Selecting Output Options >OPT<
1) Using the arrow keys, scroll to OPT (Output Options) then press ENTER.
2) The meter will prompt you for the Physical Quantities for your desired output. Use the arrow keys
to scroll up and down through the available list (Volume Flow is the most common). Press ENTER.
3) Using the arrow keys, choose your Volume In units from the available list then press ENTER.
4) Choose a Damping (averaging) time in seconds (if unknown leave as default 10s) then press ENTER
5) The meter is fitted with an internal data logger for logging time stamped data. Answer NO to Store
Meas. Data (if you wish to setup the data logger please consult the user manual) then press ENTER.
6) The meter is fitted with a Serial Output Interface. Answer NO to Serial Output (if you wish to setup
the serial interface for instantaneous output please consult the user manual) then press ENTER.
7) The meter is fitted with a 4 to 20 mA output, answer NO to Current Loop (if you wish to setup the 4
to 20mA output please consult the user manual) then press ENTER.
8) The meter is fitted with an Alarm Output, answer NO (if you wish to setup the Alarm Output please
consult the user manual) then press ENTER and you will return to the main menu.
Enter the Measure Mode >MEA<
1) Using the arrow keys, scroll back to MEA (Measuring) then press ENTER.
2) If you have setup the Store Meas. Data function then at Meas. Point No. select a file name for the
storage record then press ENTER.
3) When Sound Path appears, the number indicated below this prompt represents the number of
Sound Paths (please consult the user manual definition). This value may be manually adjusted by the
User. Choose a Sound Path that matches your needs, and then press ENTER.
4) The Transd. Distance will indicate how far apart the transducers must be separated.
5) Select an appropriate transducer Installation Site by following the guidelines in the selecting
transducer mounting location document.
6) Apply Ultrasonic Couplant to the bottom of the Transducers
7) Install Transducers with arrow pointing in the same direction and in the direction of flow.
8) Ensure Transducers are correct distance apart by pressing ENTER and reviewing the SIGNAL
STRENGTH as discussed in the user manual. Then press ENTER.
9) At Transd. Distance press ENTER.
10) You are now displaying the flow measurements!!!
Once measurements are complete, press and hold the PWR button to shut the instrument off.