Elmos AP11-37 Guide

Panel Meter AP 11
Technical documentation U-19
www.apoelmos.cz
ISO 9001
AP11 - 37 - Counter

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
january 2015, TD-U-19-25

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
1 Introduction .............................................................................................. 5
1.1 Ordering code ........................................................................................ 6
2 Technical data ........................................................................................... 7
3 Panel meter description ............................................................................. 9
3.1 Front panel ........................................................................................... 9
3.2 Dimensions of panel meter and assembly opening ....................................... 10
4 Connection ................................................................................................ 11
4.1 Description of back panel ........................................................................ 11
4.2 Instruction for making installations and connections to panel ...................... 11
4.2.1 Reducing of interference influence ....................................................................... 11
4.3 Connection to supply voltage. ................................................................... 12
4.3.1 Connection of power supply conductors in terminal board ....................................... 12
4.4 Connection of input signals ..................................................................... 13
4.4.1 Current signal 4 - 20 mA ..................................................................................... 14
4.4.2 Current signal 0/4 - 20 mA .................................................................................. 14
4.4.3 Auxiliary digital inputs ........................................................................................ 14
4.4.4 Pulse signal input (max. 10 kHz) .......................................................................... 15
5 Analogue output ......................................................................................... 16
5.1 Connection of analogue output ................................................................. 16
5.2 Block diagram of analogue output function ................................................. 17
6 Communication .......................................................................................... 18
7 Connection of contact outputs .................................................................... 19
8 Panel meter functions ................................................................................. 20
8.1 Block diagram of counter ....................................................................... 20
8.2 Panel meter switch on ............................................................................. 20
8.3 Panel meter reset ................................................................................... 20
8.4 Display setting ....................................................................................... 21
8.5 Bargraph ............................................................................................... 22
8.6 Functions of limit switch ......................................................................... 23
8.6.1 Functional characteristics of limit switches ............................................................ 24
9 Panel meter menu functions ....................................................................... 26
9.1 Functions of buttons in setting mode ......................................................... 26
9.2 Working mode ........................................................................................ 26
CONTENTS

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
9.3 Configuration mode ................................................................................ 27
9.3.1 Simplified block diagram of counter menu ........................................................... 27
9.3.2 Configuracetion of pulse input in menu CONF ....................................................... 28
9.3.3 Configuration of limit switches ALA-1 up to ALA-4 ................................................ 29
9.3.4 Configuration of current sensor SENS .................................................................. 30
9.3.5 Configuration of analogue output - DACO ............................................................. 31
9.3.6 Configuration of parameters OSTAT ..................................................................... 32
9.3.7 Configuration bargraph - BARGF ......................................................................... 33
9.3.8 Configuration of diplay - DIS-1, DIS-2 ................................................................. 34
9.3.9 Reset of sum CL-S1, CL-S2 ................................................................................. 35
9.3.10 Reseting thre measured value CL-IN .................................................................... 35
9.4 Limit values of parameters ........................................................................ 36
10 Control software ...................................................................................... 37
11 ES Declaration of conformity .................................................................... 38
12 Certificate of quality and completness ....................................................... 39
13 Guarantee conditions ............................................................................... 39
CONTENTS

5
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
AP11-37 Counter is 5 digit programmable panel meter for measuring in time dependent current signal.De-
vice is controlled with digital signal processor with A/D transducer.Counter is operated with key-board pla-
ced on front panel or with control software that is used for setting of all parameters and archiving measu-
red values. Panel meter has tobe provided with communication line, variants of communication lines are in
ordering code. Control software is included in standard delivery. Counter is equipped with three-colour dis-
play allowing for a quick check of limits within measured value oscillates. Visually interesting element is au-
xiliary horizontal bargraph. AP11 Counter inputs are assembled for unified outputs from sensors of techno-
logical processes
(0/4-20 mA, 0/2-10 V).There is an variant with analogue output that can be galvanic separated in panel me-
ter ordering code.With multi-stage types of panel meters analogue output can be assigned to any input by a
program, eventually to sum. Any of the communication line variants can be used for communication of panel
meter with PC.Communication lines RS232 or RS485 (can be galvanic separated) are in ordering code. Com-
bination of two communication lines RS232 and RS485 can be used for above standard applications. Com-
munication enables not only setting of parameters and data archiving, but regarding the possibility of ad-
dressing of individual panel meters it can be also used for interconnection of bigger number of panel me-
ters and controlling of complete technological lines.
1
INTRODUCTION

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
6
This technical documentation refers to the following chart of ordering codes
fig.1
1.1 Ordering code
1INTRODUCTION
AP 11 - XX - X - X - X - X - X - XXX
Input
37 1 x current signal input 0/4 - 20 mA
1 x pulse signal input
Contact output
0empty
12 x relay (switching contacts 250 VAC, 2A)
24 x relay (switching contacts 250 VAC, 2A
Analogue output
0empty
1current/voltage without galvanic separ
2current/voltage with galvanic separ. GS
Communication
0empty
1RS232
3RS485 with GS
6RS485 with GS + RS232
Power supply
180 - 253 VAC
218 - 36 V AC/DC
Display
4three colour
Software
001 Counter
xxx special requirement
AP 11 - 37 - 2 - 2 - 3 - 1 - 4 - 001
Example of order code

7
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
2
TECHNICAL DATA
Power supply
Power supply voltage 80 - 253 VAC, 50 Hz
18 - 36 VDC / 18 - 36 VAC, 50 Hz
Consumption max. 12 VA
Display
Display -9999 ~ 0 ~ 99999 (5-digit LED)
Height of digits 14 mm
Decimal point Adjustable with program
Bargraph 30 LED
Resolution According to position of decimal point
Auxiliary power
supply
> 18 VDC @ 25 mA for feeding of sensors
Max. load of auxiliary power supply 40 mA
Outputs
Contact 2x relay (switching contact 250 VAC, 2A or 30 VDC, 2A)
or 4x relay (switching contact 250 VAC, 2A or 30 VDC, 2A)
Analogue 13.5 bit D/A transducer without galvanic separation or with galvanic se-
paration current 0 (4) – 20 mA, loading resistance max. 400 Ω
voltage 0 – 10 V, loading resistance min. 10 kΩ
Communication
RS485 without galvanic separation or with galvanic separation,
two way communication
RS232 without galvanic separation
Input signals, accuracy
Type Input signal meas.range Accuracy of
measurement
(% of sca-
le)
Code
Process 1x current signal (AI) 1x 0/4-20mA ± 0,25% 37
Pulse 1x pulse signal (CI) 0-10KHz 37
Comparator level LOW 3,9 V/2,4 V ± 0,5 V
HIGH 7,5 V/6 V

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
8
2TECHNICAL DATA
Operating conditions
Working conditions 0 - 60 °C
Temperature coefficient 25 ppm/°C
Stabilizing time Within 5 min after activation
Shielding IP 54 (front panel)
IP 20 (terminal board)
Calibration at 25 °C and 40% relative humidity
Data back-up electrically (EEPROM), bat.CR2032SLF (RAM)
Auxiliary input Voltage free contact, digital inputs
Connection
Connector terminal board
Max. section of condu-
ctor
2.5 mm2 for power supply and contact outputs
1 mm2 for other connectors
Safety class I
Electromagnetic compatibility
ČSN EN 61326
Seismic resistibility
ČSN IEC 980: 1993, čl. 6
Electric safety
ČSN EN 61010-1: 2003
Mechanical properties
Type AP11 panel meter
Dimensions 96 x 48 x 119 mm
Opening in panel 90,5 x 43,5 (openings in corners ø 3 mm with pitch 89,5 x 42,5 mm)
Keyboard 4 keys, foil
Weight 400 g

9
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
3
PANEL METER DESCRIPTION
1
2
3
4
1
2
3
4
Menu
UP
DOWn Set
Ap 11
In Out
1
2
3456 7
8
2 - Input „IN“ indicators
No. of of selected input (functions IN-1,IN-2,SUM1,SUM2 )is indicated with respective indicator. Individual
inputs are selected with keys „UP“ and „DOWN“.
1 - Display
Five-digit display for indication of measured value. With parameter programming display show clear reports.
8 - Output „OUT“ indicators
Indicators OUT 1 to OUT 4 indicates state of individual outputs in the following manner: indicator flashing
– output switched, indicator is not flashing – output is off.
7 - Key „SET“
Key „SET“ is used for setting data, programming termination, data uploading into EEPROM and return to
operating mode.
6 - Key „DOWN“
Key „DOWN“ is used for browsing parameters and setting of numeric data when programming. Pushing and
keeping down of the key speeds up browsing and setting procedure.
5 - Key „UP“
Key „UP“ is used for browsing parameters and setting of numeric data when programming. Pushing and kee-
ping down of the key speeds up browsing and setting procedure.
4 - Key „MENU“
Key „MENU“ is used for entering configuration menu and setting mode.
fig. 2
3 - Bargraph
Bargraph relates to measured value that is determined with pre-set range of STR and END.
E.g. if start of range (STR) is 0 and end (END) 200 and measured value shall be on level 100, bargraph shall in-
dicate half of the scale. Bargraphs is formed with column of LED diodes.
3.1
Front panel

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
10
3PANEL METER DESCRIPTION
fig. 3a
Dimensions for power supply 80 - 253 VAC, 50 Hz (g. 3a)
Dimensions for power supply 18 - 36 VDC / 18 - 36 VAC, 50 Hz (g. 3b)
fig. 3b
3.2 DIMENSIONS OF Panel meter AND ASSEMBLY OPENING
Front view
Front view Side view
Side view

11
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
4
CONNECTION
4.2.1
Fix panel meter into panel with two clamps (included in delivery).
Connect conductors into screw connectors on the back panel of regulator. Connectors are designed as
separately detachable constructions blocks as follows:
connectors 11 to 18 – process and pulse inputs
connectors 21 to 28 – analogue output
connectors 31 to 38 - communication
connectors 41 to 46 – relay output
connectors 51 to 56 - relay output
connectors 61 to 63 – power supply
Pull each block with connectors out from panel meter (locking force has to be surpassed) in backward di-
rection.Then connect conductors to released blocks with connectors and then insert blocks back to appa-
ratus.
Max. cross section of conductors on relay connectors and power supply is 2,5 mm2, on other connectors
1 mm2.
41 42 43 44 45 46
51 52 53 54 55 56
61 62 63
11 12 13 14 15 16 17 18
21 22 23 24 25 26 27 28
31 32 33 34 35 36 37 38
Attention
Danger of risk
Watch for power voltage
Inputs Terminals for power supply
fig. 4
Analogue output
DI inputs
Communication
t
4 x Relay output
4.1
Description of back panel
Instructions for installation into panel and connecting 4.2
Following rules should be observed with designing of the system:
a) All power supply conductors and power lines has to be led separately from signal lines
(e.g. thermocouple, communication). Min. gap between these types of lines shouldn´t be less than 30 cm.
b) If signal line crosses power line they should intersect in right angle.
c) Lead the lines out of the potential source of interference.
d) Don‘t install relay and contactors too close to panel meter.
e) Use twisted and screened conductor for signal line.
4.2.1
Reducing of interference inuence

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
12
4CONNECTION
fig. 5
61 62 63
NL1
Alternating feeding volta-
ge 80 - 253 VAC, 50 Hz
Danger: Don‘t connect panel meter to power supply until all inputs are connected. Wrong connection
of panel meter can cause injury!
Panel meter connection
During connecting of panel meter main switch or safety circuit breaker has to be:
- part of building installation
- in the close vicinity of equipment
- easy to reach for operating personnel
- marked as equipment disconnecting element
If the equipment is used in different manner than specified by producer, protection provided with
equipment can be disturbed.
Recommended safety fuse for power supply 230 V je 1 A / 250 VAC
Recommended safety fuse for power supply 24 V je T 3,15 A / 250 V
Caution!
Connection of power supply conductors in terminal board
Feeding voltage
18 - 36 VDC
Feeding voltage
18 - 36 VAC
4.3 Connecting of power supply
4.3.1

13
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
figure.6
4
CONNECTION
Current signal input
Pulse signal
CLR Sum 1 (clear sum 1)
CLR / SET(clear / set)
LOCK
HOLD
CLR Sum 2 (clear sum 2)
COM (GND)
COM (GND)
connected inside
the device
Following diagrams show possibilities of sensors and process signal connections.
Select input signal in function menu SENS (see chapter sensor type setup - SENS)
Variants of input signals
Requirement of right measuring is necessary to set type of signal input in menu SENS du-
ring device configuration.
4.4
Connecting of input signals
Type of input signal Symbol on diplay Schematic
SENS 6,9
DIGITAL INPUT (DI)
CURRENT SIGNAL(AI)
4-20mA
Passive current transducer
4-20
7,8
0/4-20mA
Active current signal
0-20
4-20
Pulse signal (CI) Total/rate 10,11,12

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
14
4CONNECTION
4.4.1
Connect double-conductor transducer to connectors 11, 12, as
shown in schematics where connector no. 11 is power supply.
Press MENU key and enter device configuration menu, brow-
se with keys UP and DOWN to the function of sensor selecti-
on SENS . With repeated press get to submenu for selection of
sensor type TYP. Press the menu key again and enter submenu,
where with keys UP and DOWN set 4 - 20 mA for 4 - 20 double-
conductor passive transducer and press key SET to confirm the
selection. Then set range of measurement.Start of range is de-
finable in submenu SENS with function STRS, and ending of the
range with function ENDS (SET key confirms the menu icons set-
ting). Leaving the device config menu by pressing SET key will
save all settings to device EEPROM.
Current signal 4 - 20 mA (passive double-conductor transducer)
fig.7
fig.8
Connect active current signal 0/4 - 20 mA to connectors 12,
13, as shown in figure 8 where connector no. 13 is minus pole.
Press MENU key and enter device configuration menu, browse
with keys UP and DOWN to the function of sensor selection SENS
. With repeated press get to submenu for selection of sensor
type TYP. Press the menu key again and enter submenu, where
with keys UP and DOWN set 4 - 20 mA for 4 - 20 double-condu-
ctor passive transducer and press key SET to confirm the selecti-
on.Then set range of measurement. Start of range is definable
in submenu SENS with function STRS, and ending of the range
with function ENDS (SET key confirms the menu icons setting).
Leaving the device config menu by pressing SET key will save all
settings to device EEPROM.
4.4.2
4.4.3
Current signal 0/4 - 20 mA (active current signal)
Auxiliary digital inputs
K1 K2
K1 - contact resets SUM 1
K2 - contact for reseting / seting measured value
more on page 35
fig.9
AI
AI

15
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
4
CONNECTION
Pulse signal input CI (max. 10 kHz) 4.4.4
SENSOR
SENSOR
Connect pulse signal input to connectors 14 and 15 via sche-
matics, where connector 15 is minus and 14 is pulse signal. In
menu item CONF set the function of pulse input to TOTAL / RATE
(counter / frequency-meter), INPUT to NPN and the level of the
input comparator TRIG to LOW.
Example connection of sensor with NPN output level. In menu
item CONF set the function of pulse input to TOTAL / RATE (coun-
ter / frequency-meter), INPUT to NPN and the level of the input
comparator TRIG to LOW.
Example connection of sensor with PNP output level. In menu
item CONF set the function of pulse input to TOTAL / RATE (coun-
ter / frequency-meter), INPUT to PNP and the level of the input
comparator TRIG to LOW.
fig.10
fig. 11
fig. 12
CI
Attention of maximum sensor consumption

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
16
5ANALOGUE OUTPUT
Options of analogue output connections with or without galvanic separation are shown on fig. 13. Type of
output signal is selected in menu with parameter DACO and in its submenu by function AOUT.
Features of analogue output
In panel meter is necessary to select type of analogue output in menu DACO ->AOUT and setup para-
meters in DACO ->→ ASTR and AEND
Type of analogue output Symbol on display
Current
0-20 mA 0-20
4-20 mA 4-20
20-0 mA 20-0
20-4 mA 20-4
Voltage
0-10 V 0-20
2-10 V 4-20
10-0 V 20-0
10-2 V 20-4
Analogue output is connected to connectors 21 - 28 (fig. 13).
Press key MENU , make output setup and enter into configuration menu. Press keys UP and
DOWN to get to DACO option. Press menu key again to enter into setup level. Press keys
UP and DOWN to select required type of analogue output 0-20, 4-20, 20-0, 20-4. Press key
MENU to confirm selected parameter.
21 22 23 24 25 26 27 28
+
+
VmA
Napěťový
výstup
0/2-10
V
Proudový
výstup
0/4-20
mA
fig. 13
Connection of analogue output
5.1
Voltage output
Current output
LOCK
HOLD
CLR Sum 2 (clear sum 2)
COM (GND)

17
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
5
ANALOGUE OUTPUT
5.2
Current output
Voltage output
Principle of analogue output is shown on block diagrams Fig. 14 and 15. Diagrams of voltage and current
output are separated for better orientation. In real connection are inputs and setup AOUT, ASTR and END
common. For currect function of panel meter set range of analogue output ASTR (start of range), AEND (end
of range), that relates to input signal – see configuration of start and end of analogue output.
Block diagram of analogue output function
fig. 14
fig. 15

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
18
6COMMUNICATION
Panel meter AP11-37 can be equipped with communication line with protocol MODBUS, that is selected
with order of panel meter according to ordering code. Following options of communication lines are avai-
lable: RS232, RS485 with galvanic separation or both simultaneously.
Diagram of communication lines RS232 and RS485 connections
31 32 33 34 35 36 37 38
Rx D
Tx D
Komunikační linka
(svorky 35, 36, 37)
RS232
COM
Komunikační
linka
(svorky
32, 33)
RS485
31 32 33 34 35 36 37 38
Zakončení komunikační
linky RS485
(propojit
svorky 31, 32a33, 34)
RS (
-)
RS (+)
GND
+
820R
120R
820R
Diagram of communication line RS485 connection
31 32 33 34 35 36 37 38
Diagram of communication line RS232 connection to PC (connector Canon 9 pin)
fig. 16
g. 17
obr. 19
Line is terminated at the beginning
and at the end of communication
line to ensure idle conditions and to
prevent bounces in line.
Resistance termination is inside panel meter
and is the same for both lines.
Komunikační linka RS485
Communication line
(connectors 32, 33)
RS485
Communication line
RS232
(connectors 35, 36, 37)
Communication line RS485
Termination of communication
line RS485
(interconnect connectors 31, 32, 33, 34)
g. 18 connector Canon 9 pin

19
A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
7
CONNECTION OF CONTACTC OUTPUTS
Figure 19 shows options of connections of output relay contacts. Max. load of relay contacts is 250 VAC, 2 A
or 30 VDC, 2A. It is advisable to connect antijamming RC cells (e.g. 220 ohm and 0,1 μF) to respective con-
tacts to increase reliability and to reduce jamming. State of output relays 1,2,3,4 (if present) select in menu
of AP11-37 . The state of relays setup in submenu parameters ALA1 to ALA4.
41 42 43 44 45 46
RE 1RE2
51 52 53 54 55 56
RE
3R
E4
fig. 19
(ALA-3)
(ALA-1) (ALA-2) (ALA-4)

A.P.O. - ELMOS v.o.s., Pražská 90, Nová Paka 509 01
tel. +420 493 504 261, fax: +420 493 504 257, e-mail: apo@apoelmos.cz
20
8PANELMETERFUNCTIONS
8.1 BLOCK DIAGRAM OF AP11-37 COUNTER
fig.20
CURRENT SIGNAL
PULSE INPUT
CPU
COM
ANALOGUE
OUTPUT
DISPLAY
SIGNALING
KEYBOARD
BARGRAPH
INPUTS
Main switch is not included in panel meter therefore it activates immediately after connection to power
supply. LOAD briefly appears on panel meter display. During this time initialization and test of inner para-
meters is performed. After completion of test panel meter switches to working mode.
Caution
If LED diode IN 1 flashes panel meter needs to be recalibrated.
Unplug panel meter from power supply. Keep key SET pressed and connect the panel meter to power sup-
ply. Keep the key pressed until sign RESET appears on display.
Default parameters are restored after reset!
8.2 Panel meter switch on
8.3 Panel meter reset
Table of contents