EMC-PARTNER DOW3000 User manual

PARTNER
EMC
LARGEST RANGE OF IMPULSE TESTERS UP TO 100KV/100KA
User Manual
Version 3.26
Technical Data 4.8
DOW3000
•CDN-DOW-DATA-LF
•CDN-DOW-DATA-HF
•CDN-DOW-DATA-HF
•ADAPTER BOX RJ45-8L
•MF1000-1
•MF1000-2
•CN-EFT1000
•VERI-CP-EFT
•VERI1K EFT
•VERI01 OSI
•VERI50 EFT

PARTNER
EMC
The Manufacturer
EMC PARTNER AG
Baselstrasse 160
4242 Laufen
Switzerland
Phone +41 61 775 20 30
Fax +41 61 775 20 59
service@emc-partner.ch
Your Local Representative
Check www.emc-partner.ch to find your local representative
Typographical Conventions
Warnings
Safety notices that must be heeded are indicated with this warning triangle.
Additional Advice
Throughout the user manual very important or helpful advice is indicated with this symbol.
The Small Print
All products from EMC PARTNER, where appropriate, fulfil the requirements for CE
certification. A copy of the individual certificates can be viewed and downloaded from
www.emc-partner.com section company documents.
This document may only be reproduced or distributed as the complete document. Partial repro-
duction of this document requires full, written approval of EMC PARTNER.
Generated on March 31, 2020 with opensource software L
A
T
EX
Subject to change without notice.
Page 2/55

PARTNER
EMC
Safety Instructions
This warning sign is visible on the tester. Meaning: This equipment should only be operated
by trained personnel after carefully reading the user manual.
The system belongs to safety class 1.
The system fulfils the requirements of the safety standards IEC 61010 for laboratory measurement equip-
ment "Safety requirements for electrical measuring, control and laboratory equipment".
Consider always the following safety points
•Dangerous mains voltage or high voltages are present inside the test equipment and ALL devices
attached to it.
•The Protective Earth must be connected to Earth on the test equipment.
•Before removing any covers from the test equipment, remove all external connection cables.
•Before changing the mains fuse, remove all external connection cables.
•Fuses should only be replaced with the same type and value.
•People with heart pacemakers must not be in the vicinity of the test equipment when it is in
operation.
•Do not switch on or operate the test equipment if an explosion hazard exists.
•The test equipment should be operated in a dry room. If condensation is visible, the affected
unit(s) should be dried before operating.
•Never touch the Equipment Under Test (EUT), when the test system is operating.
•Establish a safety barrier around the EUT and if required connect it to the safety circuit in the
auxiliary connector on the rear panel. The cables under high voltage must not be touched during
testing.
•The EUT should be covered and/or well marked during the tests.
•If protective parts are tested, which are likely to explode during tests, they must be covered with
a protective cabinet.
•If the test equipment or any of the system components are damaged or it is possible that damage
has occurred, for example during transportation, do not apply any voltage.
•This user manual is an integral part of the test system. EMC PARTNER and its sales partners
refuse to accept any responsibility for consequential or direct damage to persons and/or goods due
to non-observance of instructions contained herein or due to incorrect use of the test equipment.
Page 3/55

PARTNER
EMC
Sicherheitshinweise
Gefahr! Dieses Warnsysmbol ist am Tester sichtbar angebracht. Es bedeutet: Diese Ausrüstung
darf nur durch ausgebildetes Personal bedient werden, nachdem diese die Bedienungsanleitung
sorgfältig gelesen hat.
Das Gerät gehört der Schutzklasse I an.
Das Gerät erfüllt die Anforderungen der Sicherheitsnormen IEC 61010 für Messungen in Laboratorien
“Sicherheitsanforderungen an elektrische Mess- und Regelgeräte sowie Laborausrüstung”.
Beachten Sie stets die folgenden Sicherheitshinweise:
•Gefährliche Netzspannung oder Hochspannung liegt im Inneren der Testausrüstung und an ALLEN
daran angeschlossenen Geräten an.
•Der Schutzerdeanschluss der Testausrüstung muss mit der Schutzerde des Netzes verbunden sein.
•Bevor Sie irgendwelche Abdeckungen von der Testausrüstung entfernen, entfernen Sie alle äusseren
Anschlusskabel.
•Bevor Sie die Netzsicherung austauschen, entfernen Sie alle äusseren Anschlusskabel.
•Netzsicherung nur mit demselben Typ und Ansprechwert ersetzen.
•Personen mit Herzschrittmacher dürfen sich nicht in der Nähe des Geräts aufhalten, wenn dieses
in Betrieb ist.
•Schalten Sie die Testausrüstung nicht ein oder betreiben Sie das Gerät nicht, wenn Explosionsgefahr
besteht.
•Die Testausrüstung darf nur in einem trockenen Raum betrieben werden. Wenn Kondensation
(Beschlag) sichtbar ist, sind die betroffenen Teile der Ausrüstung zu trocknen.
•Berühren Sie niemals die getestete Ausrüstung (EUT), wenn das Testsystem arbeitet.
•Stellen Sie eine Sicherheitsabtrennung rund um das EUT auf und verbinden Sie diese, wenn verlangt,
mit dem Sicherheitskreis im Hilfsanschluss (Auxiliary) an der Rückwand. Die unter Hochspannung
stehenden Kabel dürfen während des Tests nicht berührt werden.
•Während des Tests sollte das EUT abgedeckt und / oder gut markiert werden.
•Wenn Schutzelemente leicht explodieren können während Tests, müssen diese mit einem schützen-
den Deckel abgedeckt werden.
•Legen sie keine Spannung an, wenn die Testeinrichtung oder irgendwelche Systembestandteile
beschädigt sind oder möglicherweise ein Schaden eingetreten ist, zum Beispiel beim Transport.
•Diese Bedienungsanleitung ist ein vollständiger Bestandteil des Testsystems. EMC PARTNER und
seine Verkaufspartner lehnen jede Verantwortung für direkte wie für Folgeschäden an Personen oder
Gütern ab, die aufgrund von Nichtbeachten von hierin enthaltenen Anweisungen oder aufgrund von
unrichtigem Gebrauch dieser Testausrüstung entstehen.
Page 4/55

PARTNER
EMC
Consignes de sécurité
Ce sigle d’avertissement est visible sur le testeur. Signification: Cet équipement ne doit être
utilisé que par du personnel formé et après avoir soigneusement lu le mode d’emploi.
Le système appartient à la classe de sécurité 1
Le système répond aux exigences des normes de sécurité CEI 61010 pour les équipements de labora-
toire "Exigences de sécurité pour les appareils électriques de mesure, de contrôle et d’équipement de
laboratoire".
Considérez toujours les points de sécurité suivants:
•Des tensions secteurs dangereuses ou des tensions élevées sont présentes à l’intérieur de l’équipement
de test et de TOUS les périphériques.
•La mise à la terre de l’équipement de test doit être effectuée.
•Avant de retirer les couvercles de l’équipement, débranchez tous les câbles de connexion externes.
•Avant de changer le fusible secteur, débranchez tous les câbles de connexion externes.
•Les fusibles ne doivent être remplacés que par le même type et valeur.
•Les personnes portant un stimulateur cardiaque ne doivent pas se tenir à proximité de l’équipement
de test lorsqu’il est sous tension d’opération.
•Ne pas allumer ou faire fonctionner l’équipement de test si un risque d’explosion existe.
•L’équipement de test doit être utilisé dans un local sec. Si de la condensation est visible, les
appareils affectés doivent être séchés avant utilisation.
•Ne jamais toucher l’équipement sous test (EST), lorsque le système est sous fonction.
•Mettre en place une barrière de sécurité autour de l’EST et, si nécessaire raccorder la au circuit de
sécurité du connecteur auxiliaire sur le panneau arrière. Les câbles sous haute tension ne doivent
pas être touchés pendant le test.
•L’EST doit être couvert et / ou bien marqué pendant les essais.
•Si des éléments de protection sont susceptibles d’exploser lors de tests, ils doivent être couverts
avec un boîtier de protection.
•Si l’équipement de test ou l’un des composants du système ont été endommagés ou il est possible
que des dommages ont eu lieu, par exemple pendant le transport, ne pas appliquer de tension.
•Ce manuel fait partie intégrante du système de test. EMC PARTNER et ses partenaires commerci-
aux refusent d’accepter toute responsabilité pour des dommages indirects ou directs à des personnes
et / ou des marchandises dues au non-respect des instructions contenues dans ce document ou en
raison d’une mauvaise utilisation de l’équipement de test.
Page 5/55

PARTNER
EMC
Contents
Contents
1 System Overview 8
1.1 DOW3000 .......................................... 8
1.2 CDN-DOW-DATA-LF (PN 108260) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 CDN-DOW-DATA-HF (PN 108321) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 CDN-DOW-DATA-HF18 (PN 108495) . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.5 ADAPTER BOX RJ45-8L (PN 108322) . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.6 MF1000-1 (PN 103480) and MF1000-2 (PN 103481) . . . . . . . . . . . . . . . . . . . 10
1.7 CN-EFT1000(PN103468) ................................. 10
1.8 VERI-CP-EFT(PN104668)................................. 10
1.9 VERI1KEFT(PN103473) ................................. 10
1.10VERI01OSI(PN103570).................................. 10
1.11VERI50EFT(PN103472) ................................. 11
2 Initial Operation 12
2.1 UnpackingandChecking .................................. 12
2.2 Installation.......................................... 12
2.3 PoweringtheEquipment................................... 12
2.4 SafetyCircuit ........................................ 12
3 Using the Generator 13
3.1 FrontPanel ......................................... 13
3.2 RearPanel.......................................... 15
3.3 SoftwareUserInterface ................................... 16
4 Perform a Test 18
4.1 ProgrammingtheGenerator................................. 18
4.2 RunaTest.......................................... 19
4.3 Slow and Fast DOW Test on EUT Power Ports . . . . . . . . . . . . . . . . . . . . . . 20
4.4 Slow DOW Magnetic Field Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.5 Slow DOW Test on Unshielded and Unsymmetrical Interconnection Lines . . . . . . . . . 22
4.6 Slow DOW Test on Unshielded and Symmetrical Interconnection Lines . . . . . . . . . . 23
4.6.1 Coupling using CDN-DOW-DATA-HF according to IEC 60255-26 . . . . . . . . . 23
4.6.2 Coupling using CDN-DOW-DATA-HF18 according to IEC 61000-4-18 . . . . . . 25
4.7 Fast DOW capacitive couplig clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.8 Insulation1.2/50....................................... 28
4.9 Calibration.......................................... 28
4.9.1 SlowDOW...................................... 29
4.9.2 FastDOW...................................... 30
4.9.3 Fast DOW with CN-EFT1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.9.4 Slow DOW Magnetic Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.9.5 Insulation ...................................... 33
5 Additional Software Features 34
5.1 Protocols........................................... 34
5.2 Setuplibrary......................................... 34
5.2.1 Storesetup ..................................... 35
5.2.2 LoadSetups..................................... 35
5.3 SetupLinker......................................... 35
Page 6/55

PARTNER
EMC
Contents
5.4 FileManager......................................... 36
6 Configuration 37
6.1 Settings ........................................... 37
6.2 Generator .......................................... 38
7 Remote Operation 39
7.1 TEMA3000 ......................................... 39
8 Maintenance and Service 40
8.1 Firmwareupdate....................................... 40
8.2 Maintenance......................................... 40
8.3 Cleaning ........................................... 40
8.4 Warranty........................................... 40
8.5 CustomerService ...................................... 41
8.6 Recycling/Disposal..................................... 41
9 Technical Data 42
9.1 DOW3000 .......................................... 42
9.1.1 DOWSlow...................................... 42
9.1.2 DOWFast...................................... 44
9.1.3 InternalCDN .................................... 45
9.1.4 DOW Slow Magnetic Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
9.1.5 Insulation0.5Joule ................................. 47
9.1.6 DOW3000 Main Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
9.2 MF1000-1 .......................................... 49
9.3 MF1000-2 .......................................... 49
9.4 CDN-DOW-DATA-LF .................................... 50
9.5 CDN-DOW-DATA-HF.................................... 51
9.6 CDN-DOW-DATA-HF18................................... 52
9.7 ADAPTERBOXRJ45-8L.................................. 53
9.8 VERI1KEFT......................................... 53
9.9 VERI01OSI ......................................... 53
9.10VERI50EFT......................................... 54
9.11CN-EFT1000......................................... 54
9.12VERI-CP-EFT ........................................ 55
Page 7/55

PARTNER
EMC
1 System Overview
1 System Overview
Not all accessories are used for all tests. The chapter “Perform a test” provides more informa-
tion.
1.1 DOW3000
The DOW3000 generator is a compact system with all wave-
forms for IEC 61000-4-18, IEC 61000-4-10 and several prod-
uct standards. The Generator with an integrated 3-phase
coupling and decoupling network is available in different ver-
sions with a combination of the following circuits.
•Slow Damped Oscillation 100 kHz and 1 MHz
•Fast Damped Oscillation 3 MHz, 10 MHz and 30 MHz
•Insulation Test Impulse 1.2/50 us with 0.5 J
The 7 inch graphic colour touch panel provides an intuitive
and comfortable user interface. Interactive parameter input,
help text and graphic illustrations, guide the user through
the instrument programming process.
1.2 CDN-DOW-DATA-LF (PN 108260)
The CDN-DOW-DATA-LF is a compact coupling-decoupling
network specifically designed to couple Slow DOW signals
on up to eight datalines. Both 1 MHz and 100 kHz signals
can be coupled via 0.5 uF capacitors, either in differential
mode or in common mode according to IEC 60255-26 and
IEC 61000-4-18. Additionally, common mode coupling of
the signal lines directly to PE is possible. The CDN-DOW-
DATA-LF is capable of handling high-power EUTs with AC
or DC voltage. Each line has a decoupling impedance to
protect the auxiliary equipment from the DOW test signals.
Page 8/55

PARTNER
EMC
1 System Overview
1.3 CDN-DOW-DATA-HF (PN 108321)
The CDN-DOW-DATA-HF is a compact coupling-decoupling
network specifically designed to couple Slow DOW signals on
2-pair or 4-pair symmetrical high-speed communication lines.
Both 1 MHz and 100 kHz signals can be coupled in common
mode via 0.5 uF capacitors, according to IEC 60255-26. The
coupling is one pair versus another pair or versus multiple
other pairs. Each pair has a decoupling impedance as well
as protection devices to protect sensitive auxiliary equipment
from the DOW test signals.
1.4 CDN-DOW-DATA-HF18 (PN 108495)
The CDN-DOW-DATA-HF18 is a compact coupling-
decoupling network specifically designed to couple Slow
DOW signals on 2-pair or 4-pair symmetrical high-speed
communication lines. Both 1 MHz and 100 kHz signals can
be coupled in common mode via 0.5 uF capacitors, according
to IEC 61000-4-18. The coupling is all lines versus PE. Each
pair has a decoupling impedance as well as protection devices
to protect sensitive auxiliary equipment from the DOW test
signals.
1.5 ADAPTER BOX RJ45-8L (PN 108322)
The ADAPTER BOX RJ45-8L is an Ethernet adapter com-
patible with the CDN-DOW-DATA-HF and CDN-DOW-
DATA-HF18, which allows you to connect your Ethernet
devices directly to the CDNs without the need for special
cabling.
Page 9/55

PARTNER
EMC
1 System Overview
1.6 MF1000-1 (PN 103480) and MF1000-2 (PN 103481)
Test coil for Slow DOW magnetic field test according
IEC 61000-4-10. The MF1000-2 has more than twice the
height of MF1000-1 for larger EUTs.
1.7 CN-EFT1000 (PN 103468)
The clamp provides the ability of coupling the damped os-
cillatory waves to the circuit under test without any galvanic
connection to the terminals of the EUT’s ports, shielding of
the cables or any other part of the EUT.
1.8 VERI-CP-EFT (PN 104668)
Transducer plate for CN-EFT1000 capacitive coupling clamp
calibration. The plate has a SHV-BNC connector and a
15 cm strap to bond to the reference ground.
1.9 VERI1K EFT (PN 103473)
1 kOhm termination with SHV-BNC connector for the Fast
DOW voltage calibration.
1.10 VERI01 OSI (PN 103570)
0.1 Ohm shunt with SHV-BNC connector for the Fast DOW
current calibration.
Page 10/55

PARTNER
EMC
1 System Overview
1.11 VERI50 EFT (PN 103472)
50 Ohm termination with SHV-BNC connector for the Fast
DOW calibration together with a capacitive coupling clamp.
Page 11/55

PARTNER
EMC
2 Initial Operation
2 Initial Operation
2.1 Unpacking and Checking
1. Check the packaging for signs of damage. If damage is visible, report this to the shipping company
immediately.
2. Gently remove the equipment from the packaging.
3. Check the delivery for completeness using the delivery note and the accessory lists for the var-
ious items. If something is missing, contact EMC PARTNER or your local sales representative
immediately.
4. Check the equipment for any damage. If there is damage, immediately contact the shipping
company who made the delivery.
Retain the original packing material for a safe transport at a later date.
2.2 Installation
The equipment is designed for use under laboratory conditions. Make sure that all fan openings are
unobstructed, that the airflow perforations are unimpeded, and that the minimum distance from the wall
is 10cm. An insufficient airflow can cause the equipment to overheat, which may disturb the operation
and even cause damage.
Most equipment from EMC PARTNER can be mounted in a 19" Rack. A special rack mounting
kit is available. Contact your local sales representative for a quotation.
2.3 Powering the Equipment
Check that the local AC-supply is suitable for the equipment. Power rating of the instrument is indicated
on the type plate next to the AC power supply connector located on the rear panel. Only use the supplied
power cord to connect to your public power supply.
The equipment must always be connected to protective earth. Check the earth connection on
your power outlet before you connect and turn on the equipment.
2.4 Safety Circuit
The generator is equipped with a safety circuit. As long as the safety circuit is open, no test can be
started. To close the safety circuit
1. connect the auxiliary connector on the rear panel
2. pull out the emergency stop button on the front panel with a slight turn clockwise
A closed safety circuit is indicated by a red illuminated emergency stop button
Page 12/55

PARTNER
EMC
3 Using the Generator
3 Using the Generator
3.1 Front Panel
1
2
3
4 5
6 7 8 9 10 11
12
13
14 15
16
17
1 ON/STBY
Power-up or power-down the generator.
2 Touchscreen Monitor
User interface to perform all functions including test types selection, parameter entry, saving and
recalling tests.
3 Rotary Knob
Entry device enables faster navigation and parameter entry.
4 Run/Stop
Starts, stops or pauses the selected test.
A green blinking LED indicates a running test.
A red blinking LED indicates a stopped test due to an error.
A red LED indicates a finished test.
Page 13/55

PARTNER
EMC
3 Using the Generator
5 Emergency Stop Button
Push to open the safety circuit.
As soon as the safety circuit is open, all running tests are immediately stopped and no more tests
can be started.
Release the emergency stop button with a slight turn clockwise.
A closed safety circuit is indicated by a red illuminated emergency stop button.
6 USB Port
Connect a USB memory stick to copy files (setups, protocols, error logs) from the generator. The
USB interface is also used to update the internal firmware.
7 Auxiliary Connection
Interface for additional external equipment.
8 Measurement Output
BNC output for a direct connection to an oscilloscope to monitor the impulse of selected waveshape.
In DOW3000, only CH1 is used for the 1.2/50 us insulation waveshape voltage measurement. See
the technical data for more information.
9 Trigger Output
BNC output for a direct connection to an oscilloscope to start pulse acquisition. Level and duration
depend on the specific test. Some typical values are
•12 V primary impulse trigger
•8 V secondary impulse trigger
•5 V EUT power state
10 DOW Slow Output
Direct output for the Slow Damped Oscillation circuit.
11 1.2/50 Insulation Output
Direct output for 1.2/50 us Insulation circuit.
12 Fused EUT-Input
Fused Input of the internal CDN.
Never connect DC-Voltage. The internal fuse and switch are not compatible to DC. For
DC, use the dedicated input.
13 Fuse
14 Unfused EUT-Input
Unfused Input of the internal CDN. Use this input for DC.
15 EUT Power Output
Output of the internal CDN for the connection to the EUT. The current coupling is indicated by
a LED.
16 DOW Fast Output
Direct output for the Fast Damped Oscillation circuit.
17 Power Switch
Power-Switch of the internal CDN. A closed switch is indicated by an illuminated switch.
Page 14/55

PARTNER
EMC
3 Using the Generator
3.2 Rear Panel
1
23
4
5
6
1 AC Socket / Mains Fuse
A country specific mains cable is delivered with the equipment, this should be used to connect to
the mains power supply. A mains fuse is located in the socket housing. The value is indicated on
the equipment type plate.
Take care to ensure the power mains matches the power ratings on the equipment type
plate.
The generator must always be connected to protective earth
2 Auxiliary Connector
Fit the auxiliary connector delivered with the equipment, to close the safety circuit.
It is possible to extend the safety circuit or to connect additional external warning lamps.
Page 15/55

PARTNER
EMC
3 Using the Generator
•The green lamp indicates safety circuit open, no test can be started
•The red lamp indicates safety circuit closed, tests can be started
3 RS485 Interfaces
Interfaces to connect additional equipment that is controlled directly from the test equipment.
4 Ethernet Interface
Remote control using TEMA3000 software via a common ethernet interface.
5 EUT Failed
A falling edge to ground impressed on the BNC connector is indicated in the test protocol as an
EUT Fail condition. The input is internally pulled up to max. 24 V. Driving with an open-collector
output is recommended.
6 Trigger Input
A falling edge to ground on the BNC connector, triggers impulse release. This function must be
configured in the test parameter menu. The input is internally pulled up to max. 24 V. Driving
with an open-collector output is recommended.
3.3 Software User Interface
1
23
1 Test selection
The main window is a carousel of all tests available in the equipment. A short description is
included with each test type. A green icon indicates that the test is available, a grey icon indicates
that the test is not currently available. Select a test by bringing the relevant icon to the centre of
the screen. This is achieved by finger swipe on the touch panel or by using the rotary knob.
Page 16/55

PARTNER
EMC
3 Using the Generator
Not all tests are available all the time. Availability depends on the hardware configuration.
To simplify the menu structure, tests that are unavailable can be removed from the
carousel.
≡(General Menu) –> Settings –> Test Settings
2 Home button
Select from any point in the menu structure to return to the main window.
3 General menu
Activates a drop-down menu on the right side giving access to the equipment configuration settings.
Page 17/55

PARTNER
EMC
4 Perform a Test
4 Perform a Test
4.1 Programming the Generator
The same process is used to program the generator for different tests. In the first step, the parameters
should be set according to the requirements of the test.
1
2
3
1 Main Category
For quick access, categories at the bottom of the display contain groups of relevant parameters.
Select a category using the touch panel.
2 Parameter List
All parameters from a selected category are shown on a single screen. Click on a parameter to edit
the value.
Press and turn the rotary knob at the same time for a quick parameter change.
All parameters are checked for validity during the entry process. If necessary (for example test
level too high), the system automatically corrects the value. Corrected values are indicated by a
red background.
The system also checks for inconsistencies with other parameters and, where necessary, makes an
adjustment. It is not possible to create a test containing invalid parameters.
For each parameter, an additional help text is provided directly on the generator. Press and hold
a parameter to open the help text in a popup window.
3 Parameter Image
Individual images clarify the effect of each parameter.
Page 18/55

PARTNER
EMC
4 Perform a Test
4.2 Run a Test
Starting a test by pressing the run button on the front panel, changes the display to the run view.
1
23 4
5
6
1 Information Area
Indicates the most important information about a running test.
2 Progress Bar
Indicates progress of the whole test and, if available, about the current cycle.
3 Status Bar
Provides information about the current event.
4 Running Options
Click to open a menu for additional options while a test is running.
•Mark or Fail add this information at the current test position. This information will be
included in the protocol.
•Pause will pause the running test. With Resume, the test will continue at the current position.
•Trigger will release an impulse. This option is only available in the manual trigger mode.
Pressing the Run button during a running test activates the pause. Another press on the
run button will resume the test at the current position.
During the pause, only the disturbance source is interrupted. A possible EUT voltage will
not be switched off. Only a stopped test and an open safety circuit (pressed emergency
button) guarantees a safe operation on the connected device.
5 Parameter Change
Click to open a shutter menu containing all parameters that can be edited while a test is running.
Depending on the test type, parameter change will take place with the next cycle or after a few
seconds. Changed values are highlighted for a few seconds in the information area.
6 Pulse History
Surge tests with internal measurement circuits show the results as scrolling blocks. Browse the
blocks while a test is running. Click on a block to view complete information.
Marked pulses have an orange background, failed ones a red and correct ones a green.
The pulse history is available for the 1.2/50 us insulation waveform only.
Page 19/55

PARTNER
EMC
4 Perform a Test
4.3 Slow and Fast DOW Test on EUT Power Ports
1
2
3
1 Generator
Select the desired waveform and frequency and CDN as coupling device. The same connection
setup can be used for Slow DOW and Fast DOW.
For Fast DOW, a reference ground plane must be used. For Slow DOW, a reference
ground plane is recommended. All the equipment including generator and EUT must be
placed on a ground reference plane. The generator earth bar on the frontpanel must be
connected with the ground reference plane.
2 EUT
Connect the EUT to the output of the CDN.
•For 1-phase AC, connect L1, N and PE
•For 3-phase AC, connect L1, L2, L3, N and PE
•For DC (floating), connect DC+ to L1 and DC- to N
•For DC (when DC- = PE), connect DC+ to L1 and DC- to PE
The blank sockets are used for calibration with the calibration adapter. See the calibration
protocol for more information.
3 Auxiliary Equipment
Connect the Auxiliary Equipment to the input of the CDN.
•For 1-phase AC, connect L1, N and PE
Page 20/55
Table of contents
Popular Portable Generator manuals by other brands

Westinghouse
Westinghouse WH3250RVC owner's manual

Gopel Electronic
Gopel Electronic basicCON 4115 user manual

Firman
Firman P03603 owner's manual

jcb
jcb RENTAL Series quick start guide

HERON
HERON 8896411 Translation of the original user manual

Abicor Binzel
Abicor Binzel xFUME POWER 250 Original operating instructions

Champion
Champion 100592V quick start

Tektronix
Tektronix AWVG1 user manual

Champion
Champion 41513 Owner's Manual and Operating Instructions

Generac Power Systems
Generac Power Systems 1338-1 owner's manual

Symmetricom
Symmetricom SSU-2000 Technical reference

Porter-Cable
Porter-Cable D27245-028-0 instruction manual