WMZ-630T
STEREO ZOOM MICROSCOPE

1. INTRODUCTION 2
2. SPECIFICATION 3
2.1 Mainbody 3
2.2 Stand 3
2.3 Illuminator 3
2.4 Eyepiece 3
2.5 Accessory lens 3
3. ATTACTING WMZ-630T 4
3.1 Unpacking 4
4. ADJUSTMENT OF STAND 5
5. CHANGING THE MAGNIFICATION 6
6. FOCUSING 7
7. ATTACHING VIDEO AND PHOTOGRAPHIC SYSTEM 8
7.1 Assembly of the video/photo system 8
7.2 Assembly WMZ-630T photographic system. 8
8. PROTECTING AND MAINTAINING 9
8.1 Caution during operation 9
8.2 Maintenance 9
9. OPTICAL SPECIFICATIONS 10
Manual WMZ-630T 1
CONTENTS

TM-630 series Stereo Zoom Microscope has a side mount zoom control system, it is
easy to operate, perfect for use with bonders and prober arms. It provides excellent 3D
images with high resolution. Once brought in focus, images remains clear throughout
whole zoom range, it has large field of view, long working distance, and big depth of
field, a wide variety of optical lenses and stands are available. Designed as a modular
system, with all metal construction. It widely used in the various applications, such as
assembly and inspection industries, educational clinical and research purposes. The
microscope come with standard supper wide field, high eye point WF10X Eyepiece,
enable observation over a wide area and are convenient for operators who wear
glasses.
2Manual WMZ-630T
1. INTRODUCTION

2.1 MAINBODY
TM-630 Binocular/Trinocular Body with WF10X High Eye Point Eyepiece and Eyecup:
Zoom Range: 0.7x-4.5x
Zoom Ratio: 1:6.5
Total Magnification: 7-45x standard, 2.1-225x with optional lens
Working Distance: 108mm(without auxiliary lens), it can be 38-314mm
with optional auxiliary lens.
Field of View: 4.9-31.5mm standard 1-114.3mm with optional lens
Interpupillary Distance: Adjustable 55-75mm
Binocular Tube: inclined 45 degree.
Diopter Adjustment: 5 Diopters
Binocular Body: Can be rotated a full 360-degree.
Trinocular body: includes all of the features of TM-630B binocular and
provides a vertical phototube for CCD video and
micro-photographic system.
2.2 STAND
We also provide various stands for different usage:
2.3 ILLUMINATOR
We have developed series of illuminator for observing different kind of specimen. Main
illuminator: adjustment halogen lamp, fluorescent ring light, transmitted light base, cool
light source with double pipe fiber optics, cool light source with ring fiber optics and
stand attached illumination base.
2.4 EYEPIECE
Many kinds of eyepiece are optional,
Including: WF10X/22mm,WF10X/19mm,WF10XM, WF15X, WF20X, WF25X.
2.5 ACCESSORY LENS
Accessory lens provide wider field of view and logger working distance. Various
objectives suitable for different markets equipment: 0.3x, 0.5x, 0.75x, 1.5x and 2x.
In order to make user have good understanding of WMZ-630T microscope we make a
form of optical specifications at the end of manual.
Manual WMZ-630T 3
2. SPECIFICATION

3.1 UNPACKING
1 Open the packing material, pick out stand and put it on the working table.
2 Take out binocular body, place the TM-630 into its stand carefully, make it stable by
tightening the screw.
3 Take out the eyepiece cover and dust cover.
4 Put eyepieces into eyepiece tube and adjust eyepiece tube ().
5 If you use accessory lens, please twist the lens on main body.
4Manual WMZ-630T
3. ATTACTING WMZ-630T

1 Place the specimen sample on the objective plate.
(note :try to put the sample on the center of plate )
2 Estimate the working distance you want by your eyes. Working distance is 116mm
(without auxiliary lens detailed specification can be referred in optic specification
form)
3 Focusing hand wheel is adjustable to loosen and tighten. you can adjust it for
comfortable operation by yourself
(a) turn the focusing stop ring counter clockwise, focusing hand wheel is
tighten.
(b) turn the focusing stop ring clockwise focusing hand wheel is loosen.
Manual WMZ-630T 5
4. ADJUSTMENT OF STAND

By turning the zoom knob located in either side of the stereomicroscope. The
magnification can be changed continuously over a 1:6.5 zoom ratio. The magnification
can be extended by using different eyepieces or supplementary lenses in various
combinations.
Supplementary lens power x eyepieces power x magnification setting =total
magnification
6Manual WMZ-630T
5. CHANGING THE MAGNIFICATION

1 Set the magnification knob to the lowest power. Place and illuminate a flat
specimen within the field of view. Focus the WMZ-630T stereomicroscope by
adjusting the focusing hand wheel until a sharp image is obtained.
2 Zoom to the highest magnification by turning the zoom knob gradually. Refocus the
WMZ-630T by adjusting the focusing hand wheel until a sharp image is obtained.
The WMZ-630T is capable of scanning of at low magnification and then zooming to
a higher magnification for critical observation without losing focusing.
3 If you want higher or lower magnification, you need to choose different eyepiece or
objective, when you change eyepiece, it is not necessary to refocus WMZ-630T, but
you need to refocus WMZ-630T when you replace the objective.
Manual WMZ-630T 7
6. FOCUSING

7.1 ASSEMBLY OF THE VIDEO/PHOTO SYSTEM
1 Put WMZ-630T main body into its stand.
2 Attach video/photo adapter to the top hole located on the top of WMZ-630T main
body, and tighten it with the screws of the module.
3 Assemble video coupler to video adapter.
4 Fixed the CCD camera video couple.
5 Assembling process
WMZ-630T main body ¡û CCD Adapter ¡û CCD
Note:
Different cameras need different type of c-mount adapter.
Please use high sensitization films for sharp image.
7.2 ASSEMBLY WMZ-630T PHOTOGRAPHIC SYSTEM.
1 Put WMZ-630T main body into its stand.
2 Attach phototube to the top hole located on the top of WMZ-630T main body, and
tighten it with the screws of the module.
3 Assemble camera adapter to phototube.
4 Assembling process
WMZ-630T main body ¡û photo tube(PT) ¡û camera adapter(CMA)
8Manual WMZ-630T
7. ATTACHING VIDEO AND PHOTOGRAPHIC SYSTEM

8.1 CAUTION DURING OPERATION
1 Soft and gentle operation is necessary
2 Optical accessories should not be hit.
3 Keep all optical components clean. Dust, fingerprints or a smear on the optics will
degrade the image. If optics is dirty, it should be cleaned with a cotton swab
moistened with alcohol.
4 Microscope should be used in a clean and dry working room.
8.2 MAINTENANCE
1 The dust cover should be returned to the instrument when it is not used. It should
always be covered with dust cover when it is not in use. Put it in a dry place.
2 The optics should be placed in a container together with some dryer.
Note: service
If you experience difficulty with our products. Please contact your local representative. a simple remedy may
correct apparent the difficult without the need to send the instrument for servicing. if it becomes necessary to
send the instrument for servicing, please contact your local representative .the pack it carefully in the original
container or in a crush resistant carton with shock absorbing, dustless material surrounding it. Include a detailed
letter describing the problem. This information will help the service technician to efficiently and cost effectively
repair the instrument.
Manual WMZ-630T 9
8. PROTECTING AND MAINTAINING

Auxiliary Eyepiece Magnification Field Of View Working
Lens Distance
0.3x WF5x (24mm) 1.05-5.75x 114.3-17.8mm 314m
WF10x (22mm) 2.1-13.5x 105-16.3mm
WF15x (16mm) 3.2-20.3x 76.2-11.9mm
WF20x (11.5mm) 4.2-27x 55-8.5mm
WF25x (9mm) 5.3-33.8x 43-6.7mm
0.5x WF5x (24mm) 1.75-11.25x 68.6-10.7mm 189mm
WF10x (22mm) 3.5-22.5x 63-9.8mm
WF10x (19mm) 3.5-22.5x 54.3-8.4mm
WF15x (16mm) 5.3-33.8x 45.7-7.1mm
WF20x (11.5mm) 7-45x 33-5.1mm
WF25x (9mm) 8.8-56.3x 26-4mm
0.75x WF5x (24mm) 2.625-16.9x 45-7.1mm 122mm
WF10x (22mm) 5.3-33.8x 44-6.8mm
WF10x (19mm) 5.3-33.8x 36.2-5.6mm
WF15x (16mm) 7.9-50.6x 30.5-4.76mm
WF20x (11.5mm) 10.5-67.5x 22-3.4mm
WF25x (9mm) 13.1-84.4x 17.2-2.7mm
1x (fixed) WF5x (24mm) 3.5-22.5x 34.3-5.8mm 108mm
WF10x (22mm) 7-45x 1.3-4.9mm
WF10x (19mm) 7-4x 27.1-4.2mm
WF15x (16mm) 10.5-67.5x 22.9-3.6mm
WF20x (11.5mm) 14-90x 16.4-2.6mm
WF25x (9mm) 17.5-112.5x 13-2mm
1.5x WF5x (24mm) 5.25-33.75x 22.9-3.6mm 40mm
WF10x (22mm) 10.5-67.5x 22-3.4mm
WF10x (19mm) 10.5-67.5x 19-3mm
WF15x (16mm) 15.8-101.3x 15.2-2.4mm
WF20x (11.5mm) 21-135x 11.4-1.7mm
WF25x (9mm) 26.3-168.8x 8.6-1.3mm
2x WF5x (24mm) 7-45x 17.1-2.7mm 38mm
WF10x (22mm) 14-90x 15.7-2.4mm
WF10x (19mm) 14-90x 13.6-2.1mm
WF15x (16mm) 21-135x 11.4-1.8mm
WF20x (11.5mm) 28-180x 8.2-1.3mm
WF25x (9mm) 35-225x 6.5-1mm
10 Manual WMZ-630T
9. OPTICAL SPECIFICATIONS

W03-062
Table of contents