The ThermoCamera Connect is used to visually display thermal patterns and enables non-contact
measurement of surfaces by evaluating the radiation in the infrared wavelength range via the integrated
non-chilled microbolometer. The imaging technology of the sensor provides a visual image of the
temperatures surrounding the inspected object. The temperature differences can be visualised by colour-
coding the individual temperatures in a thermogram with false colour display. Possible applications include
locating overloads in electrical components, detecting overheating in mechanical components, pinpointing
and analysing heating lines in walls and oors, assessing refrigerant and air conditioning systems and
much more.
Function / Application
Completely read through the operating instructions, the „Warranty and Additional Information“
booklet as well as the latest information under the internet link at the end of these instructions.
Follow the instructions they contain. This document must be kept in a safe place and passed on
together with the device.
!
General safety instructions
– The device must only be used in accordance with its intended purpose and within the scope of
the specications.
– The measuring tools and accessories are not toys. Keep out of reach of children.
– Modications or changes to the device are not permitted, this will otherwise invalidate the approval
and safety specications.
– Do not expose the device to mechanical stress, extreme temperatures, moisture or signicant vibration.
– The device must no longer be used if one or more of its functions fail or the battery charge is weak.
– Use only genuine accessories. Use of inappropriate accessories will invalidate the warranty.
– The charging progress can be viewed by briey pressing the ON/OFF button.
– The residual charge of batteries with a weak charge is shown in red.
– The battery can also be charged while using the device.
– Disconnect the power pack from the mains when the device is not in use.
– Never use extension cables in conjunction with the battery charger, or similar accessories not approved
by the manufacturer, as this may result in a re risk, the risk of an electric shock or injury to persons.
Safety instructions
Using electromagnetic radiation and RF wireless radiation
– Local operating restrictions – for example, in hospitals, aircraft, petrol stations or in the vicinity of
people with pacemakers – may apply. Electronic devices can potentially cause hazards or interference
or be subject to hazards or interference.
– The measuring accuracy may be affected when working close to high voltages or high electromagnetic
alternating elds.
– The measuring device is equipped with a wireless interface.
– The measuring device complies with electromagnetic compatibility and wireless radiation regulations
and limits in accordance with the RED 2014/53/EU.
– Umarex GmbH & Co. KG hereby declares that the ThermoCamera Connect radio equipment complies
with the essential requirements and other provisions of the European Radio Equipment Directive
2014/53/EU (RED). The EU Declaration of Conformity can be found in its entirety at the following
address: http://laserliner.com/info?an=AGR
– This device complies with the CE limit values for radiation exposure as dened for uncontrolled
environments. To avoid wireless radiation, this device should be used at a minimum distance of 20 cm
from the body.

ThermoCamera Connect
3
a
c
d e
b
h
g
f
3
4
5
6
7
8
1
2
2
8
i
1 2 3 4
5
12
8
10
11
9
6
7
EN
13,2“ TFT colour display
2Hot keys
3Lens cover
4
Camera
5
Infrared camera lens
6
Trigger: Capture
7
Battery compartment
8
Shaft
aON/OFF
bMenu navigation
cMenu navigation /
select digital image,
overlay infrared / digital
image and infrared image
dOpen main menu /
exit main menu
(cancel) / save image
eControl menu
(conrmation) /
do not save image
fControl menu (cancel)
gMenu navigation /
select digital image,
overlay infrared / digital
image and infrared image
hMenu navigation
iMicro USB interface
Standard measurement view
1
Temperature at centre of image
2
WLAN active
3
Set emissivity coefcient
4
Battery charge indicator
5
Color palette
6
Time
7
Min. temperature
8
Max. temperature
9
Temperature at centre of image
10
Max. temperature
11
Thermography image
12
Min. temperature

4
13
14
15
16
17
18
19
20
21
22
23
24
25
1
3 sec
3 sec
EN
Main menu
13 Open media gallery
14 Change color palette
15
Set emissivity coefcient
16 Settings
Main menu settings
17
Auto shutdown
18 Display brightness
19 Menu language
20
Time format
21 Time settings
22 Spot (Measuring point)
23
WLAN connection ON/OFF
24 Image correction
25 Software version
Images
Color palette
Emissivity
Settings
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
ON / OFF
ON
OFF

ThermoCamera Connect
5
5.0
32
4
3 sec
EN
Settings
To charge the Li-ion
battery pack, plug
the supplied USB cable
into the charging
socket „i“ and
connect it to
the USB 2.0 power
supply unit.
Operation while
charging is possible.
Main menu
General and measurement-specic settings can be made in the main menu. The menu is controlled with
the four buttons (b, c, g, h).
Removing / inserting
li-ion battery pack
Open battery compart-
ment (7) and remove/
insert the Li-ion battery
pack. When inserting
battery ensure correct
polarity.
Charging the li-ion battery
Images
Color palette
Emissivity
Settings
Images
Color palette
Emissivity
Settings
Auto shutdown
Intensity
Language
Settings
Time format
Set time
Spot
WLAN
Version
Image correction

6
5.1
5.2
5.3
EN
Settings: Auto shutdown
Settings: Display brightness
Settings: To set menu language EN / DE / FR / NL / IT / ES / DK / FI / SE
The device switches off automatically after a set period of inactivity.
Conrmation
Exit menu
Cancel
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Auto shutdown
No
5 min
20 min
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Intensity
Low
Medium
High
Conrmation
Exit menu
Cancel
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Language
English
Deutsch
Français
Nederlands
Svenska
Conrmation
Exit menu
Cancel
Italiano
Español
Dansk
Suomalainen

ThermoCamera Connect
7
2018
10
22
14
36
25
2018
10
22
14
36
25
0
5.4
5.5
5.6
EN
Settings: Time format
Settings: Set time
Settings: Spot (Measuring point)
The temperature is shown in the centre of the image as standard. Two spots can be added.
Max: highest temperature, Min: lowest temperature
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Time format
24 hour
AM/PM
Conrmation
Exit menu
Cancel
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Set time
Year
Month
Day
Hour
Minute
Second
Conrmation
Exit menu
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Spot
Off
On
Conrmation
Exit menu
Cancel

8
MAC: MMMMMM
5.7
5.8
EN
To evaluate the data, the ThermoCamera Connect can be connected via WLAN to a terminal device
(desktop PC or mobile phone) with WLAN capabilities. For this purpose rst select the required WLAN
SSID on the device (MAC:MMMMMM). MMMMMM corresponds to the MAC address.
Settings: WLAN connection
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
WLAN
Off
On
Conrmation
Exit menu
Cancel
The connection with the corresponding SSID is then set up at the terminal
device. With any modern browser the ThermoCamera Connect makes
the data available under the IP-address 192.168.230.1 Port 80.
Settings: Image correction
Auto shutdown
Intensity
Language
Time format
Set time
Spot
WLAN
Version
Image correction
Settings
Conrmation
Cancel

ThermoCamera Connect
9
6.0
0.95
0.95
5
6.1
V 2.3.7
5.9
EN
Version
Settings
Settings: Version
Auto shutdown
Time format
Set time
Spot
WLAN
Version
Image correction
Intensity
Language
Emissivity
Before each use, check the settings for infrared measurement and adjust them to the respective
measuring situation to ensure accurate measurements. Pay particular attention to the general
parameters for the emissivity coefcient.
Emissivity: Emissivity coefcient
The level of infrared emissions given off by everything depends on the specic material and surface. This
factor is determined by the emissivity coefcient (0.10 ... 1.0). For accurate measurements, it is absolutely
essential that the emissivity coefcient is set rst. The emissivity coefcient can either be custom set or
selected based on the predened emissivity coefcients from the material list.
Images
Color palette
Emissivity
Settings
Matt (e=0.95)
Semi-matt (e=0.85)
Semi-gloss (e=0.60)
Glossy (e=0.30)
Individual
Emissivity
Matt (e=0.95)
Semi-matt (e=0.85)
Semi-gloss (e=0.60)
Glossy (e=0.30)
Individual
Emissivity
Decrease value
Increase value Conrmation
Exit menu

10
7
EN
Emissivity table (reference values with tolerances)
Colour ranges IR image
You can choose from several standard colour ranges to represent the measured infrared temperatures.
Depending on the colour palette, the measured temperatures are adjusted within the current image
section and displayed in the respective colour space. The bargraph for the respective minimum/
maximum temperatures of the entire image serves as a reference for the corresponding temperature/
colour mapping.
Alloy A3003
Oxidised
Roughened 0.20
0.20
Aluminium
Oxidised
Polished 0.30
0.05
Brass
Polished
Oxidised 0.30
0.50
Chromium oxide 0.81
Copper
Oxidised
Copperoxide 0.72
0.78
Inconel
Oxidised
Electropolished 0.83
0.15
Iron
Oxidised
With rust 0.75
0.60
Iron, cast
Non-oxidised
Molten mass 0.20
0.25
Iron, forged
Matt 0.90
Lead
Rough 0.40
Platinum
Black 0.90
Steel
Cold rolled
Ground plate
Polished plate
0.80
0.50
0.10
Steel
Alloy (8% nickel,
18% chromium)
Galvanised
Oxidised
Heavily oxidised
Freshly rolled
Rough, at surface
Rusty, red
Sheet, nickel plated
Sheet, rolled
Stainless steel
0.35
0.28
0.80
0.88
0.24
0.96
0.69
0.11
0.56
0.45
Zinc
Oxidised 0.10
Metals
Asbestos 0.93
Asphalt 0.95
Basalt 0.70
Brick, red 0.93
Carborundum 0.90
Cement 0.95
Ceramics 0.95
China
Brilliant white
With glaze 0.73
0.92
Clay 0.95
Coal
Non-oxidised 0.85
Concrete, plaster, mortar 0.93
Cotton 0.77
Earthenware, matt 0.93
Fabric 0.95
Glass 0.90
Glass wool 0.95
Graphite 0.75
Gravel 0.95
Grit 0.95
Gypsum 0.88
Gypsum cardboard 0.95
Heat sink
Black, anodized 0.98
Human skin 0.98
Ice
Clear
With heavy frost 0.97
0.98
Laminate 0.90
Lime 0.35
Lime malm brick 0.95
Limestone 0.98
Marble
Black, dull finish
Greyish, polished 0.94
0.93
Masonry 0.93
Paint
Black, matt
Heat-resistant
White
0.97
0.92
0.90
Paper
All colours 0.96
Plastic
Translucent
PE, P, PVC 0.95
0.94
Quartz glass 0.93
Rubber
Hard
Soft, grey 0.94
0.89
Sand 0.95
Screed 0.93
Snow 0.80
Soil 0.94
Tar 0.82
Tar paper 0.92
Transformer paint 0.94
Wallpaper, light-coloured 0.89
Water 0.93
Wood
Untreated
Beech, planed 0.88
0.94
Nonmetals

ThermoCamera Connect
11
8
9
Save image?
NoYes
EN
Images
Color palette
Emissivity
Settings
Spectrum
Iron
Cool
White
Black
Color palette Conrmation
Exit menu
Cancel
Image modes
There are 5 image modes to choose from.
A. IR image (IR)
B. Digital image (Visible)
C. Digital image with IR image overlay (MIX), level 1
D. Digital image with IR image overlay (MIX), level 2
E. Digital image with IR image overlay (MIX), level 3
Using the „Trigger“ button (6),
users can take images of any
measurement situation for the
subsequent documentation.
Capture image

12
20180918-093229
10
20181022-082018
20181022-081957
20181022-073111
20181022-071151
20180918-093229
EN
The image will be deleted immediately by pressing the MENU button. There are no safety prompts.
!
Information on maintenance and care
Calibration
Clean all components with a damp cloth and do not use cleaning agents, scouring agents and solvents.
Remove the battery pack before storing for longer periods. Store the device in a clean and dry place.
Do not touch the lens.
The meter needs to be calibrated and tested on a regular basis to ensure it produces accurate
measurement results. We recommend carrying out calibration once a year.
EU directives and disposal
This device complies with all necessary standards for the free movement of goods
within the EU.
This product is an electric device and must be collected separately for disposal
according to the European Directive on waste electrical and electronic equipment.
Further safety and supplementary notices at: http://laserliner.com/info?an=AGR
Media gallery / to delete recording
All images recorded
with the ThermoCamera
Connect can be accessed
in the media gallery.
Images
Color palette
Emissivity
Settings
Delete image
Next
image
Previous
image
Conrmation /
Delete image
Exit menu
Cancel / Conrm /
delete image

ThermoCamera Connect
13
EN
Technical data Subject to technical changes without notice. 19W05
IR sensor Resolution: 220 x 160 pixels,
uncooled microbolometer, 9 Hz, 8-14 μm
IR optics High-quality IR-lens,
27° x 35° (FOV),
fixed focus, working area: 0.5 m ... 20 m
Thermal sensitivity 0.07°C @ 30°C
Precision ± 2°C or ± 2% of measured value
Measurement range -20°C ... 350°C
Display 3,2“ colour TFT
Image modes Infrared image, digital image, MIX image
Digital camera Resolution: 640 x 480 pixels
Format JPEG format
Memory function Integrated SD memory (more than 20.000 images)
Interface WLAN
Ports Micro USB charging
Emissivity coefficient 0.01 - 1.0 adjustable
Protection class IP54
Operating conditions 0°C … 45°C, max. humidity 20 … 85% rH,
no condensation, max. working altitude 2000 m above sea level
Storage conditions -20°C … 60°C, max. humidity 85% rH
Radio module operating data WLAN standard IEEE 802.11 b/g/n
Frequency band 2.400 - 2.4835 GHz (IEEE 802.11 b/g/n)
Radio channels Channel 9
Transmit power 17 dBm max.
Transmission rate IEEE 802.11 b to 11 Mbps
IEEE 802.11 g/n to 54 Mbps (at 15 ± 2 dBm)
Safety Open
Local server mode IP address 192.168.230.1; HTTP; no DHCP
Port 80
Automatic switch-off Adjustable: 5 minutes / 20 minutes / no auto power off
Power supply Li-Ion pack battery 3.5V - 4.2V / 2000 mAh
Micro USB 4.75V - 5.50V
Charging time Approx. 3 - 4 h
Operating time Approx. 2 - 3 h (depending on use)
Dimensions (W x H x D) 105 x 223 x 90 mm
Weight 389 g (incl. battery pack)