GREISINGER GDUSB 1000 User manual

S56.0.01.6C-04 page 1 of 7
Installation- and operating manual
GDUSB 1000
Universal interface adapter for
GMSD/GMxD-sensors
1 General
The GDUSB1000-adaptor makes it possible to connect a standard pressure sensor of the type
GMSD/GMXD directly to the USB-port of a personal computer. Two operation modes are available:
Standard-mode: (Master-Slave)
In this mode the GDUSB1000 acts like a portable measuring instrument of the GMH3xxx-series and
gives back the wanted measurement on request (measurand, maximum or minimum data).
The communication can occur with the standard GMH/EASYBUS-software (e.g. EBS9M) and offers the
possibility to monitor long-run measurements of pressure for example.
Fast-mode: (autom. sending)
In this operating mode after a fulfilled trigger condition the GDUSB1000 sends the measuring data
automatically in the adjusted cycle time.
This mode is especially adapted for the analysis of pressure curves and the survey of pressure peaks
(with pre-trigger)
GREISINGER
electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: +49 9402 / 9383-0, Fax: +49 9402 / 9383-33, www.greisinger.de

S56.0.01.6C-04 page 2 of 7
Index
1GENERAL .......................................................................................................................................1
2SAFETY REGULATIONS ................................................................................................................2
3SPECIFICATION .............................................................................................................................3
4DISPOSAL INSTRUCTIONS ...........................................................................................................3
5OPERATING MODES......................................................................................................................4
5.1 Standard - Mode........................................................................................................................4
5.2 Fast - Mode ...............................................................................................................................4
6GENERAL HANDLING INSTRUCTIONS ........................................................................................5
6.1 To connect/change sensors.......................................................................................................5
6.2 Altitude correction of the absolute pressure sensor ...................................................................5
6.3 Zero correction of the pressure sensor ('OFFS')........................................................................5
6.4 Slope correction of the pressure sensor ('SCAL') ......................................................................5
6.5 Pressure connection to the pressure sensor..............................................................................5
6.6 Advice for calibration service .....................................................................................................5
7DRIVER INSTALLATION.................................................................................................................6
7.1 Windows Vista™, Windows XP and Windows 2000..................................................................6
7.2 Windows 98SE (Version V4.28a) ..............................................................................................6
8FREQUENTLY ASKED QUESTIONS (FAQ)...................................................................................7
8.1 How to check whether the adaptor has been installed correctly and via which COM-port
can I approach the installed adaptor? .......................................................................................7
8.2 How to change the COM-port of the adaptor? ...........................................................................7
8.3 Is there an automatic update function that updates the installed driver to the latest version? ....7
8.4 How to find out the used version of the driver?..........................................................................7
2 Safety regulations
This device was designed and tested considering the safety regulations for electronic measuring de-
vices. Faultless operation and reliability in operation of the measuring device can only be assured if the
General Safety Measures and the devices specific safety regulation mentioned in this users manual are
considered.
1. Faultless operation and reliability in operation of the measuring device can only be assured if the
device is used within the climatic conditions specified in the chapter “Specification“.
2. By faulty handling of the interface adapter a damage of the device can not be excluded. Likewise a
damage of your attached device is possible. In such case we cannot assume any warranty.
The manufacturer will not assume any warranty for damages to other devices resulting by use of the
interface adapter.
3. Standard regulations for operation and safety for electrical, light and heavy current equipment have
to be observed, with particular attention paid to the national safety regulations (e.g. VDE 0100).
4. When connecting the device to other devices (e.g. the PC) the interconnection has to be designed
most thoroughly, as internal connections in third-party devices (e.g. connection of ground with pro-
tective earth) may lead to undesired voltage potentials.

S56.0.01.6C-04 page 3 of 7
5. The device must be switched off and must be marked against using again, in case of obvious mal-
functions of the device which are e.g.:
- visible damage
- no prescripted working of the device
- storing the device under inappropriate conditions for longer time
When not sure, the device should be sent to the manufacturer for repairing or servicing.
6. Attention: Do NOT use this product as safety or emergency stopping device, or in any other appli-
cation where failure of the product could result in personal injury or material damage. Failure to
comply with these instructions could result in serious injury and material damage.
3 Specification
Measuring range: corresponding to used pressure sensor
Max. range: -19999 ... 19999 digit
Resolution: corresponding to used sensor
Pressure units: mbar, bar, kPa, MPa, mmHg, PSI, mH
2
O
can be selected corresponding the used sensor
Accuracy: (typ.) ±0,2 % FS
(at nominal temperature)
Measuring cycle: 1 ms
Connection:
PC: standard USB-plug connector (type A)
GMSD, GMxD: 6-pin screened lockable Mini-DIN-socket.
Automatic sensor detection and setting of measuring range upon plugging
in of the sensor.
Power supply: via USB-interface
Operating conditions: -25 ... 50 °C, 0 ... 95 %RH
(non condensing)
Nominal temperature: 25 °C
Storage temperature: -25 ... 70 °C
Housing:
Dimensions: approx. 56 x 31 x 24 mm
Cable length: approx. 30 cm
Weight: approx. 40 g
EMC: The device corresponds to the essential protection ratings established in
the Regulations of the Council for the Approximation of Legislation for the
member countries regarding electromagnetic compatibility (2004/108/EG).
In accordance with EN 61326
:1997
+ A1
:1998
+ A2
:2001
(app. B, class B)
4 Disposal instructions
The device must not be disposed in the regular domestic waste.
Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the de-
vice appropriate and environmentally sound.

S56.0.01.6C-04 page 4 of 7
5 Operating modes
After restart the adapter is in the standard-mode.
5.1 Standard - Mode
In this mode the GDUSB1000 acts like a portable measuring instrument of the GMH3xxx-series.
It gives back the wanted measurement on request.
For this purpose 4 measuring channels are allocated at the adresses 1 – 4:
1 Measurand: with adjustable low-pass filter (0.00 ... 10.00)
2 Average: the average between two measuring-points
3 Minimal peak: lowest measurand for the last request; unfiltrated, with highest sample rate
4 Maximal peak: highest measurand for the last request; unfiltrated, with highest sample rate
The communication can occur with every standard software for the GMH3xxx or EASY
BUS
(e.g.
EBS9M) and offers the possibility to monitor long-run measurements of pressure for example.
For the integration in your own programs, we offer you the function-library EASYBUS.DLL.
This operating mode is especially adapted for the measuring of 1 measurand per second.
For example: long-run measurements of pressure, leakage test,...
Note: for use the GDUSB1000 you must select in the software the interface USB3100.
5.2 Fast - Mode
The device can be set to the mode “fast measurand display” via command. In this mode the measure-
ment is started by a trigger condition. Then the device automatically gives back the measurands in the
adjusted cycle time without any request, until the stop condition is fulfilled.
Adjustable parameters are among other ones:
Cycle time: 1ms ... 10sec.; programmable in steps of ms
Measuring channel: cycle < 5 ms it is possible to choose the recording channel (measurand, min.
peak, max. peak)
cycle > 5 ms all four measuring channels are recorded automatically
Trigger conditions: 5 for the measuring start
6 for the measuring stop
Pre-trigger: 50 data sets (at a cycle time of ≥5ms) respectively 200 measurands (cycle
time < 5 ms)
For the communication and configuration of the device in this mode the software GSOFT-USB is nec-
essary.
This mode is especially adapted for measuring very fast pressure changes (1000 per second)
For example: detecting pressure peaks at switching operations, ....

S56.0.01.6C-04 page 5 of 7
6 General handling instructions
6.1 To connect/change sensors
Only sensors of the GMSD or GMXD-serial are allowed to be used! If other sensors are used the adapter
and the sensor may be destroyed.
To change the sensor you have to disconnect the adapter. Sensors have to be connected before acti-
vating the adapter. Otherwise they are not found by the adapter correctly.
During the connecting it is possible that the connector does not snap in the jack correctly. In this case you
have to hold the connector at the anti-kink device not at the casing. Don`t plug the connector wedged. An
accurate attached jack can be easily plugged in without great effort. To disconnect the sensors it is not
allowed to pull at the cable. In this case you have to hold the connector at its casing.
6.2 Altitude correction of the absolute pressure sensor
The device measures the absolute pressure of the ambient air. But this pressure is not to be confused
with the “pressure at sea level”, which is quoted by meteorological stations!
At this indication the decrease of the air pressure which is dependent on the level will be calculated. The
device has the ability to do this air pressure height adjustment.
Activate for this the “sea level function” via the configuration software (GSOFT_USB) and specify the
height beyond the sea level of your habitation.
6.3 Zero correction of the pressure sensor ('OFFS')
For the measurement a zero offset can be conducted:
shown vale = measured value - offset
The zero correction in combination with the slope correction (see below) are mainly used to adjust the
deviation of sensors. The entry has to be made in the configuration software. The offset information will
be stored in the sensor. So you can set it individually for each sensor.
6.4 Slope correction of the pressure sensor ('SCAL')
The measurement`s slope can be affected by this factor (factor in %):
shown value = measured value * (1 + scale / 100)
The slope correction in combination with the zero correction (see above) are mainly used to adjust the
deviation of sensors. The entry has to be made in the configuration software. The scale information will be
stored in the sensor. So you can set it individually for each sensor.
6.5 Pressure connection to the pressure sensor
The measuring adapter is conceived that way that all sensors of the GMSD/GMXD...-series can be con-
nected without new calibration. Therefore a multiplicity of fully exchangeable sensors for metering ranges
of 1.999...2.500mbar relative to 0...400.0 bar absolute for example are available.
Relative pressure sensors (type: GMSD/GMXD...MR, GMSD/GMXD...BR)
•For overpressure and respectively low-pressure measurements: fix the flexible hose with 4mm inner
diameter at the fitting "B”.
•For pressure difference measurements: fix both flexible hoses with 4mm inner diameter at the fittings
"A” and “B”, whereas the higher pressure has to be connected to the attachment “B”.
Absolute pressure sensors (type: GMSD/GMXD...BA)
Fix the flexible hose with 4mm inner diameter to the fitting “A” (“B” is without any function).
Stainless steel sensors (type: GMSD/GMXD...MRE, GMSD/GMXD...BRE, GMSD/GMXD...BAE)
For overpressure, low-pressure or absolute pressure measurements: screw the sensor in the winding
(G1/4”) or connect a flexible hose with an appropriate adapter to the pressure sensor.
6.6 Advice for calibration service
Factory calibration certificate – DKD-certificate – ministerial certifications:
If the measuring device shall have a calibration certificate you have to send it inclusive the sensor to the
manufacturer. Just the manufacturer can check the factory settings and correct them if necessary

S56.0.01.6C-04 page 6 of 7
7 Driver Installation
7.1 Windows Vista™, Windows XP and Windows 2000
For installation of the device drivers a user account with an administrator-authorisation is required. If
you haven’t an appropriate authorisation please consult your system administrator.
•
Insert driver disk into the drive of your PC.
•
Start the installation of the driver.
Path-example: E:\Treiber V5.3\Windows_2K_XP_S2K3_Vista\CP210XVCPInstaller.exe
E: means the letter of your drive.
•
Follow the instructions of installation.
•
Plug in the Device into the USB-connection.
•
Now the installation will be finished.
Now the Device is installed on your system and available for all users.
For application with our software a new additional COM-Port is available.
7.2 Windows 98SE (Version V4.28a)
Insert driver disk into the drive of your PC.
•
Plug in the USB-plug connector of the Device into the USB-connection of your PC.
•
The “Add New Hardware Wizard” for installation of
“CP2102 USB to UART Bridge Controller” appears. Click button “Next”.
•
Confirm the selection window for an appropriate driver with button “Next”.
•
In the following selection window choose “Specify a location”. In window
“Browse directory” select the path of your driver and push button “OK”.
Path-example: E:\Treiber V4.28a (Windows 98)\WIN
E: means the letter of your drive
•
Accept the suggested driver “CP210x USB Composite Device” by pushing button “Next”.
•
Now the installation will be partly finished. You must acknowledge with button “Finish”.
•
Now then the first part of the installation is completed, the hardware-assistant for installation of
“CP2102 USB to UART Bridge Controller” appears again.
•
Click on button “Next” and confirm the selection window for an appropriate driver with button
“Next”.
•
Choose the directory for your disk driver again, then click button “Next”.
•
Accept the suggested driver “CP210x USB to UART Bridge Controller” by pushing button
“Next”.
•
Now the installation will be finished. You must acknowledge with button “Finish” again.
Now the Device is installed on your system and available for all users.
For application with our software a new additional COM-Port is available.

S56.0.01.6C-04 page 7 of 7
8 Frequently asked questions (FAQ)
8.1 How to check whether the adaptor has been installed correctly and via which COM-
port can I approach the installed adaptor?
Start the Device-Manager: start \ settings \ control panel \ performance and maintenance \ system \ hard-
ware and open the submenu “Ports (COM & LPT)”
If the installation was successful, there will be the entry“CP210X USB to UART Bridge Controller (COM
x)”, in which the x after the Com indicates the available COM-port (i.e. COM 4).
Note: The adaptor will not be shown in the device manager, if it is not connected to the PC.
If the entry is occupied by an exclamation mark or the entry is not there, then the driver has not been
installed correctly.
8.2 How to change the COM-port of the adaptor?
Make sure that the adaptor is connected, start the Device-Manager start \ settings \ control panel \ per-
formance and maintenance \ system \ hardware) and open the submenu “Ports (COM & LPT)”
If the installation was successful, there will be the entry “CP210X USB to UART Bridge Controller (COM
x)”, in which the x after the Com indicates the available COM-port (i.e. COM 4).
Open the features of this connection by a double-click, go to “Port Settings” and then click there the
button “Advanced...”.
There you can choose the preferred COM-Port at the point “COM Port Number”. If this port is already
reserved for another (even not connected) device, a error message will appear:
“In principle it is possible that 2 devices are set on the same COM-port, but then you have to
take care, that not both devices are connected at the same time, because this would lead to a
conflict!“
8.3 Is there an automatic update function that updates the installed driver to the latest
version?
The driver is not updated automatically.
If there is already a suitable (for the operating system) driver during the installation, this driver will be
used automatically.
If the usages of the adapter cause problems please make sure that the latest version of the driver for
the adapter is used and installs the latest version manually if necessary.
8.4 How to find out the used version of the driver?
Make sure that the adaptor is connected and start the Device-Manager start \ settings \ control panel \
performance and maintenance \ system \ hardware).
For the COM-driver open the submenu “Ports (COM & LPT)”, open the features for this connection
“CP210X USB to UART Bridge Controller (COM x)” by double-click and change then to “Driver”.
There you can see the currently used driver version.

G38.0.01.6C-04
Operating Manual
Windows™ - Operating Software
GSOFT_USB
for USB measuring adapter GDUSB1000
as of version 1.3
Contents:
1 General Information ________________________________________________________________ 2
1.1 Software Operation _____________________________________________________________________2
1.2 Software License Agreement______________________________________________________________2
1.3 Software Registration ___________________________________________________________________2
1.4 System Requirements____________________________________________________________________2
1.5 Software Installation ____________________________________________________________________2
2 Operating Software GSOFT_USB _____________________________________________________ 3
2.1 Screen Elements ________________________________________________________________________3
2.1.1 Menu bar __________________________________________________________________________________ 3
2.1.2 Toolbar ___________________________________________________________________________________ 3
2.2 Data Recording_________________________________________________________________________3
2.2.1 Connecting the USB-Adapter __________________________________________________________________ 3
2.2.2 Sensor Data Window_________________________________________________________________________ 4
2.2.3 Start Recording _____________________________________________________________________________ 4
2.2.4 Reading in of Measurements ___________________________________________________________________ 5
2.2.5 Stop Recording _____________________________________________________________________________ 5
2.2.6 Comment __________________________________________________________________________________ 5
2.3 Measurement Diagrams__________________________________________________________________6
2.3.1 Adding Data Labels__________________________________________________________________________ 6
2.3.2 Adding Series ______________________________________________________________________________ 6
2.3.3 ‚Zoom‘: Enlarge a Area of the Display ___________________________________________________________ 6
2.3.4 Copying Data To Clipboard ___________________________________________________________________ 6
2.4 File Operation__________________________________________________________________________7
2.4.1 Export Files ________________________________________________________________________________ 7
2.5 How to Print Data ______________________________________________________________________7
This manual is subject to being changed

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 2 of 7
1 General Information
1.1 Software Operation
This operating manual postulates basic knowledge of how to handle a personal computer with the operating system
Windows™. If this is not existent it is recommended to make oneself familiar with the main features of Windows™
before installing this software (operating manuals, specialised literature, introductory courses etc.).
1.2 Software License Agreement
We kindly ask you to read the software license agreements on the disk package carefully. By opening the package
you agree to the stipulations mentioned.
1.3 Software Registration
Get your software registration now! It is profitable!
As a registered customer you will have the following advantages:
Technical support: Only our registered clients will receive support as far as technical questions on the programs
which are concerned.
Low-priced update offers: It is only our registered customers who may use our low-priced update offers.
Latest Information: We are enlarging our product range constantly. As a registered customer we will keep you
informed about our new and improved products. We want you to be up to date!
Your opinion counts: Let us know your wishes and recommendations.
Your opinion counts and may be used when we modify existing, or develop new products.
A blank registration form is included. Simply fax us the completed form.
1.4 System Requirements
IBM compatible PC ( Pentium ™ with GHz)
At least 20 MB hard disk capacity available
CD-ROM drive
At least 256 MB main memory
Windows™ 98SE, 2000, XP or Vista
mouse
One unassigned USB-port
1.5 Software Installation
Start Windows™
Insert CD-ROM in CD-ROM drive
Select “Execute“ in Menu “Start“
Enter “d:\setup“ - d stands for the letter of your CD-ROM drive –and acknowledge. In case your CD-ROM drive is
not ‚d‘, please use the corresponding letter.
Follow now the instructions given by the installation program.
A new program group with the following program icon will be created:
The software will be started by double clicking on this symbol.
Start program, choose the language you desire and the interface to which you have connected your interface
adapter.

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 3 of 7
2 Operating Software GSOFT_USB
GSOFT_USB permits the handling of an USB measuring adapter of the GDUSB1000 series. This adapter enables a
direct connection of a standard pressure sensor of the type GMSD/GMXD to the USB-interface of a personal com-
puter. With this dedicated software it is possible to start, stop, read in and display recordings. For visualisation and for
a better overview the recorded data can be shown in diagrams.
2.1 Screen Elements
2.1.1 Menu bar
You can easily have access to all software functions by using the menu bar. The content of the menu bar depends on
the program status:
2.1.2 Toolbar
To ensure comfortable software operation the most important software functions have been put together at the Tool-
bar. The content of the toolbar depends on the program status:
1.1 1.2 1.3 2.1 2.2 2.3 3.1 4.1 4.2
File functions:
1.1 open file
1.2 save file
1.3 print file
Sensor functions:
2.1 connect USB-measuring adapter
2.2 start new recording
2.3 stop recording
Data representation:
3.1 create new diagram
Further functions:
4.1 interface configuration
4.2 program information
When moving the mouse pointer over a symbol a short description of the function is shown after a short time.
2.2 Data Recording
2.2.1 Connecting the USB-Adapter
connect the USB measuring adapter to the USB-Port
click in the menu item „USB measuring adapter“ on „Connect“ or on the associated icon in the toolbar.
GSOFT_USB is now searching after connected measuring devices.
The located measuring adapter will be shown like this:
If there was no device found you have to click on „New Search“. The USB-port will be initialised and attached devices
will be located. Should still no measuring adapter be found please check the following points again:
Is the USB measuring adapter connected to the USB-port?
Was the correct interface chosen? (“Configuration / Interface“)
The marked measuring adapter will be opened by „OK“.

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 4 of 7
2.2.2 Sensor Data Window
register “data“ register “comment“
register “instrument“
register “extras“
measurements table
current status of the re-
cording;
start-/stop-condition
number of data sets and
period of recording
2.2.3 Start Recording
When recordings are restarted, all data which have been recorded until that moment will be deleted!! If
necessary read in and save important information or save the data on demand before a measurement-
restart!
To start a new recording please select “start new recording“ or click at the corresponding toolbar symbol.
Now you can choose from an assortment of start and stop conditions and define the output data.

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 5 of 7
Choice of the recording mode:
Output:
Cycle time: here you can decide in which intervals measuring points should be recorded. At the same time you can
select which value will be displayed (measured value, min/max peak).
Keep this in mind: at a cycle time of more or equal than 5ms all measured values will be selected automatically. A
direct selection is no longer possible.
Start:
You have to select the desired condition under the different possibilities. If „immediately“ is selected the measuring will
start directly after starting it. At the other conditions you have to define one or more thresholds (if these are under- or
overrun the condition depending on the selected case is fulfilled or not). The thresholds` entry is carried out in the
mask on the right side. You even can select which signal shall cause the measuring start. But this is only available
when the cycle time is higher or equal than 5ms. You can make this entry on the left side under „Start Signal“.
The preTrigger makes it possible for you to detect values before the normal values measured (after the start condition).
After the completion measurement the additional measured values will be shown in the register „Data“, too.
At a cycle time of less than 5ms 200 measured values are possible.
At a cycle time more or equal than 5ms 50 data sets are possible, which are made up of four different series of meas-
urements (50 data sets x 4 series of measurements = 200 measuring values).
Example: a cycle time of 10ms is selected. There will be 10ms long 50 data sets (=500ms) recorded before the normal
measuring will be started.
Stop: here it is possible for you, in analogy to the start conditions, to set the stop conditions. Besides the manual stop
and the measuring stop after a certain time, the stop by specified thresholds is available, too (if these are under- or
overrun the condition depending on the selected case is fulfilled or not). The thresholds` entry is carried out in the
mask on the right side. You even can select which signal shall cause the measuring stop. But this is only available
when the cycle time is higher or equal than 5ms. You can make this entry on the left side under „Stop Signal“.
The Stop Delay effects a continuous measurement until the defined time is up. After this the measuring will be
stopped automatically.
2.2.4 Reading in of Measurements
To read the measuring values the cyclic recording has to be stopped first.
After loading the data the values will be shown in the register „Data“ automatically.
The white fields for the table heading can be edited.
In the register „Comment“ you can write arbitrary texts.
Do you want to view the data later again, you have to save them now by choosing „Save“ or „Save as“ in the menu
„File“.
2.2.5 Stop Recording
If a recording is stopped, data are preserved which have been saved until that moment.
2.2.6 Comment
As soon as the data have been read
you may enter your comments here.
The comment can be printed together.
If data are saved as a file, the com-
ment will also be saved in the same
file automatically.
This is an important function to help
you describe data so that you will be
able to identify and assign them even
after a longer period of time.
Furthermore there is the possibility to
load recurrent comments in form of a
template.
To do so, enter a suitable text-file
(only .txt-format!) via button “change”.
Via button “load template” the text will
be loaded manually to the window.
If “load automatically “ is activated, the
previously selected text file is loaded
automatically, whenever data is read
out.
The comment can be edited after-
wards as usual.
When “load automatically” activated, then “overwrite existing comment” is selectable. This function overwrites the
existing comment, when a new record activate.

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 6 of 7
2.3 Measurement Diagrams
The loaded data of the measuring windows can also be displayed in diagrams.
Select for this „Create new diagram“ in the menu point „Diagram“.
Then a selection of all possible graphs will be displayed. Select the desired graphs and acknowledge your selection by
pressing ‘Ok’. The diagram will be opened.
System limitations: - maximum 2 different units (mbar, bar, etc.) per diagram
- maximum 15 graphs per unit
2.3.1 Adding Data Labels
To add a comment to a specific measuring point select menu item “Diagram/Add Data Label” or click on the corre-
sponding symbol. By moving the mouse over the diagram the symbol will appear as mousepointer. A text line
which can be filled in freely will be placed at the position desired by a mouseclick.
2.3.2 Adding Series
To be able to display additional measuring series in a diagram, select the menu item “Diagram / Add Series” or click
on the corresponding symbol of the diagram window. The data source selection will appear again and you can select
additional measuring series.
2.3.3 ‚Zoom‘: Enlarge a Area of the Display
If 2 units are displayed : use the and symbols to select the Y-axis desired.
Place the mouse at the beginning of the desired area, hold the left mouse key, move the mouse to the end of the de-
sired area and release the mouse key: the selected part will be enlarged (zoomed).
With „diagram – zoom back“ the actual area will be set back to the view before.
With „diagram – zoom all“ the whole diagram will be set back to the original view.
The enlarged view can be moved left, right, up and down with the scrollbar. To show accurate display details we rec-
ommend the function „diagram – settings“. With it a desired section can be defined by number and time declaration.
2.3.4 Copying Data To Clipboard
Diagrams can be copied to the clipboard, to insert them in other applications again:
Click on the required diagram and press the key combination Strg + C -> the diagram will be copied to the clipboard
as a bitmap.
Y-axis: displays the unit
Legend: if activated a
description of the graph
will be displayed
Add / delete series:
the number of dis-
played graphs can be
changed (see below)
Settings:
manual zoom,
grid and colour
settings
Zoom back:
cancels last zoom
action
Zoom all:
show max.
diagram area
Display points:
display (marking) of
measuring points
Change name: the diagram (here:
‚Diagram ‘) can be renamed
Cursor: if activated a crosshair cursor will be displayed to
show the individual measuring points. The measuring point
data will be displayed in the statusbar
Scrollbar: if a
zooming action
has been carried
out the scrollbar
can be used to
‚go through‘ the
diagram
Zoom left/right Y-axis:
choose here if the mouse
zoom function is to be ap-
plied to the left or right Y-
axis Add / delete labels: measuring
points can be described (see below)

G38.0.01.6C-04 GSOFT_USB –Operating Manual page 7 of 7
2.4 File Operation
If you have saved data by selecting menu „file“ function “save“ or “save as“ you can re-load them at any time into the
program to see the data or to analyse them.
In that case please select “File / open...“ and open the corresponding file.
2.4.1 Export Files
The data files can be saved as ASCII (=text) –file to view them for example in word processing or in a spreadsheet
programs. Select the required data window and choose „file / export ...“ at the menu. Pay attention particularly at
spreadsheets to make right settings in the following pop-up window „export files“.
2.5 How to Print Data
The selected window (data or diagram) can be printed via ‘File – Print’ provided a printer has been installed to the
computer. Use ‘File – Printer Setup’ to change printer parameters or to chose another printer.
Under „Pictures“ additionally header or footer pictures can be integrated to the printouts.
The pictures or other templates have to be saved as .bmp, .jpg or .icon format
(By this feature picture size must matched before).
The printout can be extended to a individual form by simple means.