IEC 60318
Calibration
Adapter for
AL0011 IEC 60711
Ear Simulator
for calibration
(DUT)
RA0067
Transmitter
Adapter
Swept sine
signal from
generator
(1 V RMS
max.)
Conditioned
output signal
to analyser
Power Module Type 12AK
Used to condition the signal
from the reference micro-
phone in the IEC 60318 Ear
Simulator
Electrostatic Actua-
tor Amplifier Type
14AA
Used to drive the
microphone as
sound source
AA0008
Cable
AA0033
Cable
AA0033
Cable
AA0033
Cable
G.R.A.S. Sound & Vibration
Coupler Calibration Stand AL0011 - Page 7
Additional parts (not included in AL0011):
• Transmitter amplifier: Electrostatic Actuator Amplifier Type 14AA.
• Transmitter Adapter for prepolarised microphone: Transmitter Adapter RA0067.
• Calibration Kit: IEC 60318 Calibration Adapter for AL0011, RA0183 (incl. Stop Washer
GR1131 and Sleeve GR0371).
• Sound source: ½˝ Prepolarised, Pressure-field Microphone Type 40AO.
• Reference microphone for measuring coupler response: ½˝ Pressure-field Microphone
Type 40AG.
• Preamplifier for reference microphone: ½˝ Preamplifier Type 26AK.
• Signal conditioning and power supply: Power Module Type 26AK
Cables:
• 1 × 7-pin LEMO to LEMO microphone cable (3 m) AA0008.
• 2 × BNC to BNC Coax.cable (1 m) for Power Module Output and Electrostatic Actuator
Amplifier Input (AA0033).
Fig. 7 Showing the AL0011 with device under test and additional G.R.A.S. equipment. Please note:
max. input to Type 14AA is 1 V RMS