°C
(77°F) (98.1°F to 102°F) (± 0.2°F)
16°C to 33°C 33°C to 42°C ± 0.2°C
(60.8°F to 91.4°F) (91.4°F to 107.6°F) (± 0.4°F)
Calibrated Accuracy Limits (after Recalibration*):
Ambient Temperature Target Temperature Accuracy
16°C to 33°C 36°C to 39°C ± 0.2°C
(60.8°F to 91.4°F) (96.8°F to 102.2°F) (± 0.4°F)
16°C to 33°C less than 36°C ± 0.3°C
(60.8°F to 91.4°F) (less than 96.8°F) (± 0.5°F)
16°C to 33°C greater than 39°C ± 0.3°C
(60.8°F to 91.4°F) (greater than 102.2°F) (± 0.5°F)
*Post recalibration accuracy using the Genius 2 Checker/Calibrator may not necessarily be equivalent to factory calibration.

en
Tympanic Thermometer and Base 11 Table of Contents
Displayed Temperature Measurement Range:
Temperature Range depends on Site Mode as follows:
Mode Range °C Range °F
Ear 33.0 to 42.0 91.4 to 107.6
Oral 33.6 to 42.0 92.5 to 107.6
Core 34.0 to 42.0 93.2 to 107.6
Rectal 34.2 to 42.0 93.6 to 107.6
Ambient Temperature Range:
16°C to 33°C (60.8°F to 91.4°F) where as CEN denes the standard as 16°C to 35°C (60.8°F to 95°F), 10 to 95 %RH, non-condensing.
Transport and Storage Temperature Range:
-25°C to 55°C (-13°F to 131°F), up to 95% RH non-condensing. If the unit is stored at extremes, it is recommended that the unit be
checked on the eld calibration checker or at the factory before returning to service.
Clinical Repeatability:
Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc onrequest.
Response Time:
Less than 2 seconds
Pulse Timer:
60 seconds
Temperature Resolution:
0.1°C or 0.1°F
Power:
Internally Powered ME Equipment
3 AAA alkaline batteries
Battery Life:
Minimum of 15,000 temperature readings
Size:
Thermometer — 17.8 cm (7”)
Base — 20.3 cm (8”)
Weight:
Thermometer (with batteries) — 160 grams
Base — 100 grams
Degree of Protection Against Electrical Shock:
Type BF
Mode of Operation:
Continuous
Degree of Protection Against Ingress of Fluids:
Not Protected-IPX0

en
Tympanic Thermometer and Base 12 Table of Contents
Regulatory and Safety standards:
The Genius 2 Tympanic Thermometer and Base follows international regulatory standards.
• EN 12470-5:2003 (E) Clinical Thermometers – Part 5: Performance of infrared ear thermometers where sections 6.3.4 and 6.5.5
were met based on Covidien llc technical report. Data is available from Covidien llc on request.
• UL 60601-1 • EN 60601-1:2005
• EN 60601-1-2:2007 • MDD 93/42/EEC
• BS EN 980: 2003 • ASTM E1965-98
• IEC 1000-4-2: 1995 • IEC 1000-4-3:1995
The device meets EN12470-5:2003 (E) and ASTM E1965-98 subject to the following conditions:
1. The software was modied during testing to increase the precision of the measurements taken from one signicant digit to two
signicant digits.
2. The increased precision numbers were then averaged to account for the known variance in measurements taken due to
humanfactors.
Section XII — Customer Service
In the event that it is necessary to return a unit for repair, please observe the following:
1. Contact Covidien technical service as shown below for the correct return procedure.
2. Ship insured parcel to your local service contact or the appropriate location below.
United States Canada Outside US & Canada
Covidien llc Covidien llc Covidien llc (UK)
Energy Based Devices 7300 Trans Canada Highway Unit 2 Talisman Business Center
5920 Longbow Drive Pointe-Claire, QC London Road
Boulder, CO 80301 H9R 1C7 Bicester, United Kingdom
USA 1-877-664-8926 or Oxon OX26 6HR
1-800-448-0190 (514) 695-1220 +44(0) 1869328065
Covidien llc (Italy) Covidien llc (Germany) Covidien llc (Spain) Covidien llc (France)
Laboratorio AssestenzaTecnica Service Centre A/A Servicio Técnico Parcd’aaires Technopolis
Via Rivoltana 2/D Raneriestr. 18, GEB.II C/Fructuós Gelabert, 6, pl. Sótano 3 Avenue du Canada,
Segrate – MI Neustadt/Donau Sant Joan Despi LP851-LES Ulis
Italy Germany Barcelona, Spain Courtaboeuf, France
20090 93333 08970 91940
(0039) 0270308131 (0049) 09445 959374 (0034) 934758669 (0033) 0810787590
Parts Listing
To order repair parts, please contact your local customer service center or sales representative for the parts listed below.
Description Order Part Number
Genius 2 Thermometer with Base 303000
Genius 2 Probe Cover 303030
Genius 2 Checker/ Calibrator 303097
Genius 2 Replacement Base F31949WT
Genius 2 Replacement Coil Cord F31950WT
Genius 2 Replacement Battery Door F31951WT
Genius 2 Locking Mount for Wall 303058
Genius 2 Cart with Locking Mount 303059

en
Tympanic Thermometer and Base 13 Table of Contents
Section XIII — Warranty
Limited Warranty: Covidien llc warrants to the original purchaser (“Customer”) that this product will be free of defects in materials
and workmanship, under normal use, (a) within the United States, for three (3) years from the date of original purchase from
Covidien llc or its authorized distributor, or (b) outside the United States for one (1) year from the date of original purchase from
Covidien llc or its authorized distributor, provided that this does not aect warranty rights under local legislation outside the
United States. If this product does not operate as warranted above during the applicable warranty period, Covidien llc may, at its
option and expense, replace the defective part or product, or, if neither replacement nor repair is reasonably available, refund to
Customer the purchase price for the defective part or product. Dated proof of original purchase will be required.
Covidien llc does not assume any liability for loss arising from unauthorized repair, misuse, neglect, chemical damage or accident.
Removal, defacement, or alteration of serial lot number voids warranty. Covidien llc disclaims all other warranties, expressed or
implied, including any implied warranty of merchantability or tness for a particular purpose or application other than expressly
set forth in the product labeling.
EXCEPT AS OTHERWISE REQUIRED OR PROHIBITED UNDER LOCAL LAW, THE WARRANTY SET FORTH IN THIS SECTION IS THE SOLE
AND EXCLUSIVE WARRANTY AS TO THE PRODUCTS, AND IS EXPRESSLY IN LIEU OF ANY OTHER WARRANTY, ORAL OR IMPLIED,
INCLUDING WITHOUT LIMITATION ANY ORAL OR IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. EXCEPT AS OTHERWISE REQUIRED OR PROHIBITED UNDER LOCAL LAW, COVIDIEN SHALL NOT BE LIABLE FOR ANY
INCIDENTAL, SPECIAL OR CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE (INCLUDING, WITHOUT LIMITATION, LOST PROFITS)
DIRECTLY OR INDIRECTLY ARISING FROM THE SALE, INABILITY TO SELL, USE OR LOSS OF USE OF ANY PRODUCT.
Section XIV – Electromagnetic Conformity Declaration
The Genius 2 Tympanic Thermometer and Base has been built and tested according to IEC60601-1,
CAN/CSA C22.2 No. 60601-1-08, and EN60601-1-2 Standards.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Genius 2 Tympanic Thermometer and Base is intended for use in the electromagnetic environment specied below. The
user of the Genius 2 Tympanic Thermometer and Base should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
(CISPR 11)
Group 1 The Genius 2 Tympanic Thermometer and Base uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
(CISPR 11)
Class B The Genius 2 Tympanic Thermometer and Base is suitable for
use in all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domesticpurposes.
Harmonic emissions
(IEC 61000-3-2)
Not applicable

en
Tympanic Thermometer and Base 14 Table of Contents
Guidance and manufacturer’s declaration – electromagnetic immunity
The Genius 2 Tympanic Thermometer and Base is intended for use in the electromagnetic environment specied below. The
user of the Genius 2 Tympanic Thermometer and Base should assure that it is used in such anenvironment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance
Electrostatic discharge
(ESD) (EN 61000-4-2 per
EN 60601-1-2: 2007)
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient /
burst IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
not applicable
not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 ± 1 kV dierential mode
± 2 kV common mode
not applicable
not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
< 5 % UT (>95 % dip in UT )
for 0,5 cycle
40 % UT (60 % dip in UT )
for 5 cycles
70 % UT (30 % dip in UT )
for 25 cycles
< 5 % UT (>95 % dip in UT )
for 5 sec
not applicable Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Genius 2 Tympanic Thermometer and
Base requires continued operation during power
mains interruptions, it is recommended that
the P-STIM be powered from an uninterruptible
power supply or abattery.
Power frequency (50/60
Hz) magnetic eld
(EN 61000-4-8 per EN
60601-1-2: 2007)
3 A/m 3 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note UT is the a. c. mains voltage prior to application of the test level.

en
Tympanic Thermometer and Base 15 Table of Contents
Recommended separation distances between portable and mobile RF
communications equipment and the Genius 2 Tympanic Thermometer and Base
The Genius 2 Tympanic Thermometer and Base is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Genius 2 Tympanic Thermometer and Base can help prevent
electromagnetic interference by maintaining the minimum distance between portable and mobile RF communications
equipment (transmitters) and the Genius 2 Tympanic Thermometer and Base recommended below, according to the maximum
output power of the communication equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d= 1,2√P
80 MHz to 800 MHz
d=1,2√P
800 MHz to 2,5 GHz
d = 2,3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection
from structures, objects and people.

en
Tympanic Thermometer and Base 16 Table of Contents
Guidance and manufacturer’s declaration – electromagnetic immunity
The Genius 2 Tympanic Thermometer and Base is intended for use in the electromagnetic environment specied below. The
customer or the user of the Genius 2 Tympanic Thermometer and Base should assure that it is used in such an environment.
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment –guidance
Conducted RF IEC
61000-4-6
Radiated RF (EN
61000-4-3 per EN
60601-1-2: 2007)
3 Vrms 150 kHz to 80
MHz
3 V/m 80 MHz to 2,5
GHz
Not applicable
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Genius 2
Tympanic Thermometer and Base, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
Not applicable
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey, ashould be less than the
compliance level in each frequency range. bInterference
may occur in the vicinity of equipment marked with the
following symbol:
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection
from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the
measured eld strength in the location in which the Genius 2 Tympanic Thermometer and Base is used exceeds the applicable
RF compliance level above, the Genius 2 Tympanic Thermometer and Base should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Genius 2
Tympanic Thermometer and Base.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.

Manual No. SMF0212-009
U.S. Patents 7,108,419; 7,237,949; 7,354,194; 7,434,991; 7,478,946; 7,507,021;
7,520,671; 7,536,568; 7,556,424; 7,686,506; 7,731,418; 7,841,767.
COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and internationally
registered trademarks of Covidien AG.
™* Trademark of its respective owner.
Other brands are trademarks of a Covidien company.
©2011 Covidien.
Made in China.
Covidien llc, 15 Hampshire Street, Manseld, MA 02048 USA.
Covidien Ireland Limited, IDA Business & Technology Park, Tullamore.
www.covidien.com [T] 1-800-448-0190
REV 11/2012
0123
Identification of a substance that is not
contained or present within the product
orpackaging.

GeniusTM 2
Thermomètre auriculaire et sa base
Manuel de fonctionnement
Other manuals for Genius 2
4
Table of contents
Languages:
Other Covidien Thermometer manuals