HUATEC HG904 User manual

USER’S MANUAL
HG904
A Portable Dual-Channel
Microprocessor-Based
On-site Balancer
Spectrum Analyzer
Vibration Data Collector
E-mail:[email protected]
Fax: 8610 82916893
Tel : 8610 82921131
Address: Rm. 730, Chengyuan Building, the Mid. Road of Jiancaicheng
Haidian Dist. BeiJing, CHINA
www.huatecgroup.com

HUATEC Group Corporation www.huatecgroup.com
P a ge 1 o f 37
The only difference between HG904 and HG907 is that HG904 can be used
together with Machinery Condition Monitoring software (WaveSoft) .In the following
contents of this manual we use the word ―HG904‖ in most case for both of HG904A
and HG907
The information contained in this document is proprietary. No part of this manual
may be reproduced or transmitted in any form or by any means without the
permission of HUATEC Group Corporation.
Copyright 2001 by Beijing HUATEC Group Corporation. All rights reserved.
The information provided in this Manual is believed to be
reliable. However, HUATEC Group Corporation no responsibility
for inaccuracies or omissions. HUATEC Group Corporation
assumes no responsibilit y for the use of this information, and
all use of such information shall be entirel y at the user's own
risk. Specifications are subject to change without notice. No
patent rights or licences to any of the functions described are
implied or granted to any third party. HUATEC Group Corporation
does not authorise or warrant any HUATEC Group Corporation
product for use in life support devices or systems.

HUATEC Group Corporation www.huatecgroup.com
P a ge 2 o f 37
CONTENTS
CHAPTER 1 GENERAL ……………………………………………………………………… 3
CHAPTER 2 BALANCE PRINCIPLES……………………………………………………… 4
CHAPTER 3 BALANCING PREPARATION …………………………………………….. .. 7
CHAPTER 4 BALANCING OPERATIONS………………………………………………… 8
1-plane balancing…………………………………………………………………………… 8
without influence coefficients…………………………………………………………. 8
with influence coefficients……………………………………………………………..13
2-planes balancing………………………………………………………………………… 15
without influence coefficients………………………………………………………….15
with influence coefficients……………………………………….. ………………… 21
CHAPTER 5 Vibration Analyzer / Data Collector…………………………………………….21
APPENDIX BALANCE GRADE OF RIGID ROTORS ……………………………….. 31

HUATEC Group Corporation www.huatecgroup.com
P a ge 3 o f 37
1 GENERAL
Imbalance is one of the main causes of machine vibration. It is estimated that
about 50 percent of all the faults are caused by imbalance of rotors. The best mean for
getting rid of mechanical imbalance is on-site balancing. HG904 Machinery
Analyzer/Balancer is especially useful for this purpose. It can measure the intensity of
the vibration; diagnose vibration cause by analyzing the frequency spectrum. It can
also be used to measure the rotation speed and vibration phase, and to calculate the
imbalance weight and location.
Dual channel vibration analyzer Phase GDT-2 sensor
Features:
-Easy to use
-Vector illustration, the balancing process and result are shown clearly at a
glance
-Storage of 10 rotors balancing data
-Selectable of trial mass remove or remain
-Decompose of balancing result to two assigned location
-Trial weight range calculated according to the weight of rotor, rotation speed,
radius and required balance grade
-Trial mass validity judged automatically
-Measuring RPM, amplitude and phase
-On-site 400 lines FFT spectrum and diagnosis function
-Dual-channel simultaneous data collection
-Hardware envelop demodulation for bearing and gear diagnosis
-Transfer function for measuring natural frequency
-Waveform and spectrum display by large LCD
-Storage of vibration values and vibration waveform
Operation condition
-Temperature range: from 5C to 50C
-Relative humidity: < 85%, unsaturated
-Without caustic gases
-Without strong electric-magnetic field & strong impact
Safety
The HG904 are not permitted to contact with running part of machine.
Specifications

HUATEC Group Corporation www.huatecgroup.com
P a ge 4 o f 37
Rotation speed range for dynamic balancing: 70-60,000 r/min
Auto-range and manual-range selectable
Measuring accuracy of vibration: 5 %
Sensors: Piezoelectric accelerometer, Magnetoelectricity velocity, Eddy
displacement and etc.
8th-order ellipse anti-aliasing filter, RPM band-pass filter
Input signal: Accelerometer, and voltage
Amplitude spectrum analysis: 100 lines to 400 lines (zoom), Hanning windowed
Frequency span of spectrum analysis: 100, 200, 500, 1K, 2K, 5K, 10KHz(only for
1-channel sampling)
Data storage: 400 waveforms of 1024-points and 400 data sets
Amplitude ranges & Frequency Response for overall vibration measurement:
Amplitude ranges
Frequency Response
Displacement
0.003 –5 mm peak-peak
10 –500 Hz
Velocity
0.2 –200mm/s true RMS
10 –1000 Hz
Acceleration
0.5 –250m/s2peak
20 –5000 Hz
Envelope
0.1 –20unit true RMS
5-1000Hz from 15-40 KHz
Voltage
0.1 –10V peak-peak
10 –10000 Hz
Notepad: 10 condition codes for visual inspection
Output: USB for communication with PC
Power: Ni-MH rechargeable battery for 8 hours continuous operation, low battery
warning
Operating Environment: 0~55 oC, 90% humidity non-condensing
Rotating speed measurement with photocell sensor
Dynamic Range: 60dB with 48dB adjustable gain range
Dimensions: 21×13×4 cm; Weight: 1.2 kg (Include batteries)
2 BALANCE PRINCIPLES
2.1 What is 1-plane rotor imbalance?
Rigid rotor works under the speed far less than its first-order critical rotate speed
and its deformation can be neglected. When the quality of the rigid rotor nearly
focuses on a disc, namely the ratio of its axial length and its diameter is less than 0.5;
we can do 1-plane balance on it and gain satisfactory result.
But for the flexible rotor, the deformation cannot be neglected. So its method of
balance differs from rigid rotor. However, the rotor with single imbalance plane can
also be balanced according to the method of 1-plane balance of rigid rotor. For above,
whether the rigid or flexible rotor, its quality of imbalance always focuses on one disc,
so both of them can be balanced according to the method of 1-plane rotor balance. In
the plant and manufactory, such rotating machines are ubiquitous, such as pump,
ventilator etc.
2.2 Principle of 1-Plane Balancing
HG904’s 1-plane balance adopts the method of influence coefficient, which is also
called balance method of 1-plane phase measuring. Just as its name implies, it’s
desirous to do the measurement with the phase of rotation-speed vibration when
measuring the vibration amplitude of rotor. The rotation-speed vibration can be
indicated as vector. The process of 1-plane balance is as follows:
(1) Measure and gain the initial rotation-speed vibration vector (A0) under the
normal operation condition.

HUATEC Group Corporation www.huatecgroup.com
P a ge 5 o f 37
(2) Load the proper trial mass (M) on the rotor and then measure vibration vector
(A01) under the condition of the same rotate speed.
(3) Calculate the balance mass (Q) which should be load on the rotor according to the
following formula:
Q = - M xA0 / (A01 –A0)
During the balancing, all the vibration should be measured under the same rotate
speed because of the imbalance force of rotor relates with the rotate speed.
2.3 Principle of 2-planes balancing
Almost all the balance of single span rotor can be achieved through the method of
2-planes dynamic balance. As a matter of fact, 1-plane dynamic balance is only one
special example of 2-planes dynamic balance.
When doing 2-planes dynamic balance, two planes of adding mass and two points
of vibration measure are needed. HG904’s 2-planes dynamic balance also adopts the
method of influence coefficient. But the difference from 1-plane dynamic balance is
that the vibration of two measurement points should be measured when adding
trial-mass to one of the planes. That’s so-called interact effect. 2-planes dynamic
balance has four influence coefficients.
The steps to do 2-planes dynamic balance is as followings:
(1) Measure initial value of two measurement points.
(2) Load trial mass to first plane, then measure the vibration of two measurement
points respectively.
Load trial mass to second plane, and then measure the vibration of two
measurement points respectively.
(3) Get conclusion of the correction mass.
If the influence coefficient is known, it can be input directly and the above step (2)
can be omitted.

HUATEC Group Corporation www.huatecgroup.com
P a ge 6 o f 37
1- Plane balancing

HUATEC Group Corporation www.huatecgroup.com
P a ge 7 o f 37
2-Plane balancing

HUATEC Group Corporation www.huatecgroup.com
P a ge 8 o f 37
3 BALANCING PREPARATION
Before using HG904 for balancing, the following must be done firstly:
<1> Confirm Dynamic Balance Needed
Affirm that machine has fault of imbalance and that it can be eliminated by the
method of balancing. Once the machine has heavy vibration, we should do frequency
analysis with the vibration signals firstly. Generally, the probability of imbalance is
higher. So, when the vibration includes a high rotation-frequency spectrum peak and
less harmonic frequency peak, meantime without evidence of other faults, very
possible it is the fault of imbalance.
<2> Mount Sensors
Vibration Sensors: It should be disposed in the horizontal direction (or vertical if
horizontal not possible) on the bearing base, or on the base frame. Because dynamic
balance is vector calculus, the position and direction of the sensor should keep
unchanged in the balancing process.
Photoelectric sensors: Paste a reflecting or unreflecting adhesive paper (according
to the condition of the rotor’s surface) to a visible place of the rotor or its axis.
Photoelectric sensor can give HG904 1 electric pulse per turn of the running rotor.
Not only the rotate speed can be measured but also the vibration signals can be
compared with pulse signals. So the phase of the rotation-frequency vibration can be
obtained. During the balancing process, the position of the reflecting adhesive paper,
the position and direction of the photoelectric sensor should keep unchanged.
<3> Connect Sensor Cables
The photoelectric sensor should be connected to the Tacho/Trigger socket of HG904.
For 1-plane balancing, 1 vibration sensor should be connected to the Accelerometer
Channel A or Voltage Channel A socket. For 2-plane balancing, a second
accelerometer should be connected to the Accelerometer Channel B or Voltage
Channel B socket.
<4> Start Machine
Start the machine and wait till it reaches its normal rotation speed (could be
measured with the RPM measurement function).
NOTICE:
1Because of the strong pull force of the magnetic mount, when remove the
sensor from the measurement surface, use your hand carefully. Do not
remove the sensor by dragging the cables.
2Before measurement, the sensitivity of the sensors must be inputted to
HG904 correctly.
3The internal/external trigger switch should be set to “external”.

HUATEC Group Corporation www.huatecgroup.com
P a ge 9 o f 37
4. BALANCING OPERATIONS
4.1 1-Plane Balancing
4.1.1 WITHOUT INFLUENCE COEFFICIENT:
Step 1:Setting Parameters
Press 【On/Off】on the panel for a moment, the balancing page appears. Then
you can set the parameters.
Line 1 on the left: Change rotor number. The note displays ―Which Rotor to
Balance‖(It means which rotor you will choose to Balance,HG904/HG907 contains
200 arrays ,each arrays describe kinds of data during the balancing ). You can
press 〖〗or〖〗to change rotor number ,or you can input rotor number .The
method of input numbers is that you press 【In】at first, then input numbers
with number keys, press 【ENTER】or 【↓】at last to affirm it and the
cursor goes to the next position automatically. Press or key to move the
cursor between lines on the left (the following are the same).
Line 2 on the left: To set parameters, the note displays ―Enter Menu Key”,
press 【MENU】to right to set parameters.
Line 2 on the right: To choose the operation mode,the note displays ―How
Many Plane to Balance”. You press 〖〗or〖〗to choose 1 or 2(indicating
that 1-plane balancing or 2-plane balancing is to be performed).Here please
choose “1”.
Line 3 on the right: To choose the measuring mode, there are four modes such as
ACC(acceleration),VEL(velocity),DISP(displacement) and VOL(voltage).The note
displays ―Enter to Select”, press
〖〗or〖〗to choose the measuring mode,
The displayed is the selected.
Line 4 on the right: Input the weight of a trial unit. This is useful if some fixed
weight unit is used during the balancing. The method of input numbers is the same as
above.
Line 5 on the right: Tell whether the influence coefficients are known. The note
displays ―to S elect”, press 〖〗or 〖〗to choose YES or

HUATEC Group Corporation www.huatecgroup.com
P a ge 10 o f 37
NO(indicating that the influence coefficients are known or unkown), Here please
choose NO.
The displayed is the selected.
Line 6 on the right: Input sensitivity of sensor A. The note displays ―Input
Sensitivity of Sensor A”.The sensitivity can be usually obtained from the sensor
certificate. The method of input numbers is the same as above.
Line 7 on the right: To Clear Data of the Rotor. The note displays‖Enter to
Clear”.Press 【ENTER】, fomat all is shown in the 7th line and only erase
this rotor is shown in the 8th line.Press 【↑】or【↓】to choose fomat all
or only erase this rotor ,then Press【ENTER】, the following warning is shown:
Are you sure
to format all?
YES NO
Press 〖〗or〖〗to choose YES or NO.
If you want to clear the rotor datas displayed,c hoose only erase this
rotor and press ENTER, then choose Y ES and press ENTER to erase the
rotor datas displayed , the measured data will be cleared, the balance grade and the
sensor sensitivities will change to default values at the same time ―successfully
eras e th is ro tor ”is showed.
If you want to clear all the rotor datas measured,c hoose f omat all
and press ENTER, t he n choose YES and press ENTER to clear all the rotor
datas measured, the measured data will be cleared, the balance grade and the sensor
sensitivities will change to default values at the same time ―successfully fomat”is
shown.
If you don’t want to clear any data,c hoose NO and then press ENTER to
return directly without clearing anyting.
Before a balancing operation is performed, each line of this page must be rightly
configured. You may skip a line by press or . Press to go to the next item.
Step 2: Estimate weight of trial, and then fix a trial mass on plane I
Line 3 on the left: Calculate the weight range of trial mass according to the
formula recommended by the ISO. The note displays ―Enter Menu Key”, press

HUATEC Group Corporation www.huatecgroup.com
P a ge 11 o f 37
【MENU】to right.
Line 2 on the right: The note displays‖ Input How Heavy‖. The method of
input numbers is the same as above.
Line 3 on the right: The note displays‖ Maximal Rotation Speed ‖. The
method of input numbers is the same as above.
Line 4 on the right: The note displays‖ Input Radius of Trial Mass‖. The
method of input numbers is the same as above.
Line 5 on the right: The note displays‖ Enter to Select‖. Choose the
required balance grade according to ISO1940 with 〖〗or〖〗.4000→1600
→630→250 →100 →40→16 →6.3 →2.5 →1.0 →0.4→4000 are shown
circularly. The displayed is the selected.
Line 6 on the right: The note displays‖ Calculate Trial Range‖. Press
〖Enter 〗to calculate weight range of trial mass and the result is shown in the 7th line.
Stop the machine, fix a trial mass on plane I according to the calculation, and then
restart the machine.
Step 3: Initial measurement:
Line 5 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗to
right.
Line 2 on the right:All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 2nd line .When the cursor is in the 2nd line,the note
displays‖Enter to Measure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 3rd line will update continually. And
Amplitude&Phase is shown in the 2nd line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 3rd line (rotation speed) and in the 5th line (the left is amplitude while right is
phase) will update continually. When the rotation speed, amplitude and phase all
become stable, press ENTER for a moment to finish the measurement.
Step 4: Trial 1 measurement:

HUATEC Group Corporation www.huatecgroup.com
P a ge 12 o f 37
Line 6 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗
to right.
Line 2 on the right: The note displays‖ Input Angel‖. The method of input
numbers is the same as above.
Line 3 on the right: T he note displays‖ Input Trial Mass Weight‖. The
method of input numbers is the same as above.
Line 4 on the right: The note displays ―to Select”, press〖〗or 〖〗
to choose Remove or Remain The displayed is the selected.
Line 5 on the right: All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 5th line .When the cursor is in the 5th line,the note
displays‖Enter to Measure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 6th line will update continually. And
Amplitude&Phase is shown in the 5th line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 6th line (rotation speed) and in the 8th line (the left is amplitude while right is
phase) will update continually. When the rotation speed, amplitude and phase all
become stable, press ENTER for a moment to finish the measurement.And the valid
of trial is shown in the 9th line,Valid means you can goto the next step, Remove ,add
indicating that you should add a new trial at another angle, Remove indicating that
you should add this trial at another angle.Stop the machine, adjust the trial mass
accordingly to the note. Repeat the trial I measurement.
//Line 10 on the right: The note displays ―Enter to See”.
Step 5: calculate influence coefficients and balancing mass

HUATEC Group Corporation www.huatecgroup.com
P a ge 13 o f 37
Line 7 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗to
right.
Line 2 on the right: The note displays‖ Enter to Calculate‖,all the places
showing values are vacant. If you hope to see some measurement, move the cursor to
there, and then the value is displayed. You can also input the value. The method of
input numbers is the same as above.Press 【ENTER 】, the balancing mass is
calculated and shown in the 5th line. The location of balancing mass is also shown
with vector chart. Press 【↑】or【↓】to move cursor on the right.
Line 3 on the right: The note displays‖ Enter to See
Coeffient‖.Press 【ENTER】to Coeffient page to see influence
coefficients .The cursor is in the 1st line, the note displays‖ Enter to
Return‖.Press 【ENTER】to the former page.
Step 6: discompose vector
Line 8 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗to
right.
Firstly, the vector of balancing mass can be found in the vector graph just on the
circle.Then find two proper angles at each side of the vector of balancing mass .
Line 2 on the right: The note displays‖ Input Angle‖. Input the first angle PI1
(the first angle is at the anti-clockwise side of the balancing mass vector). The method
of input numbers is the same as above.
Line 3 on the right: The note displays‖ Input Angle‖. Input the first angle PI1
(the first angle is at the anti-clockwise side of the balancing mass vector). The method
of input numbers is the same as above.
Line 4 on the right: The note displays‖ Enter to Calculate‖. Press
【ENTER】then HG904 calculates. The weights of the balancing masses to fix on
the selected angles are shown on the right side of the 2nd and 3rd lines. Assure that the
sums of the two angles depart from the balancing mass should be less than 180
degree.
Stop the machine, fix balancing mass on the rotor according to the result
presented by HG904. The balancing mass is removed or remained should consist with
the settings. Restart the machine. When the rotation speed becomes steady, press

HUATEC Group Corporation www.huatecgroup.com
P a ge 14 o f 37
【↓】to the next item.
Step 7: Verification
Line 9 on the left: The note displays‖ Enter Menu Key ‖. Press〖MENU〗to
right.
Line 2 on the right: All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 2nd line .When the cursor is in the 2nd line,the note
displays‖Enter to Measure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 3rd line will update continually. And
Amplitude&Phase is shown in the 2nd line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 3rd line (rotation speed) and in the 5th line (the left is amplitude while right is
phase) will update continually. When the rotation speed, amplitude and phase all
become stable, press ENTER for a moment to finish the measurement. The 6th line
displays how much the vibration of point I has changed. The user could see whether
the result is usable. If the result is not acceptable, repeat the operation. The influence
coefficients could be borrowed from the last measurement.
4.1.2 WITH INFLUENCE COEFFICIENTS:
Step 1: Setting Parameters
The operation of lines except line 5 is the same as 1-plane balancing without
influence coefficients.
Choose YES in the 5th line.

HUATEC Group Corporation www.huatecgroup.com
P a ge 15 o f 37
Step 2:Initial measurement:
The operation of lines is the same as 1-plane balancing without influence
coefficients.
Step 3: input influence coefficients
The operation of lines except line 3 is the same as 1-plane balancing without
influence coefficients.
Line 3 on the right: The note displays‖ Enter to See Coeffient ‖.Press
【ENTER 】to Coeffient page to input influence coefficients . The
method of input numbers is the same as above.The cursor is in the 1st line, the
note displays‖ Enter to Return‖.Press 【ENTER】to the former page.
Step 4: calculate
The operation of lines is the same as 1-plane balancing without influence
coefficients.
Step 5: discompose vector

HUATEC Group Corporation www.huatecgroup.com
P a ge 16 o f 37
The operation of lines is the same as 1-plane balancing without influence
coefficients.
Step 6: verification
The operation of lines is the same as 1-plane balancing without influence
coefficients.
4.2 2-PLANES BALANCING
Illustration of accelerometer location and cable connection
4.2.1:WITHOUT INFLUENCE COEFFICIENTS:

HUATEC Group Corporation www.huatecgroup.com
P a ge 17 o f 37
Step 1: Setting Parameters
The operation of lines except line 2 is the same as 1-plane balancing without
influence coefficients.
Choose 2in the 2ndline.
Step 2: Estimate trial mass, and then fix a trial mass on plane I
The operation of lines is the same as 1-plane balancing without influence
coefficients.
Step 3: Initial measurement:
Line 5 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗to
right.
Line 2 on the right:All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 2nd line .When the cursor is in the 2nd line,the note
displays‖Enter to Measure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 3rd line will update continually. And
Amplitude&Phase is shown in the 2nd line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 3rd line (rotation speed) ,in the 5th line and in the 6th line (the left is amplitude
while right is phase) will update continually. When the rotation speed, amplitude and

HUATEC Group Corporation www.huatecgroup.com
P a ge 18 o f 37
phase all become stable, press ENTER for a moment to finish the measurement.
Step 4: fix a trial mass on plane I and measure the vibration
Line 6 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗
to right.
Line 2 on the right: The note displays‖ Input Angel‖. The method of input
numbers is the same as above.
Line 3 on the right: T he note displays‖ Input Trial Mass Weight‖. The
method of input numbers is the same as above.
Line 4 on the right: The note displays ―to Select”, press〖〗or 〖〗
to choose Remove or Remain The displayed is the selected.
Line 5 on the right: All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 5th line .When the cursor is in the 5th line,the note
displays‖Enter to Measure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 6th line will update continually. And
Amplitude&Phase is shown in the 5th line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 6th line (rotation speed) ,in the 8th line and in the 9th line (the left is amplitude
while right is phase) will update continually. When the rotation speed, amplitude and
phase all become stable, press ENTER for a moment to finish the measurement.And
the valid of trial is shown in the 10hline,Valid means you can goto the next step,
Remove ,add indicating that you should add a new trial at another angle, Remove
indicating that you should add this trial at another angle.Stop the machine, adjust the
trial mass accordingly to the note. Repeat the trial I measurement.
//Line 11 on the right: The note displays ―Enter to See”.
Step 5: fix a trial mass on plane I and measure the vibration

HUATEC Group Corporation www.huatecgroup.com
P a ge 19 o f 37
Line 6 on the left: The note displays‖ Enter Menu Key‖. Press〖MENU〗
to right.
Line 2 on the right: The note displays‖ Input Angel‖. The method of input
numbers is the same as above.
Line 3 on the right: T he note displays‖ Input Trial Mass Weight‖. The
method of input numbers is the same as above.
Line 4 on the right: The note displays ―to Select”, press〖〗or 〖〗
to choose Remove or Remain The displayed is the selected.
Line 5 on the right: All the places showing values are vacant If you hope to see
some measurement, move the cursor to there, and then the value is displayed. You can
also input the value. The method of input numbers is the same as above. Rotation
Speed is shown in the 5th line .When the cursor is in the 5th line,the note
displays‖Enter to Meas ure‖. Press ENTER, to measure the rotation speed. The
RPM value on the right side of the 6th line will update continually. And
Amplitude&Phase is shown in the 5th line .When the rotation speed becomes
stable, press ENTER for a moment, the HG904 begins measuring vibration. The data
in the 6th line (rotation speed) ,in the 8th line and in the 9th line (the left is amplitude
while right is phase) will update continually. When the rotation speed, amplitude and
phase all become stable, press ENTER for a moment to finish the measurement.And
the valid of trial is shown in the 10hline,Valid means you can goto the next step,
Remove ,add indicating that you should add a new trial at another angle, Remove
indicating that you should add this trial at another angle.Stop the machine, adjust the
trial mass accordingly to the note. Repeat the trial I measurement.
//Line 11 on the right: The note displays ―Enter to See”.
Step 6: calculate influence coefficients and balancing mass
Table of contents
Other HUATEC Measuring Instrument manuals