DV Power TRT30C User manual

TRT30C
THREE-PHASE TRANSFORMER
TURNS-RATIO TEST INSTRUMENT
Manual

TRT30C Contents
M-T030CN-203-EN 2
Contents
1 Introduction............................................................................................................................................3
1.1 Safety Instructions..............................................................................................................................3
1.1.1 Safety Terms and Symbols.........................................................................................................3
1.1.2 Terms of Use..............................................................................................................................3
1.1.3 Orderly Practices and Procedures..............................................................................................4
1.1.4 Instrument Maintenance.............................................................................................................4
1.1.5 Operator Qualifications...............................................................................................................4
1.1.6 Safe Operating Procedures ........................................................................................................4
1.2 Power Supply.....................................................................................................................................5
1.3 Measurement Category......................................................................................................................5
1.4 Intended Use......................................................................................................................................5
2 Description.............................................................................................................................................7
2.1 Front Panel Components....................................................................................................................7
3 Getting Started.......................................................................................................................................9
3.1 Connecting the TRT30C to a Test Object...........................................................................................9
3.2 Setting the Measurement Parameters..............................................................................................10
4 Error Messages....................................................................................................................................13
4.1 Error Message “Excitation current too high”......................................................................................13
4.2 Error Message “Turns ratio too low” .................................................................................................13
4.3 Error Message “Malfunction” ............................................................................................................13
4.4 Error Message “Error printer”............................................................................................................14
4.5 Error Message “Check paper” ..........................................................................................................14
4.6 Error Message “USB flash drive” ......................................................................................................14
4.7 Error Message “Connect both neutral cables” ..................................................................................14
4.8 Error Message “Emergency Stop” ....................................................................................................15
4.9 Error Message “Connections”...........................................................................................................15
5 Technical Data.....................................................................................................................................16
5.1 Mains Power Supply.........................................................................................................................16
5.2 Output data ......................................................................................................................................16
5.3 Measurement...................................................................................................................................16
5.4 Environmental conditions .................................................................................................................16
5.5 Dimensions and Weight....................................................................................................................16
5.6 Applicable Standards .......................................................................................................................17
6 Accessories..........................................................................................................................................18
Manufacturer Contact Information................................................................................................................19
Manual Version: M-T030CN-203-EN
This Manual refers to the firmware version TR3C-5.15

TRT30C Introduction
M-T030CN-203-EN 3
1 Introduction
The purpose of this Manual is to provide helpful instructions on how to use the TRT30C instrument safely,
properly and efficiently.
The following instructions will help the user avoid unsafe situations, reduce maintenance costs and will
ensure the reliability and durability of the TRT30C instrument.
The TRT30C must be used in accordance with all existing safety requirements and regulations based on
national/local standards for accident prevention and environmental protection. In addition, the relevant
international standards are listed in paragraph 5.6 of the “Technical Data” section of this document.
1.1 Safety Instructions
Safety is the responsibility of the user. Before operating the TRT30C, please read the following safety
instructions carefully.
It is not recommended that the TRT30C is used (or even turned on) without careful observation of the
instructions listed in this Manual. The TRT30C should only be operated by trained and authorized
personnel.
1.1.1 Safety Terms and Symbols
Terms in this Manual
These terms may appear in the Manual:
WARNING: Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or
to other property.
Terms on the Device
The following warning terms used in this document may appear on the device:
WARNING: Warning indicates that potential hazard may occur.
CAUTION: Caution indicates that potential damage may occur to the instrument or to the test object
connected to the instrument.
Symbols on the Device
The following symbols may appear on the device:
Refer to Protective
Manual Earth
Terminal
1.1.2 Terms of Use
•The TRT30C shall be used only if it is in good technical condition. Its use shall be in accordance with
local safety and industrial regulations. Adequate precautions must be taken to avoid any risks related to
high voltages associated with this equipment and nearby objects.
•The TRT30C shall be used only for the application purposes described in the ”Intended Use” section.
The manufacturer and distributors are not liable for damage resulting from wrong usage. The user bears
responsibility for not following the instructions defined in this document.

TRT30C Introduction
M-T030CN-203-EN 4
•Do not remove the protective casing of the TRT30C.
•All service and maintenance work must be performed by qualified personnel only.
1.1.3 Orderly Practices and Procedures
•The Manual shall always be available on the site where the TRT30C is used.
•Before using the TRT30C, all personnel (even personnel who only occasionally, or less frequently, work
with the TRT30C) assigned to operate the TRT30C should read the operations Manual.
•Do not make any modifications, extensions, or adaptations to the TRT30C.
•Use the TRT30C only with the original accessories provided by the manufacturer.
•Use the TRT30C and its original accessories for the device’s intended use only.
1.1.4 Instrument Maintenance
The device should be kept clean in order to prevent excessive cases of dust or other contaminants affecting
its operation. It should be cleaned with water/isopropyl alcohol after any dirt/contaminants are noticed on its
surfaces.
1.1.5 Operator Qualifications
•Testing with the TRT30C should only be carried out by authorized and qualified personnel.
•Personnel receiving training, instruction or education on the TRT30C should remain under the constant
supervision of an experienced operator while working with the test set and the test object.
1.1.6 Safe Operating Procedures
•Hazardous voltages of up to 400 V can occur inside the TRT30C. Therefore, it is not permitted to
remove the protective casing of the TRT30C.
•Hazardous voltages exist on the terminals of the TRT30C when the “Red”LED is lit. Never assume that
the connections are safe even if this LED is off. Switch off and unplug the TRT30C before touching the
connections, particularly if a fault is suspected.
•Before putting the TRT30C into operation, check the test set for any visible damage.
•Do not operate the TRT30C under wet or moist conditions (condensation).
•Do not operate the TRT30C if explosive gas or vapors are present.
•Only the external devices which meet the requirements for SELV equipment according to EN 60950 or
IEC 60950 should be connected to the TRT30C through the serial interface.
•Removing theTRT30C protective casing will void the warranty. Any work inside the instrument without
prior authorization from DV Power will also void the warranty.
•If the TRT30C seems to be malfunctioning, please contact the DV Power Support Team (refer to the
“Manufacturer Contact Information” section) after previously checking the “Error Messages” section.
•Prior to connecting the TRT30C, ensure that the transformer (object) to be tested is completely de-
energized and isolated from both the line and the load. Every terminal should be checked and verified
before connecting the TRT30C. Ground connections may be left in place.
•Do not use the TRT30C without the extra protective ground cables supplied with the TRT30C. It must
never be operated in a non-grounded configuration as this may result in an electrical shock to the user
or damage to the TRT30C. Always establish this connection first before establishing any other
connections and remove this connection as the very last one.

TRT30C Introduction
M-T030CN-203-EN 5
•Wherever possible, the outer casing of the transformer under test should also be connected to a safety
earth to prevent the risk of shock. Where this cannot be achieved, adequate precautions should be
taken to prevent access to the transformer (e.g. barriers).
•Never exchange connections to the HV and LV sides of the transformer. The “H”connections must
always be connected to the High Voltage side of the transformer. The “X”connections must always be
connected to the Low Voltage side of the transformer. Interchanging any or all of the connections may
result in damage to the transformer or the TRT30C and is a significant safety hazard to personnel.
1.2 Power Supply
•Supply the TRT30C only from a power outlet which is equipped with the protective ground.
•Besides being supplying the TRT30C from phase –neutral (L1-N, A-N), it may also be supplied from
phase to phase (e.g., L1-L2; A-B). However, the voltage must not exceed 264 V AC. Please refer to the
section “Technical Data”.
•The TRT30C should be positioned in such a way that it is possible to safely disconnect it from the power
supply at any moment.
WARNING / AVERTISSEMENT
This is a class A product. In a domestic environment this product may cause radio interference in which
case the user may be required to take adequate measures.
Il s'agit d'un produit de classe A. Dans un environnement domestique, ce produit peut provoquer des
interférences radio, auquel cas l'utilisateur peut être amené à prendre des mesures adéquates.
1.3 Measurement Category
The TRT30C is intended to be used for measurements in Measurement Category I (CAT I) for voltages up
to 250V. The device is also designed to withstand occasional transient overvoltage up to 1000Vpk.
WARNING / AVERTISSEMENT
This equipment is classified as measurement category I, and must not be used within measurement
category II, III and IV.
Cet équipement est classée dans la I catégorie de mesure, et ne doit pas être utilisé pendant les
catégories de mesure II ,III et IV.
1.4 Intended Use
The Three-phase Transformer Turns Ratio Tester TRT30C is designed specifically for performing
measurements on power, distribution and measurement transformers as follows:
•turns ratio measurement,
•phase shift measurement,
•excitation current measurement.
These three tests are performed at the same time. The TRT30C applies test voltage to HV transformer
terminals and measures induced voltage at LV transformer terminals. The ratio of these voltages is actually
the turns ratio of a transformer. At the same time, the TRT30C measures the excitation current at the HV
transformer side, as well as the phase angle between the applied test voltage and induced voltage.

TRT30C Introduction
M-T030CN-203-EN 6
There are three ways to perform the measurements depending on the test object and test voltage. If the test
object is a single-phase transformer, a single-phase autotransformer or a current transformer, the TRT30C
provides a single-phase test voltage to the transformer’s primary side and measures the induced voltage at
the transformer’s secondary side.
If the test object is a three-phase transformer or a three-phase autotransformer, the TRT30C can apply
either three single-phase test voltages in turns or a true three-phase test voltage to the transformer primary
windings. Three induced voltages at the transformer’s secondary side are measured.
Please note that any uses of the TRT30C other than the ones described in this Manual are considered
improper and will void the device warranty.

TRT30C Description
M-T030CN-203-EN 7
2 Description
2.1 Front Panel Components
1. Mains Power Connector and Power Switch
Mains power connector
Connect the TRT30C to the mains power supply with a power cord.
Power switch –Double pole switch
•IIn this position, the TRT30C is connected to the mains power supply.
•0In this position, the TRT30C is separated from the mains power supply.
2. Protective Earth Connector
For protection against parasitic currents or voltages, always connect the TRT30C protective earth
connector to the protective ground (PE). Use only the original cable.
For safety reasons, always establish this connection before establishing any other connection, and
remove this connection as the very last step.
3. Operator Control
Display
Displays the settings during the programming of the device as well as the measured values during a
test operation.

TRT30C Description
M-T030CN-203-EN 8
Keyboard
Used to control the device.
•Use the UP/DOWN button to navigate and set the test parameters.
•Use the ENTER button to confirm the defined test parameters, language, time and date.
•Use the STOP button to stop a test, to acknowledge the alarm buzzer and to return to the
previous menu.
•Use the SET button to scroll between the two main menus and the START button to start the
test.
•Use the PRINT button to print the results with the built-in thermal printer. The built-in thermal
printer is an optional accessory.
Green LED
•Lights continuously when the TRT30C is turned on.
•Flashes when a test can be started.
Red LED
•Lights continuously in case of an operational error.
•Lights continuously when a test is started and a test voltage is applied.
4. H and X Terminals
H cable terminal
The terminal for connecting the H test cable.
X cable terminal
The terminal for connecting the X test cable.
5. Emergency Stop
Turns off output voltage in case of an emergency.
6. Interface and Flash Drive Connectors
Interface
The TRT30C is equipped with an USB serial interface to connect to an external computer if
required.
Flash Drive
The TRT30C is equipped with a USB flash drive connector to save the test results to the flash drive
for additional analysis if desired.
7. TC Control
TC Connector
Output for the remote control of the transformer tap changer.
•Use the UP button to raise the tap changer position.
•Use the DOWN button to lower the tap changer position.

TRT30C Getting Started
M-T030CN-203-EN 9
3 Getting Started
3.1 Connecting the TRT30C to a Test Object
Before the TRT30C has been connected to a test object (e.g. power transformer), the following steps have
to be verified:
•The test object is disconnected from its circuit in accordance with the national safety regulations and
is properly grounded to the protective earth.
•The transformer is completely de-energized.
•The TRT30C itself should be properly grounded. To do this, the grounding screw on the top of the
TRT30C should be connected to PE using the provided grounding cable.
Note: Always connect measuring cables to the TRT30C first and then to the test object
terminals. Additionally, measuring cables can be connected to the test object terminals
first, but only under condition that the test object terminals are grounded.
When disconnecting, always disconnect cables from the test object terminals first and
then from the TRT30C. Additionally, measuring cables can be disconnected from the
TRT30C first, but only under condition that the test object terminals are grounded.
The grounding wire PE should be disconnected last.
Not following these instructions may cause life-threatening situations.
Cable test leads are terminated with the specially adjusted clips. The colors and markings of the test cables
are shown in table 1.
Table 1: Test lead colors and markings
If the primary neutral connection does not exist, the 1N (H0; N) lead should be left isolated from the
transformer (i.e. the test object) and the other connection leads. It must be placed in such a way that it is
completely isolated from other connections or earth. No personnel should be in a position to touch it!
To maximize the accuracy and measurement repeatability, make sure that all clamps have a good
connection to the test object and avoid any crossing between the measuring cables.
If the three-phase transformer has a tertiary winding, it should be tested in the same manner as the
secondary. The transformer is energized using the high voltage windings. Test leads 2U, 2V, 2W and 2N
(X1, X2, X3 and X0; a, b, c and n)are in this case connected to the tertiary winding of the transformer.
IEC Test
Lead
Marking
ANSI
Test Lead
Marking
Australian
Test Lead
Marking
Transformer
Terminal
Voltage
Test Lead
Color IEC
Australian
Test Lead
Colors
Malaysian
Test Lead
Colors
1N
H0
N
Neutral
Blue
Black
Black
1U
H1
A
High
Red
Red
Red
1V
H2
B
High
Black
White
Yellow
1W
H3
C
High
Yellow
Blue
Blue
2N
X0
n
Neutral
Blue
Black
Black
2U
X1
a
Low
Red
Red
Red
2V
X2
b
Low
Black
White
Yellow
2W
X3
c
Low
Yellow
Blue
Blue

TRT30C Getting Started
M-T030CN-203-EN 10
3.2 Setting the Measurement Parameters
Five seconds after the first (initial) message on the display, the TRT30C automatically changes the display
to the CONFIGURATION menu and the green LED lights up. By pressing SET the user can switch between
the CONFIGURATION menu and the MEMORY menu.
Press UP/DOWN in the HV CONFIG field to select the test mode to be performed. A list of the available test
modes is shown in tables 2 and 4.
Table 2: Available options in the configuration menu
HV
Config.
LV
Config.
Ph
Displacement
Comment
Test mode
D
displayed
displayed
Delta winding
Phase by phase test
using single-phase
voltage in turns
&
the true 3-phase test
using 3-phase voltage
Y
displayed
displayed
Star winding
YN
displayed
displayed
Star winding with neutral point
Z
displayed
displayed
Zigzag
ZN
displayed
displayed
Zigzag with neutral point
AVGD
hidden
hidden
Automatic vector group
detection
SINGLE
hidden
hidden
Single-phase transformer
Single phase test
using single-phase
voltage
CT
hidden
hidden
Current transformer (CT)
TRUE 3~
TEST
hidden
hidden
Only true three-phase test
True 3-phase test
using 3-phase voltage
AUTO
hidden
hidden
Autotransformer menu
See table 4
Press ENTER to choose one of the available options in the configuration menu.
1. If the vector group is selected, the TRT30C will perform both the true three-phase test by applying true
three-phase test voltage and the phase by phase test by applying single phase test voltages to all phases
in turn.
2. If SINGLE is chosen, the TRT30C will test a single phase transformer by applying the single phase test
voltage between 1U/H1/A and 1N/H0/N.
3. If CT is chosen, the TRT30C will test a current transformer by applying the single phase test voltage
between 1U/H1/A and 1N/H0/N.
4. If AUTO is chosen, the TRT30C will display the special menu for autotransformers to the user.
5. If the AVGD option is selected, the TRT30C will automatically detect the vector group of the transformer
under test.
6. If TRUE 3~ TEST is selected, the TRT30C will perform the true three-phase test by applying the true
three-phase test voltage.
Table 3 lists the available vector groups in this menu.
Table 3: Available vector group configurations
Configuration
Displacements
D-D, D-Z, D-Zn, Z-D, Zn-D, Y-Y, Y-Yn, Yn-Y, Yn-Yn
0-2-4-6-8-10
D-Y, D-Yn, Y-D, Y-Z, Y-Zn, Yn-D, Yn-Zn, Z-Y, Z-Yn, Zn-Y, Zn-Yn
1-3-5-7-9-11

TRT30C Getting Started
M-T030CN-203-EN 11
Available options in the AUTOTRANSFORMER menu are listed in Table 4.
Table 4: Available options in the autotransformer menu
Config
PH Displacement
Comment
Test mode
Da
displayed
Delta winding
Phase by phase test using
single-phase voltage in turns
or true 3-phase test using 3-
phase voltage, depending on
particular vector group.
Ya
displayed
Star winding
Yna
displayed
Star winding with neutral point
Za
displayed
Zigzag
Zna
displayed
Zigzag with neutral point
AVGD
hidden
Automatic vector group
detection
TRUE 3~ TEST
hidden
Only true three-phase test
True 3-phase test using 3-
phase voltage
SINGLE
hidden
Single-phase autotransformer
Single-phase test using
single-phase voltage
Table 5: Available autotransformer vector groups
Configuration
PH displacement
Ya, Yna
0-4-8
Da, Za, Zna
0-1-2,…11
If ratio deviation calculation is enabled, the following menu is displayed for entering the name plate voltages
and the tap changer positions:
If the name plate voltages are not entered, the TRT30C will perform the test and display the results, but it
will not calculate or display the turns ratio deviation. After the name plate voltages and tap changer
position(s) are entered, press ENTER to confirm selection. The next display allows the selection of the test
voltage.
The TRT30C generates both single-phase and true three-phase test voltages. When generating a single-
phase excitation voltage, the TRT30C supplies a single-phase excitation test voltage to each of the three
transformer primary windings in turn. The induced voltages across each of the corresponding unloaded
transformer windings are then measured. The ratio of these voltages is calculated and presented on the
display. It is repeated for all three phases.
Figure 3-1:
Input name plate
voltages values
High Voltage Tap Changer
position
Low Voltage Tap Changer
position
Figure 3-2:
The test voltage
selection

TRT30C Getting Started
M-T030CN-203-EN 12
When generating a true three-phase test voltage, the TRT30C generates and provides a true three-phase
excitation test voltage to three transformer primary windings. The induced three-phase voltages across the
unloaded transformer windings are measured and the transformer turns ratio is calculated.
Test voltage of 8 V is recommended for current transformers (CT) testing. For testing distribution or power
transformers it is recommended to use the highest test voltage of 80 V. In addition, test voltage of 40 V can
also be used. Pressing UP/DOWN the user selects the value of his/her choice and then confirms it by
pressing ENTER.
The TRT30C automatically detects if it is connected to a DC power supply. It also automatically detects if
the power supply frequency is 50 Hz or 60 Hz. The frequency of the output voltage is the same as the
frequency of the power supply of the TRT30C. In case the TRT30C is connected to a DC power supply, the
user has to choose the frequency of the output voltage to be either 50 Hz or 60 Hz.
After the test voltage (and frequency) is selected, the user can select a memory location manually (Test
Reading and Test Record). SET is used to scroll between Test Reading and Test Record use. Selection of
Test Reading or Test Record is performed by using UP/DOWN. Pressing ENTER confirms a selection.
Test Reading can contain, for example, all the test results obtained in one tap changer position of the
transformer. Test Record can contain the results of all tap changer positions, or the complete results
obtained from one transformer.
Note: Manual selection of record/reading will delete results from the existing memory locations.
Once these parameters are defined, pressing ENTER changes the status to the READY state.
If one of the values has to be changed, press STOP to return to the previous menu.
The flashing green LED indicates that the TRT30C is now ready to start the test. Press START to run a test.
During the test, the red LED glows continuously signaling that the test voltage is applied. Pressing STOP
(the CONFIGURATION menu is displayed) will interrupt (abort) the test. When the test is completed, the
TRT30C displays the test results. To scroll between all the results press ENTER. Once all the results are
viewed, press ENTER to test the next tap or to finish the test.
If the AVGD option in the CONFIGURATION menu is selected, the TRT30C will try to automatically detect a
vector group of the transformer being tested. The user will be asked whether to continue the test with a
detected vector group or to select one of the available vector groups. If for some reason the vector group
cannot be detected, the device will display the message “Unable to detect vector group” and the test will be
aborted. Once the vector group is detected and confirmed, the TRT30C continues with the test. After the
test is finished, the TRT30C displays the test results. To scroll between all the results, the user will press
ENTER. Once all the results are viewed, the user can choose whether to test the next tap or to finish the
test.
Figure 3-3:
The READY state
before the test

TRT30C Error Messages
M-T030CN-203-EN 13
4 Error Messages
Any operational error is indicated by a red LED and accompanied by an audio alarm. Furthermore, the
display indicates an error status message. To remove the status message on the display and return to the
main menu, press STOP.
4.1 Error Message “Excitation current too high”
The Error message “Excitation current too high” is displayed if the excitation current exceeds expected
values. This value is set to 500 mA. The excitation current can be reduced by selecting lower test voltages
from the device menu.
Possible reasons for this are that the transformer is drawing too much current. In that case, the test should
be repeated with a lower test voltage. Another reason for an excitation current which is too high is that the
connections are reversed. If there are no connections errors found, there may be a short circuit within the
transformer itself.
4.2 Error Message “Turns ratio too low”
This message is displayed if the turns ratio of the transformer under the test is lower than 0.8. In this case
the user should check whether H and X cables are misplaced (reversed).
4.3 Error Message “Malfunction”
In the case of an internal error, the TRT30C will display the massage “Malfunction”. In this case, the
TRT30C device should be restarted.
If the error occurs repeatedly, please contact the DV POWER Support Team (refer to the section
”Manufacturer Contact Information”).
Figure 4-2:
Error message
"Turns ratio too low’’
Figure 4-3:
Error message
“Malfunction”
Figure 4-1:
Error message
"Excitation current too high"

TRT30C Error Messages
M-T030CN-203-EN 14
4.4 Error Message “Error printer”
The message “Error printer” is displayed if the hardware error concerns the printer. In this case, please
contact DV POWER Support Team (refer to the section ”Manufacturer Contact Information”).
This message is also displayed if the user tries to print the results from the device which has no built-in
thermal printer installed.
4.5 Error Message “Check paper”
The message “Check paper” is displayed if the printer is out of paper.
4.6 Error Message “USB flash drive”
If the USB flash drive is not plugged in while trying to export the data, the error message shown below will
be displayed.
4.7 Error Message “Connect both neutral cables”
If one of the neutral cables is not connected to the autotransformer under test during AVGD test, the error
message shown below will be displayed. User needs to connect both neutral cables to avoid this message.
Figure 4-5:
Error message
"Check paper"
Figure 4-4:
Error message
"Error printer"
Figure 4-6:
Error message “USB flash
drive”
Figure 4-7:
Error message “Connect
both neutral cables”

TRT30C Error Messages
M-T030CN-203-EN 15
4.8 Error Message “Emergency Stop”
This message is displayed when the “Emergency stop” button on the front panel of the device is pressed.
Release the “Emergency stop” button by rotating it clockwise.
4.9 Error Message “Connections”
The Error message “Connections” is displayed if the H or X cables are not connected properly to the
TRT30C. There are also green LED lights near the connectors which light up to indicate that these
connectors are connected properly.
Figure 4-9:
Error message
"Connections"
Figure 4-8:
Error message
"Emergency stop"

TRT30C Technical Data
M-T030CN-203-EN 16
5 Technical Data
5.1 Mains Power Supply
- Connection: According to IEC/EN60320-1; UL498, CSA 22.2
- Voltage: 90 –264 V AC
- Frequency: 50 –60 Hz
- Input power: 200 VA
- Fuse: 2 A / 250 V, type F, but not user replaceable
5.2 Output data
- Test Voltage 8 V AC, 40 V AC, 80 V AC
3 x (8, 40, 80)3 V AC
5.3 Measurement
- Ratio Measuring Range 0.8 to 15 000 (5-digit Resolution)
- Typical Ratio Accuracy:
0.8 - 999: ±0.05 %
1000 - 3999: ±0.05 %
4000 - 15000: ±0.1 %
@80 V AC
0.8 - 999: ±0.05 %
1000 - 3999: ±0.1 %
4000 - 15000: ±0.2 %
@40 V AC
0.8 - 999: ±0.05 %
1000 - 3999: ±0.1 %
4000 - 15000: ±0.2 %
@8 V AC
- Guaranteed Ratio Accuracy:
0.8 - 999: ±0.125 %
1000 - 3999: ±0.125 %
4000 - 15000: ±0.25 %
@80 V AC
0.8 - 999: ±0.125 %
1000 - 3999: ±0.25 %
4000 - 15000: ±0.5 %
@40 V AC
0.8 - 999: ±0.125 %
1000 - 3999: ±0.25 %
4000 - 15000: ±0.5 %
@8 V AC
- Excitation Current Range 0 –2 A
- Typical Excitation Current Accuracy ± (0.25% + 500 μA)
- Guaranteed Excitation Current Accuracy ± (0.25% + 1 mA)
- Excitation Current Resolution
0 –9.9999 mA 0.1 μA
10 –99.999 mA 1 μA
100 –999.99 mA 10 μA
1 –2 A 100 μA
- Phase Angle Range 360 Degrees
- Typical Phase Angle Accuracy ± 0.05 Degrees
- Guaranteed Phase Angle Accuracy ± 0.1 Degrees
- Phase Angle Resolution 0.01 Degree
5.4 Environmental conditions
- Operating temperature -10 °C –+55 °C / +14 °F –+131 °F
- Storage temperature -40 °C –+70 °C / -40 °F –+158 °F
- Maximum relative humidity 95 % non-condensing
5.5 Dimensions and Weight
- Dimensions (W x H x D) 480 x 197 x 395 mm / 18.9 x 7.75 x 15.55 in
- Weight 8 kg / 17.5 lbs

TRT30C Technical Data
M-T030CN-203-EN 17
5.6 Applicable Standards
- Installation/overvoltage category: II
- Pollution degree: 2
- Safety LVD 2006/95/EC (CE Conform)
Standard EN 61010-1:2001
- EMC Directive 2004/108/EC (CE Conform)
Standard EN 61326-1:2006

TRT30C Accessories
M-T030CN-203-EN 18
6 Accessories
Included
Article No
DV-Win PC software including USB cable
Built-in Tap Changer Control Unit
Tap Changer Control cable 5 m
Mains power cable
Ground (PE) cable
Recommended
Article No
H winding test cable 5 m, three-phase connection, with TTA clamps
HC-05-03MCWC
H winding test cable extension, 5 m, shielded
HE-05-03MCWC
X winding test cable, 5 m, three-phase connection, with TTA clamps
XC-05-03FCWC
X winding test cable extension, 5 m, shielded
XE-05-03FCWC
Cable bag
CABLE-BAG-01
Optional
Article No
H winding test cable extension, 10 m, shielded
HE-10-03MCWC
X winding test cable extension, 10 m, shielded
XE-10-03FCWC
H winding test cable extension, 15 m, shielded
HE-15-03MCWC
X winding test cable extension, 15 m, shielded
XE-15-03FCWC
H winding test cable 3 m, 3~ connection, with TTA clamps
HC-03-03MCWC
X winding test cable 3 m, 3~ connection, with TTA clamps
XC-03-03FCWC
H winding test cable 10 m, 3~ connection, with TTA clamps
HC-10-03MCWC
X winding test cable 10 m, 3~ connection, with TTA clamps
XC-10-03FCWC
H winding test lead, 3 m, 1~ connection, with TTA clamps
HC-03-01MCWC
X winding test lead, 3 m, 1~ connection, with TTA clamps
XC-03-01FCWC
H winding test lead, 5 m, 1~ connection, with TTA clamps
HC-05-01MCWC
X winding test lead, 5 m, 1~ connection, with TTA clamps
XC-05-01FCWC
Cable plastic case –medium size
CABLE-CAS-02
Cable plastic case with wheels –medium size
CABLE-CAS-W2
Transport case
HARD-CASE-D1
Built-in thermal printer
PRINT-112-00
Bluetooth communication module
BLUETOOTH-00
Inverter 12 V DC to 230 V AC, 50 Hz
IN650-12-230
Verification Calibrator TRTC
TRTC-05-4800
H winding test lead, 1 m, 3~ connection with banana plugs
HC-01-03MCBP
X winding test lead, 1 m, 3~ connection with banana plugs
XC-01-03FCBP
IBEKO Power AB 2013
This Manual is a publication of IBEKO Power AB, 181 50 Lidingö, Sweden. These documents are protected by
Swedish Copyright law and international contracts as intellectual property of the IBEKO Power AB. The documents
contain confidential information of IBEKO Power AB which is protected by patent, copyright, trademarks or otherwise
as inventions, trademarks or creations of IBEKO Power AB. The reproduction, duplication, transmission or use of these
documents or its contents is not permitted without express prior written consent of the IBEKO Power AB.
IBEKO Power AB shall not be liable for any incidental or consequential damages resulting from the performance or use
of this document or its product. This document has undergone extensive technical approval before being released.
IBEKO Power AB reviews this document at regular intervals, and includes appropriate amendments in subsequent
issues. While every effort has been made to keep the information herein as accurate and up to date as possible,
IBEKO Power AB assumes no responsibility for errors or omissions or for damages resulting from the use of the
information contained herein. IBEKO Power AB cannot take over liability resulting in any way from the use of this
document or parts thereof. The product information, pictures, drawings and all technical data contained within this
manual are not contractually binding and IBEKO Power AB reserves the right to make modifications at any time to the
technology and/or configuration without prior notice.
Insofar as any information, software or documentation is made available, any liability for defects as to quality or title of
the information, software and documentation especially in relation to the correctness or absence of defects or the
absence of claims or third party rights or in relation to completeness and/or fitness for purpose are excluded except for
cases involving willful misconduct or fraud.
In case of a disagreement between the translation and the original English version of this Manual, the original English
version will prevail.
Table of contents
Popular Transformer manuals by other brands

ATL Transformers
ATL Transformers ECO-RAIL T3067 O & M Manual

Pulsar
Pulsar AWT050 Assembly instructions

Heatit Controls
Heatit Controls 45 126 48 quick guide

Campingaz
Campingaz 203 164 Instruction leaflet

Bender
Bender W.S Series manual

Zenith
Zenith 122C-A - Heath - Lock-Nut Transformer installation instructions

Torus Power
Torus Power WM AVR-2 ELITE Series owner's manual

Data Video
Data Video DAC-80 quick start guide

Green Brook
Green Brook DAT01A instructions

Power Innovations
Power Innovations Chameleon Installation and operation manual

Gossen MetraWatt
Gossen MetraWatt WZ12A operating instructions

Altronix
Altronix T1618300 installation instructions