Dentmate Ledex WL-090+ User manual

Instructionsfor use
SVĚTELNÁ POLYMERAČNÍ LAMPA A DETEKTOR PLAKU A KAZU
Návod k použití

Dear Customer,
Thank you for choosing DENTMATE LEDEX
TM
WL-090+ Dental curing light.
A lot of researches & developments have gone into the manufacturing of this product.
We sincerely hope that it will give you many years of trouble-free use. Please read and
understand all the instructions before using this equipment, and save this instruction
for use for your reference.
The instruction for use is subject to change without further notice.
One Touch, And You Will Enjoy
The Powerful Light

1. Symbols Used 1
1.1. In these instructions for use 1
1.2. On the product/packaging 1
2. Product Information 2
2.1. Indications for use 2
2.2. Procedure 2
2.3. System components 3
2.4. Features 4
2.5. Display 4
2.6. Installation and charging 5
2.7. Operating modes 6
3. Contraindications 10
4. Warnings 11
4.1. User 11
4.2. Ambient conditions 11
4.3. To avoid electric shock (shock hazard) 12
4.4. Heat development (burn hazard) 12
4.5. Battery 12
4.6. Accessories 13
4.7. Repairs and defects 13
4.8. Transport 13
5. Precaution 14
6. Troubleshooting 14
7. Daily Maintenance 15
8. Disposal 16
9. Warranty 16
10. Product Specifications 17
11. Emc Declaration Of Conformity 18
12. Manufacturer’s Declaration 19
Table of Contents

1. Symbols Used
1.1.
In these instructions for use
If the instructions are not followed properly, the operations
may lead to hazards for the product or the user/patient.
WARNING
1.2. On the product/packaging
Serial number Catalogue number
Manufacturer Date of manufacture
Class II (AC Adapter) Consult the
instructions for use
This shows the Type
B applied part.
Do not dispose
it with normal
household waste
Recycling Ordinary equipment
EU-representative
Temperature limits
CE marking Keep dry
Atmospheric
pressure for storage
Storage humidity range
95%
10%
106kPa
50kPa
10° C
40° C
1

2. Product Information
The light has been manufactured with a super-high luminosity 10 W LED. The light
wavelength of LEDEX
TM
WL-090+ is between 390-480nm and the light intensity
is up to 2400mW/cm
2
. It can cure the composite over 2mm in 2 seconds. These
characteristics especially enable to phenyl propanedione (PPD) and Lucirine TPO
photoinitiator.
LEDEX
TM
WL-090+ dental curing light is characterized by :
•
8 powerful extensive modes including Low, Ramp, Standard, High, Fast Ortho,
Turbo, Plaque inspect and Caries inspect modes.
•
The light guide rod is made from genuine optical fiber and not inferior acrylic
that optimizes light conduction and minimizes loss of light from source to tip. It
ensures the highest possible intensity of light at the light guide tip.
•
Advanced and high efficient cooling heat sink are designed and accompanied
with over-heating prevention. A thermal protection circuit and safety mode are
also designed to prevent the light from overheating.
•
The automatic memorization of the last operation is another unique feature of the
light.
•
There is a built-in radiometer displaying light intensity.
•
Automatic sleep mode saves battery and extends usage time.
2.1. Indications for use
LEDEX
TM
WL-090+ is a visible curing unit programmed for polymerization of
dental light cured materials only by dental professionals.
FOR DENTAL USE ONLY!
2.2. Procedure
Please follow the instruction before use. As for handpiece instructions, please
refer to 2.6-2.7. It is required to use the disposable sleeves to prevent cross
infection.
Put the sleeve and cover the handpiece before operation. After conducting a
surgery on a patient, please take off the sleeve and throw away, as the sleeve is
prohibited to reuse.
2
EN

2.3. System components
abc
defg
hij
l
k
LEDEX™ WL-090+ Handpiece
Battery (3.7 V/2500 mAh)
Cradle
1
1
1
Item Description Quantity
a
b
c
m
Power supply(Input AC100~240 V,
50-60 Hz, output DC5 V/2 A)
Optical fiber light guide rod-
(Ø 11>8 mm)
Optical fiber light guide rod-
(Ø 11>5 mm)
Filter
1
1
1
1
d
e
f
Anti-glare shield
*FIND
TM
Optical filter
*FIND
TM
Mirror
2
1
1
g
h
i
j
Disposable curing light sleeves
Disposable light guide sleeves
Instructions for use
10
20
1
l
k
m
*FINDTM: Fluorescence Inspect and Determine system
Manual
WL-090+
3

2.4. Features
MODE :Pressing this button sequentially toggles the
unit through the 8 curing modes.
TIME :Pressing this button sequentially toggles the unit
through the serial curing time.
ON-OFF :Press this button to start the handpiece.
Press this button again to turn on the curing light.
Pressing this button during the curing cycle will
interrupt and end the cycle. If the unit is in sleep mode,
press this button will wake the unit and it will display
with the last mode chosen.
a
b
c
a
c
b
d
The display shows different information to the user.
As shown above, it comprises different zones identified
from top to bottom as follows:
A display of the curing mode selected.
The output mode icons represent the type of light
emission.
This is the display of duration seconds of the selected
curing cycle. During operation, this countdown display
indicates the remaining activation time until the current
cycle is completed.
A battery charge level indicator is symbolized by a
ladder with 0 to 5 levels and charging status.
a
b
c
d
2.5. Display
ab
c
WL-090+
4
EN

2.6. Installation and charging
Startup
When receiving the unit, any damages may occur
during transportation. If necessary, contact your
supplier.
Handpiece
First of all, this product must be cleaned as normal
preparation for each patient to prevent cross infection
(see the chapter 7).
Remove the protective caps from the handpiece which
should be kept whenever not in use to prevent liquid
from damaging the LED. Next, insert the light guide into
the handpiece and ensure the light guide is properly
inserted.
Battery
We recommend you to charge the battery fully before
the first use.
Put the battery into the handpiece and rotate clockwise
until you hear a “click” sound and feel the battery
clicked into the right place.
Ensure all the segments of the display are shown. The battery supplied is only charged
to about 60% prior to shipment. Each time, charge it fully before using it.
Click
WL-090+
5

1. Connect the power supply to AC100~240 V electronic
socket and plug-in the connector to the cradle. Put the
handpiece into the cradle to charge the battery.
2. When the battery is low, the display of the handpiece will
glow and show the of low battery sign.
3. When the battery is charging, the display of the handpiece
will glow and show the sign of charging battery.
4. When the battery is fully charged, the display of the
handpiece will be turn off to save power.
2.7. Operating modes
Select curing programs and curing time. The curing modes and the curing time
can be individually set. LEDEX
TM
WL-090+ is equipped with the following 8 curing
modes for different indications. Use the mode selection button to choose the
curing programs. The display will change accordingly (see Indicators on the
handpiece). The device comes equipped with the following pre-set modes :
6
EN

Icon Mode Curing time duration
(sec) Intensity of light
LOW 10,20,30,40 600 mW/cm
2
(+/-15%)
RAMP 5,10,15,20,25,30,35,40 1000 mW/cm
2
(+/-10%)
STANDARD 5,10,15,20,25,30,35,40 1000 mW/cm
2
(+/-10%)
HIGH 2,4,6,8,10 1800 mW/cm
2
(+/-10%)
FAST ORTHO 3,4,5 repeat 10 times 1800 mW/cm
2
(+/-10%)
TURBO 2,3,4,5 2400 mW/cm
2
(+/-10%)
PLAQUE
INSPECT
no timer selected
30 seconds setting only 250 mW/cm
2
(+/-15%)
CARIES
INSPECT
no timer selected
30 seconds setting only 350 mW/cm
2
(+/-15%)
LOW
10
RAMP
05
STANDARD
05
HIGH
02
FAST ORTHO
03
Factory settings
INSPECT
PLAQUE
INSPECT
CARIES
TURBO
02
7

Light intensity
Recommended Curing Time (on STANDARD mode)
Fill Materials Curing time
Universal composite (2 mm depth) 10 seconds
Universal composite (4 mm depth) 20 seconds
**Generally, these recommendations apply depends on situations. The emission
window of the light probe is placed directly over the material in order to be polymerized.
Extend the curing time accordingly to increase the distance between the light source
and the material.
Recommended curing modes
Mode Application
LOW Tooth and composite resin
RAMP Wide area of composite resin, avoid shrinkage.
STANDARD Most cases.
HIGH For orthodontic or pediatric dentistry,
FAST ORTHO Band materials
TURBO Dental cement, porcelain veneer, fiber post.
PLAQUE INSPECT Ideal for dental plaque inspect. The location of plaque
or tartar will be changed to red of fluorescence.
CARIES INSPECT
Ideal for caries inspect. The location of caries which has
caries bacteria will be changed to red of fluorescence.
If you see the dark spot, it also is caries.
8
EN

•
While using plaque or caries inspect, please put on 5 mm light guide rod, it
is the best size to check plaque and caries in dark environment and observe
the pure violet wavelength output through FIND
TM
optical filter, you can see
easily and immediately where excavation of caries is.
•
While using plaque or caries inspect, please turn off dental unit lamp and
reduce ambient light of clinic, it will avoid other sources of light to ensure that
the distinction between red and green fluorescence is clearly maintained.
The plaque & caries area will be displayed on red fluorescent results.
•
While using plaque or caries inspect, patient can hold the FIND
TM
Mirror,
patient is easy to see their plaque or caries via the mirror.
Tips
RedFluorescence
PlaqueorCaries
GreenFluorescence
Healthydentin
Cliniclamp
Reduceambientlight
Dentalunitlamp
Turnoff
FINDTM
Mirror
FINDTM
OpticalFilter
DarkSpot
Caries
Dentist
Patient
9

When the temperature becomes quite high, the display of
handpiece will glow and show the sign of “Over Heat”.
Over
Heat
When LEDEX
TM
WL-090 is operated frequently for long
period of time, the temperature may become too high.
Therefore, the “Safety” mode function will then be activated
automatically to prevent the light from overheating. The
safety mode cuts the light intensity in approx. half and
extends the irradiation time.
STANDARD
20
SS
Safety status :
LEDEX
TM
WL-090+ will sleeps automatically if no operations are performed for three
minutes, the display will be turn off.
Auto sleep designed :
1. Press button M to switch to sound setting.
2. In sound setting, press button T to switch between Mute and Unmute.
3. Press button M to return to other mode.
Sound mode
MUTE
3. Contraindications
For patients who are prone to photobiological reactions :
Do not use the LEDEX
TM
WL-090+ dental curing light for patients with a history of
photobiological reactions (including patients with Urticaria solaris or erythropoietic
protoporphyria) or those who currently have treatments with photosensitising
pharmaceuticals.
10
EN

4. Warnings
The handpiece is intended for the polymerization of light-cured
materials and is used by trained and qualified professionals, such
as dentists.
Do not place the device in humid surroundings or any places
which are close to any liquids.
Do not expose the device to any heat sources. Store the device in
a safe environment.
•
The device could be operated up to a maximum temperature of
35 °C and up to an altitude of 2,000 m above sea level.
•
Do not use the device in the presence of free oxygen,
anesthetics or flammable substances.
•
The device may interference or interfere with the radio or the
operation of the equipment nearby. If this happens, reduce
the interference by reorienting and repositioning the device or
screening off the immediate environment. The electromagnetic
radiation emitted from this device is below the recommended
limits specified by the applicable relevant provisions(EN 60601-
1-2:2007 & EN 60601-1:2006).
•
The device requires special precautions with regard to
electromagnetic compatibility (EMC) and it must be installed
and operated in strict compliance with the EMC information.
Especially, do not use the device in the vicinity of fluorescent
lamps, radio transmitters, remote controls, portable or mobile
RF communication devices, even if they meet CISPR 8
requirements.
•
Do not charge, operate or store the device at high
temperatures. Comply it with the specified operating and
storage conditions.
4.1. User
4.2. Ambient conditions

The LEDEX
TM
WL-090+ Dental Curing Light is an electric device
designed to meet the worldwide electrical safety standards, which
includes U.S.A. and Europe, so it’s safe and effective for all dental
applications.
To avoid electric shock:
•
Do not attempt to open or alter the unit in any way. Only the
service centers authorized by DENTMATE can open the unit
housing and repair the device.
•
Do not put any foreign objects into the housing of the unit.
•
Use only the LEDEX
TM
WL-090+ cradle when recharging this
product. Never attempt to use any other devices for recharging.
•
Connect the power plug into a suitably grounded and approved
outlet. When you use an extend cable, make sure the grounded
line is not interrupted.
•
Always unplug the charging dock before disinfecting.
•
Never use the power supply if the cord has been damaged.
4.3. To avoid electric shock (shock hazard)
As it is the case with all high-performance lights, the high light
intensity results in a certain heat development. Prolonged
exposure near the pulp and soft tissues may result in irreversible
or reversible damage. Therefore, this high-performance curing
light must be operated by trained professionals.
4.4. Heat development (burn hazard)
Use only original spare parts, particularly DENTMATE batteries
and charging bases. Do not short circuit battery. Do not store at
temperatures above 40 °C / 104 °F(or 60 °C / 140 °F for a short
period). Always store batteries charged. The storage period must
not exceed 6 months. It may explode if it’s disposed of in fire.
4.5. Battery
Note : At least 10 mm gap between soft tissues and optical fiber
light guide rod.
12
EN

Only use original DENTMATE components/accessories and spare
parts :
Original DENTMATE accessories DENTMATE REF
LEDEX™ WL-090+ Handpiece 3102-1001
Battery (3.7 V/2500 mAh) 2101-0010
Cradle 3102-2001
Power supply (Input AC100~240 V, 50-60
Hz, output DC5V/2 A) 2103-0008
Optical fiber light guide rod-(Ø 11>8 mm) 2402-0007
Optical fiber light guide rod-(Ø11>5 mm) 2402-0009
Filter 2101-0013
FIND
TM
Optical filter 2102-0011
FIND
TM
Mirror 2102-0013
Anti-glare shield 1220-0004
Disposable light guide sleeves 2401-0001
Disposable curing light sleeves 2401-0005
Using other accessories/spare parts may lead to increased
emission of electromagnetic interference or to reduced
electromagnetic interference immunity.
4.6. Accessories
4.7. Repairs and defects
Do not use the device if you suspect its damage or defect.
4.8. Transport
Intact devices can be transported by land freight or air freight in
the original packaging. The applicable requirements must be met.
Defective devices can also be transported by air freight or land
freight in the original packaging. If the battery is defective, the
device won’t be able to be transported by air freight under any
circumstances.
13

5.1. During operation, the light should be aimed straightly on the resin to ensure
solidification effectively.
5.2. Never aim the light directly at unprotected soft tissues because this may lead to
injury or irritation. Do not aim the light at eyes. Light reflected from the tooth surface
may also injure eyes. Use the protective shield supplied with the unit or suitable,
light filtering safety glasses.
5. Precaution
Problem Resolution
Can not turn on the handpiece
Remove the battery and insert it again. If the
error
persists, please plug-in the power supply to
cradle and charge the battery at least 10 mins.
Then push the ON/OFF button again.
Can not charge the battery Please clean battery contacts. If the error
persists, please change a new battery.
Display shows “overheat” sign If the temperature rises up to a really level,
please wait a moment and then use it again.
Display shows “Error 1” sign LED module is unnormal, please call your
qualified technician.
The intensity is too low
If the result of intensity test in Standard mode
is under 700 mW/cm
2
, and the output is too
low,please contact with your dealer.
6. Troubleshooting
14
EN

7. Daily Maintenance
Read this entire section before cleaning the unit. This product must be cleaned as
normal preparation for each patient to prevent cross infection.
The use of the sleeve is an additional precautionary measure against contamination
and does not substitute disinfection of the device.
After using, remove the sleeve. Clean the optical light guide rods and the handpiece
with cloth moisture with alcohol or cleaner. Keep solvents or flammable liquids from
the unit because they may damage its plastic housing.
Always keep the charger, handpiece and light guide well. Moisture may cause
electrical short-circuit or dangerous malfunction.
1. The curing time interval should exceed 5 seconds
for each cycle.
2. Verify the radiometer sensor which can impact the
accuracy of the measurement. The surface of sensor
area can be wiped with a cotton swab with alcohol.
3. Carefully hold the unit, so the Light Guide
Attachment is aligned with the radiometer sensor
and centered within the white circle provided.
STANDARD
05
Tips:
Optical fiber light guide must be horizontal
alignment in the middle of the sensor area.
Test the Light Guide Attachment with the Radiometer
Verify the LEDEX
TM
WL-090 performance each time before using the radiometer
which is built into the cradle.
15

4. While holding the unit, press and release the ON-
OFF Button. The radiometer will then provide a reading
of the light intensity on the cradle display.
5. Wait till the light is off before moving the Light Guide
Attachment away from the radiometer sensor.
1020
Notice:
The WL-090 built-in radiometer is only suitable
for measuring 8 mm optical fiber light guide rod.
It will cause incorrect measurement if it uses in
other sizes of optical light guide rod.
8. Disposal
Comply with your national regulations, guidelines and requirements for the disposal
of end-of-life electrical equipment and batteries. Specialized dental dealers will be
pleased to provide you with country-specific information concerning disposal.
This device is provided with a Li-ion battery. For environmental reasons, please
dispose of the device according to local environmental guidelines or regulations.
Make sure the product or the battery is not mixed with other types of waste when it
is disposed of. Prior to disassembly and disposal, your device has to be completely
reprocessed and must not be contaminated.
9. Warranty
DENTMATE TECHNOLOGY Co., Ltd. warrants the product to be free of
manufacturing defects for a period of one year from the date of purchase; this
is deemed as the date of the invoice. It could be repaired or replaced at its own
discretion all equipment failures due to manufacturing defects. However,the
followings are expressly excluded from the warranty:
1.
Damage and/or failure of the equipment caused by falling and/or jolting during
transportation after purchase and/or during the normal use.
2.
Damage and/or failure of the equipment caused by natural disasters, such as
earthquakes, floods, lightning, pollution, incorrect electrical voltage and voltage
spikes.
3. Any attempts to open the hand piece will invalidate the warranty.
16
EN

10. Product Specifications
Type of Information Specifications
Dental curing light Medical equipment
Device name LEDEX
TM
Model number WL-090+
Power supply Input: AC100~240 V, 50-60 Hz
Output: DC 5 V/2 A
Battery 3.7 V, 2500 mAh, type: Li-ion
Light source 10 W LED
The range of wavelength 390 ~ 480 nm, peak: 405, 460 nm
Radiant intensity Up to 2400 mW/cm² (± 10%)
Hand piece dimensions Ø38 (max.) x L190 mm
Hand piece weight 180 g (with battery & light guide rod)
Cradle dimensions Ø115 (max.) x H68 mm
Cradle weight 140 g
Equipment class (AC Adapter) Class II
Safety IEC 60601-1
EMC(Electro-Magnetic Compliance) IEC 60601-1-2
Protection from electric shock Type B applied part
Protection from ingress of liquids IPX0
Operation
Continuous operation patient application, duty
cycle 40 seconds ON / 120 seconds OFF on
STANDARD mode.
Operating environment
Ambient temperature:
Relative humidity:
Atmospheric pressure:
Storage and transport environment
Ambient temperature:
Relative humidity:
Atmospheric pressure:
75%
30% 1.0-atm
0.5-atm
10° C
40° C
50° F
104° F
106kPa
50kPa
95%
10% 1.0-atm
0.5-atm
10° C
40° C
50° F
104° F
106kPa
50kPa
17