FERVE F-1860
WARNING
♦ Engines produce carbon monoxide which is odourless, causes slower
reaction time, and can lead to serious injury through inhalation. Keep service
areas WELL VENTILATED while the engine is running!
♦ Activate the handbrake and block the wheels be ore testing or repairing the
vehicle. It is especially important to block the wheels on ront-drive vehicles,
as the handbrake does not lock the drive wheels.
♦ Wear an eye shield when testing or repairing vehicles. Exceeding the limits
o the F-1860 is dangerous. It will expose you to serious or possibly atal
injury. Care ully read and take heed o the precautions described in this
instruction manual, and make sure that the established limits o the F-1860
are never exceeded.
♦ Under absolutely no circumstances must voltage exceed 1000 VDC or 750
VDC between any terminal or ground.
♦ Extreme caution must be exercised when measuring voltage above 25 VAC
or VDC.
♦ Circuits to be tested must be protected by a 20A use or circuit breaker.
♦ Do not use the F-1860 i it has been damaged.
♦ Do not use the test leads i the insulation is damaged or the cable is
exposed.
♦ Use adequate current clamps when measuring circuits in excess o 10A.
♦ To avoid electrical shock, never touch the test leads, tips or circuit while
testing.
♦ Do not make a voltage measurement with the test leads in the 20A or mA
terminal.
♦ When testing or voltage or current, make sure that the corresponding
unction o the F-1860 is the correct one. We recommend you take a test
reading o a known voltage or current be ore carrying out a speci ic unction.
♦ Choose the proper range and unction or the measurement. Do not try to
take voltage or current measurements that may exceed the ratings marked
on the unction /range switch or terminal.
♦ When measuring current, connect the F-1860 in series with the load.
♦ Never connect more than one set o leads to the F-1860.
♦ Disconnect the live test lead be ore disconnecting the common test lead
(COM).
♦ The mA and 20A terminals are protected by uses. In order to avoid possible
injury or damage, use only in circuits not exceeding 400mA, 10A continuous
or 20A or a maximum15 seconds.
IMPORTANT
♦ In order to maintain the accuracy o the F-1860, replace the battery as soon
as the battery symbol appears on the display.
♦ Avoid measuring errors rom external inter erence: keep the F-1860 away
rom spark plug cables or wire coils.