GEO Calibration HYGRO MAX User manual

GEO Calibration Inc.
2190 Smithtown Avenue
Ronkonkoma, NY 11779
USA
HYGRO MAX
User Manual
English
Deutsch
Francais

2
TABLE OF CONTENTS
INTRODUCTION .............................................. 4
Product Warranty.................................................................................................................5
Unpacking Instructions .....................................................................................................6
Calibrator Applications......................................................................................................6
Standard Packing Check List Table ..............................................................................7
Visual Item Check List .......................................................................................................8
Accessories Bag Contents...............................................................................................9
Available Accessories........................................................................................................10
Available Accessories Tables..........................................................................................11
Unit Diagram And Parts Listing......................................................................................13
QUICK START GUIDE..................................... 15
Filling The Reservoir ..........................................................................................................16
Powering The Unit...............................................................................................................18
Unit Operation (PID Controller)..........................................................................19
Unit Operation (Main Display) ............................................................................21
Opening The Door ...............................................................................................22
Error Codes And Descriptions........................................................................................23
General Specications.......................................................................................................24
Consumables ........................................................................................................................25
Mechanical And Elecreical ...............................................................................................25
CALIBRATION.................................................. 26
UUTs (Units Under Test)....................................................................................................26
System Recalibration..........................................................................................27
System Uncertainty .........................................................................................27
Auto-Calibration | PC Software ......................................................................27
Reference Standard Recalibration.....................................................................28
Control Probe / Reference Sensor.................................................................28
Entering Hygro Max Osets ...............................................................................30
Unit Calibration Overview...............................................................................30
Accessing the Device Manager......................................................................30
Reading the COM Port ....................................................................................31
Installing PuTTY ..............................................................................................31
PuTTY Setup ....................................................................................................31
Connecting Through PuTTY...........................................................................32
Reading Temperature and Humidity Osets.................................................32
Changing Humidity and Temperature Osets ......................................................33
SAFETY WARNING ......................................... 34
General Safety Information ..............................................................................................34
Disposal Safety Information ............................................................................................35
TECHNICAL SUPPORT .................................. 36

3
REPAIRS........................................................... 37
Unit Repair Procedure.........................................................................................37
MAINTENANCE................................................ 38
Service Schedule .................................................................................................................38
Replacing The Desiccant ..................................................................................................39
Condensation........................................................................................................................41
Draining The Unit.................................................................................................................42
Probe Removal .....................................................................................................................43
Probe Replacement.............................................................................................................44
HYGRO MAX DESICCANT SYSTEM............. 45
System Overview .................................................................................................................45
Desiccant - LED Status Indications...............................................................................45
GEO Calibration Inc
2190 Smithtown Avenue, Ronkonkoma, NY 11779, USA
Tel.: +001 (631) 471 - 6157 ● Fax: +001 (631) 471 - 6158
[email protected] ● www.geocalibration.com

4
INTRODUCTION
MESSAGE FROM GEO CALIBRATION
Thank you for purchasing the GEO Calibration Hygro Max humidity and temperature generator/
calibrator. We look forward to providing you the highest quality technical support as you become
familiar with your new humidity and temperature calibrator.
To better familiarize yourself with the Hygro Max, please visit our Youtube Channel by searching
GEO Calibration for user friendly videos.
To start using your GEO Hygro Max immediately, you may proceed to our Quick Start Guide (page
16).
For a deeper review of the Hygro Max, See our Calibration Technical Recommendations (page
25).
Regards,
GEO Calibration Service Team
WARNING As you read through this product manual, please familiarize yourself with
our recommended best practices. By following the proper procedures,
you will ensure your unit consistently performs to its highest potential.
Once you have removed the Model Hygro Max from its external
packaging, please visually inspect the unit for damage. If damage is
found, please immediately contact your supplier.

5
PRODUCT WARRANTY
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each GEO Calibration product is warranted to be free from defects in material and workmanship under
normal use and service. The warranty period is one year and begins on the date of shipment. Parts,
product repairs, and services are warranted for 90 days. This warranty extends only to the original
buyer or end-user customer of a GEO Calibration authorized reseller, and does not apply to fuses,
disposable batteries, desiccants, distilled water, or to any product which, in GEO Calibration’s opinion,
has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions
of operation or handling. GEO Calibration does not warrant that software will be error free or operate wi
thout interruption.
GEO Calibration authorized resellers shall extend this warranty on new and unused products to end-
use customers only but have no authority to extend a greater or dierent warranty on behalf of GEO
Calibration. Warranty support requires activation and registration at:
https://www.geocalibration.com/register
and is purchased through a GEO Calibration authorized sales outlet or Buyer has purchased unit
directly from GEO Calibration. GEO Calibration reserves the right to invoice Buyer for importation
costs of repair/replacement parts when product purchased in one country is submitted for repair in
another country. GEO Calibration’s warranty obligation is limited, at GEO Calibration’s option, to
refund of the purchase price, free of charge repair, or replacement of a defective product which is
returned to a GEO Calibration authorized service center within the warranty period.
To obtain warranty service, contact your nearest GEO Calibration authorized service center to obtain
return authorization information, then send the product to that service center, with a description of the
diculty, postage and insurance prepaid (FOB Destination). GEO Calibration assumes no risk for dam-
age in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid
(FOB Destination). If GEO Calibration determines that failure was caused by neglect, misuse, con-
tamination, alteration, accident, or abnormal condition of operation or handling, including over voltage
failures caused by use outside the product’s specied rating, or normal wear and tear of mechanical
components, GEO Calibration will provide an estimate of repair costs and obtain authorization before
commencing the work. Following repair, the product will be returned to the Buyer, shipped Ex Works
FOB Suolk County NY.
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WAR-
RANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. GEO CALIBRA-
TION SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion
or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may
not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or
other decision-maker of competent jurisdiction, such holding will not aect the validity or enforceability
of any other provision.
GEO Calibration Inc.
2190 Smithtown Avenue
Ronkonkoma, NY, 11779
U.S.A.

6
Once you have removed the Hygro Max from its external packaging, please visually inspect the unit
for damage. If damage is found, please immediately contact your supplier.
CALIBRATOR APPLICATIONS
UNPACKING INSTRUCTIONS
The GEO Hygro Max Humidity Calibrator generates and maintains a controlled humidity and tem-
perature environment for the purpose of testing or calibrating humidity and temperature sensors, also
known as hygrometers. The humidity calibration range is 1% up to 99%, while the temperature range
is from 0°C to 60°C.
See the full technical specications (page 25).
The system’s accuracy is achieved via a NIST traceable internal control and reference probe.
The Hygro Max can calibrate many hygrometer types:
• Probes
• Data-loggers
• Chart Recorders
• Additional Assorted Hygrometers
Please review the dimensions of the calibration chamber to ensure compatibility with the size require-
ments of any hygrometers you intend to calibrate. You may require an Expansion Chamber. Your
GEO representative should have reviewed this with you at time of purchase, if not, please contact our
support team and they will size an Expansion Chamber accordingly.
The Hygro Max allows for multiple types and sizes of hygrometers, for a complete list please contact
GEO Calibration Inc
2190 Smithtown Avenue, Ronkonkoma, NY 11779, USA
Tel.: +001 (631) 471 - 6157 ● Fax: +001 (631) 471 - 6158
[email protected] ● www.geocalibration.com

7
STANDARD PACKING CHECK LIST
Contents of GEO Accessories Kit
QTY Part Number Description
101-200-82-0001 GEO Accessories Bag
101-450-36-0003 Fill Syringe (50ml)
201-200-85-0001 4 AMP Fuses (1 pack)
101-200-46-0001 Mains Power Cord
101-250-07-0005 Desiccant Tool
102-053-07-0006 Shelf
101-410-01-0045 Hygro Max Bung Kit
101-040-01-0010 Spare Ultrasonic Transducer Replacement
101-450-01-0045 Heating Element
Within the Shipped Case
QTY Part Number Description
101-410-00-0000 Hygro Max Humidity Calibrator
601-250-11-0001 Desiccant Canister (3 Spares)
101-200-36-0002 Reference Standard / Control Probe
101-410-01-0044 GEO Accessories Kit
101-410-01-0015 Hygro Max Door (10 Port Door with Plugs)
101-400-01-0021 Hygro Max Chilled Mirror Door
101-400-82-0001 Hygro Max Accessories Kit
1N/A Manilla Envelope with Calibration Docs
101-410-01-0003 Hygro Max Chamber Insert
101-011-01-0002 1850-2 Chilled Mirror (Optional)
101-011-01-0001 Chilled Mirror Sensor with Cable (Optional)
101-410-36-0006 Chilled Mirror Temperature Sensor (Optional)
Calibration Documents
QTY Part Number Description
1NO REORDER Factory Calibration Report
101-999-99-0001 3rd Party Calibration Certication (IF ORDERED)

CALIBRATION
Hygro Max Accessories
Kit
Contents:
See Page 9
P/N: 01-200-01-0035
M4000 Chamber Insert
P/N: 01-400-01-0023
Hygro Max Chilled Mirror
Door
P/N: 01-400-01-0021
8
VISUAL ITEM CHECK LIST
Listed below are standard contents included with the purchase of a new Hygro Max.
Humidity Generator Desiccant Canister
Calibration Documents
GEO Calibration
Hygro Max
P/N: 01-400-00-0000
Pre-Filled with molecular sieve
P/N: 01-250-11-0001
Control Probe
HC2-S HygroClip
control probe (pre-installed)
P/N: 01-200-36-0002
Hygro Max Door
10 Port Door with Plugs
P/N:01-410-01-0015
1850-2 Chilled Mirror
Chilled Mirror Sensor
Chilled Mirror Temperature Sensor
(Optional)

Desiccant Tool
P/N: 01-250-07-0005
Hygro Max Bung Kit
P/N: 01-410-01-0045
4 AMP Fuse
P/N: 01-200-85-0001
Fill Tube
(50ml)
P/N: 01-450-36-0003
ACCESSORIES KIT CONTENTS
Shelf
P/N: 02-053-07-0006
Power Cord
P/N: 01-200-46-0001 US
01-200-46-0002 EUR
Spare Ultrasonic
Transducer Replacement
P/N: 01-250-07-0005
Heating Element
P/N: 01-450-01-0045 9

10
Replacement
Desiccant
P/N: 01-250-11-0001
PID Temperature
Controller
P/N: 01-012-00-0000
Chilled Mirror
Internal Door Insert
P/N: 01-400-01-0021
Replacement
Control Sensor
HC2-S HygroClip
control probe
P/N: 01-200-36-0002
Silicone
Adapter Kits
P/N: 01-400-01-0032
Replacement
Fill Tube
(50ml)
P/N: 01-450-36-0003
AVAILABLE ACCESSORIES
ISO 17025 System
Recalibration
External Desiccant
Canister
Pre-Filled with molecular sieve
P/N: 01-450-01-0034 P/N:
01-999-99-0004

11
AVAILABLE ACCESSORIES
Humidity Generator
QTY Part Number Description
101-250-11-0001 Desiccant Canister
101-200-36-0002 Control Probe
101-400-01-0001 Hygro Max Accessories Kit
101-400-01-0023 Hygro Max Chamber Insert
101-200-46-0001 US Spec Power Cord / Mains Cable
101-200-46-0002 EUR Spec Power Cord / Mains Cable
101-450-36-0003 Fill Tube (50ml)
201-200-85-0001 4 AMP Fuse (Fast Acting)
101-410-01-0045 Hygro Max Bung Kit
AVAILABLE SERVICE
QTY Part Number Description
101-999-99-0001 ANAB ISO 17025 Calibration Certication

12
AVAILABLE ACCESSORIES
Grommets / Bungs
M36 are for Standard Hygrometers
PG36 are for Chilled Mirror Adapters
Part Number Type Bung Size Port Size
mm in mm in.
01-004-09-0001
Regular M36 31.00 1.22”
0.000 0.000”
01-004-09-0002 3.175 0.125”
01-004-09-0003 6.350 0.250”
01-004-09-0004 9.525 0.375”
01-004-09-0005 12.700 0.500”
01-004-09-0006 15.875 0.625”
01-004-09-0007 19.050 0.750”
01-004-09-0008 22.225 0.875”
01-004-09-0009 25.400 1.000”
01-004-09-0020
Slimline M36 31.00 1.22”
3.175 0.125”
01-004-09-0021 6.350 0.250”
01-004-09-0022 9.525 0.375”
01-004-09-0023 12.700 0.500”
01-004-09-0024 15.875 0.625”
01-004-09-0025 19.050 0.750”
01-004-09-0026 20.950 0.825”
01-004-09-0027 25.400 1.000”
01-004-09-0034
Regular PG36 44.45 1.75”
00.000 0.000”
01-004-09-0028 12.700 0.500”
01-004-09-0029 15.875 0.625”
01-004-09-0030 19.050 0.750”
01-004-09-0031 25.400 1.000”
01-004-09-0032 31.750 1.250”
01-004-09-0033 34.290 1.350”
01-004-09-0035
Slimline PG36 44.45 1.75”
9.525 0.375“
01-004-09-0036 15.875 0.625“
01-004-09-0037 25.400 1.00“

13
Below you will nd a diagram of the Hygro Max’s various operational parts.
FRONT
REAR
Probe Port
Cooling Fan
Heat Sync
External Desiccant Port
Desiccant
Mains Power
Master Power
PC USB Inputs
Primary Display
Control panel
Door Latch
Drain Port
UNIT DIAGRAM AND PARTS LISTING
Chilled Mirror Ports

14
SIDE
Chilled Mirror
(Optional)
Water Fill Level
Indicator
Drain Port
Fill Port

15
QUICK START GUIDE
This guide is intended to give you the necessary information to quickly get up and running with your
Hygro Max humidity and temperature generator.
It includes the following instructions:
• Filling the Reservoir
• Powering On the Unit
• Operating the Front Panel Controller
• Operating the Main Display
• Opening the Door

16
Turn on the unit and set the temperature to 30 °C and
humidity to 30 % to let the unit settle, approximately for 20
minutes (rst time only).
After receiving the unit, open the door to let the chamber dry
for 10 minutes. Use soft paper tower to wipe the water in the
chamber if needed.
PLEASE NOTE:
1. Do not use alcohol inside the chamber.
2. The above instruction must be followed to ensure unit is dry
from any changes during shipping.
1.
2.
Read Before Turning On The Unit

17
3.
The port is located on the center top area of the right panel
of the unit.
Attach Fill Syringe to Fill Port
Press the ll syringe tip into the ll port, then rotate the cap
clockwise to secure.
1.
Remove the Fill Cap
Rotate Counter-Clockwise to remove.
2.
Supplies Needed
Fill Syringe
Locate the Fill Port on the right
side panel
- Pour distilled water into the elevated tube body.
- Take care not to insert any air into the reservoir.
- Monitor the overow while lling.
- If water exits the overow, then you have reached the
maximum ll level.
4. Elevate and Fill Syringe
FILLING THE RESERVOIR
Distilled Water
Only

DO NOT BLOCK OR PUT WATER IN VENT HOLE.
DO NOT PUT WATER IN “IN” OR “OUT” PORT.
PLEASE NOTE:
DO NOT FILL WATER ABOVE MAXIMUM FILL LINE.
Max.Fill Line
18
Loosen and Remove Syringe
Turn the ll syringe tip counter-clockwise to loosen.
5.
Re-Install the Fill Cap
Turn the cap clockwise to tighten the cap and seal the ll
port.
6.
PLEASE NOTE:
FILLING THE RESERVOIR
External Dryer“In” Port
External Dryer“Out” Port

19
Locate the Power Input
1.
Set Power Switches to “ON”
4.
Plug Power Supply into Unit
3.
Plug Power Supply into Wall
2.
Supplies Needed
Power Cord
POWERING THE UNIT

20
Boot Screen
Manufacturer Name
Model Number
Serial Number
Firmware Version Number
1.
Observe the Display
Current Temperature Top Right
Current Relative Humidity Middle Right
Programmed Temperature Top Left
Programmed Relative Humidity Middle Left
Calculated Dew Point Bottom Left
2.
UNIT OPERATION (PID CONTROLLER)
Raise the Desired Set-point
Pressing the “UP” key will raise the selected set-point.
Holding the “UP” key will rapidly raise the selected set-
point.
4.
Changing Humidity & Temperature
Pressing the “Next” button activates the set-point
interface. The selected eld will repeatedly ash its current
programmed value. Repeated presses will toggle
between the Humidity and Temperature elds.
3.
Lower the Desired Set-point
Pressing the “DN” key will lower the selected set-point.
Holding the “DN” key will rapidly lower the selected set-
point.
5.
Table of contents