Fixing the data logger
Fix the HD33LMT.GSM data logger to the
lower part of the tripod central mast by
using the bracket provided on the back of
the data logger box.
A SIM card must be inserted in the data
logger, as described in the “SIM CARD”
section of the data logger user manual.
HD52.3D… installation
•Pass the HD52.3DT147 (HDMCS100) or
HD52.3DP147 (HDMCS200) connection
cable inside the tripod central mast.
The cable has 19-pole M23 connector
on the sensor side and 8-pole M12
connector on the data logger side.
•Connect the cable to the M23 connector
at the bottom of the sensor. Ensure
connection stability by tightening the
connector external nut.
•Fix the sensor on the top of the mast
by tightening the cable tie at the
bottom of the sensor housing.
•Connect the other end of the cable to
the 8-pole M12 connector at the bottom
of the data logger.
Warning: The sensor should be installed vertically and in an open area. The pyranometer
should not be obscured or overshadowed. The rain gauge should be exposed to open air and
away from items that create turbulences.
Fixing the solar panel
•Secure the solar panel to the tripod by
using the supplied support. The panel
should face SOUTH if the installation is
in the Northern Hemisphere and
NORTH if the installation is in the
Southern Hemisphere.
•Adjust the inclination of the panel
according to the place of installation.
•Pass the panel connection cable
through the cable gland at the bottom
of the data logger.
Power supply
•Connect the cables of the data logger internal
battery to the +BATT (1) and GND (2) terminals.
•Connect the solar panel to the +Vpanel (4) and
GND (5) terminals of the data logger.
Pay attention to the correct polarity of the
connection.
7
6
5
4