Gentec 882VR User manual


IMPORTANT SAFETY INSTRUCTIONS.
READ AND UNDERSTAND THESE INSTRUCTIONS
COMPLETELY BEFORE OPERATING THIS
EQUIPMENT.
If you do not understand any of these instructions, or if you have any
questions regarding the use of this product, please contact your
facility’s training manager, your supervisor, the medical equipment
dealer from whom the product was purchased, or the manufacturer
before operating the equipment.
Do not attempt to repair this device if you have not been properly
trained. Doing so may create a hazardous situation that may result in
death or serious injury. Attempted repair by anyone other than a duly
authorized repair/service center of Genstar Technologies Co., Inc. voids
any and all warranties, express or implied.
Carefully inspect and test this product before each use to ensure proper
operation. Do not use the product if there are signs of damage or if it
does not pass the initial suction test.
Should this product require repair or service that will require shipping
the product to another location, bear in mind that United States
Federal law prohibits shipping contaminated products, as they
constitute a biohazard. This product must be properly sterilized and
securely packaged prior to shipping.
Genstar Technologies Co., Inc. (GENTEC®) manufactures continuous suction
regulators in two ranges, 0 to 300mmHg (882VR-300) and 0 to 760mmHg (882VR-
760). These suction regulators provide three modes: REG (regulated vacuum), OFF
(no vacuum) and FULL (whatever is available at the wall inlet). Please take a few
minutes to familiarize yourself with the product by reviewing Figure 1 on the next
page.
The mode is selected by moving the lever at the top of the regulator to the left
(REG) for regulated suction, the center (OFF) to turn off the regulator, or the right
(FULL) to allow full line suction.
The REG mode provides regulated suction levels as set by the user. The suction
level is set by occluding the suction tubing, then adjusting the regulator knob on
the front of the suction regulator to achieve the desired suction level, up to the
designed range. Suction is increased by turning the regulator knob clockwise,
decreased by turning it counter-clockwise.
The OFF mode turns off the suction regulator, allowing no suction at the tubing.
1 6
# Problem Probable Cause Corrective
Measures
1
The vacuum gauge
needle does not move
off “0” when regulator
is connected to
vacuum source.
The regulator is in the “OFF”
mode, or is not fully in the
“REG” or “FULL” position.
Move the mode
selector to the
“REG” or “FULL”
position.
The adjustment knob is in the
fully counter-clockwise, closed,
position.
Turn the
adjustment knob
clockwise to open.
The collection bottle or suction
tube is leaking.
Check the
collection bottle
and tubing for leaks.
2
The vacuum regulator
gauge shows a reading,
but there is no suction
at the tubing.
The collection bottle is too full,
causing the float to shut off the
suction.
Empty the
collection bottle.
The filters or suction tubing are
clogged.
Change the filters
and tubing.
3
Under REG mode the
regulator cannot reach
the specified suction
level (i.e., 200mmHg or
760mmHg, depending
upon model).
Suction source cannot provide
sufficient suction.
Increase source
equipment vacuum
settings.
Regulator
Internal
Problems
Relief valve (8) is
damaged. Replace relief valve.
O-ring on vacuum
gauge (7) is
damaged.
Replace o-ring.
O-ring in regulator
assembly (4) is
damaged.
Replace o-ring.
Regulator assembly
(4) is damaged.
Replace regulator
assembly.
Mode selector
assembly (3) is
loose.
Tighten or replace
mode selector
assembly.
4
Under REG mode the
suction is too strong
and cannot be reduced.
Clean gas assist
port.
Gas assist port (6) is clogged.

No. Part No. Description No. Part No. Description
1 300660110 Screw Kit (Set of 5) Shaft
2 200260010 Back Plate Spring
3 203060040
Mode Selector
Assembly 6 204880020
Filter
Assembly
4 203060010 Regulator Assembly Protecting Cap
5 204880010
Adjustment Knob
Assembly Sponge Filter
Logo Sticker Orifice Union
Screw Orifice Seat
Knob 7 141742301 Gauge
Washer 8 304360010 Relief Valve
Spring 9 304660010 Main Cover
C-Clip
5
The FULL mode bypasses the regulator, and is designed for use only when high
levels of suction are needed. Care should be exercised when using this setting, as
the patient will be exposed to whatever level of suction is available at the wall inlet
(typically above 480mmHg).
A suction filter or vacuum trap assembly (GENTEC catalog #880VT) should be
used to prevent aspirate from entering the suction regulator. Typically, the suction
catheter is connected to the suction tubing, which is then connected to the inlet
fitting on the suction canister.
The canister can be connected directly to the suction regulator via DISS
connection, or, as is recommended, connected to a filter or vacuum trap, which is
then connected to the suction regulator via direct, threaded connection, suction
tubing, or DISS connector (see Figure 2).
The appropriate outlet connector (located on the back of the suction regulator)
must be used for connection to the wall inlet. The use of converting adapters (e.g.,
DISS connection to Ohio connection) should be avoided. If the suction regulator is
connected via tubing or hose assembly to the wall inlet (as occurs when the
suction regulator is attached to a mobile stand), a minimum inside hose diameter
(ID) of 5/16” (7.9mm) should be used to prevent loss of flow.
DO NOT connect the suction tubing directly from the patient to either the
Mode Selector
Gauge
Regulator
Adjustment Knob Outlet (in back)
Inlet (on bottom)
Figure 1 - Main Features
2

vacuum trap assembly or the suction regulator. Doing so may permanently damage
the suction regulator and void any and all warranties, express or implied.
A collection canister (reusable or disposable) MUST be used between the patient
and the suction regulator or vacuum trap, if used. If a vacuum trap is not used it is
recommended that a disposable hydrophobic bacterial filter be used between the
suction canister and the suction regulator to prevent overflow of the canister into
the suction regulator. Use of these filters may also prevent build up of aerosolized
particulate inside the suction regulator, thus reducing maintenance requirements
and extending the life of the unit.
VERIFYING REGULATOR OPERATION
NOTE: The proper operation of the suction regulator must be verified prior to each
use. Should the regulator not operate in accordance with the following, it must be
repaired by authorized personnel.
1) Ensure that the Mode selector is in the OFF (center) position.
2) Connect the regulator to a vacuum source (normally the wall-mounted inlet).
3) Occlude the regulator inlet, and turn the adjustment knob one full turn
clockwise.
4) Verify that the gauge needle does not move from the “0” position.
5) Move the Mode selector to the “REG” position.
6) Occlude the regulator inlet, and turn the adjustment knob counter-clockwise
until the gauge needle is at 0.
7) Keeping the inlet occluded, turn the adjustment knob one full turn clockwise.
The gauge should read between 120 and 170 mmHg.
8) Keeping the inlet occluded, turn the adjustment knob fully clockwise. The
gauge needle should move clockwise to the highest vacuum setting.
9) Move the Mode selector to the “FULL” position.
10) Occlude the regulator inlet. The gauge needle should move to the highest
vacuum setting.
11) Turning the adjustment knob should have no effect on the gauge reading when
the selector knob is in the FULL position.
12) Move the Mode selector to the “OFF” position. The needle should return to 0
If the regulator passes all of the above, it is ready for patient use. Failure of any of
the above requires that the unit be serviced by authorized personnel.
Once the operation of the regulator has been verified, it can be prepared for patient
use as follows:
1) Connect the regulator to the wall inlet, ensuring proper latching.
2) Move the Mode selector to “REG”.
3) Occlude the inlet or tubing.
4) Adjust the regulator knob to achieve the desired level of vacuum.
5) Attach the suction catheter and proceed.
NOTE: the suction regulator should always be turned OFF when not in use. This
3
882VR Disassembly Instructions
1) Disconnect the 882VR from the suction source
2) Sterilize the suction regulator using ETO or autoclave, for self-protection
3) Remove the four screws on the back of the unit
4) Turn the regulator adjustment knob counterclockwise to release the faceplate
5) The gauge can be gently lifted out of the socket
6) The regulator can be gently lifted out of the socket
7) One screw holds the Mode selector lever in place.
Reassembly is accomplished by performing the above steps in reverse order.
The following repair assemblies can be purchased from your Authorized GENTEC
distributor (refer to the exploded view on the next page):
Possible Back
Fittings
Ohio/Ohmeda
Chemetron
DISS Nut & Nipple
DISS Hand-Tight
Puritan-Bennett
Hose Barb
Possible Inlet Fittings
Vacuum Trap Assembly or Fitting Only
Hose Barb DISS Male
Figure 2 - Suction Regulator Configurations
DO NOT connect suction tubing directly to the vacuum trap, filter or suction regulator!
Vacuum Filter
4
The regulator can be cleaned by wiping the outside surfaces with disinfectants,
Virox (accelerated hydrogen peroxide), Clorox (sodium hypochlorite), and Cavi-
wipes. The internal flow path can be cleaned by suctioning a cold sterilant
through the unit, then allowing the unit to run for 30-45 minutes to dry the
interior.

vacuum trap assembly or the suction regulator. Doing so may permanently damage
the suction regulator and void any and all warranties, express or implied.
A collection canister (reusable or disposable) MUST be used between the patient
and the suction regulator or vacuum trap, if used. If a vacuum trap is not used it is
recommended that a disposable hydrophobic bacterial filter be used between the
suction canister and the suction regulator to prevent overflow of the canister into
the suction regulator. Use of these filters may also prevent build up of aerosolized
particulate inside the suction regulator, thus reducing maintenance requirements
and extending the life of the unit.
VERIFYING REGULATOR OPERATION
NOTE: The proper operation of the suction regulator must be verified prior to each
use. Should the regulator not operate in accordance with the following, it must be
repaired by authorized personnel.
1) Ensure that the Mode selector is in the OFF (center) position.
2) Connect the regulator to a vacuum source (normally the wall-mounted inlet).
3) Occlude the regulator inlet, and turn the adjustment knob one full turn
clockwise.
4) Verify that the gauge needle does not move from the “0” position.
5) Move the Mode selector to the “REG” position.
6) Occlude the regulator inlet, and turn the adjustment knob counter-clockwise
until the gauge needle is at 0.
7) Keeping the inlet occluded, turn the adjustment knob one full turn clockwise.
The gauge should read between 120 and 170 mmHg.
8) Keeping the inlet occluded, turn the adjustment knob fully clockwise. The
gauge needle should move clockwise to the highest vacuum setting.
9) Move the Mode selector to the “FULL” position.
10) Occlude the regulator inlet. The gauge needle should move to the highest
vacuum setting.
11) Turning the adjustment knob should have no effect on the gauge reading when
the selector knob is in the FULL position.
12) Move the Mode selector to the “OFF” position. The needle should return to 0
If the regulator passes all of the above, it is ready for patient use. Failure of any of
the above requires that the unit be serviced by authorized personnel.
Once the operation of the regulator has been verified, it can be prepared for patient
use as follows:
1) Connect the regulator to the wall inlet, ensuring proper latching.
2) Move the Mode selector to “REG”.
3) Occlude the inlet or tubing.
4) Adjust the regulator knob to achieve the desired level of vacuum.
5) Attach the suction catheter and proceed.
NOTE: the suction regulator should always be turned OFF when not in use. This
3
882VR Disassembly Instructions
1) Disconnect the 882VR from the suction source
2) Sterilize the suction regulator using ETO or autoclave, for self-protection
3) Remove the four screws on the back of the unit
4) Turn the regulator adjustment knob counterclockwise to release the faceplate
5) The gauge can be gently lifted out of the socket
6) The regulator can be gently lifted out of the socket
7) One screw holds the Mode selector lever in place.
Reassembly is accomplished by performing the above steps in reverse order.
The following repair assemblies can be purchased from your Authorized GENTEC
distributor (refer to the exploded view on the next page):
Possible Back
Fittings
Ohio/Ohmeda
Chemetron
DISS Nut & Nipple
DISS Hand-Tight
Puritan-Bennett
Hose Barb
Possible Inlet Fittings
Vacuum Trap Assembly or Fitting Only
Hose Barb DISS Male
Figure 2 - Suction Regulator Configurations
DO NOT connect suction tubing directly to the vacuum trap, filter or suction regulator!
Vacuum Filter
4
The regulator can be cleaned by wiping the outside surfaces with disinfectants,
Virox (accelerated hydrogen peroxide), Clorox (sodium hypochlorite), and Cavi-
wipes. The internal flow path can be cleaned by suctioning a cold sterilant
through the unit, then allowing the unit to run for 30-45 minutes to dry the
interior.

No. Part No. Description No. Part No. Description
1 300660110 Screw Kit (Set of 5) Shaft
2 200260010 Back Plate Spring
3 203060040
Mode Selector
Assembly 6 204880020
Filter
Assembly
4 203060010 Regulator Assembly Protecting Cap
5 204880010
Adjustment Knob
Assembly Sponge Filter
Logo Sticker Orifice Union
Screw Orifice Seat
Knob 7 141742301 Gauge
Washer 8 304360010 Relief Valve
Spring 9 304660010 Main Cover
C-Clip
5
The FULL mode bypasses the regulator, and is designed for use only when high
levels of suction are needed. Care should be exercised when using this setting, as
the patient will be exposed to whatever level of suction is available at the wall inlet
(typically above 480mmHg).
A suction filter or vacuum trap assembly (GENTEC catalog #880VT) should be
used to prevent aspirate from entering the suction regulator. Typically, the suction
catheter is connected to the suction tubing, which is then connected to the inlet
fitting on the suction canister.
The canister can be connected directly to the suction regulator via DISS
connection, or, as is recommended, connected to a filter or vacuum trap, which is
then connected to the suction regulator via direct, threaded connection, suction
tubing, or DISS connector (see Figure 2).
The appropriate outlet connector (located on the back of the suction regulator)
must be used for connection to the wall inlet. The use of converting adapters (e.g.,
DISS connection to Ohio connection) should be avoided. If the suction regulator is
connected via tubing or hose assembly to the wall inlet (as occurs when the
suction regulator is attached to a mobile stand), a minimum inside hose diameter
(ID) of 5/16” (7.9mm) should be used to prevent loss of flow.
DO NOT connect the suction tubing directly from the patient to either the
Mode Selector
Gauge
Regulator
Adjustment Knob Outlet (in back)
Inlet (on bottom)
Figure 1 - Main Features
2

IMPORTANT SAFETY INSTRUCTIONS.
READ AND UNDERSTAND THESE INSTRUCTIONS
COMPLETELY BEFORE OPERATING THIS
EQUIPMENT.
If you do not understand any of these instructions, or if you have any
questions regarding the use of this product, please contact your
facility’s training manager, your supervisor, the medical equipment
dealer from whom the product was purchased, or the manufacturer
before operating the equipment.
Do not attempt to repair this device if you have not been properly
trained. Doing so may create a hazardous situation that may result in
death or serious injury. Attempted repair by anyone other than a duly
authorized repair/service center of Genstar Technologies Co., Inc. voids
any and all warranties, express or implied.
Carefully inspect and test this product before each use to ensure proper
operation. Do not use the product if there are signs of damage or if it
does not pass the initial suction test.
Should this product require repair or service that will require shipping
the product to another location, bear in mind that United States
Federal law prohibits shipping contaminated products, as they
constitute a biohazard. This product must be properly sterilized and
securely packaged prior to shipping.
Genstar Technologies Co., Inc. (GENTEC®) manufactures continuous suction
regulators in two ranges, 0 to 300mmHg (882VR-300) and 0 to 760mmHg (882VR-
760). These suction regulators provide three modes: REG (regulated vacuum), OFF
(no vacuum) and FULL (whatever is available at the wall inlet). Please take a few
minutes to familiarize yourself with the product by reviewing Figure 1 on the next
page.
The mode is selected by moving the lever at the top of the regulator to the left
(REG) for regulated suction, the center (OFF) to turn off the regulator, or the right
(FULL) to allow full line suction.
The REG mode provides regulated suction levels as set by the user. The suction
level is set by occluding the suction tubing, then adjusting the regulator knob on
the front of the suction regulator to achieve the desired suction level, up to the
designed range. Suction is increased by turning the regulator knob clockwise,
decreased by turning it counter-clockwise.
The OFF mode turns off the suction regulator, allowing no suction at the tubing.
1 6
# Problem Probable Cause Corrective
Measures
1
The vacuum gauge
needle does not move
off “0” when regulator
is connected to
vacuum source.
The regulator is in the “OFF”
mode, or is not fully in the
“REG” or “FULL” position.
Move the mode
selector to the
“REG” or “FULL”
position.
The adjustment knob is in the
fully counter-clockwise, closed,
position.
Turn the
adjustment knob
clockwise to open.
The collection bottle or suction
tube is leaking.
Check the
collection bottle
and tubing for leaks.
2
The vacuum regulator
gauge shows a reading,
but there is no suction
at the tubing.
The collection bottle is too full,
causing the float to shut off the
suction.
Empty the
collection bottle.
The filters or suction tubing are
clogged.
Change the filters
and tubing.
3
Under REG mode the
regulator cannot reach
the specified suction
level (i.e., 200mmHg or
760mmHg, depending
upon model).
Suction source cannot provide
sufficient suction.
Increase source
equipment vacuum
settings.
Regulator
Internal
Problems
Relief valve (8) is
damaged. Replace relief valve.
O-ring on vacuum
gauge (7) is
damaged.
Replace o-ring.
O-ring in regulator
assembly (4) is
damaged.
Replace o-ring.
Regulator assembly
(4) is damaged.
Replace regulator
assembly.
Mode selector
assembly (3) is
loose.
Tighten or replace
mode selector
assembly.
4
Under REG mode the
suction is too strong
and cannot be reduced.
Clean gas assist
port.
Gas assist port (6) is clogged.

Other Gentec Controllers manuals