Actuator Calibration Stand Type AL0010 - Page 5
G.R.A.S. Sound & Vibration
Setting up
Note: the terms generator and measurement amplier refer to a set up which simultaneously
provides the test signal to the electro-static actuator and measures the signal picked up by
the microphone.
Proceed as follows:
1) Power Module Type 12AK
Connect the free end of the preamplier cable to the LEMO Input socket.
Connect, via a suitable cable, the BNC Output to the input of the measuring amplier.
Select Lin. or Direct.
2) Actuator Supply Type 14AA
Connect the thin lead (with banana plug) from the AL0010 to the socket marked
+ 800 V DC.
Connect the output from the signal generator to the BNC socket marked AC Input.
3) Adjust the signal output level as required.
Applying a Swept Signal and Measuring the Output from the Microphone
The following describes a typical procedure for applying the swept signal.
It is assumed that the generator and measuring amplier work to produce constant-condence
results (i.e. maintaining a constant β T product) in real time throughout the frequency range of
interest and make the measurement data available graphically and numerically.
With everything set up as described above, proceed as follows:
a) set the generator to oscillator mode
b) set the measuring amplier to at response
c) initiate a constant-level logarithmic sweep on the generator.
The measuring amplier will follow the response of the microphone throughout the sweep
and record and display the results accordingly.
Curves showing the upper and lower tolerance levels for the frequency range of interest could
be superimposed on the graphical display.