
Operating InstructionsComparison Test Pump Version 1.0 13
4. Troubleshooting measures .
If faults cannot be repaired, the system must be put out of operation
immediately and this information is to be given to the manufacturer.
Repairs must only be carried out by the manufacturer. Interventions and
changes on the appliance are not allowed.
In case of faults caused by defects of the hydraulic equipment the operators must inform their
superiors immediately and call in the qualified and authorised technical staff for maintenance.
Table: Fault description and measures .
Type of fault Measures
I. Unable to build up pressure / leak in the system
Close outlet valve correctly
Attention: Do not tighten the fine
adjustment valve more than finger tight.
Otherwise the valve seat could be
damaged.
Check whether the seals have been
placed in the quick clamp for the test
specimen and the reference instrument
and whether they are properly positioned.
Check whether there is enough operating
fluid in the tank.
II. Unable to build up pressure, or range cannot
be reached
After the clamping of the test specimen
and the reference instrument, air may be
trapped in the system.
Please note: The system should be
vented before beginning with calibration.
For this purpose, proceed according to
section 3.1.3.
Afterwards, build the pressure back up.
III. Slow lowering of the pressure
Leak in the system, see fault I.
If small amounts of air are compressed in
the system together with medium, the test
pressure that has been generated first
drops off slightly. After this it must be
readjusted.
If the system is pressurised quickly, it
takes a certain time (< 1 minute) until it
can stabilise thermally. The pressure
should then be adjusted accordingly.
After the clamping of the test specimen
and the reference instrument, air may be
trapped in the system (hydraulic design
only), see point II.
Afterwards, build the pressure back up.
Comparison Test Pump GB
LSP 1000-BM and LSP 1600-BM