Denstply Sirona Sim Intego User manual

Table of Contents Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
266 09 437 D3649
D3649.201.01.04.0206.2019
Table of Contents
1General Information ................................................................................................ 7
1.1 Dear customer, ... ........................................................................................ 7
1.2 Contact data................................................................................................. 7
1.3 Notes on these operating instructions.......................................................... 8
1.3.1 General information about these operating instructions ................ 8
1.3.2 Scope of these operating instructions............................................ 8
1.3.3 Other valid documents................................................................... 8
1.4 Intended use ................................................................................................ 9
1.5 Formats and symbols used.......................................................................... 9
2Safety instructions .................................................................................................. 10
2.1 Identification of the danger levels ................................................................ 10
2.2 Information on the unit ................................................................................. 10
2.3 On-site installation ....................................................................................... 11
2.4 Installing the simulation unit......................................................................... 11
2.5 Media quality................................................................................................ 11
2.6 Warranty and Liability .................................................................................. 12
2.7 Maintenance ................................................................................................ 12
2.8 Trouble-free operation ................................................................................. 12
2.9 Vacuum system ........................................................................................... 13
2.10 Intermittent operation ................................................................................... 13
2.11 Ventilation slots............................................................................................ 13
2.12 Fresh water bottle ........................................................................................ 14
2.13 User interface............................................................................................... 14
2.14 Care, cleaning, and disinfecting agents ....................................................... 14
2.15 Wireless phone interference ........................................................................ 14
2.16 Modifications and extensions to the device ................................................. 15
2.17 Electrostatic discharge................................................................................. 15
2.18 Dismantling/Installation ................................................................................ 16
3Unit description ....................................................................................................... 17
3.1 Technical data.............................................................................................. 17
3.2 System overview.......................................................................................... 20
3.3 Treatment position LEFT ............................................................................. 22
3.4 Foot control .................................................................................................. 23
3.4.1 Pneumatic foot switch.................................................................... 23

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
Table of Contents
66 09 437 D3649
D3649.201.01.04.0206.2019 3
3.5 Dentist element ............................................................................................ 24
3.5.1 TS dentist element hanging hoses................................................. 24
3.5.2 Instrument positions....................................................................... 25
3.5.3 Standard EasyPad user interface .................................................. 26
3.5.3.1 EasyPad display and status displays ........................... 26
3.5.3.2 Overview - Fixed keys .................................................. 27
3.6 Assistant element, suction handpiece holder............................................... 28
3.6.1 Option: Comfort assistant element................................................. 28
3.6.2 Option: Suction handpiece holder.................................................. 30
4Operation ................................................................................................................ 31
4.1 Putting the simulation unit into operation ..................................................... 31
4.1.1 Initial Operation.............................................................................. 31
4.1.2 Selecting a user profile .................................................................. 33
4.2 Concept of the user interface....................................................................... 34
4.2.1 Standard EasyPad user interface .................................................. 34
4.3 Foot control .................................................................................................. 36
4.3.1 Pneumatic foot control ................................................................... 36
4.4 Dentist element ............................................................................................ 37
4.4.1 Maximum load capacity ................................................................. 37
4.4.2 Positioning the dentist element...................................................... 37
4.4.3 Fixed keys and displays................................................................. 38
4.4.4 Placing the instruments in their holders......................................... 42
4.4.5 General instrument functions......................................................... 43
4.4.5.1 Instrument functions on the EasyPad........................... 43
4.4.5.2 Saving instrument settings ........................................... 44
4.4.5.3 Setting the amount of spray water................................ 44
4.4.6 Standard and Sprayvit E 3-way syringe......................................... 45
4.4.6.1 Safety instructions ........................................................ 45
4.4.6.2 Using the standard 3-way syringe ................................ 46
4.4.6.3 Operating the SprayvitE 3-way syringe ....................... 46
4.4.7 Turbine / air motor / other air-driven instruments........................... 47
4.4.7.1 Using the highspeed handpiece ................................... 47
4.4.7.2 Setting the light of the highspeed handpiece................ 47
4.4.8 Electric motor................................................................................. 50
4.4.8.1 Motor versions.............................................................. 50
4.4.8.2 Setting the speed on the EasyPad ............................... 51
4.4.8.3 Setting the direction of rotation..................................... 52

Table of Contents Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
466 09 437 D3649
D3649.201.01.04.0206.2019
4.4.9 Endo function................................................................................. 53
4.4.9.1 Switching Endo function on/off ..................................... 54
4.4.9.2 Calibrating the burr drive .............................................. 54
4.4.9.3 Setting the torque ......................................................... 55
4.4.9.4 Setting the speed.......................................................... 56
4.4.9.5 Switching AutoReverse ON/OFF.................................. 57
4.4.9.6 Changing direction of rotation....................................... 57
4.4.10 SiroSonic L scaler.......................................................................... 58
4.4.10.1 Safety instructions ........................................................ 58
4.4.10.2 Setting the intensity on the EasyPad............................ 58
4.4.11 Cavitron scaler............................................................................... 60
4.4.11.1 Safety Instructions........................................................ 60
4.4.11.2 Setting spray................................................................. 61
4.4.11.3 Adjusting the intensity of the dentist element ............... 61
4.4.12 Mini LED curing light...................................................................... 62
4.4.13 SiroCam F/AF/AF+ intraoral camera ............................................. 62
4.4.14 Timer function ................................................................................ 62
4.4.14.1 Timer function on the EasyPad .................................... 62
4.4.15 USB interfaces............................................................................... 63
4.5 Assistant element / Suction handpiece tray ................................................. 64
4.5.1 Maximum load capacity ................................................................. 64
4.5.2 Positioning ..................................................................................... 64
4.5.3 Suction handpieces........................................................................ 64
4.5.4 Standard and Sprayvit E 3-way syringe......................................... 65
4.5.5 Mini L.E.D. curing light................................................................... 66
4.5.5.1 Safety instructions ........................................................ 66
4.5.5.2 Symbols on the mini LED ............................................. 67
4.5.5.3 Connecting the Mini L.E.D............................................ 67
4.5.5.4 Functional description................................................... 67
4.5.5.5 Operating the Mini LED ................................................ 69
4.5.5.6 Technical data .............................................................. 70
4.6 Supply module ............................................................................................. 71
4.6.1 Stand-alone water supply .............................................................. 71
4.6.1.1 Water supply with fresh water bottle............................. 72
4.7 Operating light.............................................................................................. 74
4.7.1 Switching the operating light on/off................................................ 75
4.7.2 Setting the LEDlight brightness...................................................... 75
4.7.3 Setting the brightness, color temperature, and sensor control of
the LEDview Plus........................................................................... 76

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
Table of Contents
66 09 437 D3649
D3649.201.01.04.0206.2019 5
4.8 Video system Sivision Digital ....................................................................... 78
4.8.1 Sivision monitor.............................................................................. 78
4.8.2 SiroCam F/AF/AF+ intraoral camera ............................................. 79
4.8.2.1 Safety instructions ........................................................ 79
4.8.2.2 Functional description................................................... 79
4.8.2.3 Connecting the SiroCam F / AF / AF+ intraoral
camera.......................................................................... 80
4.8.2.4 Operating SiroCam F / AF / AF+ intraoral camera ....... 80
4.8.2.5 Technical camera data ................................................. 84
4.9 Configurating the simulation unit (setup) ..................................................... 85
4.9.1 Configurating the simulation unit on the EasyPad ......................... 85
4.9.1.1 Access the Setup program ........................................... 85
4.9.1.2 Setting the time and date.............................................. 86
4.9.1.3 Presetting the timer ...................................................... 87
4.9.1.4 Setting the purging time for the purge function............. 87
4.9.1.5 Switching afterblow on/off............................................. 87
4.9.1.6 Switching the key sound on/off..................................... 87
4.9.1.7 Opening the service function........................................ 87
5Care, cleaning and maintenance by the user ......................................................... 88
5.1 Basics .......................................................................................................... 88
5.1.1 Intervals ......................................................................................... 88
5.1.2 Care, cleaning, and disinfecting agents......................................... 90
5.1.3 Conduct a microbiological water test ............................................. 90
5.1.3.1 Biofilm removal by the service technician..................... 91
5.1.4 General handling instructions ........................................................ 92
5.1.5 Inspection, maintenance and testing ............................................. 93
5.2 Surfaces....................................................................................................... 94
5.2.1 Clean/disinfect surfaces................................................................. 94
5.2.2 Disinfect user interfaces................................................................. 95
5.2.3 Sterilize silicone mats and handle covers...................................... 96
5.2.4 Clean the foot switch...................................................................... 96
5.3 Instruments and instrument hoses............................................................... 97
5.3.1 Rinse water paths .......................................................................... 97
5.3.2 Purging water paths....................................................................... 97
5.3.2.1 Purge function on the EasyPad.................................... 98
5.3.2.2 Purging water paths manually ...................................... 100
5.3.3 Lubricate, disinfect/sterilize the treatment instruments.................. 101
5.3.3.1 Treatment instruments.................................................. 101
5.3.3.2 Cleaning and sterilizing the standard 3-way syringe .... 102

Table of Contents Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
666 09 437 D3649
D3649.201.01.04.0206.2019
5.3.3.3 Cleaning, disinfecting/sterilizing the Cavitron scaler .... 102
5.3.3.4 Disinfecting/sterilizing Mini L.E.D. curing light.............. 103
5.3.3.5 Cleaning/disinfecting the SiroCam F / AF / AF+
intraoral camera............................................................ 103
5.3.4 Change the cotton wool roll on the turbine hose and oil collector.. 104
5.4 Vacuum system ........................................................................................... 105
5.4.1 Purge the vacuum system. ............................................................ 105
5.4.2 Cleaning the suction hoses............................................................ 105
5.4.3 Cleaning suction hoses and suction system.................................. 106
5.4.4 Sterilize/disinfect the suction handpieces ...................................... 108
5.4.5 Cleaning and disinfecting the suction hoses.................................. 109
5.4.6 Empty the central suction sieve. .................................................... 109
5.5 Components of the supply module .............................................................. 110
5.5.1 Service block.................................................................................. 110
5.5.2 Amalgam separator........................................................................ 112
5.5.2.1 Changing the amalgam rotor........................................ 112
5.5.3 Automatic separator....................................................................... 116
5.5.4 Air jet pump.................................................................................... 118
5.5.4.1 Air jet pump without drain tank ..................................... 118
5.5.4.2 Air jet pump with drain tank .......................................... 120
5.5.5 Sanitization .................................................................................... 123
5.5.5.1 Sanitizing with the fresh water bottle............................ 123
5.5.5.2 Cleaning the suction hoses .......................................... 127
6Maintenance by the service engineer ..................................................................... 129
6.1 Inspection and maintenance ........................................................................ 129
6.2 Safety checks............................................................................................... 129
7Spare parts and consumables ................................................................................ 130
8Disposal .................................................................................................................. 131
Index ....................................................................................................................... 132

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
1 General Information
1.1 Dear customer, ...
66 09 437 D3649
D3649.201.01.04.0206.2019 7
1General Information
1.1 Dear customer, ...
We are pleased that you have purchased the Dentsply Sirona Sim Intego
simulation unit.
Our aim is to identify our customers' needs early and create innovative
solutions. Together with your trade partner, you have configured the
treatment center most suitable for you. The new hub of your training
center is tailored to your personal needs.
With Sim Intego, you have selected a simulation unit that stands for
easy operation, high-quality design, and cost effectiveness.
These operating instructions are designed to assist you prior to initial
use and whenever you require information later on.
We wish you a great deal of success and enjoyment with Sim Intego.
Your Sim Intego team
1.2 Contact data
Customer Service Center In the event of technical queries, please use our online contact form at
the following address:
http://srvcontact.sirona.com
Manufacturer's address Sirona Dental Systems GmbH
Fabrikstrasse 31
64625 Bensheim
Germany
Tel.: +49(0) 6251/16-0
Fax: +49(0) 6251/16-2591
e-Mail: [email protected]
www.dentsplysirona.com

1 General Information
1.3 Notes on these operating instructions
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
866 09 437 D3649
D3649.201.01.04.0206.2019
1.3 Notes on these operating instructions
1.3.1 General information about these operating instructions
Follow the operating instructions Please familiarize yourself with the unit by reading through these
Operating Instructions before putting it into operation. It is essential that
you comply with the specified warning and safety information.
Keep documents safe Always keep the operating instructions handy in case you or another
user require(s) information at a later point in time. Save the operating
instructions on the PC or print them out.
If you sell the unit, make sure that the operating instructions are
included with it either as a hard copy or on an electronic storage device
so that the new owner can familiarize himself with its functions and the
specified warning and safety information.
Online portal for technical documents We have set up an online portal for the Technical Documents at http://
www.sirona.com/manuals. From here, you can download these
Operating Instructions along with other documents. Please complete the
online form if you would like a hard copy of a particular document. We
will then be happy to send you a printed copy free of charge.
Help If you reach an impasse despite having thoroughly studied the operating
instructions, please contact your dealer.
1.3.2 Scope of these operating instructions
Device variants These operating instructions are valid for the following simulation unit:
● Simulation workstation Sim Intego
Equipment options This document describes the full version of your unit. It may therefore
cover components that are not included in the unit you purchased.
1.3.3 Other valid documents
Other relevant documents The simulation workstation can be equipped with additional components
that are described in separate sets of operating instructions. The
instructions as well as all warning and safety information they contain
must also be followed.
A separate set of operating instructions exists for each of the following
products:
● Treatment instruments
● LEDlight Plus operating light
● LEDview Plus operating light
● Monitor
● Phantom heads
● Curing light
● Intraoral camera

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
1 General Information
1.4 Intended use
66 09 437 D3649
D3649.201.01.04.0206.2019 9
1.4 Intended use
The simulation workstation is used for education and training of dental
students and for further training at universities, clinics, and continuing
education institutes.
Any use that goes beyond this, e.g. reasonably foreseeable improper
use due to inattentiveness, lack of concentration, reflexive behavior in
the event of a breakdown/malfunction or misuse is considered to be use
not as intended. The manufacturer is not liable for damage resulting
from such use; the user is solely responsible for the risk.
WARNING
Risk of injury or hazard/damage for material and equipment through
use not as intended.
This is a laboratory workstation that may not be used to treat living
persons or animals. Before each use, the user (trained dental
professional or persons/students instructed in its use) must ensure the
functional safety and good condition of the equipment.
To prevent injuries or damage due to improper use, the product may only
be operated in accordance with the manufacturer's Operating Instructions.
If the unit is not used in accordance with the operating instructions, this may
have a negative effect on the recommended protective measures.
CAUTION
Follow the instructions!
Intended use also includes compliance with these operating
instructions. If the instructions for operating the unit described in this
document are not followed, the intended protection of the user may
be impaired.
This unit is not intended for operation in areas subject to explosion
hazards and is permitted only for use indoors.
1.5 Formats and symbols used
The formats and symbols used in this document have the following meaning:
Prerequisite
1. First action step
2. Second action step
or
➢ Alternative action
Result
➢ Individual action step
Requests you to do something.
See "Formats and symbols used
[→9]"
Identifies a reference to another
text passage and specifies its
page number.
● List Designates a list.
"Command / menu item" Indicates commands / menu
items or quotations.

2 Safety instructions
2.1 Identification of the danger levels
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
10 66 09 437 D3649
D3649.201.01.04.0206.2019
2Safety instructions
2.1 Identification of the danger levels
To prevent personal injury and material damage, please observe the
warning and safety information provided in these operating instructions.
Such information is highlighted as follows:
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily
injury or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTE
A possibly harmful situation which could lead to damage of the
product or an object in its environment.
IMPORTANT
Application instructions and other important information.
Tip: Information for simplifying work.
2.2 Information on the unit
Accompanying documents Meaning: Note accompanying documents!
Meaning: The accompanying documents are available on the Dentsply
Sirona website.
Electrostatic discharge (ESD) Connector pins or sockets bearing ESD warning labels must not be
touched or interconnected without ESD protective measures. See also
"Electrostatic discharge" [→15] and "Electromagnetic compatibility".

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
2 Safety instructions
2.3 On-site installation
66 09 437 D3649
D3649.201.01.04.0206.2019 11
2.3 On-site installation
The on-site installation must be carried out according to our
requirements. Details are described in the document "Installation
Instructions".
2.4 Installing the simulation unit
Installation must be carried out by authorized technicians in accordance
with the installation instructions.
2.5 Media quality
The air and water supply must meet the requirements in the installation
instructions or Sect. 3.
Use only drinking water and dry, oil-free, hygienically clean air in this
simulation workstation.
As the owner of the treatment center, you are generally responsible for
the water quality.
The bacterial count should comply with the national drinking water
regulations, but in no case exceed 500 CFU / ml (CFU: colony-forming
unit).
If the bacterial count is too high, the building water system must be
checked and the cause of contamination eliminated.
Alternatively, a self-contained water supply can also be installed.
Depending on the equipment of the simulation unit, the fresh water
bottle can optionally be used as a water tank, see "Self-contained water
supply".
Bottled water can be used for the fresh water bottle without any
restrictions.
Before installing the simulation unit, the microbiologically sound water
quality of the domestic water supply should be ensured and
documented in the form of a bacterial count. Sampling and germ counts
should be performed by a competent laboratory.
Test the quality of the water from the simulation unit at regular intervals
and after it has not been used for > 1 week, see Conduct a
microbiological water test [→90].
CAUTION
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one’s
health.
➢ Mix fresh water for the self-contained water supply every day. At
the end of the work day, the fresh water bottle must be emptied.

2 Safety instructions
2.6 Warranty and Liability
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
12 66 09 437 D3649
D3649.201.01.04.0206.2019
2.6 Warranty and Liability
Maintenance and
repair
Maintenance must be performed to ensure the operational and
functional reliability of your product and to protect the safety and health
of users or third parties.
The operator must ensure the implementation of the maintenance work.
Authorized technical personnel
and spare parts
As a manufacturer of laboratory equipment, we can consider ourselves
responsible for the safety characteristics of the device only if
maintenance and repairs are carried out only by us or by companies
authorized explicitly by us for this purpose and if components are
replaced with original spare parts in case of failure.
We suggest that you request a certificate, showing the nature and
extent of the work performed, from those who carry out such work, and
specify that the certificate show any changes in rated parameters or
working ranges, as well as the date, the name of the firm, and a
signature.
Disclaimer of liability In the event that the system owner fails to fulfill their obligation to
perform maintenance activities or ignores error messages, Sirona
Dental Systems GmbH cannot assume any liability for any damage thus
incurred.
2.7 Maintenance
Despite the outstanding quality of your simulation workstation and
regular care by the user, in the interest of operational safety it is
essential to perform preventive maintenance at scheduled intervals.
In order to ensure operational and functional reliability and to avoid
damage due to wear, you as the operator must have your unit checked
regularly by an authorized service engineer.
Please contact Sirona for a maintenance offer.
2.8 Trouble-free operation
Use of this unit is permissible only if it works properly without
malfunctions. If trouble-free operation cannot be ensured, the unit must
be taken out of service, checked by authorized technicians for
malfunctions and, if necessary, repaired or replaced.

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
2 Safety instructions
2.9 Vacuum system
66 09 437 D3649
D3649.201.01.04.0206.2019 13
2.9 Vacuum system
The suction removal of aluminum and other metal oxides from blasting
devices via the automatic separator built into the simulation unit is
prohibited! This would result in extreme wear and clogging of the
vacuum and drain lines.
A separate vacuum system must be used in connection with metal
oxide blasting devices. Simulation units equipped with a central
standard wet suction are generally suitable for suction removal of the
above material. However, make sure to observe the instructions
provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in connection with
Sirona simulation units. However, in such cases, make sure that the
system is subsequently flushed with an adequate amount of water.
To clean the vacuum system, a suction hose connection is optionally
integrated in the supply module. Water is pumped into a container
behind the suction hose intake and suctioned off from there, see
“Vacuum system”.
The unit can be optionally operated with an air jet pump (Air Venturi)
instead of a suction device.
2.10 Intermittent operation
Continuous operation with intermittent loading corresponding to the
dental mode of working.
IMPORTANT
The procedures required for the following treatment instruments are
described in the separate pertaining operating instructions. The
instructions as well as all warning and safety information they contain
must also be followed.
2.11 Ventilation slots
Under no circumstances may the ventilation slots on the unit be
covered, otherwise the air circulation will be obstructed. This can cause
the unit to overheat.
Do not spray liquids such as disinfectants into the ventilation slots or
openings in the supply module or dentist element. This may lead to
malfunctions. Use only wipe disinfection in these areas.

2 Safety instructions
2.12 Fresh water bottle
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
14 66 09 437 D3649
D3649.201.01.04.0206.2019
2.12 Fresh water bottle
LEMON
LEMON
!
!
!
The simulation unit can be equipped with a fresh water bottle for stand-
alone water supply of the instruments.
Use only Sirona fresh water bottles, see "Spare parts" ! Compressed air
is added to the bottle when it is screwed into the holder. Beverage
bottles could burst.
Misshapen, scratched, or discolored fresh water bottles must be
replaced immediately. Do not use fresh water bottles after the expiry
date. The date is marked on the bottle.
2.13 User interface
The dentist element of your simulation unit is equipped with a control
panel with seven-segment displays (EasyPad)
2.14 Care, cleaning, and disinfecting agents
Unsuitable care and cleaning agents or disinfectants may corrode the
surface of the unit or impair its functioning.
Please therefore only use care, cleaning, and disinfecting agents
approved by the manufacturer. For more information, please refer to the
section "Care, cleaning agents, and disinfectants" [→130].
NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by
Dentsply Sirona!
A continuously updated list of approved products can be obtained via
the Internet at
"www.sirona.com"
. In the navigation bar, follow the menu
path
"SERVICE"
/
"Technical Documentation"
/
"General documents"
and then open the document
"Care, cleaning and disinfection agents"
.
If you do not have access to the Internet, please contact your customer
service center (see Sect. 1.2) to order the list (REF5970905).
2.15 Wireless phone interference
To ensure safe operation of the unit, the use of possible wireless
phones (cell phones) must be prohibited in the operating area.

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
2 Safety instructions
2.16 Modifications and extensions to the device
66 09 437 D3649
D3649.201.01.04.0206.2019 15
2.16 Modifications and extensions to the device
Modifications to this unit which might affect the safety of the operator,
users or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with
original Dentsply Sirona parts or parts expressly approved by Dentsply
Sirona. The user assumes the risk of unforeseen modifications and of
using non-approved parts and accessories. Dentsply Sirona is not liable
for resulting damages.
All units connected to this product must comply with the applicable
standards:
● IEC 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use
● IEC61326-1 Electrical equipment for measurement, control and
laboratory use – EMC requirements
● IEC 62368-1-1 or IEC 60950-1 Safety of information technology
equipment (e.g. computer, monitor)
The simulation unit monitor must fulfill the requirements of the IEC
60950-1 or IEC60950-1 standard.
If a system is created during the installation, then the creator is
responsible for the conformity of the system. In accordance with
Directive 2006/42/EC.
2.17 Electrostatic discharge
Protective measures
ESD stands for ElectroStatic Discharge.
Electrostatic discharge from people can damage electronic components
when the components are touched. Damaged components usually have
to be replaced. Repairs must be performed by qualified personnel.
Measures to protect against ESD include:
● Procedures to avoid electrostatic charging via
– air conditioning
– air humidification
– conductive floor coverings
– non-synthetic clothing
● discharging the electrostatic charges from your own body through
contact with
– a metallic unit casing
– a larger metallic object
– any other metal part grounded with the protective earth
● Wearing an antistatic band that creates a connection between the
body and a protective ground wire.
Areas at risk are indicated on the unit with the ESD warning label:
We recommend that all persons working with this system are made aware
of the significance of the ESD warning label. A training course should also
be held to inform users about the physics of electrostatic charges.

2 Safety instructions
2.18 Dismantling/Installation
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
16 66 09 437 D3649
D3649.201.01.04.0206.2019
Physics of electrostatic charges
An electrostatic discharge requires prior electrostatic charging.
There is a danger of electrostatic charges building up whenever two
bodies rub against each other, e.g. when:
● walking (soles of shoes against the floor) or
● moving (chair casters against floor).
The amount of charge depends on several factors. The charge is:
● higher at low air humidity than at high air humidity, and
● higher with synthetic materials than with natural materials (clothing,
floor coverings).
The following rule of thumb can be applied to assess the transient
voltages resulting from an electrostatic discharge.
An electrostatic discharge is:
● perceptible at 3,000 V or higher
● audible at 5,000 V or higher (cracking, crackling)
● visible at 10,000 V or higher (arc-over)
The transient currents resulting from these discharges have a
magnitude of over 10 amps. They are not hazardous for humans
because they last for only several nanoseconds.
Note: nanosecond = 1/1,000,000,000second = 1billionth of a second
Voltage differentials exceeding 30,000 volts per centimeter may lead to
a charge transfer (electrostatic discharge, lightning, spark-over).
Integrated circuits (logical circuits and microprocessors) are used in
order to implement a wide variety of functions in a device. The circuits
must be miniaturized to a very high degree in order to include as many
functions as possible on these chips. This leads to structure thicknesses
as low as a few ten thousandths of a millimeter. Integrated circuits that
are connected to wires leading externally are therefore particularly at
risk from electrostatic discharge.
Even voltages that are imperceptible to the user can cause breakdown
of the structures, thus leading to a discharge current that melts the chip
in the affected areas. Damage to individual integrated circuits may
cause malfunction or failure of the unit.
2.18 Dismantling/Installation
When dismantling and reinstalling the simulation unit, proceed
according to the installation instructions for new installation in order to
guarantee its proper functioning and stability.
Dismantling or installation must be done by technicians authorized by
Dentsply Sirona. Before dismantling, disconnect the simulation unit from
the grid (e.g. by removing the building fuse).

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
3 Unit description
3.1 Technical data
66 09 437 D3649
D3649.201.01.04.0206.2019 17
3Unit description
3.1 Technical data
Model designation Simulation workstation Sim Intego
Power connection 100-240V AC ± 10%
50/60Hz
Rated current 3.3 - 1.5A at 100-240V AC
Main building fuse TypeB automatic circuit
breaker
100 - 115 VAC: 20 A medium-blow
220 - 240 VAC: 16 A medium-blow
Overvoltage category II according to IEC 60664-1
Protection class I
Media Water Quality Drinking water
Water hardness 1.4mmol/l (=8°dH) -2.1mmol/l
(=12°dH)
Air quality oil-free, dry, non-contaminated and free
of dirt
Power ratings Average power consump-
tion (for dimensioning an air
conditioning system):
100 W
Power consumption in
switched off state
0W (power switch present)
On-site water pressure 2.5-6bar (≥2.5l/min)
On-site air pressure 5.5-7.5bar (≥ 50l/min)
On-site suction air vacuum P ≤0.18bar (≥ 500l/min)
Waste water max. 3l/min
Operating conditions
+10
+40 1060
700
85
30
Intended use For indoor use
Ambient temperature 10°C-40°C (50°F-104°F)
Relative humidity 30%-85%
Air pressure 500hPa-1060hPa
Installation site ≤3000 m above sea level
The simulation unit is not suitable for
operation in explosion hazard areas.
Degree of contamination 2 acc to. IEC 60664-1
Dimensions WxHxD approx. 2500 cm x 1900 cm x 700 cm
Weight approx. 104 kg

3 Unit description
3.1 Technical data
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
18 66 09 437 D3649
D3649.201.01.04.0206.2019
Transport and storage conditions
-40
+70 1060
500
95
10
Temperature -40°C-+70°C (-40°F-+158°F)
Relative humidity 10%-95%
Air pressure 500hPa-1060hPa
Degree of protection against ingress of
water:
Ordinary device (without protection against ingress of water)
The foot control has IP X1 degree of protection against liquids (drip-
proof)
Operating mode Intermittent operation for
electric motor instruments
1 min operation ‒ 7 min pause
(at a motor current of 1.5A)
Year of manufacture
20XX
(on the rating plate)
Testing/approval This work station complies with the
requirements of
DIN EN ISO 12100:2011-03
DIN EN 61010-1:2011-07
DIN EN 61326-1:2013-07
This product bears the CE marking in
accordance with Directive 2014/30/EU
of the European Parliament and of the
Council of February 16, 2014 and Di-
rective
2006/42/EC of the European Parlia-
ment and of the Council of May 17,
2006.
The contents of the operating instruc-
tions may be changed at any time with-
out notice. The German version of the
operating instructions is
binding for the translation into other
languages.

Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
3 Unit description
3.1 Technical data
66 09 437 D3649
D3649.201.01.04.0206.2019 19
Sim Intego simulation
workstation rating plate
TYP:
SN:
REF:
D3649
XXXXXXX
XXXXXXX
100 - 240 V ~ 3,3 - 1,5 A 50/60 Hz
Made in Germany
Sirona Dental Systems GmbH
Fabrikstr. 31, 64625 Bensheim
Germany
YYYY
Device type
SN
Serial Number
REF
Material number
CE marking
Observe accompanying documents!
For disposal information, see the section “Dis-
posal [→131]”
YYYY
Date of manufacture
Manufacturer
The accompanying documents are available on
the Dentsply Sirona homepage.

3 Unit description
3.2 System overview
Dentsply Sirona
Operating Instructions Simulation Unit Sim Intego
20 66 09 437 D3649
D3649.201.01.04.0206.2019
3.2 System overview
The Sim Intego simulation workplace comprises the following main
components:
Overview of Sim Intego with patient simulator on the rod and track,
height-adjustable dentist element, assistant element, and LEDview Plus
B
H
G
A
C
E
F
D
J
A Operating light LEDview Plus I
B Removable handle covers (left/right)
C Assistance element
D Clamping lever for inclination and height adjustment of the patient simulator
E Patient simulator
F Pneumatic foot control
G High-speed handpiece hose dentist element
H Supply module
I Fresh water bottle
J Main switch
IOptionally, the unit can be equipped with an LEDlightPlus
Table of contents
Other Denstply Sirona Dental Equipment manuals
Popular Dental Equipment manuals by other brands

smartdent
smartdent smartgun PERIO Quick Maintenance Manual

GNATUS
GNATUS EasySonic owner's manual

Florida Probe
Florida Probe VoiceWorks user manual

KaVo
KaVo SMARTmatic PROPHY S31 Instructions for use

Sirona
Sirona Heliodent plus installation instructions

Ivoclar Vivadent
Ivoclar Vivadent Bluephase Style Instructions for use