
10 Chapter 4: Installation and Operation GWMT Transmitter Manual
If entering the value directly, enter the value using the numeric keys, including the decimal point
and fraction (KHz), followed by the Enter key. The display will automatically adjust your entry to
the nearest valid frequency setting that is less than or equal to the value entered.
If scrolling, press the up or down arrow keys repeatedly until the desired value is obtained. The
display will automatically increment (decrement) to the next (previous) valid frequency each time
one of the keys is pressed.
Note: The entry of an out-of-band or invalid frequency value will result in the display
returning to the previous valid entry.
Valid Frequencies
The selection of valid frequencies is dependent on the frequency range and the minimum
frequency tuning step for the specific transmitter. For example, the RF 1900 PCS transmitter can
implement any frequency from 1850 to 1990 MHz provided that it is an integer multiple of either
30KHz or 50 KHz (the minimum frequency tuning step). If an invalid frequency such as 1860.02
MHz is entered, the frequency synthesizer will automatically adjust the entry to 1860.00 MHz
which is the nearest valid frequency that is less than or equal to the value entered.
The value 1860.02 MHz is invalid since 1860.02 MHz divided by 50 KHz (1860020 / 50 =
37200.4) results in a value that is not an integer and 1860.02 MHz divided by 30 KHz (1860020 /
30 = 62000.666) also results in a value that is not an integer. The value 1860 MHz is valid since
1860 MHz divided by 50 KHz (1860000 / 50 = 3720) results in an integer. The nearest valid
frequencies above 1860 MHz are 1860.03 MHz, which is an integer multiple of 30 KHz and
1860.05 MHz which is a multiple of 50 KHz. Refer to Chapter 5, Performance Specifications, for
the valid frequency range and minimum frequency tuning step for specific transmitters.
Amplitude Control
The desired amplitude can be entered either directly using the numeric keys, or it can be scrolled
to by using the up and down arrow keys. The keypad sequence is as follows:
Push the AMP (amplitude) button. Notice that the flashing cursor moves to the left-most digit
of the amplitude value.
If entering the value directly, enter the desired amplitude from 13 to 43 dBm, by using the
numeric keys, including the decimal point and fraction, followed by the Enter key.
If scrolling, press the up or down arrow keys repeatedly until the desired value is obtained.
The display will automatically increment (decrement) by 0.1 dB each time one of the keys is
pressed.
Note: The entry of an amplitude value that is below 13 dBm or above 43 dBm will
result in the display returning to the previous valid entry.
Modulation Control
Not currently implemented
Transmit control
The RF Transmitter toggles on and off with each pressing of the Transmit key. Prior to turning
on the transmitter, the unit checks PA temperature and the Battery Voltage, and will not allow the
transmitter to turn on if either is out of range. The unit also checks the Phase-locked Loop to
ensure that the Synthesizer is locked, and will not allow the transmitter to turn on if it is not. After
turning on the transmitter, the unit checks the level control to make sure the transmit level is
within spec, and will turn off the transmitter if it is not. The RF entry on line 3 of the display is