Kanomax 3522 User manual

Piezobalance Dust Monitor
MODEL 3522
Operation Manual
Read this manual carefully and understand the warnings
before operating the instrument.
Keep this manual handy for future reference.
01001
08 .11
PIEZOBALANCE DUST
MONITOR MODEL3522

Important Safety Information
Types and definitions of warning signs used in this operation manual are shown below.
Danger: To prevent serious injury or death.
Items under this heading show measures to prevent serious injury or death,
which may result if the instructions in this manual are not observed and the
instrument is operated inappropriately.
Caution: To prevent damage to the product.
Items under this heading show measures to prevent damage to the product
and conditions that affect our product warranty.
[Description of Symbols]
This symbol indicates a condition (including danger) that requires caution.
The subject of each caution is illustrated inside the triangle (e.g., high
temperature caution symbol shown on the left).
This symbol indicates prohibition. Do not take a prohibited action shown inside
or near this symbol (e.g., disassembly prohibition symbol shown on the left).
This symbol indicates a mandatory action. A specific action is given near the
symbol.
Danger
*Never expose the instrument to flammable gas.
*Never disassemble or heat the battery pack,
or discard the battery pack in fire.
…… The battery pack may explode.
*Never perform the cleaning procedure
with instrument set to high voltage (if
high-voltage is active, the pointer will
be located at the center of the indicator.)
…… Performing cleaning while high-voltage is active may cause electrical shock.
Do not use near
flammable gas.
Explosive Handle Properly

Caution
*Do not drop or hit the instrument.
…… Dropping or hitting the instrument may
cause damage or malfunction.
*Never disassemble, modify or repair the instrument.
…… Failure to observe the above may cause short circuit
and malfunction.
*Do not clog the impactor inlet with a dirty hand.
…… Failure to observe the above may cause pump failure.
Prohibition
*Do not perform a measurement in an environment with extreme
temperature/humidity change.
…… Use in an environment with extreme temperature/humidity change
may cause error in measurement value.
The allowable rate of change during a 2 minute measurement is:
Temperature: ± 1°C (0°C to 40°C), Humidity: ± 1%RH (30%RH to 80%RH).
Handle Properly
*Do not wipe the instrument with a volatile solvent.
…… The body may deform or deteriorate. Use soft dry cloth to remove stains.
If stains persist, soak the cloth in a neutral detergent and wipe the
instrument with a soft cloth.
Do not use volatile solvents such as thinner and benzene.
Prohibition
*Make sure there is no turbulent flow near the sampling inlet.
…… Turbulent flow may cause measurement error.
Handle Properly
*Do not attempt cleaning in an environment where temperature is
below 0 °C.
…… Cleaning sponge may freeze and cause damage to the sensor. Prohibition

Table of Contents
1. Part Names and Functions......................................................................................1
1.1 Part Names....................................................................................................................... 1
1.2 LCD Screen ..................................................................................................................... 2
1.3 Operation Keys ................................................................................................................ 2
2. Battery Indicator......................................................................................................3
3. Battery Charge.........................................................................................................4
3.1 AC/DC Adapter................................................................................................................ 4
3.2 Battery Charging Display ................................................................................................ 4
4. Parts Installation and Measurement Preparation................................................5
4.1 Removal and Installation of Rechargeable Battery Pack................................................. 5
4.2 Removal and Installation of Cleaning Device ................................................................. 6
4.3 Confirming the Sensor Condition.................................................................................... 7
4.4 Cleaning the Sponge........................................................................................................ 8
4.5 Cleaning the Sensor ......................................................................................................... 9
4.6 Adjusting the Impactor Position .................................................................................... 10
4.7 Sampling Time............................................................................................................... 10
5. Measurement Method............................................................................................11
5.1 High Voltage Supply.......................................................................................................11
5.2 Measurement Start ........................................................................................................ 12
5.3 Measurement End .......................................................................................................... 13
6. Setting the Menu....................................................................................................14
6.1 Main Menu .................................................................................................................... 14
6.2 Sampling Time Menu .................................................................................................... 14
6.3 Data Output Menu ......................................................................................................... 15
6.4 Time Adjustment Menu ................................................................................................. 15
6.5 Communication Menu ................................................................................................... 16
6.6 Data Clear Menu............................................................................................................ 17
7. Printing...................................................................................................................18
7.1 Connecting the Printer ................................................................................................... 18
7.2 Print Output Menu......................................................................................................... 18
8. Regular Maintenance and Impactor Nozzle Replacement.................................19
8.1 Impactor Cleaning ......................................................................................................... 19
8.2 Removing and Cleaning the Needle .............................................................................. 20
8.3 Installing the Needle...................................................................................................... 21
8.3 Installing the Needle...................................................................................................... 21
8.4 Impactor Nozzle ............................................................................................................ 21
9. Main Specifications................................................................................................22

10. Troubleshooting....................................................................................................23
11. Warranty and After-sales Service.......................................................................24
12. Contact Information............................................................................................25

1. Part Names and Functions
1.1 Part Names
(unit: mm)
65
180
Display screen Needle
Impactor
Operation keys
PIEZOBALANCE DUST
MONITOR MODEL3522
Cleaning
device
Batter
y
p
ack release
Cleaning device lock
Sponges
Display screen
contrast adjustment
AC adapter socket
Cleaning port
Rechargeable battery pack
150
Communication port
Power switch
1

1.2 LCD Screen
a. Date & Time:
Indicates current date & time.
ca
b
d
g e
f
h
b. Sampling Time:
Sampling time can be quick set to either 24s or 120s or custom set from 1-60min.
c. Data Record:
Indicates current number of data records (Max.500). When measurement is complete,
“STORE” will flash here. To save data, press the [MEMORY] key.
d. Measurable Range:
Remaining measurable range is indicated by a bar indicator, which moves towards the right
as the sensor absorbs dust and the measurable range decreases. When the sampled dust
exceeds approx. 10μg(>Fundamental Frequency + 2000Hz), “CLEANING” will flash here.
The cleaning procedure must be performed when the remaining measurable range is too
small or “CLEANING” flashes.
e. Concentration Reading:
The concentration reading flashes during a measuring.
f. Operating Status:
“WAIT” flashes during cleaning and power-up.
“READY” flashes when high voltage is supplied and the instrument is ready for
measurement.
Remaining sampling time is displayed during a measurement.
g. High Voltage Indicator:
High voltage status indicates when high voltage is supplied. When the pointer moves to the
center of the frame (under the black block), the operating status changes to “READY”, and
the instrument is ready for measurement.
h. Battery Indicator:
Indicates remaining battery power.
1.3 Operation Keys
M
enu key
MENU
MEAS
CLN MEMOR
Y
START
Up key
Set key
Start key
High voltage
supply/cut key
Down key Data storage key
SET
2

2. Battery Indicator
Remaining battery life displays while the instrument is powered on.
The battery indicator changes as shown below.
Full Half
Requires
recharge
Battery indicator
flashes
Automatic
power off
¾The battery indicator flashes when the battery voltage drops below 9.3V. The battery must be
charged immediately to continue a measurement.
¾If the battery is left uncharged, the instrument will be powered off automatically.
3

3. Battery Charge
3.1 AC/DC Adapter
For prolonged measurements, the instrument should be powered using the supplied AC/DC
adapter where AC100-240V power is available. Insert the adapter jack into the adapter socket of
the instrument first, and then to a power outlet.
When AC power is supplied, the battery charging status bar will display on the screen.
While the battery is being charged, the bar indicator will be moving.
When the battery is fully charged, the bar indicator will stop moving.
3.2 Battery Charging Display
The battery can be recharged while the instrument is
turned OFF or ON. The bar indicator shown on the right
displays the charging status while the battery is being
charged.
When the battery is fully charged as shown on the right,
unplug the adapter, and charging is complete.
Caution:
1) The input of the AC/DC adapter is AC 100-240v and the output is DC 12.5v. Do not use an
adapter other than the one supplied with the instrument. Use of unspecified adapters may
cause charging failure or damage to the instrument.
2) The estimated continuous operating time using the rechargeable Ni-MH battery is approx. 4.5
hours. Charging time is approx. 5-8 hours.
3) If the instrument is not to be used for an extended period, the battery should be charged once a
month in order to extend the service life of the battery.
4) Before starting a measurement, if the adapter plug is inserted into the socket while the
instrument is turned off, the charging status will be displayed on the screen. A measurement
can be performed under this condition by turning on the power. To perform a measurement
without using the adapter, it is recommended that the battery be fully charged before use.
5) Turn off the instrument after a measurement is complete. If the adapter is left inserted into the
socket, the charging status will be displayed on the screen. If there is sufficient battery
remaining, the adapter must be unplugged as charging is not necessary.
In charge
Char
g
e ove
r
Adapter
socket
Adapter
Ada
p
ter
j
ack
4

4. Parts Installation and Measurement Preparation
4.1 Removal and Installation of Rechargeable Battery Pack
1) Press the rechargeable battery pack release, and slide the battery pack along the guiding
groove away from the release as shown below.
Rechargeable battery pack
Battery pack release
Guiding groove
2) Align the tabs on the rechargeable battery pack with the slots in the guiding groove and lift the
battery pack free from the instrument.
3) To reinstall the battery pack, reverse the above procedure.
5

4.2 Removal and Installation of Cleaning Device
1) Hold the cleaning device, and slide the lock away from the cleaning device as shown below.
Cleaning device
Cleaning device
lock
Rechargeable battery pack
2) After the lock is released, remove the cleaning device by holding the front end.
Front end of
the cleaning device
3) To install the cleaning device, place the rear end of the cleaning device onto the battery pack
first, and then push the front end of the cleaning device into the locker. After the cleaning
device is installed, push the lock to its original position.
2
3
1
4
6

4.3 Confirming the Sensor Condition
When the instrument is powered on, the screen below will display with a flashing “WAIT”
message in the operating status field. After approx. 10 seconds the sensor will stabilize and the
“WAIT” message will stop flashing.
The measurable range of the sensor can be confirmed by the bar indicator and the frequency
displayed on the screen.
1) Confirming the Measurable Range using the Bar Indicator:
The bar indicator will be at full length immediately after cleaning. The bar will move to the
right, decreasing in length as dust accumulates on the sensor. When the remaining length of
the bar indicator is short, or when “CLEANING” is flashing, the sensor should be cleaned
immediately. (Refer to section 4.4 and 4.5 for cleaning.)
2) Confirming the Measurable Range using the Frequency display:
The frequency can be displayed by pressing the key, and hidden by pressing the
key again.
The frequency will increase as dust accumulates on the sensor. When the frequency exceeds
the Fundamental Frequency + 2000Hz, “CLEANING” will flash. Cleaning should be done
immediately.
* The frequency increases as the length of the measurable range bar indicator decreases.
Operating status
Frequency:
Displayed/Hidden by
key.
Measurable range
Bar indicator
Caution:
Confirm that the bar indicator has sufficient length before starting a measurement. Never leave
dust accumulated on the sensor as it may prevent the bar indicator from recovering even after
providing cleaning. It can also affect the frequency display, which will affect the measurement
accuracy.
7

4.4 Cleaning the Sponge
Preparation Before Cleaning
1) The sponge on the cleaning device must be moistened
before use. If the sponge is unclean or stained, put the
cleaning device into a vessel filled with clean water with a
few drops of neutral detergent, and leave it for a few
minutes (this is not necessary for a new product).
2) When the stain is removed, wash the sponge with running
water and completely remove all detergent.
3) Use a folded tissue or paper towel on the cleaning device to
remove all excess water from the sponge. This process
should be repeated several times to ensure removal of all
moisture.
4) Put 1-2 drops of detergent (blue) on the SMALLER sponge.
5) Put 2-3 drops of purified water on the LARGER sponge.
Caution:
1) Note that the two sponges differ in size. The proper amount
of detergent must be dropped on the smaller sponge.
2) If the sponge is dirty or stained with oil that cannot be
removed by washing in clean water, put the cleaning device
into a vessel filled with clean water with 3-4 drops of
neutral detergent. After the stain is removed, wash the
sponge with clean water.
3) If the stain cannot be removed from the sponge or if the
sponge is damaged, the sponge must be discarded. Please
contact your local distributor or our service center to
purchase a new sponge.
4) Any water remaining on the cleaning device other than the
sponge itself must be removed completely using a folded
tissue or paper towel. If the cleaning device is inserted in
the instrument with water drops, they may attach to the
sensor or needle, causing recovery failure of the measurable
range or failure when applying high voltage.
5) Use of a sponge with excessive moisture may require a very
long time to dry. In addition, the instrument reading may
become unstable, and the “WAIT” message may not
display.
6) Do not use a detergent other than the one specified. Use of
other detergents may result in an oily film to remain or
accumulation of sediment, which will affect the
measurement.
8

4.5 Cleaning the Sensor
First, confirm that the high voltage is turned OFF. Then align the front end of the cleaning device
with the cleaning port, and slowly insert the cleaning device with the sponge side facing down.
The cleaning device must be inserted up to the scale line indicated on the back side of the cleaning
device. When the scale line reaches the port, stop inserting and wait for approx. 20 seconds. Then,
insert the cleaning device further, and pull it out from the other side. The sensor will then be dried,
and a flashing “WAIT” message will be displayed on the screen. Approximately 3 minutes is
required for drying. When the sensor is dried, the “WAIT” message will stop flashing. The bar
indicator of the measurable range must return to the initial position. If the indicator is not returned
to its initial position, the above cleaning process must be repeated again as there may be dust
remaining on the senor.
Scale line Cleaning device
Caution:
1) Ensure that the high voltage is turned OFF before cleaning the sensor.
Residual high voltage is not fully discharged right after power off. Before starting cleaning,
confirm that the pointer of the high voltage indicator is returned to the left end.
2) While the sensor is cleaned, the flashing “WAIT” message will be displayed. During this
period, high voltage cannot be supplied even if the [MEAS] key is pressed. This is not a
malfunction. When the sensor is dried, the “WAIT” message will stop flashing. Excess
moisture on the cleaning device could lead to a prolonged drying time. If that is the case, the
moisture on the sponge should be absorbed using a folded tissue or paper towel and cleaning
must be performed again.
3) Do not force the cleaning device in if the sponge is torn or protruding from the cleaning
device. Forcing in the cleaning device could apply excessive pressure to the sensor causing
damage.
4) Use of an unclean sponge could prevent the measurable range indicator from returning to its
initial position. Ensure that the sponge is completely cleaned before performing a sensor
cleaning.
5) Do not perform a sensor cleaning if the ambient temperature is below freezing.
9

4.6 Adjusting the Impactor Position
Impactor may be set in three positions. A stopper is provided for fixing the impactor in three
angles: 0°, 45°and 90°. Fix the impactor at the desired angle.
Rotation angle: 45° Rotation angle: 90°
4.7 Sampling Time
For how to set the sampling time, see Section 6. Setting the Menu”.
The initial setting of the sampling mode is 120s.
Change the setting according to the dust concentration of the environment to be measured.
When performing measurements in a high concentration environment, such as a work
environment, select 24s. For indoor measurement in a low concentration environment, select 120s.
For measurements in a high concentration environment, the cleaning frequency can be reduced by
setting a short sampling time. For measurements in a low concentration environment, the
measurement accuracy can be increased by extending the sampling time.
Sampling time
10

5. Measurement Method
5.1 High Voltage Supply
On turning on the instrument, the flashing “WAIT” message will be displayed. After a brief delay,
the “WAIT” message will stop flashing, and high voltage can be supplied. Press the [MEAS/CLN]
key to supply high voltage. The pointer at the left end of the high-voltage indicator frame will
move toward the right. The pointer will stop at the center of the frame under the black block.
When the sensor becomes stable, the operating status will change from “WAIT” to “READY” and
the instrument will be ready for measurement.
Press MEAS/CLN key
The high voltage pointer moves
to the center, and the operating
status changes to “READY”.
Operating status
11

5.2 Measurement Start
To start a measurement, press the [START] key after the operating status has changed to
“READY”. Then the “READY” message will change to an elapsed time indicator. The bar
indicator will extend as the time elapses. The concentration reading will be blinking during a
measurement. The measurement will stop automatically when set time has elapsed. The reading
also stops flashing and the final concentration value is displayed.
Press START key
Concentration reading
during a measurement
Elapsed measurement time indicator
Measurable range
Bar indicator
Caution:
1) When [START] key is pressed during a measurement, the measurement will stop and the
concentration reading will change to “- -.- -”.
2) When [MEAS/CLN] key is pressed during a measurement, the measurement will stop. The
high voltage supply will stop and the concentration reading will change to “- -.- -”. This
procedure allows emergency stop during a measurement or canceling a measurement.
3) If the dust sampled during a measurement exceeds the measurable range, the measurable
range indicator will change to a flashing “CLEANING” message. If a sampling time of 24s or
120s was selected, the measurement will continue. After the measurement is complete,
high-voltage supply will stop automatically to prevent further measurement. Perform a sensor
cleaning to restore the instrument to a measurable condition. When an arbitrary sampling time
was set using TSET, the measurement will stop immediately if the measurable range is
exceeded. The data sampled up to that point will not be saved in this case.
4) After the completion of each measurement, “STORE” will flash. This is asking whether you
want to save the data or not. You can ignore this message if you do not want to save the data.
5) The reading of 0.00mg/m³ will be flashing for 5 seconds after starting a measurement. This is
because the reading is updated every 5 seconds.
12

5.3 Measurement End
When the measurement is complete, the concentration reading will stop flashing. The displayed
value is the measurement result for the specified sampling time. The operating status display will
change from the elapsed time indicator to the “READY” message, and the number of data records
will change to a flashing “STORE” message. If you want to save the measurement result, press the
[MEMORY] key. Ignore the message if you do not want to save the result.
To start another measurement, press the [START] key.
To end the measurement, press the [MEAS/CLN] key to stop the high voltage supply. Then, turn
off the power.
“STORE” blinks
when measurement
is complete
Measurement result is saved by
pressing the MEMORY key.
Operating status changes to “READY”.
Number of data records is increased
by one when data is stored.
Caution:
When a measurement is complete, it is recommended that the sensor is cleaned before powering
down the instrument for the convenience of future measurements.
The accumulated dust left on the sensor will become hard to clean, and will lead to a gradual
reduction of the measurable range.
13

6. Setting the Menu
6.1 Main Menu
The following menu will be displayed by pressing the [MENU] key when high voltage is not
supplied to the instrument. Use the keys to select each item.
To enter the sub-menu of each item, press the [SET] key.
To switch to the measurement screen, select “1. NORMAL” and press the [SET] key.
6.2 Sampling Time Menu
To set the measurement sampling time
Move the cursor to item “2. SAMPLING TIME” of the
main menu by using the key, and press the
[SET] key. The screen on the left will be displayed.
Caution: If [MENU] key is pressed before pressing the
“2. SAVE INFO”, the set time will NOT be saved.
Select “1” and
press key.
SET
Change the setting time by using the key.
The setting changes as below:
120s 24s TSET
When [SET] key is pressed on 120s or 24s, the cursor
will return to the “1.”
When [SET] key is pressed on “TSET”, the following
screen will be displayed.
Select “1” and
press key.
SET
Set the time by using the key.
The settable time range is 1-60min.
When [SET] is pressed, the cursor will return on “1.”
To enter the set time, move the cursor to the “2. SAVE
INFO” by using the key, and press the [SET]
key.
01m
Caution:
After completing the time setup, confirm the setting on the measurement screen. When you have
made the TSET setting, the display will simply show “TSET” (the specified time will not be
displayed.) The elapsed time indicator will be in full scale.
14

6.3 Data Output Menu
To select and view stored data
Move the cursor to item “3. DATA OUTPUT” of the main menu by using the key and
press the [SET] key. The screen as below will be displayed.
When there is no data stored, the message “There is no record. Please set MENU key” will be
displayed.
The data storage number can be selected by using the
key. On releasing the key, the contents
(concentration, date/time, sampling time and sampling
status) will be displayed.
No: Data storage number
MASS CONCENTRATION: Measured concentration value
DATE / TIME: Date/Time the measurement was performed
SAMPLING TIME: Sampling mode: 24s, 120s or TSET (1-60min).
STATUS: Sampling status. 0 or 1
0: Normal measurement condition
1: Indicated when the measurable range is exceeded during a
measurement in sampling mode 24s or 120s.
6.4 Time Adjustment Menu
To set the data and time
Move the cursor to item “6. TIME ADJUST” of the main
menu by using the key and press the [SET]
key. The screen on the left will be displayed.
When “3. SAVE INFO” is pressed, the set date/time will
be saved, and the screen will return to the menu.
On pressing the [SET] key on “1”or “2”, the following
screen will be displayed.
Set the year, month and day by using the key.
Press the [SET] key to proceed to the next item.
When day is set, the cursor will return to the “1.DATE”.
Move to item “2. TIME” and press the [SET] key.
The next screen will be displayed.
Set the hour, minute and second by using the key,
and press the [SET] key. Set the time for approx. 30-60sec
in the future, then press “3.SAVE INFO” when the set time
comes. The set time will be entered, and the screen will
return to the menu.
Select “1.DATE”
and press key
Select “2.TIME”
and press key
SET
SET
15
Table of contents