Edaphic AZ8215 User manual

OPERATION MANUAL
Spsi and 30psi
Manometer
CE
Edaphic
Scientific
Pty
Ltd
www.edaphic.com.au
1300 430 928

Congratulations on your purchase of
the Manometer ! This instrument is
portable, battery operated pressure
measuring device.
The Manometer is ideal for HVAC/R
technicians measuring pressure
level ,Medical equipment ,Computer
peripherals , Pneumatic Controls.
The meter will display all LCD
segments when it is first turned on
for approx. 3 seconds. Though you
might see DATALOGGER, Y/M/D,
REL, AVG ..these are not available
for the meter but for the meters
with datalogger features named
8205D/8215D/8230D/82100D.
Please contact the store or the
place you purchased.
The LCD is divided into two distinct
sections : One large (Primary) top
screen and one smaller right-
bottom screens (Relative Clock).
The 2 display areas keep you
constantly updated of the pressure
measurements.
The Meter measures Guage
pressure-a measure of pressure in
psi that is reference to ambient
pressure and Differential pressure
-a measure of the difference between
two pressures .
1
INTRODUCTION

11 pressure units are selectable
for Imperial and Metric in the
2
different area .: bar ,mmHg, ozin ,
2
kgcm , psi, inH2O, kPa, ftH2O,inHg
,cmH2O,mbar.
4 models Manometer for your choice:
Pressure range :0~ 5 psi ( #8205)
" :0~ 15 psi (#8215)
" :0~ 30 psi (#8230)
" :0~ 100psi(#82100)
Please check the tubing is not
leakage or damaged before using.
CONVERSION & RESOLUTION
2
Quick Conversion Sheet
AZ8205
Psi inH2O
mbar 2
kg/cm
5
138
345 0.35
unit
model

4
3
10
12
1
5
9
8
7
6
11
15
14
2
13
HOL D REL DIF
DATALOGGER
REC AVG
MAX MIN BAT
AZ8215
AZ8230
AZ82100
15
30
100
415
830
2768
1034
2068
6895
1.05
2.10
7.03
3
CONTROLS AND INDICATORS
1.
2. "-" . Minus pressure display.
3. MAX MIN presssure recorded.
4. REC starts recording mode and
displays max./min. pressure
recorded.
5. AVG. Average records
6.
7.
8.
9. . Pressure unit indication.
10.
11. DIF. Differential pressure mode.
Primary Data Screen displays
pressure value.
(N/A).
DC Power input jack.
RS232 output port.
H/M/S 88:88:88 displays data for
Hour / Minute / Second.
BAT. Low battery indicator.

"-" Negative
pressure
"+" Positive
pressure
4
12. "+" Positive pressure hose plug.
13. "-" Negative pressure hose plug.
14. HOLD. Freezes pressure reading.
15. REL. Establish a relative zero
for the primary screen
information.(N/A)
This instrument shuts off automatically
in approx. 20 minutes of inactivity.
For recording or operating over longer
periods of time, you can disable the
sleep mode by pressing and
simultaneously before power on.
An "n" will appear in the middle of the
screen at which time you can release
the button. (See Fig. A) The
disabled sleep mode will be invalid after
power off.
Fig. A
I
HOLD
AUTO POWER OFF (SLEEP
FUNCTION)
HOLD
MODE OPTIONS
There are six function keys for mode
selection and measuring operation.

HOLDHOLD DIFDIF
RECREC
UNITUNIT
ll
++
==
==
==
DISABLE SLEEP MODE
ZERO SETTING
SELECT UNIT / SKIP CAL.
HOLD
HOLD
Hold 2 sec.
Turns instrument on (Default set-
ting) and off.
I
REC
Press momentarily and relative
clock starts in the lower right
screen.
REC is displayed in the middle
left of (Fig. B) other button
functions are locked out except
Power and Backlight.
Press momentarily again and the
unit cycles through MAX (Fig. C)
and MIN (Fig.D) and back to
current pressure ; the record
mode is displayed on the LCD.
Press and hold for 3
seconds to turn off the record
function to the normal mode.
5
For your
convenience, the
meter defaults the
setting used in
the last operation.
The following lists
the operation for
each function key.
REC
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. B

HOLD
Press momentarily to freeze the
pressure reading . (Fig.E)
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. C
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. D
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. E
6
Press momentarily ,DIF appears
on top of the LCD and the display
indicates the relative zero (
Relative zero causes the value of
the display to show as "0.0")-only
the amount of pressure change
will be indicated. Press
momentarily again and the unit
returns to the normal mode of
pressure differential (see Fig.F).
DIFDIF
DIF
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. F

Press momentarily and the unit
will cycle through "bar" ,
"mmHg","ozin2","kgcm2" ,"
psi","inH2O","kPa","ftH2O" ,"
inHg" ,"cmH2O" ,"mbar" ,which
indicated on the bottom of the
display (See Fig.G &H).
UNIT
Differential Pressure : A measure of
the difference between two pressures
, i.e. use differential pressure sensor
to measure guage pressure by
leaving one process connection open
to atmosphere and connecting the
second sensor port to your system.
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. G
7
Press momentarily and the back-
light illuminates for approx. 30
seconds then turns off
automatically.
Or press momentarily to
decrease the figure when
calibration is being performed.
2
bar ozin psi kPa inHg mbar
2
mmHg kgcm inH2O ftH2O cmH2O
Fig. H

MAINTENANCE
The meter is calibrated in house
before shipping.
To maintain the meter in the good
condition for use , recommend to
calibrate the meter after long time
using.
When properly maintained , the
meter will maintain an accuracy
specification, to ensure your meter
is performing at its peak, send it to
the factory or a qualified instrument
calibration facility for annual
calibration.
Recommend always to set zero
before measurement .Refer to the
zero setting procedure in page 10.
Cleaning:
Ue a damp cloth and mild soap to
clean the case of the Manometer, do
not use harsh detergents or abrasives
as these may mar the finish or
damage the unit's case with an
adverse chemical reaction.
8
CALIBRATION MODE
Calibration mode is only applicable
for a standrad Manometer calibrator
or any qualified meter calibration
facility for annual calibration.

Fig. I
psi
4. The meter has defaulted as 80 psi
calibration point , the adjustable
pressure range is from 78.0 to 82.0
. if calibration pressure source is
not 80psi , to increase the figure by
pressing key , or decrease the
figure by pressing key to set
calibration point as required.
5. Save the calibration point by
pressing key , "SA" and small
"CA"appears on the display (See
Fig. J ) in 2 seconds , the meter
auto-skip to the negative pressure(
-) point for next calibration mode.
RECREC
RECREC
9
1. First , please manually set the
display to zero ( no pressure
applied to the connector ), refer to
the Manual zero procedure.
2. Turn the meter off.
3. Press & simultaneously
,"CA" appears on the display, (See
Fig.I )the meter enters to the
calibration mode ,make sure the
pressure unit to be pointed under
the arrow mark " " is "psi" to
start positive (+) pressure
calibration
DIFDIF

6. Follow the same procedure as step
4 for the negative pressure
calibration point by pressing
key , the LCD now displays" -80.0 "
and small"CA" (See Fig.K ) , do the
necessary calibration figure refer to
your pressure standard if needed.
7. Again save the calibration point by
pressing key, "SA" and "CA"
appears in 2 seconds and then "
End"and "CA" appears in another
2 seconds, the meter turns back to
the normal mode (See Fig. L ).
UNITUNIT
RECREC
RECREC
psi
Fig. J
psi
Fig. K
psi
Fig. L
10
If you can't save by press key, i.e.
no "SA" appeared , please check : (a)
The cabliration pressure source is
between 75.0 and 85.0, or check (b) if
you enter the right positive pressure
(+) or negative pressure (-) .

11
UNITUNIT
If you want to skip positive (+)
calibration when entered to the
Calibration mode , press to skip
to negative (-) calibration point.
Above calibration is an example for
model 82100 , i.e. the pressure range
is from 0 to +100psi (Positive
pressure ) or from 0 to -100 psi (
Negative pressure).
Calibration point reference
Model:
psi
range
Calibration
point
Reommend
4.000 3.900~4.100
12.00 11.70 ~12.30
24.00 23.40~24.60
AZ8205
AZ8215
AZ8230
AZ82100 80.00 78.00~82.00
0~ 5
0~ 15
0~ 30
0~ 100
( )( )
HOLDHOLD
MANUAL ZERO SETTING
When you set the display to zero( no
pressure applied to the connector), press
button for 2 seconds , now the
meter display "0 .0.0.0" from right to left
(See Fig.M) and then disappear each "0"
from left back to right , the LCD display
shows a normal mode .(See Fig.N)
Fig. M

12
2
bar oz in psi kP a inH g mba r
2
mmHg kgcm inH 2O ftH 2O cmH2O
Fig. N
TROUBLESHOOTING
?Power on but no display. Check
the battery are in place and making
good contact or correct polarity ,
replace a new battery or attach
optional AC adaptor for the weak
battery caused.
BAT indication. Replace with a
new battery when LCD display BAT
at the middle bottom.
?
No Display. Make sure battery is
not empty , if the display disappear,
check sleep mode is active . Refer
to the Disable sleep mode function
for a long time using the measure
ment. Or check the tubing is
connected to the meter tightly.
?
?Err.1. For the pressure value
exceed the maximum range , "Err.1
" appears on the display ( See Fig.
O ) ,please change the sensor ,
otherwise, the sensor will be hurt
for going on the overrange
measurement.
Fig.O

?Err.2. For the measurement
pressure is less than minimum
range ,"Err. 2" will appear (See
Fig.P ),recommend to change the
sensor (meter).
Fig. P
13
?
Err.4. When you set zero ,make
sure you have disconnected the
tubing ,no pressure applied to the
connector. Then if you see an
Err.4 appears on the display , it
means the sensor or meter
damaged (See Fig.R ).Return the
unit to the store you purchased for
repaired .
Err.3.For operating the DIF function
, the differential pressure value is
larger than maximum display digit ,
Err.3 appears on the display (See
Fig.Q ).
?
Fig. Q
P.S. Err.4 will be also appeared while
the tube or hose is connecting during
setting zero mode.
Fig.R

?
Replace your 9-volt battery when:
The BAT icon appears on the right
of the screen.
The meter will not power on.
Use of the back-light causes the
BAT icon to appear.
14
E1OL or E2UL. When you see the
errors while operating Rs232
software, it means pressure source
is less or over than the range of the
instrument.
REPLACING THE BATTERY
Remove battery from instruments that
you do not plan to use for a month or
more.
Do not leave battery in instrument.
1. Remove the tubing of the
instrument.
2. Lay the instrument face-down on a
clean, flat surface.
3. Remove the battery by screw driver
and observe indicated polarity and
close the cover after replacing with
a new battery.
Even if the battery was recently
replaced, check its voltage level if you
get no response from your instrument.
To replace the battery:

OPERATING CONDITIONS
Compensated temperature range:
o
0~50 C.
o o
Operating temperature 0 ~50 C (
o
32~122 F)
Storage temperature range:
o
-20~60 C
Operating Humidity Max. 80% RH
Power : One 9.0 volt battery
Exceeding Maximum pressure will
cause permanent sensor damage.
Model Max.Pressure
AZ8205
AZ8215
AZ8230
AZ82100
20 psi
30 psi
60 psi
150 psi
15
SPECIFICATION
Pressure
Range
Resolution
Accuracy
Dimension
Unit Weight
Response
time
Format
0~ (5 or 15 or 30 or 100 psi)
See Page 8 data sheet
o
0.3% of full scale at 25 C
72 x 182 x 30 mm (meter)
Approx. 220 gram (with
battery)
0.5 seconds
Baud Rate : 2400 bit/sec
Data Bit : 8 , Stop Bit : 1
P XXXXX , P - XXXXX (unit)
The meter Pressure measurement
instruments are not suitable for the
absolute pressure measurement.

OPTIONAL ACCESSORY:
RS232 software disk or CD with D-
sub connector
DC Adaptor
The meter x 1
Battery x 1 (9.0 volt)
Operation manual
Hard carrying case
Connection hose 4mm(ID)x6mm
(OD) x 500mm length x 2 pcs
MATERIAL SUPPLIED
This package contains:
The meter instruments are fitted with
two 4.8mm lugs. Before you connect
the instruments to a pressure sources
, check carefully the security of all
fitting.
16
RS232 OUTPUT:
The meter can link with personal
computer to capture on-line datas ,
display pressure records with real-
time output, you can retrieve file ,
save the datas for operating data
analysis, records statistic ,multi-files
display in the screen,....versatile
functions for your choice.
Connection procedures:
1.Plug the optional accessory RS232
cable onto the DC jack port ( at the
right side of the meter)

2.Instert the D-sub 9P type connector
onto computer's Com.1 or 2 port or...
3.Start to set up RS232 software by
inserting the CD-ROM or Floppy
diskette.
4. When installing the RS232 software
,please follow the operation manual
procedure in the software package.
17