
TheLumaxElectronicFlowMeterfeaturesamenuwithwhichtheusercanselectasthemainunitmeasurement:Quarts(Qts),Pints(Pts),Litres(Lit)orGallons(Gal).Thecombinationoftheunitof
measurementofthePartialregisterandthatoftheTotalsispredenedaccordingtothefollowingtable:
Combination UnitofMeasurement RegisterUnitofMeasurement
No. ofthePartialRegister oftheTotalsRegister
1 Litres (L) Litres (L)
2 Gallons (Gal) Gallons (Gal)
3 Quarts(Qts) Gallons(Gal)
4 Pints(Pts) Gallons(Gal)
Maintenance: The Lumax Electronic Flow Meter has been designed to require a
minimumamountofmaintenance.Theonlymaintenancerequiredarethefollowing:
1. BatteryChange
2. Cleaning of the Turbine
1. Battery Replacement: The Lumax Electronic Flow Meter comes complete with
2x1.5V.AlkalineBatteries(SizeAAA).LumaxElectronicFlowMeterfeatures
twolow-batteryalarmlevels:
a. WhenthebatterychargefallsbelowtherstlevelontheLCD,thexedbatterysymbolappears.Inthiscondition,
Flow Meter continues to operate correctly, but the fixed icon warns the user that it is advisable to change the batteries.
b. If Flow Meter operation continues without changing the batteries, the second battery alarm level will be reached
which will prevent operation. In this condition the battery icon starts to flash and is the only icon that remains
visibleontheLCD.TochangetheBatteries,withreferencetotheexplodeddiagrampositions,proceedasfollows:
• PressRESETtoupdateallthetotals
• Loosenthe4screwsonthebackcover
• RemovetheoldBatteries
• Placethenewbatteriesinthesamepositionastheoldones
• Closethecoveragain,bypositioningtherubberprotectionasagasket
• FlowMeterwillswitchonautomaticallyandnormaloperationcanberesumed.
TheFlowMeterwilldisplaythesameResetTotal,thesameTotalandthesamePartialindicatedbeforetheBatterieswerechanged.
AfterchangingtheBatteries,themeterdoesnotneedcalibratingagain.
Malfunctions
4. Direct modification of K Factor: This procedure is especially useful to correct a “mean error” obtainable on the basis of several performed dispensing operations. If normal Flow Meter operation
showsameanpercentageerror,thiscanbecorrectedbyapplyingtothecurrentlyusedcalibrationfactoracorrectionofthesamepercentage.Inthiscase,thepercentagecorrectionofthe“USER”K
Factormustbecalculatedbytheoperatorinthefollowingway:
NewKFactor=OldKFactorx(100-E%)
100
If the meter indicates less than the real dispensed value (negative error) the new calibration factor must be higher than the
old one as shown in the example. The opposite applies if the meter shows more than the real dispensed value (positive error).
(
(
Example:
ErrorpercentagefoundE% -0.9%
CurrentCalibrationFactor 1,000
New“USER”KFactor 1,000*[(100–(-0,9))/100]=
1,000*[(100+0,9)/100]=1.009
Action Display
6 SHORT/LONG CAL KEY KEYING
Theindicatedvaluechangesinthedirectionindicatedbythearrow:
-oneunitforeveryshort“MENU”keykeying
-continuallyifthe“MENU”keyiskeptpressed.Thespeed
increase rises by keeping the key pressed. If the desired
value is exceeded, repeat the operations from point (5).
7 LONG RESET KEY KEYING
Flow Meter is informed that the calibration procedure is
nished.Beforeperformingthisoperation,makesure
the indicated value is that required.
8 NO OPERATION
Attheendofthecalculation,thenew“USER”KFactorisshownfor
a few seconds, after which the restart cycle is repeated to finally
achieve standby condition. Attention: From now on, the indicated
factor will become the calibration factor used by the meter and will
continue to remain such even after a battery change.
9. NO OPERATION
The Flow Meter stores the new work calibration factor and
isreadytobegindispensing,usingthe“USER”KFactor
that has just been calculated.
1.003
QTS
cal
DIRECT
-----
QTS
cal
DIRECT
1.003
QTS
cal
end
0.000
QTS
13456
TOTALGAL
Action Display
1 NONE
FlowMeterinStandBy(notincountingmode)
2 LONG “CAL” KEY KEYING
FlowMeterenterscalibrationmode,shows“CAL”anddisplaysthe
Calibration Factor being used instead of the Partial. The words
“Fact”and“USER”indicatewhichofthetwoFactors(factoryoruser)
is currently being used.
3LONG RESET KEY KEYING
FlowMetershows“CAL”andthePartialatzero.FlowMeterisready
to perform on-site calibration by dispensing.
4LONG RESET KEY KEYING
WenowgoontoDirectchangeofthecalibrationfactor:Theword
“Direct” appears together with the Currently used Calibration Factor.
In the bottom left part of the display, an arrow appears (upwards or
downwards) defining the direction (increase or decrease) of change
ofthedisplayedvaluewhensubsequentoperations5or6areperformed.
5SHORT RESET KEY KEYING
Arrowdirectionchanges.Theoperationcanbe
repeated to alternate the direction of the arrow.
12.345
QTS
1234.5
TOTALGAL
1.000
(user)
cal
FACT
12.345
QTS
cal
field
1.000
QTS
cal
DIRECT
1.000
cal
DIRECT
12.345
Q
TS
23412.3
TOTAL
G
AL
Problem
LCD:Noindicationcontact
Precision Measurement not enough
ReducedorZeroFlowRate
Themeterdoesnotcount,buttheFlowRateiscorrect
Possible Cause
BadBattery
WrongKFactor
ThemeterworksbelowminimumacceptableFlowRate
TurbineBlocked
Incorrect Installation of Gears after Cleaning
Possible Electronic Card issues
Corrective Action
CheckBatteryContacts
ChecktheKFactoraccordingtotheInstructions
IncreasetheFlowRateuntilanacceptableFlowRaterangehasbeenachieved
Clean the Turbine
RepeattheReassemblyProcedure
Contact your Dealer
WARNING!:
Do not discard the old batteries in the environment.
Refertolocaldisposalregulations.
2. Cleaning:
Only one operation is necessary to clean the Flow Meter.
AfterremovingFlowMeterfromtheplantwhereitwas
built in, any residual elements can be removed by
washing or mechanically-handling. If this operation
does not restore a smooth rotation of the turbine, it will
have to be replaced.
WARNING!:
Do not use compressed air onto the turbine in order to
avoid its damage because of an excessive rotation
Form:
LX-1371-IS
© 2016 Lumax. All Rights Reserved. Lumax®and the Products described in it are Copyrighted by Lumax®,withAllRightsReserved.
ThisFormmaynotbeCopiedinWholeorPart,withouttheExpressWrittenConsentofLumax®.
Viton® ; Ryton®and Teflon®are registered trademarks of their respective companies
Tochoosebetweenthe4availablecombinations:
• WaitforFlowMetertogotoStandby
• Pressthe“MENU”and“RESET”keystogetherandholdforveseconds.Thecurrentunitofmeasurewill
begintoash.Press“RESET”tochoseadifferentunitofmeasure,thenpressMENUtoconrm.
• FlowMeterwillpassthroughthestartcycleandwillthenbereadytodispenseinthesetunits.
Note: TheResettableTotalandTotalregisterswillbeautomaticallychangedtothenewunitof
measurement.NOnewcalibrationisrequiredafterchangingtheUnitofMeasurement.
unit
QTS
GAL
unit
PTS
GAL
unit
L
L
unit
GAL
GAL
R
R
RR