Lumax LX-1371 User manual

Electronic Flow Meter
Thank you for purchasing Lumax®Electronic Flow Meter. Please read this Operational Manual carefully. It is important that you read the
entire manual to become familiar with the unit before you begin assembly and/or operate the unit. Lumax Electronic Flow Meter is ideal
for metering the transfer of fluids. It is simple and easy to use. It can be installed in-line or at the end of the delivery hose before the nozzle. Using precision turbine measuring technology, Lumax
Electronic Flow Meter measure liquids
accurately and conveniently. Large 1” (25mm) LCD display can be calibrated in Gallons, Quarts, Pints and Liters. It has an easy-to-read five-digit LCD dis
play
for accurate readings while dispensing fluids, with a Totalizer (resettable) to track total usage. The display can be rotated at any angle which simplifies the reading in any position. Field replaceable
batteries provide power and are easily accessible - simply loosen four screws on the back of the unit.
Technical Specifications:
• Construction:Ryton®PolyphenyleneSulde(PPS) • ConnectionsSize:1”NPTInlet/Outlet • FlowRange:3-30GPM(10L-120L/min)
• LCDDisplay:Upto5digits • MaximumPressure:300PSI(20Bar) • Accuracy:±1%
• MaximumLiquidTemperature:140°(60°C) • SingleCount:0.00-9999.99 • WorkingVoltage:2.3-3.3V
• AccumulativeCount:0.00-99999.9 • StandbyTime:2years • Measuresin-Quarts-Pints-Liters-Gallons
WARNING: Please read instructions carefully before using the device. To ensure operator safety and to protect the device from potential damage, you must be fully acquainted with this instruction
manual before performing any operation. The Instructions contained in this Manual must be read and understood by the operator prior to operating the device. The meter installation work shall
be performed in strict observance of the procedures and rules described in this manual. If instructions are not followed, it could result in malfunctions and damage to the device! Please store this
Operational Manual, for later reference.
• Donotexposetheelectronicmetertodirectsunlight.UseCoverorShieldforoutdoorinstallation.
• Theapparatusenclosuremaycontainaluminumandisconsideredtoconstituteapotentialriskofignitionbyimpactorfriction.Caremustbetakenintoaccountduring
installation and use to prevent impact or friction.
• Partoftheenclosureisconstructedfromplastic.Topreventtheriskofelectrostaticsparkingtheplasticsurfaceshouldonlybecleanedwithadampcloth.
• Whenmeteringammableliquids,observeprecautionsagainstreorexplosion.Donotmeterinthepresenceofanysourceofignitionincludingrunningorhotengines,
lighted cigarettes, or gas or electric heaters.
• Ifhandlinghazardousliquids,alwaysfollowthemanufacturer’ssafetyprecautions.Wearprotectiveclothingsuchasgoggles,glovesandrespiratorsasinstructed.
Alwaysdisposeofusedcleaningsolventsinasafemanneraccordingtothesolventmanufacturer’sinstructions.
• Duringmeterremoval,liquidmayspill.Followtheliquidmanufacturer’ssafetyprecautionstocleanupminorspills.
• Donotblowcompressedairthroughthemeter. • Donotsubmergethemeter.
• Donotallowliquidstodryinsidethemeter. • Donotuseawrenchtoinstallplasticmeters.Handtightenonly.
WARNING: NOT APPROVED FOR USE IN EXPLOSIVE ATMOSPHERES OR HAZARDOUS LOCATIONS
Wetted Components: Ryton®PolyphenyleneSulde(PPS);StainlessSteel(SS316);Viton®(Fluoropolymer Elastomer); Teflon®(PTFE)
Materials (Can be Pumped): AntiFreeze(Alcohol&GlycolBase);DetergentSolutions;FruitJuices;SaltWater;SeaWater(Brine);SoapSolutions;Urea/DEF/AdBlue;Veg
etableJuices;Water:Fresh
and Distilled (Deionized), Water-basedMildChemicals,LowViscosityOilorPetroleum-basedproductssuchasKerosene,Diesel,and Gasoline,Wines.
Warning: AlthoughGasolineisoilbased,itisExplosive&ExtremelyDangeroustopump.GroundWiresareessentialandrequiredforGasolineandFlammables
Warning:Thisproductisnotintendedforpotablewaterapplications(humanconsumption–drinkingandcooking)andhasnotbeendesignedtobecompliantwiththe“SafeDrinkingWaterAct”
requirements for low lead in potable water applications. This item is for use only in non-potable (non-human consumption) water applications.
Materials (Cannot be Pumped): Acetone,Ammonia,BleachSolutions,BrakeFluid(HexyleneGlycol)Chlorine,HydraulicOil,Lacquers&LacquerSolventsandMethanol
Caution: The above Materials List is to be referred for guidance only in selecting equipment for appropriate chemical compatibility. ALWAYS test your equipment under the specific conditions of your
applicationbeforepermanentinstallation.Ratingsofchemicalbehaviorlistedinthismanualapplyata48hourexposureperiod.Lumaxhasnoknowledgeofpossibleeffectsbeyondthisperiod.All
recommendationsarebasedonroomtemperature(72°F/22°C)exposure.Compatibilityresultsmaydeviatefromtheserecommendationsatelevatedtemperatures.Recommendationsarebasedon
presence of listed chemical only. Mixtures containing more than one chemical can greatly affect chemical compatibility. Polypropylene products and components should not be used with low flash
point chemicals, regardless of chemical compatibility results. Failure, improper selection, or improper use of the products referred to and described herein or related items can cause severe personal
injury or property damage. This document and other information provides product options for further investigation by users having technical expertise. It is important that you analyze all aspects of
your application and review the information concerning the product or system.
Due to the variety of operating conditions and applications for these products, the user, through his or her own analysis and testing, is solely responsible for making the final selection of the products
and assuring that all performance, safety and warning requirements are met. The materials described and referred to herein are subject to change at any time without notice. Lumax does not warrant
(neither express nor implied) that the information in this chart is accurate or complete or that any material is suitable for any purpose. If there is any question about any Material that can or cannot be
pumped, do not operate the device. Check with Lumax as to whether it is safe or unsafe, to operate the device.
Caution: Always checklocalareaLawsandOrdinancesregardingHandling,PumpingandDisposingofLiquids.Disposeofpumpeduidproperly.Do Not drainintoStreetDrains,HouseDrains,or
Backyards.ContactyourlocalMunicipalWasteManagementofcialsforproperdisposalmethods.If there is any question about a condition being safe or unsafe, do not operate the meter.
Before Installation: Uponreceipt, examineyour meterfor visibledamage. Iftheitemisdamaged, contactyour distributor.Make surethe metermodel meetsyour specicneeds. Referto the
SpecicationsSectionandconrmthefollowing:
1. The flow-rate is within the limits of your model. 2. Theliquidiscompatiblewithyourmeter’smaterial.
3. Thesystem’spressuredoesnotexceedthemeter’smaximumpressurerating.
Connections: 1. To protect against leakage, make sure all threads are sealed with two or three turns of thread tape or a sealing compound compatible with the liquid being metered.
Caution: Make sure the thread tape or sealing compound does not interfere with flow.
2. Make sure the arrow on the outlet is pointing in the direction of the flow. 3. Tightenthemeterontothettings.Handtighten.Usingawrenchcoulddamagethemeter.
Display Positioning: Square shape housing allows the LCD Display to be rotated in any position for easy display readings. The LCD Display Card is easily accessible, by removing the faceplate cover
and unscrewing the four screws. The LCD Display can be rotated in any direction to suit the piping system.
Caution: Make sure you remove the LCD Display Card very slowly and
carefullytoensurethatanydamagetotheWires,Connections,
or the LCD Display Card is avoided.
Operating Modes:Choosebetweentwomodesofoperation:
-NormalMode:DisplayofPartialandTOTAL amount dispensed
-FlowRateMode:DisplayofFlowRateandPartialdispensedquantities
The meter features a non-volatile memory, for storing the dispensing data
for longer periods in the case of power outage.
The electronic measurements are displayed in the upper part of
the LCD Display. They are isolated from the liquid measuring Chamber.
LCD Display:TheLCDDisplayoftheLumaxElectronicFlowMetersfeaturestwonumericalregistersandseveralfunctionorstatusindicators:
Key:
1. PartialRegister(5digitfrom0.1to9999.9)indicatingthevolumedispensedsincetheresetbuttonwaslastpressed.ForPartialRESET-Press“RESET”Buttonfor1Second
2.BatteryChargeIndicator
3. Calibration Mode Indicator
LX-1371
RESET
MENU
RESET
MENU
RESET
MENU
RESET
MENU

4.TotalizerRegister(6guresfrom0.1to99999.9),thatcanindicatetwotypesofTotals:
4.1 General TOTALthatcannotbereset(TOTAL)
4.2 ResettableTOTAL(ResetTOTAL)
ForTotalizerRegister-Press“RESET”Buttonfor3Seconds
5.IndicatorofTotalMultiplicationFactor(x10/x100)
6. Display of type of TOTAL,(TOTAL/RESETTOTAL)
7.DisplayofMeasurementofTotals:L=Liters;Gal=Gallons;Qts=Quarts;Pts=Pints
8.DisplayofFlowRate:3-30GPM(10L-120L/min)
9. DisplayofUnitofMeasureoftheSubset:L=Liters;Gal=Gallons;Qts=Quarts;Pts=Pints
User Buttons: LumaxElectronicFlowMetersfeaturesTwoButtons(RESETandMENU)whichindividuallyperform
twomainfunctionsandtogether,othersecondaryfunctions.Themainfunctionsperformedare:
- “RESET”Key:ResettingthePartialRegisterandresettableTOTAL (RESETTOTAL)
- “MENU”Key:EnteringCalibrationMode.
Usedtogether(PressbothButtonsfor5Seconds),thetwokeyspermitenteringCongurationmode,usefulforchangingtheunitsofmeasurementsandcalibrationfactor.
Battery Housing:LumaxElectronicFlowMeterusestwostandard1.5VpowerBatteries(size:AAA).BatteryHousingiseasilyaccessible-simplyloosenfourscrewsonthebackoftheDevice.Caution:
Make sure you remove the BatteriesveryslowlyandcarefullytoensurethatanydamagetotheWires,Connections,ortheLCDDisplayCardisavoided.
Installation: TheinletandoutletfortheFlowMeteris1“NPT.ItcanbeeasilyconnectedtoaPipeorNozzle.Itcanbeinstalledinanydesiredpositionandissuitableforstationaryormobileapplications.
Donotuseawrenchtoinstallthemeter.Handtightenonly.Tip: To extend the life of the device, it is recommended to install a filter on inlet side of the meter.
1. Daily Use: The only operation that is necessary for daily use are, partial and/or resettable total register resetting. Occasionally the meter may be necessary to be configured or calibrated. For this,
pleasefollowtheinstructionsinthismanual.Below,arethetwotypicaldisplayguresdisplayedinnormaloperation.
FigureontheleftdenotesthePartialandtheResettableTotalregisters.Thegureontheright
denotesPartialandNon-resettableTOTAL.Resettabletotalamountaddstonon-resettableTOTAL
amount automatically. This is preset at the manufacturing plant and cannot be modified.
NOTE:6digitsareavailableforTotals,plustwoiconsx10/x100.Theincrementsequenceasfollows:
0.0-99999.9-999999-100000x10-999999x10-100000x100-999999x100
Dispensing in Normal Mode: Normalmodeisthestandarddispensing.Whilethecountis
made,thepartialandresettabletotalaredisplayedatthesametime:ResetTotal(Fig. 1).
Should one of the keys be accidentally pressed during dispensing, this will have no effect.
Afewsecondsafterdispensinghasended,onthelowerregister,thedisplayswitchesfrom
ResettableTOTAL to General TOTAL:thewordResetabovethewordTOTAL disappears, and
the RESETTOTAL is replaced by the General TOTAL (Fig. 2). This situation is called standby
and remains stable until the user operates the Flow Meter again.
Partial RESET:ThePartialRegistercanberesetbypressingtheRESET key when the meter
isinstandby,meaningwhenthedisplayscreenshowstheword“TOTAL”(Fig. 3).
AfterpressingtheRESET key, during reset, the display screen first shows all the lit-up digits
and then all the digits that are not lit up (Fig. 4).
Attheendoftheprocess,adisplayisrstshownwiththeRESET Partial and the RESET Total (Fig. 5)
and, after a few moments, the RESETTOTALisreplacedbytheNon-ResettableTotal(Fig. 6).
Resetting the RESET TOTAL: The RESET TOTAL resetting operation can only be performed after
resetting the partial register. The RESETTOTAL can in fact be reset by pressing the RESET key at
length while the display screen shows RESETTOTAL (Fig. 7).Schematically,thestepstobetakenare:
1. WaitforthedisplaytoshowNormalStandbydisplaypage-withtotalonlydisplayed(Fig. 8).
2. Press the RESET key quickly
3. The meter starts to reset the Partial
4.WhilethedisplaypageshowingtheRESETTOTAL is displayed (Fig. 9). Press the RESET key
again for at least 1 second
5. The display screen again shows all the segments of the display followed by all the
switched-off segments and finally shows the display page where the reset
“RESETTOTAL” (Fig. 10) is shown.
Dispensing with Flow Rate Mode Display: It is possible to dispense fluids,
displaying at the same time, as shown (Fig. 11):
* the dispensed partial
*theFlowRatein[PartialUnit/Minute]
Procedure for entering this Mode:
-waitfortheRemoteDisplaytogotoStandby,meaningthedisplayscreenshowsTOTAL only
-quicklypresstheMENUkey.
- Start dispensing
TheFlowRateisupdatedevery0.7seconds.Consequently,thedisplaycouldberelativelyunstableatlowerFlowRates.ThehighertheFlowRate,themorestablethedisplayedvalue.IMPORTANT -The
FlowRateismeasuredwithreferencetotheunitofmeasurementofthePartial.Forthisreason,incaseoftheunitofmeasurementofthePartialandTOTAL being different, as in the example shown
below,itshouldberememberedthattheindicatedFlowRaterelatestotheunitofmeasurementofthepartial.Intheexampleshown,theFlowRateisexpressedinQts/min.Theword“Gal”remaining
alongsidetheFlowRatereferstotheregisteroftheTotals(ResetorNONReset)whichareagaindisplayedwhenexitingfromtheFlowRatereadingmode(Fig. 12).
Toreturnto“Normal”mode,presstheMENUkeyagain.IfoneofthetwokeysRESETorMENUisaccidentallypressedduringthecount,thiswillhavenoeffect.
Important: Even though in this mode they are not displayed, both the RESETTOTAL and the General Total (TOTAL) increase. Their value can be checked after dispensing has terminated, returning to
“Normal”mode,byquicklypressingMENUkey.
Partial RESET:ToresetthePartialRegister,nishdispensingandwaitfortheRemoteDisplaytoshowaFlowRateof0.0asindicatedintheillustrationthen
quicklypressRESET(Fig. 13).
2. Calibration
Definitions: Calibration factor or “k factor”: Multiplication factor applied by the system to the electrical pulses received, to transform these into measured fluid units.
FACTORY “K FACTOR”:Factory-setdefaultfactor.Itisequalto1,000.Thiscalibrationfactorensuresutmostprecisioninthefollowingoperatingconditions:
Fluid: Diesel Fuel Temperature:20°C(68°F) Flow Rate:10-120litres/min.(2.65-31.75Gal.)
Even after any changes have been made by the user, the factory k factor can be restored by means of a simple procedure.
USER K FACTOR: Customized calibration factor, meaning modified by calibration.
Why Calibrate?: Whenoperatingclosetoextremeconditions,suchasforinstancewithuidsclosetoacceptablerangeextremes(likeDieselFuelatLowTemperatures)orinextremeFlowRate
conditions (close to minimum or maximum acceptable values), an on-site calibration may be required to suit the real conditions in which the Flow Meter is required to operate.
Fig. 8 Fig. 9
Fig. 7
Fig. 11 Fig. 12
Fig. 10
4 5 6
9
7
1
2
3
8
88,8,8,8
88888,8
RESET
TOTAL
x100
FLOW RATE
QTS
PTS
GAL
L
GAL
L
cal
12,345
QTS
12,3
RESET
TOTAL GAL
12,345
QTS
12,3
TOTAL GAL
Partial Register
GENERAL
TOTAL
Register
(TOTAL)
RESET TOTAL
(Register Total)
RESET TOTAL
Register
Fig. 2
12.345
QTS
12.3
TOTAL GAL
Fig. 3
12.345
QTS
TOTAL GAL
23412.3
Fig. 1
12.345
QTS
12.3
RESET
TOTAL GAL
Fig. 4
88,8,8,8
88888.8
RESET
TOTAL
x100
FLOW RATE
QTS
PTS
GAL
L
GAL
L
c
al
Fig. 5
0.000
QTS
TOTAL GAL
23412.3
Fig. 6
23412.3
0.000
QTS
RESET
TOTAL GAL
23412.3
0.000
QTS
RESET
TOTAL
GAL
23412.3
12.345
QTS
12.3
TOTAL GAL
2345.67
0.000
QTS
RESET
TOTAL
GAL
0.0
0.000
QTS
RESET
TOTAL GAL
12.345
12.3
FLOW RATE L
QTS
GAL
12.345
12.5
FLOW RATE
Fig. 13 L
L
12.345
0.0
FLOW RATE

Calibration Procedure: The Lumax Electronic Flow Meter permits making quick and precise electronic calibration by changing the calibration factor (k factor). There are 2 dif
ferentwaysofcalibration:
1. On-site calibration, performed by means of a dispensing operation. 2. Direct calibration, performed by directly changing the k factor.
To enter the calibration phases it is necessary to press and hold down the “Menu” button.
Whyenterthecalibrationphases?
• Displaythecurrentlyusedcalibrationfactor • Returntofactorykfactorafterapreviouscalibrationwithuserkfactor
• Changethecalibrationfactorusingoneofthetwopreviouslyindicatedprocedures.
In calibration mode, the partial and total dispensed quantities indicated on the display screen take on different meanings according to the calibration procedure phase. During the calibration, the Flow
Meter cannot perform any normal dispensing operations. In calibration mode, the totals are not increased.
Caution: The Lumax Electronic Flow Meter features a non-volatile memory. It keeps the calibration and dispensing
data stored even after replacing new batteries or long periods of inactivity.
3. Display of Current “K Factor” and Restoring “Factory K Factor”:
BypressingtheMENUkeywhiletheapplianceisinstandby(PressButtonfor3Seconds),
the display page appears showing the current calibration factor used. If you are using the Flow Meter with
“factory k factor”, the display page shown in the diagram will be displayed, with the word “fact” (Fig. 14).
Ifone“userkfactor”hasbeenset,thecalibrationfactorsetbytheuser(inourexample0.998)willbedisplayed.
The word “user” indicates a calibration factor set by the user is being used (Fig. 15).
The Flow chart alongside shows the switchover logic from one display page to another. In this condition,
theResetkeypermitsswitchingfrom“User”factorto“Factory”factor.
To confirm the choice of calibration
factor,quicklypressMENUwhile“User”or“Fact”aredisplayed.Aftertherestartcycle,themeterusesthe
calibration factor that has just been confirmed.
Attention:Whenthe“FactoryFactor”isconrmed,theold“UserFactor”isdeletedfromthememory.
In-field Calibration: This procedure calls for the fluid to be dispensed into a graduated sample
containerinrealoperatingconditions(FlowRate,Viscosity,etc.)requiringabsoluteprecision.
WARNING:
ForcorrectFlowMetercalibration,itismostimportantto:
• Completelyeliminateairfromthesystembeforecalibrating
• UseapreciseSampleContainerwithacapacityofnotlessthan5liters,featuringanaccurate
Graduated indicator.
• Ensurecalibrationdispensingisdoneataconstantowrateequivalenttothatofnormaluse,
until the container is full
• Notreducetheowratetoreachthegraduatedareaofthecontainerduringthenaldispensing
stage (the correct method during the final stages of sample container filling consists in making
short top-ups at normal operation flow rate)
• Afterdispensing,waitafewminutestomakesureanyairbubblesareeliminatedfromthesample
container;onlyreadtheRealvalueattheendofthisstage,duringwhichthelevelinthecontainercoulddrop.
• Ifnecessary,carefullyfollowtheprocedureindicatedbelow.
Fig. 14
1.000
FACT
cal
Fig. 15
0.998
user
cal
12.345
QTS
23412.3
TOTALGAL
c
al
1.000
FACT
c
al
12.345
USER
c
al
1.000
FACT
12.345
QTS
23412.3
TOTALGAL
R
R
C+C
C
STAND BY
STAND BY
C
R+R Long RESET
RShort RESET
C+C Long CAL
CShort CAL
Time Out
LEGEND:
In-field Calibration Procedure
Action Display
1 NONE
FlowMeterin“StandBy”(notincountingmode)
2 LONG MENU KEY KEYING
FlowMeterenterscalibrationmode,shows”MENU”
and displays the calibration factor in use instead
oftotal.Thewords“Fact”and“USER”indicate
which of the two factors is currently in use.
3LONG RESET KEY KEYING
FlowMetershows“MENU”andthepartialatzero.
Flow Meter is ready to perform on-site calibration.
4DISPENSING INTO SAMPLE CONTAINER
WithoutpressinganyKEY,startdispensingintothesamplecontainer.
Dispensing can be interrupted and started again at will. Continue dispensing until the
level of the fluid in the sample container has reached the graduated area. There is no
need to reach a preset quantity.
5SHORT RESET KEY KEYING
Flow Meter is informed that the calibration dispensing
operation is finished. Make sure dispensing is correctly
finished before performing this operation. To calibrate
the Flow Meter, the value indicated by the partial totalizer
(example9.800)mustbeforcedtotherealvaluemarked
on the graduated sample container. In the bottom left part of the display an
arrow appears (upwards and downwards), that shows the direction (increase
ordecrease)ofthe“USER”KFactorvaluechangewhentheoperations6or7
are performed.
Action Display
6 SHORT RESET KEY KEYING
Arrowdirectionchanges.Theoperationcanberepeated
if necessary.
7 SHORT/LONG CAL KEY KEYING
The indicated value changes in the direction indicated
bythearrow-oneunitforeveryshort”MENU”key
keying-continuallyifthe”MENU”keyiskeptpressed
(for the first 5 units slowly and then quickly). If the desired
valueisexceeded,repeattheoperationsfrompoint(6).
8 LONG RESET KEY KEYING
Flow Meter is informed that the calibration procedure is
nished.Beforedoingthis,makesuretheDisplayed
FactoristheActualFactor.
FlowMetercalculatesthenew“USER”KFactor.Thiscalculationcouldrequirea
few seconds, depending on the correction to be made. During this operation
thearrowdisappearsbutthe”MENU”indicationremains.Ifthisoperationis
performedafteroperation(5),withoutchangingtheindicatedvalue,the“USER”
KFactorwouldbethesameastheFactoryKFactor,thusitisignored.
9 NO OPERATION
Attheendofthecalculation,thenew“USER”KFactorisshownfor
a few seconds, after which the restart cycle is repeated to finally
achieve standby condition. ATTENTION: From now on, the indicated
factor will become the calibration factor used by the meter and
will continue to remain such even after a battery change.
10 NO OPERATION
Flow Meter stores the new calibration factor and is ready
fordispensing,applyingthenewlydened“USER”KFactor
12.345
QTS
12.5
TOTAL GAL
1.000
QTS
GAL
(user)
cal
FACT
0.000
QTS
cal
field
9.800
QTS
cal
field
9.800
QTS
cal
field
9.800
QTS
cal
field
9.800
QTS
cal
field
-----
QTS
cal
END
9.860
QTS
c
al
FACT
Indicated Value Real Value
9.86
9.800
QTS
cal
0.0000
GAL
Indicated Value Real Value
9.86
1.015
QTS
cal
END
0.000
QTS
1234.5
TOTALGAL
cal

TheLumaxElectronicFlowMeterfeaturesamenuwithwhichtheusercanselectasthemainunitmeasurement:Quarts(Qts),Pints(Pts),Litres(Lit)orGallons(Gal).Thecombinationoftheunitof
measurementofthePartialregisterandthatoftheTotalsispredenedaccordingtothefollowingtable:
Combination UnitofMeasurement RegisterUnitofMeasurement
No. ofthePartialRegister oftheTotalsRegister
1 Litres (L) Litres (L)
2 Gallons (Gal) Gallons (Gal)
3 Quarts(Qts) Gallons(Gal)
4 Pints(Pts) Gallons(Gal)
Maintenance: The Lumax Electronic Flow Meter has been designed to require a
minimumamountofmaintenance.Theonlymaintenancerequiredarethefollowing:
1. BatteryChange
2. Cleaning of the Turbine
1. Battery Replacement: The Lumax Electronic Flow Meter comes complete with
2x1.5V.AlkalineBatteries(SizeAAA).LumaxElectronicFlowMeterfeatures
twolow-batteryalarmlevels:
a. WhenthebatterychargefallsbelowtherstlevelontheLCD,thexedbatterysymbolappears.Inthiscondition,
Flow Meter continues to operate correctly, but the fixed icon warns the user that it is advisable to change the batteries.
b. If Flow Meter operation continues without changing the batteries, the second battery alarm level will be reached
which will prevent operation. In this condition the battery icon starts to flash and is the only icon that remains
visibleontheLCD.TochangetheBatteries,withreferencetotheexplodeddiagrampositions,proceedasfollows:
• PressRESETtoupdateallthetotals
• Loosenthe4screwsonthebackcover
• RemovetheoldBatteries
• Placethenewbatteriesinthesamepositionastheoldones
• Closethecoveragain,bypositioningtherubberprotectionasagasket
• FlowMeterwillswitchonautomaticallyandnormaloperationcanberesumed.
TheFlowMeterwilldisplaythesameResetTotal,thesameTotalandthesamePartialindicatedbeforetheBatterieswerechanged.
AfterchangingtheBatteries,themeterdoesnotneedcalibratingagain.
Malfunctions
4. Direct modification of K Factor: This procedure is especially useful to correct a “mean error” obtainable on the basis of several performed dispensing operations. If normal Flow Meter operation
showsameanpercentageerror,thiscanbecorrectedbyapplyingtothecurrentlyusedcalibrationfactoracorrectionofthesamepercentage.Inthiscase,thepercentagecorrectionofthe“USER”K
Factormustbecalculatedbytheoperatorinthefollowingway:
NewKFactor=OldKFactorx(100-E%)
100
If the meter indicates less than the real dispensed value (negative error) the new calibration factor must be higher than the
old one as shown in the example. The opposite applies if the meter shows more than the real dispensed value (positive error).
(
(
Example:
ErrorpercentagefoundE% -0.9%
CurrentCalibrationFactor 1,000
New“USER”KFactor 1,000*[(100–(-0,9))/100]=
1,000*[(100+0,9)/100]=1.009
Action Display
6 SHORT/LONG CAL KEY KEYING
Theindicatedvaluechangesinthedirectionindicatedbythearrow:
-oneunitforeveryshort“MENU”keykeying
-continuallyifthe“MENU”keyiskeptpressed.Thespeed
increase rises by keeping the key pressed. If the desired
value is exceeded, repeat the operations from point (5).
7 LONG RESET KEY KEYING
Flow Meter is informed that the calibration procedure is
nished.Beforeperformingthisoperation,makesure
the indicated value is that required.
8 NO OPERATION
Attheendofthecalculation,thenew“USER”KFactorisshownfor
a few seconds, after which the restart cycle is repeated to finally
achieve standby condition. Attention: From now on, the indicated
factor will become the calibration factor used by the meter and will
continue to remain such even after a battery change.
9. NO OPERATION
The Flow Meter stores the new work calibration factor and
isreadytobegindispensing,usingthe“USER”KFactor
that has just been calculated.
1.003
QTS
cal
DIRECT
-----
QTS
cal
DIRECT
1.003
QTS
cal
end
0.000
QTS
13456
TOTALGAL
Action Display
1 NONE
FlowMeterinStandBy(notincountingmode)
2 LONG “CAL” KEY KEYING
FlowMeterenterscalibrationmode,shows“CAL”anddisplaysthe
Calibration Factor being used instead of the Partial. The words
“Fact”and“USER”indicatewhichofthetwoFactors(factoryoruser)
is currently being used.
3LONG RESET KEY KEYING
FlowMetershows“CAL”andthePartialatzero.FlowMeterisready
to perform on-site calibration by dispensing.
4LONG RESET KEY KEYING
WenowgoontoDirectchangeofthecalibrationfactor:Theword
“Direct” appears together with the Currently used Calibration Factor.
In the bottom left part of the display, an arrow appears (upwards or
downwards) defining the direction (increase or decrease) of change
ofthedisplayedvaluewhensubsequentoperations5or6areperformed.
5SHORT RESET KEY KEYING
Arrowdirectionchanges.Theoperationcanbe
repeated to alternate the direction of the arrow.
12.345
QTS
1234.5
TOTALGAL
1.000
(user)
cal
FACT
12.345
QTS
cal
field
1.000
QTS
cal
DIRECT
1.000
cal
DIRECT
12.345
Q
TS
23412.3
TOTAL
G
AL
Problem
LCD:Noindicationcontact
Precision Measurement not enough
ReducedorZeroFlowRate
Themeterdoesnotcount,buttheFlowRateiscorrect
Possible Cause
BadBattery
WrongKFactor
ThemeterworksbelowminimumacceptableFlowRate
TurbineBlocked
Incorrect Installation of Gears after Cleaning
Possible Electronic Card issues
Corrective Action
CheckBatteryContacts
ChecktheKFactoraccordingtotheInstructions
IncreasetheFlowRateuntilanacceptableFlowRaterangehasbeenachieved
Clean the Turbine
RepeattheReassemblyProcedure
Contact your Dealer
WARNING!:
Do not discard the old batteries in the environment.
Refertolocaldisposalregulations.
2. Cleaning:
Only one operation is necessary to clean the Flow Meter.
AfterremovingFlowMeterfromtheplantwhereitwas
built in, any residual elements can be removed by
washing or mechanically-handling. If this operation
does not restore a smooth rotation of the turbine, it will
have to be replaced.
WARNING!:
Do not use compressed air onto the turbine in order to
avoid its damage because of an excessive rotation
Form:
LX-1371-IS
© 2016 Lumax. All Rights Reserved. Lumax®and the Products described in it are Copyrighted by Lumax®,withAllRightsReserved.
ThisFormmaynotbeCopiedinWholeorPart,withouttheExpressWrittenConsentofLumax®.
Viton® ; Ryton®and Teflon®are registered trademarks of their respective companies
Tochoosebetweenthe4availablecombinations:
• WaitforFlowMetertogotoStandby
• Pressthe“MENU”and“RESET”keystogetherandholdforveseconds.Thecurrentunitofmeasurewill
begintoash.Press“RESET”tochoseadifferentunitofmeasure,thenpressMENUtoconrm.
• FlowMeterwillpassthroughthestartcycleandwillthenbereadytodispenseinthesetunits.
Note: TheResettableTotalandTotalregisterswillbeautomaticallychangedtothenewunitof
measurement.NOnewcalibrationisrequiredafterchangingtheUnitofMeasurement.
unit
QTS
GAL
unit
PTS
GAL
unit
L
L
unit
GAL
GAL
R
R
RR
Table of contents