Dürr Dental XR 24 Easy User manual

EN
9000-600-84/30
2010/03
Dürr Dental XR 24 Easy
Installation and Operating Instructions


EN
3
12. Start-up and first use � � � � � � � � � � � � � 20
12�1 Check temperature of the develo-
per and carry out commissioning
test � � � � � � � � � � � � � � � � � � � � � � � � 20
13. Transport � � � � � � � � � � � � � � � � � � � � � � � 21
13�1 Remove inlet and drainage hoses � � 21
13�2 Protection from damage � � � � � � � � 21
Use
14. Operation � � � � � � � � � � � � � � � � � � � � � � � 22
14�1 Mornings or before surgery start-
up � � � � � � � � � � � � � � � � � � � � � � � � � 22
14�2 Insertion of films � � � � � � � � � � � � � � 23
14�3 In the evening or after surgery
closing � � � � � � � � � � � � � � � � � � � � � 23
14�4 Every week � � � � � � � � � � � � � � � � � � 23
14�5 Every 3 weeks � � � � � � � � � � � � � � � � 23
14�6 Every 3 months � � � � � � � � � � � � � � � 23
14�7 Surgery or clinic closures � � � � � � � � 23
14�8 Film recommendation list for Dürr
developer unit XR24Easy � � � � � � � 24
15. Maintenance � � � � � � � � � � � � � � � � � � � � 25
15�1 Every 3 weeks
chemical change interval
without regeneration unit � � � � � � � � 25
15�2 Every 3 weeks chemical change
interval with regeneration unit � � � � 25
15�3 Every 3 months Cleaning the rol-
ler set � � � � � � � � � � � � � � � � � � � � � � 25
15�4 Replacing chemicals � � � � � � � � � � � 26
Trouble-shooting
16. Tips for Operators and Technicians � � 32
17. Tips for Service Technicians � � � � � � � � 34
18. Tips for Technicians during
Troubleshooting � � � � � � � � � � � � � � � � � � 35
18�1 Before starting � � � � � � � � � � � � � � � 35
18�2 Remove front panel � � � � � � � � � � � � 35
18�3 Remove rear panel � � � � � � � � � � � � 35
18�4 After completing work � � � � � � � � � � 35
18�5 Check unit fusing � � � � � � � � � � � � � 36
18�6 Setting developer temperature
using capillary tube controller � � � � 36
18�7 Check capillary tube controller � � � � 36
18�8 Check solenoid valve for water in-
flow � � � � � � � � � � � � � � � � � � � � � � � � 37
18�9 Replace water level sensor � � � � � � 38
Content
Important Information
1. General � � � � � � � � � � � � � � � � � � � � � � � � � 4
1�1 Note on Conformity � � � � � � � � � � � � � 4
1�2 General Notes � � � � � � � � � � � � � � � � � 4
1�3 Appliance Disposal � � � � � � � � � � � � � � 4
1�4 Notes on medical products � � � � � � � � 4
1�5 Notes on EMC and
medical products � � � � � � � � � � � � � � � 5
1�6 Correct Usage � � � � � � � � � � � � � � � � � 5
1�7 Incorrect usage � � � � � � � � � � � � � � � � 5
1�8 Connecting peripheral appliances � � � 5
2. Safety � � � � � � � � � � � � � � � � � � � � � � � � � � 5
2�1 General Safety Notes � � � � � � � � � � � � 5
2�2 Electrical safety instructions � � � � � � � 6
3. Warnings and Symbols � � � � � � � � � � � � � 6
3�1 Model identification plate � � � � � � � � � 7
4. Delivery Contents � � � � � � � � � � � � � � � � � 8
4�1 Special accessories � � � � � � � � � � � � � 8
4�2 Disposable materials � � � � � � � � � � � � 8
5. Technical Data � � � � � � � � � � � � � � � � � � � 9
5�1 X-ray film developer XR24Easy � � � � 9
5�2 Environmental conditions � � � � � � � � � 9
6. Functional Diagram XR 24 Easy � � � � � 10
7. Functional description � � � � � � � � � � � � 11
7�1 Brief details � � � � � � � � � � � � � � � � � � 11
7�2 Warm-up stage � � � � � � � � � � � � � � � 11
7�3 Operating phase � � � � � � � � � � � � � � � 11
Mounting
8. Set-up � � � � � � � � � � � � � � � � � � � � � � � � � 12
8�1 Room for set-up � � � � � � � � � � � � � � � 12
8�2 Unpack the appliance � � � � � � � � � � � 13
8�3 Prepare set-up location � � � � � � � � � 13
8�4 Set-up � � � � � � � � � � � � � � � � � � � � � � 13
9. Installation � � � � � � � � � � � � � � � � � � � � � 14
9�1 Water connections � � � � � � � � � � � � � 14
9�2 Waste water connections � � � � � � � � 14
9�3 Waste water hose connection
from developer and fixer � � � � � � � � � 15
9�4 Cleaning the Appliance before
Commissioning � � � � � � � � � � � � � � � � 16
10. Electrical connection � � � � � � � � � � � � � 17
11. Circuit diagram type 1700-10 � � � � � � � 18
11�1 Key to circuit diagram � � � � � � � � � � 19

EN
4
1. General
1.1 Note on Conformity
This product has been tested for conformity ac-
cording to the appropriate directives as laid
down by the European Union and has been
found to conform to all the required
specifications�
1.2 General Notes
• These Installation and Operating Instructions
form an integral part of the unit� They must be
kept close to the unit at all times� Precise ob-
servance of these instructions is a precondi-
tion for use of the unit for the intended purpo-
se and for its correct operation�
New personnel must be made aware of the
contents, and they should be passed on to fu-
ture operating staff�
• The safety of operators and fault-free operati-
on of the appliance are only guaranteed when
using original appliance parts� Only accessori-
es and consumable materials that are indica-
ted in the installation and operating instruc-
tions or are specifically approved by Dürr
Dental are allowed for use as well� If other ac-
cessories or consumable materials are used,
Dürr Dental cannot guarantee safe operation
or safe functionality� No liability on the part of
the manufacture will be accepted in the case
that damage arises through the use of non-
approved accessories�
• Dürr Dental are only responsible for the equip-
ment with regard to safety, reliability and pro-
per functioning where assembly, resettings,
changes or modifications, extensions and re-
pairs have been carried out by Dürr Dental or
an agency authorized by Dürr Dental and if
the equipment is used in conformity with the
Installation and Operating Instructions�
• These Installation and Operating Instructions
conform to the relevant version of the equip-
ment and the underlying safety standards va-
lid at the time of going to press� All circuits,
processes, names, software and appliances
quoted are protected under industrial property
rights�
• The translation of these Installation and
Operating Instructions has been carried out in
good faith� Dürr Dental will not accept any
claims arising from incorrect translation� The
enclosed German version of these Installation
and Operating Instructions is the final refe-
rence - in case of doubt contact the manufac-
turer or their agent�
• Any reprinting of the technical documentation,
in whole or in part, is subject to prior approval
of Dürr Dental being given in writing�
• Retain the packaging for possible return of the
product to the manufacturers� Only the origi-
nal packaging provides adequate protection
during transport of the unit�
Should return of the product to the manufac-
turers be necessary during the guarantee peri-
od, Dürr Dental accepts no responsibility for
damage occurring during transport where the
original packaging was not used!
Ensure that the packaging is kept out of the
reach of children�
1.3 Appliance Disposal
• EU directive 2002/96/EG - WEEE (Waste
Electric and Electronic Equipment) of 27th
January 2003 and its current application in
national law clearly states that dental products
are covered by the above-mentioned directive
and, as such, are covered by special waste
disposal requirements within the European
Union�
• If you have any questions regarding the cor-
rect waste disposal procedures for this pro-
duct please contact Dürr Dental or your dental
supplier who will be pleased to assist�
1.4 Notes on medical products
• This product is a technical medical appliance
and, as such, may only be operated by such
persons who, as a result of training or expe-
rience, can be confidently expected to ope-
rate it correctly and safely for patients and
staff alike�
• Multi-socket power outlet extensions may not
be placed on the floor� Where multi-socket
power strips are used then the requirements
under EN60601-1-1 must be strictly
observed�
• Other appliances should not be connected to
the same multi-socket power outlet
extensions�
Important
Information

EN
5
1.5 Notes on EMC and
medical products
Medical products should be treated with respect
when it comes to electro-magnetic compatibility
and special safety measures must be taken�
Notes on EMC for medical products can be
found in the information obtained with order no�
9000-606-67/30 or the information on the
Internet (www�duerr�de) in the download area for
technical documentation�
1.6 Correct Usage
The X-ray film-developing-machine XR24Easy
is designed for the automatic development of
extraoral and intraoral X-ray films of formats 2 x
3 cm to 24x30cm, according to the Dürr list of
recommended films�
1.7 Incorrect usage
Any use of this appliance/these appliances abo-
ve and beyond that laid down in the Installation
and Operating Instructions is deemed to be in-
correct usage� The manufacturer cannot be held
liable for any damage resulting from incorrect
usage� The operator will be held liable and bears
all risks�
This appliance may NOT be used in
operating rooms.
1.8 Connecting peripheral
appliances
• Only auxiliary devices (computers, monitors,
printer) are to be connected to the appliance if
they conform with standard IEC 60950-1
(EN 60950-1)�
The computer must be located outside the vi-
cinity of the patients in accordance with
IEC 60601-1-1 (EN 60601-1-1)� If this is not
possible, the computer must be constructed
in accordance with IEC 60601-1-1
(EN 60601-1)�
• Appliances may only be connected together
or connected to any other assemblies where
complete and utter safety of the patients,
operators and staff and of the environment will
not be affected in any way�
Where any doubts remain concerning the sa-
fety when connecting to other units then the
operator is obliged to ascertain that such con-
nection can in no way affect the safety of ope-
rator, patient or other staff by referring to the
manufacturer or a fully qualified and compe-
tent expert�
2. Safety
2.1 General Safety Notes
This appliance has been designed and
constructed by Dürr Dental that any danger ari-
sing from correct usage is virtually impossible�
In spite of this, we feel it is our duty to mention
the following safety measures in order to pre-
vent any possible danger�
• When using this appliance all local and rele-
vant regulations must be observed!
Converting or modifying the appliance in any
way is strictly prohibited� In such cases, any
and all guarantees immediately become inva-
lid� The operation of modified appliances can
be punishable by law� In the interests of trou-
ble-free operation the operator is responsible
for observing these regulations�
• Installation must be carried out by a technical
expert�
• Before every use the operator must check the
functional safety and the condition of the
appliance�
• The operator must be knowledgeable in the
operation of the appliance�
• The product is not designed to be used in
medical treatment areas where there exists
the danger of explosion� Areas where explosi-
ons could occur are those where flammable
anaesthetic material, skin cleansers, oxygen
and skin disinfectants are present� This appli-
ance is not to be used in areas where the at-
mosphere could cause fire�

EN
6
3. Warnings and Symbols
In the operating instructions the following war-
nings and symbols have been used:
Information and/or mandatory re-
gulations or prohibitions for the
prevention of personal injury or sub-
stantial property damage
Information and/or instructions or prohi-
bitions regarding personal safety or ex-
tensive material damage�
Warning - dangerous electrical
voltage.
Each morning - open the water tap
Each evening - turn water tap off
Main power supply - On
Main power supply - Off
Yellow LED - warming-up phase of
appliance
Green LED - Device
ready for operation
After approximately 20 minutes, the set
temperature has been reached and the
yellow and green LEDS light up�
Orange LED
Processing (do not insert another film)
Fuse
Heating
2.2 Electrical safety instructions
• The appliance may only be operated using a
correctly installed power socket�
• Before connecting to the electricity supply the
appliance must be inspected and checked
that the supply voltage and the supply fre-
quency correspond to that of the local electri-
cal supply�
• Before initial use and start-up the appliance
and all supply lines must be checked for any
signs of damage� Damaged supply lines and
connections must be replaced immediately�
• Never come into contact with patients and
open plug-in connections on the appliance at
the same time�
• When using the appliance observe all the rele-
vant electrical safety procedures�

EN
7
3.1 Model identification plate
A model identification plate (fig� 1) can be found,
after removing the top cover, to the right below
the main power unit�
A second model identification plate can be
found to the rear of the machine, near to the
power cord�
REF Order No� / Model No�
SN Serial No�
Observe accompanying documentation!
Dispose of correctly in accordance with
EU-guidelines (2002/96/EG-WEE)
1

EN
8
4. Delivery Contents
The parts listed as special accessories
are not part of the standard scope of
delivery but can be ordered separately�
X-ray film developer XR24Easy
Model 1700-10 (230 V AC, 50/60 Hz)
Dürr Cleaning Instructions ������� 9000-600-19/01
Accessories I ������������������������������� 1700-001-00
Starter pack �������������������������������� 1702-050-00
4.1 Special accessories
The following parts are not included�
Order separately as required!
XR-Clean-Set for cleaning the roller set,
Contents: 1 x Cleaning bath,
2 x Universal Cleaner,
2 x Sponges ................. 1700-007-00
Water hose with Aqua-Stop ........... 1330-001-51
Fine filter ........................................1700-002-50
Inspection jumper for
maintenance work������������������������ 1700-012-06
Darkroom lighting ������������������������� 1494-000-00
Regeneration unit RU 24 Easy ������ 1700-880-00
Retrofit set for connecting
the RU 25 to the XR 24 Easy �������� 1700-845-00
When installing an existing
regeneration unit RU 25 to the
XR24Easy, this retrofit set must also
be ordered�
Container for developer
(black cover without opening) ������� 1330-015-00
Container for fixer
(red cover without opening) ���������� 1330-016-00
Large collector, 20 Liter
for developer�������������������������������� 1416-021-00
Large collector, 20 Liter
for fixer ���������������������������������������� 1416-011-00
Daylight cover DL 24 �������������������� 1700-910-00
suitable for use with following film sizes:
2 x 3 cm to max� 15 x 30 cm
Daylight cover DL 26 ������������������� 1700-920-00
suitable for conventional film sizes
to max� 24 x 30 cm
Darkroom adapter DA 24 ������������� 1700-930-00
suitable for use with following film sizes:
2 x 3 cm to max� 24 x 30 cm
Conversion kit XR�� for 60 Hz ������� 1700-005-50
4.2 Disposable materials
Dürr Automatic XR
developer and fixer concentrates
for 2 applications �������������������������CXB140C99��
Universal Cleaner, for cleaning
roller sets, 2 x Pack ..................... CCB600C55..
Spray cleaner WR 2000, for
transport roller sets, 6 x Pack ������ CCB800C55��
DÜRR Cleaning film, clear x 100 ... 1410-007-00

EN
9
5. Technical Data
5.1 X-ray film developer XR24Easy
Model 1700-10
Voltage V 230 / 1~
Electrical frequency Hz 50/60
Power output W 950
Heat output W 850
Bath heating W 400
Drying W 450
Warm-up time 0�5°C/min ca� 20
Film processing
time
min ca� 6
Bath agitator for
developer
l/min ca� 1,5
Bath agitator for
fixer
l/min ca� 2,0
Water circulation
rate
l/min 2
Water pressure,
min..
bar 2
Water pressure,
max.
bar 6
Tank volume
Developer, Fixer
l 5 each
Fuse type IP 20
Protection class I
Medical device I
Surge voltage
category
II
Declaration of
conformity
CE
EMC Conforms
with
standards
Weight kg 25
Dimensions DxWxH 48x43x43
5.2 Environmental conditions
Storage and transport conditions
Temperature (°C) � � � � � � � � � � � � � � � � -10 to +60
Rel� humidity (%) � � � � � � � � � � � � � � � � � � max� 95
Operating conditions
Temperature (°C) � � � � � � � � � � � � � � � +10 to +40
Rel� humidity (%) � � � � � � � � � � � � � � � � � � max� 70

EN
10
6. Functional Diagram XR 24 Easy
1 Film feed
2 Start flap
3 Roller set - film feed
4 Roller sets
developer/fixer
5 Cover
6 Roller set - water
7 Water bath level sensor
8 Film outlet
9 Roller set - dryer
10 Drive shaft
11 Thermofusing
12 Dryer heating
13 Fan
14 Roller belt
15 Geared motor
16 Rotatory pump
17 Solenoid valve for water inflow
18 Capillary tube controller
19 Circulation heater
21 Drainage plug for developer
22 Drainage plug for fixer
23 Bath
24 Seal profile
A Main power supply
E 5 Litre tank for developer bath
F 5 Litre tank for fixer bath
W Water bath
AE Developer drainage
AF Fixer drainage
KE Developer circulation
KF Fixer circulation
WZ Water inflow
WA Waste water drainage
12 3 A 4 5 6 7 8 9 10
11
12
13
14
WA
15
KF
16
KE
17
WZ1819AFAE18
21
22
23
24
A Main power supply
B Yellow LED
warm-up stage of approx. 20
minutes,
C On stand-by mode green and
yellow LEDs
D Orange LED - appliance in opera
tion, do not insert further films!
Once the LED has switched off,
insert the next film.
AD C B
2

EN
11
7. Functional description
7.1 Brief details
The X-ray film developer XR24Easy is fitted
with a Start-Stop-Automatic�
The appliance on stand-by mode will start auto-
matically on the insertion of a film�
The film will be transported through both the de-
veloper and fixer baths, and then be carried
through the water bath to the drying zone�
Approximately 1 minute after the film run
through process has ended, the machine auto-
matically switches into stand-by mode�
7.2 Warm-up stage
After switching on the main power supply (A) the
yellow LED (B) lights up� As long as the yellow
LED is displayed, the temperature in the develo-
per bath and the fixer bath is being raised�
Heating is effected by a heat cartridge in the
process heater (19)� The temperature, factory-
set, is regulated by a capillary tube controller
(18)�
During the warm-up stage, no film can be pulled
in and developed�
The bath circulation processes (KE) and (KF) of
the developer bath and fixer bath start after a
delay of ca� 60 seconds after the XR24Easy
has been switched on�
The rotary pump (16) feeds developer and fixer
from 5 litre tanks via the process heater (19) into
the bath (23) through two separated hose
systems�
7.3 Operating phase
The XR24Easy is on stand-by mode when the
yellow and green LEDs both light�
When a film is placed in the film insertion slot (1)
it triggers the film feed flap (2) and the orange
LED (D) lights�
The following activities are initiated:
Solenoid valve, motor, heating and rotary
pump
• The solenoid valve (17) serves to open the
water inflow (WZ)� The connection pressure
causes the water to flow to the water bath (W)
and from there to the waste water drainage
(WA)� If the waste water drainage is blocked,
the level will rise in the water bath (W) until the
level sensor (7) closes the solenoid valve (17)
and prevents further water inflow�
• The motor (15) operates the drive shaft (10)
via the roller belt (14)� The toothed cogs on
the roller sets are directly connected to the
worm screw cogs of the drive shaft� This con-
nection serves to rotate the rollers of the roller
sets in order to transport the film through the
various process steps�
The various film processing steps are:
1� Film insertion slot, 2� Developer bath and
Fixer bath, 3� Water bath, 4�Dryer�
• The ventilation fan (13) serves to extract air�
On its way to the drying zone the air is war-
med by the heating system�
The heating temperature is set to a stationary
value�
• The rotary pump (16) activates the circulation
of the developer and fixer baths� During the
operational cycle the capillary tube controller
(18) regulates the temperature of the bath(s)�
After insertion of the film the film feed flap (2)
closes� A further film can be inserted as soon as
the orange LED (D) switches off�
The film travels through the developer bath (E),
then through the fixer bath (F) and finally through
the water rinse bath (W) and always at a con-
stant steady speed� Afterwards the film is dried
by passing through warm air and arrives at the
film outlet slot (8) and leaves the developer�
Approximately 1 minute after the film processing
cycle has ended, the machine is automatically
returned to stand-by mode�
7.3.1 Temperature control
The temperature of the bath(s) is regulated by
the capillary tube controller in the process
heater�
7.3.2 Sealing profile of roller sets
The roller sets developer/fixer (4) and the roller
set water (6) are fitted with extra sealing profiles
(24)� These sealing lips prevent fluids from es-
caping� The level of fluids in the roller sets devel-
oper/fixer are kept to a set level�

EN
12
8. Set-up
8.1 Room for set-up
• The XR24Easy developer may only be ope-
rated in a clean, dry and well-ventilated room�
• Room temperature must not fall below +10°C
in winter and must not exceed +28°C in
summer�
Never expose the machine to direct
sunlight! Danger of the chemicals over-
heating or pre-exposure of film
materials�
• The room chosen for set-up must have ac-
cess to a water connection (water tap) (28)
R3/4" with pipe anti-vacuum device, waste
water drainage (29) with aeration (anti-odour)
via plastic siphon and an electrical safety so-
cket (27)� Having a sink for washing next to
the developer is also recommended�
The water tap (28) and electrical safety so-
cket (27) must be easily accessible to the
operating personnel.
• We strongly recommend that set-up of the
XR24Easy is effected in a darkroom�
If the appliance cannot be set up in a dar-
kroom, then a daylight cover (24) (DL24 or
DL26) must be fitted�
• When operating an appliance with daylight co-
ver (25) please note:
Do not store films in the vicinity of a
room used to take X-rays!
When storing film material in the film de-
pot (26) of a daylight cover there is the
danger of pre-exposure� The pre-expo-
sure can be caused by the X-ray
appliance�
If necessary impose a separating wall or barrier
equipped with lead protection between�
28
27
29
25
26
3
4
5
Mounting

EN
13
30
31
25
8.2 Unpack the appliance
• Cut the plastic ties from the packaging� Lift
the box upwards and remove�
• Remove all small parts�
8.3 Prepare set-up location
• The appliance should be set-up on a stable,
flat and suitably large working surface at an
appropriate height for working at�
• In order to have access to the hoses the wor-
king surface should have at least one opening
of ca� 10x7cm�
• Take into account the need for additional
space for servicing and for sufficient ventilati-
on for the appliance:
At least 10 cm free space behind, appro-
ximately 50 cm free space to the left and
right.
Leave enough space at the front in case a day-
light cover (25) is to be used�
8.4 Set-up
• Lift the appliance at the sides and place in its
set-up position�
• The distance between the machine and the
water connection or the water drain and the
electrical safety socket should be maximum
1�5 m�
• Set up the appliance so that it has a slight tilt,
ca� 0�3° to the fixer bath (fig� 7)�
Adjust the feet of the appliance (30) using the
spanner provided (31) as required�
6
7

EN
14
9. Installation
9.1 Water connections
• Check that clean, clear water flows from the
water tap (28)�
When installing for the first time make sure
the connection is allowed to flow and rinse
first!
• Turn off the water tap (28)�
• Connect the 90° elbow piece (34) to the water
tap (28), mounting the fine filter (33) (arrow in
direction of flow) and the double nipple piece
(32) at the same time�
Because the quality of water varies from
region to region we strongly recommend
fitting a fine filter (33) (order no� 1700-
008-00) between the water tap (28) and
the water hose (35)�
The fine filter (33) protects the flow regu-
lator in the water intake valve of the ap-
pliance from becoming dirty�
• Check, whether the coarse filter (35a) and
both seals (35b) are present in the threaded
pipe fittings for the water hose (35)�
• Connect the water hose (35) to the water in-
flow (36) of the appliance (threaded pipe fitting
with coarse filter) and connect to double nipp-
le piece (32)�
35
36
37
28
32
33
34
35
35a
35b
9.2 Waste water connections
• The waste water connection must include a
siphon trap�
• Lay the waste water hose (37) with a constant
incline to the siphon trap�
Shorten the waste water hose to the correct
length�
43
8
9
10

EN
15
41
42
43
43a
42
43
max. 3 cm
38
39
40
Problems with the waste water
flow can be avoided by keeping the
following in mind when laying the
waste water hose:
Do not let the hose hang!
Avoid squeezing or reducing the
pressure on hose!
• Apply a small amount of UHU-Plast or similar
to the outside of the hose end of the waste
water hose (37)�
• Screw the screw socket (38) onto the waste
water hose (37)�
• Fasten the drainage hose onto the connector
of the siphon trap (40) with hose clamp (39)�
• Fix the waste water hose to the wall using
hose clamp, screws and dowels�
9.3 Waste water hose connection
from developer and fixer
• Place the 10l-collector tanks (43) for develo-
per (black level indicator marker) and fixer (red
level indicator marker) below the appliance in
a clearly visible position and free from the
possibility of tilting� See also fig� 11�
• Lay the hose (41) with constant incline
• Shorten the hoses to the correct length�
• Slide the hose for developer (black marking)
through the opening of the black cover (42)�
(Fig� 13�)
• Screw the cover (42) of the collector (43) for
developer on tightly�
• Fix the hose for fixer (red markings) in the
same way to the fixer collector�
As soon as the level in the collector
reaches the level indicator markings
(43a) then the collector must be
emptied.
Danger of spilling!
37
11
12
13

EN
16
45
44
9.4 Cleaning the Appliance before
Commissioning
• Remove the cover and the transportation sa-
fety devices (2 styrofoam strips)�
• Open the green catch (44) and locking de-
vices (45)�
• Lift the bath (23) out�
• Clean the roller sets, bath and tanks using a
damp sponge�
• When finished replace the parts in reverse
order�
9
6
4
3
• Swing the roller set for the dryer (9) towards
the rear�
• Remove roller sets (3), (4) and (6)�
23
14
15
16

EN
17
49
10. Electrical connection
This appliance is designed for use with
supply voltage of 230 V, 50/60 Hz (see mo-
del identification plate, fig. 1).
Before connection of the main po-
wer cable it is important to check
whether the required voltage is
available, otherwise the appliance
can be damaged.
Use the Dürr mains cable supplied.
The appliance may only be connected to
the mains supply when all panels and the
cover are correctly in place.
• Connect the power plug (49) into the appli-
ance and into the main power socket�
• Replace the side covers (46)
(ventilation slots to rear)�
• Place the front panel with the label
XR24Easy in place and secure with three
cross-headed screws (48)�
Do not mix up the front and rear panels�
Only the front panel has the necessary
bored holes for attaching the daylight
cover�
• Pass the quick-release catches (46a) through
the holes provided on the front panel�
46
47
48
• Similarly fix the rear panel in place with the
three cross-headed screws (47)�
• Fill chemicals, see 14�4�3
and install roller set, see 14�4�4�
46a
17
18
19

EN
18
11. Circuit diagram type 1700-10

EN
19
Name of appliance or part
B1 Thermic switch for bath heating
B2 Level sensor for water bath
F3 Temperature control fuse for dryer
H1 Gl=LED (red)
H2 Gl=LED (green)
H3 Gl=LED (yellow)
M1 Rotary pump for bath
M2 Dryer
M3 Drive motor for rollers
R1 Bath heating
R2 Dryer heater
S1 Power supply safety fuse
S2 Main power supply
X1 Appliance terminal clamp
X2 Mains plug
Y1 Water intake valve
PCB connections
X1�1 - X1�3 N
X1�4 Roller drive motor capacitor
X1�5 Capacitor to fan motor
X1�6 LED red
X1�7 solenoid valve for water,
drive motor,
dryer heater,
dryer
X1�8 NC=not used
X1�9 NC=not used
X2�10 Thermic switch 4
LED green
X2�11 Regeneration
LED yellow
X2�12 L1
X2�13 Thermic switch 1
X1�14 NC=not used
X2�15 PE
X2�16 Bath rotary pump
X2�17 Thermic switch 2
X2�18 Bath heating
light barrier, flap
X3�1 OUT
X3�2 +5V
X3�3 LED
X3�4 GND=Earth
11.1 Key to circuit diagram

EN
20
12. Start-up and first use
12.1 Check temperature of the de-
veloper and carry out commis-
sioning test
• Switch the unit on and wait for the warm-up
stage of the developer bath, see also
"13�1 Mornings or before surgery start-up"�
• To check the temperature of the developer,
switch off the unit�
• Unplug at the mains and remove all power�
• Close the water tap�
• Remove the cover�
• Measure the temperature of the developer
(front left to a depth of ca� 20 cm)�
The developer temperature must be, accor-
ding to the film list, e�g� +28 °C +/- 0�5 °C�
• Carry out a Commissioning Test according to
national regulations�
20