If it is necessary to enter a decimal point in the scaling factor you must use Switch A to step to the
most significant decade (even if the value for the higher decades is to remain at zero) and then press
Switch A once again to move on to the decimal point setting function.
At this point the decimal point will start to blink to show that you have reached this stage of the
scaling procedure. The decimal point should then be positioned as required by use of Switch B. At
this stage, each press of Switch B will move the decimal point one decade to the left.
On an Instrument that has been previously scaled
If a Tachometer has been supplied with factory programmed settings as requested, or once the user
has programmed the settings and returned the instrument to the RUN mode, on power up the display
will show run for one second followed by -sig (indicating NO SIGNAL) or another value if there is
already an input signal of sufficient amplitude.
If it is necessary to change the scaling factor, hold down Switch A and Switch B at the same time until
the display changes to show SET and then release the switches. The display should now show the
values previously programmed and the least significant (right hand) decade should be blinking.
To alter the scaling factor, simply press Switch B to alter the value for the decade that is blinking until
the correct figure is displayed. To move to the next highest decade in turn just press Switch A to
move from right to left.
If it is necessary to enter a decimal point in the scaling factor you must use Switch A to step to the
most significant decade (even if the value for the higher decades is to remain at zero) and then press
Switch A once again to move on to the decimal point setting function.
At this point the decimal point will start to blink to show that you have reached this stage of the
scaling procedure. The decimal point should then be positioned as required by use of Switch B. At
this stage, each press of Switch B will move the decimal point one decade to the left.
Increasing the discrimination of the display
In some applications it may be an advantage to increase the discrimination of the displayed value D.
This can be achieved by moving the decimal point when setting the scaling factor S.
If M1 is OPEN then moving the decimal point to the right will increase discrimination. If M1 is
GROUNDED then moving the decimal point to the left will increase discrimination.
See the following example:
Assume an input frequency of 50 Hz and a calculated scaling factor of
1.00, then with M1 OPEN D = S x R = 50