Programming the Instrument
Setting a scaling factor
Once a scaling factor has been set it is retained in memory upon power down. When power is
reinstated the unit will immediately be ready for operation without need for a reset command. If no
scaling factor has been set, when power is applied the unit will enter the SET mode (operation of the
unit without a pre-set scaling factor is not possible).
To access the SET mode when a scaling factor has already been set press the RUN/SET button.
The display will show SET …. SCALE (.... signifies a one second delay before the display
changes) and then show the scaling factor that is currently contained in memory. The digits can be
changed by pressing the buttons beneath each decade. The position of the decimal point is changed
by pressing the bottom right hand button to step to the position required.
Calculating the scaling factor
The displayed count in the 2000C is calculated D = Sc where Sis the Scaling Factor and cis the
number of input pulses counted — ignoring the decimal point in the display.
For example, if for each fifteen (15) pulses received you want the display to count once, the scaling
factor should be set as one fifteenth (0.06667).
Similarly, if for each pulse received you want the display to increment by ten (10) then a scaling factor
of 10 should be set.
Increasing the discrimination of the display
Once you have calculated the required scaling factor it may be advantageous to increase the
discrimination of the display. This can be achieved by moving the decimal point when setting the
scaling factor.
For example, in a system where metres need to be displayed and where twenty (20) pulses = one (1)
metre the scaling factor would be one twentieth (0.05). If this factor was used the display would
indicate metres only (displayed as 123456).
The discrimination can be increased by setting a scaling factor of 0.5 or 5, effectively multiplying the
displayed value by 10 or 100 respectively and then altering the position of the displayed decimal
point using the OFFSET facility (displayed as 12345.6 or 1234.56).