Electronics Technology ES2010E User manual

-0-

-1-
CONTENT
ISafetyPrecautions and Procedures ………………………………………2
II Introduction …………………………………………………………………3
III.Basic Function Introduction ……………………………………………4
IV ElectricalSymbols …………………………………………………………4
VTechnical Specifications……………………………………………………4
1 Basic WorkingCondition……………………………………………………4
2Range and Accuracy Error……………………………………………………5
3 General Specification………………………………………………………5
VI Structure………………………………………………………………………7
VII Operation………………………………………………………………………7
1Switchon/off …………………………………………………………………7
2Data Hold/Delection…………………………………………………………7
3Backlight Control ……………………………………………………………8
4Phase Measurement………………………………………………………………8
5AC Current、Leakage Current Measurement…………………………………9
6AC Voltage Measurement………………………………………………………10
7Perceptual andCapacitive Circuit Discrimination……………………10
8Frequency Measurement………………………………………………………10
VIII Various Test Wiring Reference…………………………………………11
IX Battery Replacement…………………………………………………… 14
XOther Explanations and Considerations……………………………… 13
XI Accessories………………………………………………………………… 14

-2-
ISafetyPrecautionsAnd Procedures
Thankyouforpurchasingour digitaldoubleclampphase voltammeter.Before
youusetheinstrumentforthefirsttime,inordertoavoidpossibleelectric
shockor personal injury, please be sure to: read and strictlyobserve the safety
rules and precautions listed in this manual.
uThisinstrumentisaliveworkingequipment.Foryoursafety, pleasecomply
withtherelevantnationalsafetyregulationsandstrictlyfollowtheelectric
safety work procedures.
uPayattentiontothetextandsymbolsoftheinstrumentpanelandbackplate.
uBeforeuse,makesurethattheinstrumentandaccessoriesare ingoodcondition
and thattheyare not damaged,bare orbroken.Donotuseitwhenthe instrument
backcoverandbatterycover arenotcoveredproperly.Whenthe instrumentis
inuseandtheenclosure ortest wireisbroken andthemetalisexposed,please
stopusingit.
uCannot be used to test voltages higher than 600V. Do not use under strong
electromagnetic environment to avoidaffectingthe normal operation ofthe
instrument. Do nottest in flammable and hazardousareas.
uMakesure the connecting plugoft
hewireis tightlyinserted intotheconnector.
Please pay attentionto the direction whentestingthephase. Thetest line must
beevacuatedfromthemeterbefore beingpulledoutofthemeter.Donot touch
the input jack toavoid electric shock.
uDonotoperate the instrumentwith wet handsor exposingit to rain.Donot
use theinstrumentwhen it is wet.
uDonotplaceandstoretheinstrumentforlongperiodsoftimeunderhigh
temperature and humidity, condensation, and directsunlight.
uKeepthesurfaceoftheproductcleananddry.Donotcleanitwithcorrosives
orcrudematerials.Useasoftcloth(suchaseyeglasscloth),dipalubricant
thatpreventsrustand dehumidification, andgentlywipe the currentclamp.
uDonothitor drop the instrumenttoavoid impact on the instrumentand the
currentclamp, damagethe instrument, especially theJointsurface ofjaw.

-3-
uPay attention to the measuring range and use environment specified by this
instrument.
uIt is forbidden to use, disassemble, calibrate and repair thisinstrument
withoutpermission. It mustbeoperated by authorized personnel.
uThe""inthe manualisasafetywarningsign,andtheusermustoperate
safely in accordance with the contentsofthis manual.
uThe""andotherdangeroussignsinthemanualmustbefollowedbythe
userinstrict accordance withthecontents of thismanual.
II Introduction
Digitaldoubleclamp phase voltammeter isthe latestproductofourcompany's
clamp-typemeterseries.Itisa hand-held, dual-channel input all-digital,
multi-function, high-precision measurement instrument.The instrument uses the
latest microprocessortechnology and digitalsignalprocessing technology to
achievepowerfrequency measurement(suchasvoltage, currentRMS, power
frequency, phase relationship, etc.), Determine the transformer wiring group,
inductive,capacitivecircuit,testthe secondarycircuitandbusdifferential
protectionsystem,read outthe phase relationshipbetweenthe differential
protectionCTgroups,checkthepower meter wiring iscorrector not,themeter
usestheclamp-typecurrenttransformerconversionmethod to inputthemeasured
current,sothereisnoneedtodisconnect thecircuit under test. Provideasafe,
accurate, andconvenientnew typeof power meter forelectricityinspection
personnel.
Digitaldouble clamp phase voltammeter is suitableforelectricpower,
petrochemical,metallurgy, railway, industrial and mining enterprises,
scientificresearchinstitutions,andmeasurementdepartments.Especially
suitable for energy billingsystemsand relay protection systems.

-4-
III.BasicFunction Introduction
1. Largescreenhighbacklightdisplay,cleardisplay oftheinstrument'sworking
status andtest parameters,the operationisvery convenient.
2. Measuretwo wire’svoltage, current,phase, and measure grid frequency.
3. Low-current5mAmeasurementphaseis very suitablefornewusersto checkthe
connection statusafterno-load operation, the instrument has highmeasurement
accuracy and widemeasurement range.
4. Automatic identification of transformer windings, capacitive and inductive
loads.
IV Electrical Symbols
Extremely dangerous! The operator must strictly abideby thesafety
rules, otherwise therewill bedanger ofelectric shock and cause
personal injury orcasualties.
Danger! Theoperator muststrictlyabideby the safety rules,
o
therwise there will be dangerofelectricshockandcausepersonal
injury or casualties.
Caveat!The safetyrules mustbestrictlyobserved,otherwise
personal injury orequipment damagemay result.
AC
DC
Double insulation
VTechnicalSpecifications
1BasicWorking Condition
Influencing factors
Baseline
conditions Working
conditions Remark
Ambient
temperature 23℃1℃-10℃~40℃
Ambienthumidity 40%~60% <80%
Signal waveform Sine wave Sine wave β=0.05
Signal frequency 50Hz 1Hz 45Hz~65Hz

-5-
Instrument working
voltage 9V 0.1V 9V 1V
Voltageamplitude
when measuring
phase 220V 20V 30V~500V
Currentamplitude
when measuring
phase 2A 0.2A 10mA~20.00A
External electric
field, magnetic
field Should avoid
Wireposition tobe
measured The wire under testisintheapproximategeometriccenterof
the jaw
2Rangeand AccuracyError
classification
Range Resolution intrinsic error
Voltage AC 0.00V~600V 0.01V ±(1.5%rdg+3dgt)
Current AC 0.0mA~20.0A
0.1mA ±(1.5%rdg+3dgt)
Phase 0.0°~360°0.1°±1°
Frequency 45Hz~65Hz 0.01Hz ±0.5Hz
Remark:The phaseerror is ±3°under workingconditions.
3.GeneralSpecification
Function
Measure twoAC voltages, currents, phasebetween voltages,
phasebetween currents,phasesbetweenvoltagesandcurrents,
frequencies, de
terminetransformerwiring groups,inductive,
capacitivecircuits,testsecondaryloopsand busbar
protectionsystems,readoutthephaserelationship between
the CTsofthe differentialprotectiongroups,check the
wiring of thepower meter, check the lineequipment,etc.
Power DC9VAlkalinedrybattery
Power
consumption
Turnonthebacklightandconsumeupto30mA.The battery can
workcontinuouslyformore than 12hours.
Displaymode LCD display,60×33mm
Instrument
size L/W/T:192*92.5*36mm
Voltagerange
AC 0.00V~600V
Currentrange
AC 0.0mA~20.0A

-6-
Phase range 0.0°~360.0°
Frequency
range 45.00Hz~65.00Hz
Resolution
Voltage:AC 0.01V
Current:AC 0.1mA
Phase:0.1°
Frequency:0.01Hz
Detection rate
About 2seconds/time
Data hold Press HOLD keytoholddataduringtest,"H"symbol display
Automatic
shut-down
About15minutes afterpoweron,themeter willautomatically
shutdown to reduce batteryconsumption
Backlight
function Suitable for dimplaces andnightuse
Voltage
detection
When the batteryvoltage islowerthan 7V, the low battery
voltagewill display
Instrument
weight
Hose:362g(including battery)
Tip-shapedcurrentclamp:180g×2
Testline:190g
Test line
length 1.5m
Line length of
the current
clamp 2m
Working
temperature
and humidity -10℃~40℃;below80%Rh
Storage
temperature
and humidity -10℃~60℃;below70%Rh
input
impedance Testvoltage inputimpedance is:1MΩ
Pressure
resistance
The withstandvoltageof1000V/50Hz sinewaveACvoltage
betweentheinstrumentcircuit and thehousinglasts 1minute
insulation Betweeninstrumentline andhousing≥100MΩ
structure Double insulation
Suitable for
safety
regulations
IEC61010-1CAT Ⅲ600V,IEC61010-031,IEC61326,
Pollution
level 2

-7-
VI Structure
1.Insulation anti-vibrationsheath2.LCD display3.Clampjaw
4.Currentclamp 5.Boot-strapPOWRRkey 6.Hz key
7.DataholdHOLD key/backlight8.Functional rotary switch
9.Current clamp inputhole (2wires) 10.Voltageinputhole(2wires)
11.Currentclamplead12. Test wire(Red and blackeach 2)
VII Operation
Beforeuse,carefullycheck allpartsoftheinstrumentfordamage,
and use themwithoutanydamage.
Itisforbiddentouse thisinstrumentinhazardousareas
Followthemanualinstructionsto installthebattery.
1Switchon/off
PressthePOWER button to turn on, the LCDdisplay. Press the POWER button
to shutdown,themeter will automatically shutdownafterabout15minutes.
2Data Hold/Delection
Pressthe HOLD button to keepthedisplay dataor cancelthe hold。

-8-
3Backlight Control
Afterpoweron,longpressthe HOLD buttontocontrolthebacklight,suitable
for dimplaces.
Danger:Electricity! Mustoperatedby thequalifiedpersonnelthat
trained andauthorized .The operator must strictly abidebythe
safetyrules,otherwisethereis
adangerofelectricshock,
resulting inpersonal injury or equipmentdamage.
Danger!Cannot beused tomeasure voltagelinesexceeding 600V,
otherwise there is adangerof electric shock, resultingin personal
injury or equipment damage.
Danger!Cann
ot beused tomeasure thelinesmorethan20A. Otherwise,
there isadanger of electric shockthat maycause personal injury
or equipmentdamage.
Before testing,you mustfirstswitchthe functionrotaryswitchtothe
correspondingfunction indicatingposition,thenconnectthe testline to
the instrument,andthenconnectthetestlinetothetestedcircuitfor
testing.
Muststrictlyfollowthe manualinstructionstoconnect
Afterthetest iscompleted,thetestlinemustbeevacuated fromtheline
beforeremoved fromthemeter.
4PhaseMeasurement
The phasetestrelationshipofthisinstrument:U1U2,I1I2,U1I2,I1U2,
the phaseofeachrelation measuredis the phaseof1signal advanced 2
signals.
The U1red voltagejackandthe U2redvoltage jackandthe startofthe
arrowsymbol are the same.
The directionofthecurrentinputclampin phasetestingisthesameas
the directionofthearrowsign ontheclamp.
For 3phaselinephaserelationship,pairwisecorrespondencecanbe
tested.
(1).Phase testingofU1U2
RotatetherotaryswitchtoU1U2position,connectthered and black4test
leadstotheU1andU2oftheinstrument,voltage inputjacks, andthenconnectthe
test leadstoU1andU2lines.Thetest displayvalue isthephasebetween the two
voltage lines, thatisU2isthephaseangle ofU1.When testingtheU1U2phase,the

-9-
twocircuitinputcircuitsarecompletelyisolatedandinsulatedtoavoid shortcircuit
and burnoutof the instrument.
(2).Phase testingofI1I2
Rotatethe rotaryswitchtothe I1I2position.Connectthe twocurrentclampsto
theI1andI2currentinputjacks onthefrontoftheinstrument. Thenclampthe
currentclamptotheI1andI2lines. The test displayvalueisthe phasebetween the
twocurrents.
(3).Phase testingofU1I2
RotatetherotaryswitchtotheU1I2position.Connect theredand black test
leadstotheU1voltage inputjackofthemeter.Connect one currentclamptotheI2
currentinput jack onthefront of theinstrument,andthen connect thetestleadsand
currentclampstoU1,I2line,testdisplayvalue isthe phasebetween voltage and
current.
(4).Phase testingofI1U2
RotatetherotaryswitchtotheI1U2position.Connectonecurrentclamptothe
I1currentinputjack onthefrontoftheinstrument, andconnect theredandblack
test wirestotheU2voltageinputjack ofthe instrument.Then connectthecurrent
clampandthetestwire totheI1.,U2lines, testdisplayvalue isthephasebetween
currentand voltage.
5ACCurrent、LeakageCurrentMeasurement
Rotatetherotaryswitchtothe20AofI1.AcurrentclampisconnectedtotheI1
currentinputjackonthefrontoftheinstrument. Thenuseacurrentclamptoclamp
thecircuitundertest. The testdisplayvalue isthecurrentorleakagecurrentinthe
circuitundertest. If thecurrenttobetestedisrelativelysmall,asmallergearcan be
selectedandtestedtoimprovetheaccuracyofthetest.Itisalsopossibletorotate

-10-
the rotaryswitchtothe appropriatelimitofI2and useI2totestthe currentor
leakagecurrent.Notethatthepositionsofthe I1andI2oftherotaryswitchmust
correspondtotheinputjacks ofthecurrentclampsI1andI2.
6ACVoltageMeasurement
RotatetherotaryswitchtoU1's 600Vrange.Theredandblacktestleadsare
connectedtothemeter'sU1voltageinputjack.Then connectthe testlinetothe
testedcircuit,andthe testdisplayvalue isthe voltage ofthe tested circuit.Ifthe
voltage tobetestedisrelativelysmall,asmallergearcanbe selected andtestedto
improvethe accuracyofthetest.Itisalsopossibletorotatetherotaryswitchtothe
appropriatelimitof U2andtestthevoltage withU2.NotethatthepositionsoftheU1
and U2oftherotaryswitchmust correspondtotheinputjacksofthevoltagesU1
and U2.
7PerceptualandCapacitiveCircuitDiscrimination
RotatetherotaryswitchtoU112,inputcircuitvoltage toU1jack, circuitcurrent
inputtoI2jack.Ifthe phaseisdisplayed in the range of00 to900,the measured
load isinductive.Ifthephaseisdisplayed in therangeof2700to3600,the
measuredloadis capacitive.
8.FrequencyMeasurement
Rotatetherotaryswitchtothecurrentandvoltage levels ofU1orU2andpress
Hztomeasurethe currentfrequency.

-11-
VIII VariousTestWiring Reference:

-12-
IX BatteryReplacement
Pleasepayattentiontothe polarityofthebattery,itmustbeinstalled
withthe correctpolarity, otherwisetheinstrumentwill be damaged.
Itis forbiddentoreplacethebatteryin hazardouslocations
Mustuseaqualified9Valkaline drybattery
1.Whentheinstrument power supplyvoltage islower than7V,the symbol
“”isdisplayed, indicatingthatthebatterypower isinsufficient.Please
replacethe batteryin time.

-13-
XOther Explanationsand Considerations
1Currentclampspecificity
The twocurrentclampsofthemeterarededicated tothismeter and cannotbe
used for another meter.Currentclamptopreventtheimpact,the jawplane must
be kept clean,completelyclosedtest is reliable.
2. Currentclamp maintenance
Afterthecurrentclampisused,thedustintheplaneofthejawsshouldbe
promptly removed.The planeofthejawsshouldnotbecleaned withroughobjects
or corrosion.Itisbesttouseasoftcloth(suchas:WD-40 lubrication) togently
wipe.
3.Beforethemeasurement,itshouldbepreheatedfor3to5minutestoensurethe
measurement accuracy。
4Thisinstrumentisusedforthedetectionofsecondaryloopsandlowvoltage
loops.Itcannotbeusedtomeasurethecurrentinhighvoltage linestoprevent
electric shock.
5Three-phasefour-wire (thephasewhen three-phaseload is balanced):
Phase
relationship
Phase
value
Phase
relationship
Phase
value
Ua-Ub 120°Ia-Ib 120°
Ub-Uc 120°Ib-Ic 120°
Uc-Ua 120°Ic-Ia 120°
6Three-phasethree-wire (the phase whenthree-phase loadis balanced):
Phase
relationship
Phase
value
Phase
relationship
Phase
value
Uab-Ucb 300°Ia-Ic 240°
Uab-Ia 30°Ucb-Ic 330°
Ifthecurrentclampisreversed orthecurrentlineisreversed,the
displayed phasevalue will be 180°outofphase,whichisan increase
of 180°fromthe abovestandard value.

-14-
XI Accessories
Host 1PC
Instrumentbox 1PC
Currentclamp 2PCS
Testline 4(2eachforred andblack)
Battery 9VAlkalinebattery1PC
Manual,certificate 1SET

-15-
The contentsofthisusermanualcannot be used asareason tousethe product for specialpurposes.
The companyisnot responsiblefor other losses caused byuse.
The companyreserves the righttomodifythe contentsofthe usermanual.Ifthereisanychange,itwill notbe
notified.