TROUBLESHOOTING INSTRUMENT
December, 2019
2019 Gatrik Innotech Indonesia, PT. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
C
Page-3
December, 2019
2019 Gatrik Innotech Indonesia, PT. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
C
COMPACT
PRECISION
Reliability Design
Resistivity Meter
R12-800SC
User Manual
UNREADABLE
POTENTIAL
VOLTAGE INPUT
THE SYMPTOMS POSSIBLE CAUSE PROBLEM SOLVING
1. Low Current Inject
2. Low Power Battery Input
3. Bad Connection Cable from
terminal or elekrtodes
1. Switch On some additional
converters
2. Change Battery Connection /
Replace New Battery
3. Check cable & Electrodes
connection
ALARM OVER CURRENT
ACTIVATED
High Current Inject Switch Off some converters
ALARM OVER VOLTAGE
ACTIVATED
Out of Range Setting Measurement Chek Manual Range Setting
Converter Module Fail
NO OUTPUT POWER
INSTRUMENT
If one of the converter switches
is pressed, but voltage output
display is not increase, the
converter module is fails
1.
2.
MCB DC POWER TRIP Short Circuit Connection Please Contact Authorized Dealer
for Repair the Instrument
The measurement process can
still be done, but the maximum
output voltage (800V) cannot be
achieved
1 2 3 4 5 6 7 8
DC POWER
100 - 800 VDC
RS12-800SC comply with IEC 61010-1, IEC61010-2-030 CAT III 600 V, CAT II 1000 V, Pollution Degree 2 IEC
60998-2-1, U L 1059, UL 486E. See Specifications. Use the Product only as specified in this manual, otherwise the
protection provided by the Product may be impaired. In this manual.
WARNING identifies conditions and actions that pose hazards to the user.
CAUTION identifies conditions and actions that may damage the Profuct or the equipment under test.
International symbols used on the device.
1
1. Terminal Battery Input
2. Handle Instrument
3. DC Power Input
4. Push ON Meter Voltage Input
5. Push ON Meter Current Inject
6. Backlight Function
7. Hold Function
8. Manual Range Function
9. REL Function
10. Over Voltage Alarm
11. Display Battery Voltage
12. Display Voltage Input
13. Display Current Inject
14. Display Voltage Output
15. Terminal Current Inject
16. Switch ON, Converter
17. Alarm Over Current
18. Terminal Voltage Input
19. Push On, Current Inject
: input connection from main power source (Battery 12V)
: for handling process or moving instrument and protect main metering while
operation
: Switch On/Off Main Power Instrument (50A MCB DC)
: to activation/power ON display instrument voltage input
: to activation/power ON display instrument current inject
: to activation/power ON/OFF backlight
: to hold the present reading measurement
: for accuration settings (point reference)
: stores existing reading (a delta) and resets display to zero
: Monitor Alarm Over Voltage measurement
: monitor condition battery voltage while instrument operation
: monitor potential voltage reading
: monitor current inject reading
: monitor output voltage while instrument operation
: terminal for connection cable inject current (“+”, and “-”) from Steel (SS) stake
: for Switch On/Off each converter ( 1 - 8 Converter)
: monitor information over current injection (while Inject current more than 1200mA)
: terminal for connection cable voltage input ( “+”, and “-”) from copper (Cu) stake
: for activation injection current into the earth
1. Make sure all switch instrument in Off Condition
2. Connect Positive (+) battery cable, from Pole (+) RED, to Positive (+) Terminal Battery Input instrument
3. Connect Negative (-) battery cable, from Pole (-) BLACK, to Negative (-) Terminal Battery Input instrument
4. Connect Cable Current Inject (+/C1/A) & (-/C2/B) from terminal current inject instrument to the
stainless stee (SS) stake refer to measurement mode.
5. C onnect Cable Voltage Input (+) & (-) from terminal instrument to copper (Cu) stake base on configuration
measurement
6. Make sure all all cable & electrodes connection properly
1. Switch On Main MCB DC Power Instrument, (display battery voltage will show the voltage value).
2. Push On, voltage input meter to activation/power on display instrument voltage input
3. Push On, current inject meter to activation/power on display instrument current inject
4. Read & write for SP Value
5. Setting manual Range for meter Voltage Input Instrument
6. Make sure REL function was activated to zero
7. Switch ON some converter for generate current inject (display output converter will show the output
voltage value),
Note : You can configure the output depend on your requirement, ach converter 100V, and up to 800V
while 8 converter activated.
8. Current Inject & Voltage Input will show while current injector release
9. Hold the reading and write the data
1. Switch Off all converter (see the LED on the switch has been off)
2. Switch Off MCB DC Power (display battery voltage, display voltage output, and display main metering will
be off)
3. Dismantle cable battery from terminal battery input
4. Dismantle cable current inject & voltage input from terminal and the electroda/stake
5. Make sure the instrument clean and dry, before the hardcase is closed
1. Terminal Battery Input
2. Handle Instrument
3. DC Power Input
4. Push ON Meter Voltage Input
5. Push ON Meter Current Inject
6. Backlight Function
7. Hold Function
8. Manual Range Function
9. REL Function
10. Over Voltage Alarm
11. Display Battery Voltage
12. Display Voltage Input
13. Display Current Inject
14. Display Voltage Output
15. Terminal Current Inject
16. Switch ON, Converter
17. Alarm Over Current
18. Terminal Voltage Input
19. Push On, Current Inject
15
Safety Information
BATTERY
INPUT
CURRENT
INJECT
VOLTAGE
INPUT
I
G
I
N
R
A
L
O
ORIGINAL
PRODUCT
12 13
11 14
18
HOLD RANGE REL HOLD RA NGE
BATTERY VOLTAGE VOLTAGE OUTPUT
VOLTAGE INPUT CURRENT INJECT
ALARM
OVER
CURRENT
VOLTAGE
METER
CURRENT
METER
SWITCH ON CURRENT
INJECT
INSTRUMENT OVERVIEW
17
LEGEND :
FUNCTION :
OPERATING INSTRUMENTS
A. Installation Instrument
B. Power Up & Measurement Process
1 2 3 4 5 6 7 8
DC POWER
100 - 800 VDC
BATTERY
INPUT
CURRENT
INJECT
VOLTAGE
INPUT
I
G
I
N
R
A
L
O
ORIGINAL
PRODUCT
HOLD RANGE REL HOLD RANGE
BATTERY VOLTAGE VOLTAGE OUTPUT
VOLTAGE INPUT CURRENT INJECT
ALARM
OVER
CURRENT
VOLTAGE
METER
CURRENT
METER
SWITCH ON CURRENT
INJECT
BATTERY 12V
Page-1 Page-2
C. Power Off the Instrument
2
3
5
4
6
16
19
C1/A
C2/B
P1/M
P2/N
9
8
710

Popular Measuring Instrument manuals by other brands

YOKOGAWA DAQSTATION DX100 technical information

YOKOGAWA

YOKOGAWA DAQSTATION DX100 technical information

Toshiba LF470 Specification sheet

Toshiba

Toshiba LF470 Specification sheet

Tektronix BERTScope BSA85C Quick start user manual

Tektronix

Tektronix BERTScope BSA85C Quick start user manual

Applent Instruments Anbai AT4808 user manual

Applent Instruments

Applent Instruments Anbai AT4808 user manual

Siemens FUS1010 operating instructions

Siemens

Siemens FUS1010 operating instructions

Emerson Daniel 3812 installation manual

Emerson

Emerson Daniel 3812 installation manual

AKO AKO-59820 instructions

AKO

AKO AKO-59820 instructions

ISKRAEMECO ME371 Technical description

ISKRAEMECO

ISKRAEMECO ME371 Technical description

Hydrofarm Active Eye quantum PAR meter instruction manual

Hydrofarm

Hydrofarm Active Eye quantum PAR meter instruction manual

Eastron SDM320-C user manual

Eastron

Eastron SDM320-C user manual

Velleman NETBSEM user manual

Velleman

Velleman NETBSEM user manual

Rohde & Schwarz ZNL6-COM Getting started

Rohde & Schwarz

Rohde & Schwarz ZNL6-COM Getting started

McPower EM-70 instruction manual

McPower

McPower EM-70 instruction manual

SIGRIST StackGuard 2 instruction manual

SIGRIST

SIGRIST StackGuard 2 instruction manual

Mannesmann 82520 instruction manual

Mannesmann

Mannesmann 82520 instruction manual

SkyScan Field Pro Operation manual

SkyScan

SkyScan Field Pro Operation manual

Cannon SimpleVIS quick start guide

Cannon

Cannon SimpleVIS quick start guide

KYORITSU 3131A instruction manual

KYORITSU

KYORITSU 3131A instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.