Maxcare KYX8012-USB User manual

User Manual
Model: KYX8012(D)-USB
Thank for purchasing Automatic Digital Blood Pressure
Monitor. It is mainly designed for general home use.
Before using the device, please read this manual carefully
to ensure proper and safe operation.
Please take good care of the manual for future reference.
Welcome your advice and support.
Automatic Digital
Blood Pressure Monitor

Table of Contents
About the Products
1.Brief Introduction of the Product..............................................01
2.The Constitution of the Product...............................................01
How to Use
1.Batteries Installation......................................................................02
2.Unit(mmHg/ kPa)Display Setup and Memory Deletion ............02
3.Cuff Connection.............................................................................03
4.Fitting the Arm Cuff........................................................................03.
5.Adjust Your Sitting Posture............................................................04
6.Taking a Measurement..................................................................05
7.Reading Memory...........................................................................06
8.Manual and Auto Inflation..............................................................07
9.Advice for Self-measurement........................................................07
.
Common Sense of Blood Pressure
1.About Blood Pressure...................................................................07
2.Table for Classifying Blood Pressure Values(mmHg)...............08
3.Life Guide for Hypertension Patients............................................09
Table of Units Conversion...................................................................09
Safety Instruction................................................................................10
Abnormal Phenomenon and Solutions................................................11
Packing List.........................................................................................11
Maintenance and Storage...................................................................12
Technical Specifications......................................................................13
Guidance and Manufacturer’s Declaration..........................................14

About the Products
1.Brief Introduction of the Product
This new digital blood pressure monitor is an automatic measuring
device to be used on the upper arm. It enable simple, accurate and
fast measurement of blood pressure and pulse rate by use of the
oscillometric method and people-orientated design.
The advantages of this product:
1.Fully automatic
2.Motion detection
3.Irregular heartbeat detection
4.60 Memories places
5.WHO indicator
2. The Constitution of the Product
01
Display window
USB charging
ON/OFF Button
Memory Button
Air tube connection
MEM SET
ON
OFF
SET Button

How to Use
1.Batteries Installation
Press the indicator ▼ on the battery cover and slide the cover off in
the direction of the arrow
Insert 4 “AA” size batteries, align the polarities of the battery.
Note:
If the low battery indicator appears in the display, the batteries must be
replaced.This monitor can also be operated with USB charging
WARNING
Pls purchase AC adapter (USB port) with CE certification.
Do not plug or unplug the power cord into the electrical outlet with wet hands.
02
2. Unit (mmHg/ kPa) Display Setup and Memory Deletion.
In power off condition, press SET button for 3 seconds. Enter into
the setup interface, the content of setup will be displayed by flashing.
Press the SET button again, then it will enter into the switch of setup
content “memory deletion ->unit display->speaker function(option).”
Press MEM button to revise setup value.
Press ON/OFF button to store the setup content.
Note:
This monitor displays two units: mmHg/ kPa, the user manual takes
mmHg as an example to illustrate.The default unit is mmHg.

3.Cuff Connection
Insert the air tube plug into the socket by revolving it.
4.Fitting the Arm Cuff
Take off thick clothes, wear thin clothes or unclothe your upper arm
for measurement. Lay the cuff on the upper arm. Turn the palm of
your hand upward. Put one side of the cuff (with Velcro) through the
metal ring to form a cylinder.
03
Tighten the cuff by affixing the Velcro.
The cuff should be snug on your upper arm.
Make sure that the bottom of the cuff lies approximately 1 to 2 cm
above the elbow and the air tube lies on the inner of the arm. The
tube should be aligned with your middle finger after it has been
straightened.

Note:
Both of your arms can be measured, but there are about 10mmHg
differences between them. Each measurement should be made
with the same arm.
5. Adjust Your Sitting Posture
Sit on a chair with your feet flat on the floor. Put your arm on a table
with your hand upward, relax and keep the cuff parallel with your
heart.
The arm cuff should be at the
same level as your heart.
Sit upright with
your back straight.
Take off thick clothes, wear
thin clothes or bare your
upper arm for measurement
6.Taking a Measurement
(1). Press ON/OFF button to power on, The pump starts to inflate
the cuff.
04
ON
OFF

(2). After having reached the anticipated value of inflation, the
monitor starts to deflate automatically with a constant speed to detect
your blood pressure and pulse.
(3). When the measurement is completed, your
blood pressure and pulse rate will be displayed.
Example:
SYS: 128mmHg
DIA: 78mmHg
Pulse: 68/min
The result will be stored automatically. The storage series
number is “0”. The previous measurement serial number is
“1”, and so forth.
(4). To shut off the monitor, press ON/OFF buttom.
05

Note:
Once you feel uncomfortable with inflation or the inflation becomes
ceaseless, please loosen the cuff and pull out the air tube
immediately.
Don’t talk, move your arm or body during the measurement.
06
The monitor will be automatically shut off in 3 minutes without
operation. When the monitor can’t measure the value correctly, the
screen will display “E” icon.
Press ON/OFF button to shut off the monitor and wait for 4 to 5
minutes to start a new test. At the same time, follow the following
suggestions during the measurement to avoid “E” icon display again.
1. Keep quiet and correct posture.
2. Fit the cuff correctly.
3. Connect the air tube correctly.
4. Avoid the electromagnetic interference.
7. Reading Memory
Press MEM button to check memory.
Press MEM and SET button to read more stored memories.
Note:
The monitor can memorize 60 data. The monitor will delete the earliest
data automatically when the number of data is beyond 60.
SETMEM

8. Manual and Auto Inflation
● In most of cases, choose auto inflation. If your systolic pressure is
more than 220mmHg, choose manual inflation.
9. Advice for Self-measurement
● Avoid eating, smoking, drinking alcohol, exercising, bathing as
well as forms of exertion activities before measurement.
● Keep quiet and relax for measurement.
● Keep correct posture during the measurement.
●Take measurement on the same arm (normally left) and posture
regularly at the same time of the day.
● Do not take the measurement in an extreme temperature.
The digital blood pressure monitor could pre-detect your blood
pressure during the process of inflation, and set specific inflation
value for your measurement. The measuring time is shortened
and every user could enjoy quick and comfortable measurement.
Common Sense of Blood Pressure
1. About Blood Pressure
● In the process of inflation, press MEM button for manual inflation.
The preset value will be 260mmHg.
The preset value will add 20mmHg by each touch.
07
Blood is carried from the heart to all parts of your body through
vessels called arteries. Blood pressure is the force of the blood
pushing against the walls of the arteries.

Each time the heart beats, it pumps out blood into the arteries.
Blood pressure is at its highest when the heart pumps or ejects blood.
This is called Systolic Blood Pressure.
Blood pressure is at its lowest when the heart rests ( in- between beats).
This is called Diastolic Blood Pressure.
The blood pressure is constantly changing. The different cuff fitting
and measurement posture can cause the blood pressure to change.
You should take the measurement at the same time with the same
arm and posture everyday.
2. Table for Classifying Blood Pressure Values
(mmHg): WHO
08
120 130 140 160 180
110
105
100
95
90
85
80 Normal BP
High-normal BP
Mild Hypertension
Moderate Hypertension
Severe Hypertension
Systolic
(mmHg)
Diastolic
(mmHg)
Blood Pressure
Classification
SBP
mmHg
DBP
mmHg
Color
Indicator
Optimal 120
120-129
80 Green
Normal 80-84 Green
130-139High-normal 85-89 Green
140-159Grade 1 Hypertension 90-99 Yellow
160-179Grade 2 Hypertension 100-109 Orange
180Grade 3 Hypertension 110 Red

3. Life Guide for Hypertension Patients
※High blood pressure (Hypertension);
A blood pressure of 140/90 or higher is considered high blood
pressure. High blood pressure (HBP) is a serious condition that can
lead to coronary heart disease, heart failure, stroke, kidney failure,
and other health problems. Blood pressure tends to rise with age.
Following a healthy lifestyle helps some people delay or prevent this
rise in blood pressure. The following measures can be regarded as a
life guide for hypertension patients:
1) Consult your doctor about your blood pressure in daily life.
2) Maintain a normal weight.
3) Avoid excessive consumption of common salt. Avoid high cholesterol
foods and fatty foods. Eat more fruits, vegetables and high-fiber foods.
4) Avoid smoking, alcohol and reviver.
5) Moderated exercise.
6) relax and reduce stress. Stress can increase the blood pressure
suddenly. Relax and reduce stress is a good habit for your health.
Table of Units Conversion(mmHg and kPa)
10mmHg =1.3KPa 150mmHg =20.0KPa
20mmHg =2.7KPa 160mmHg =21.3KPa
30mmHg =4.0KPa 170mmHg =22.7KPa
40mmHg =5.3KPa 180mmHg =24.0KPa
50mmHg =6.7KPa 190mmHg =25.3KPa
60mmHg =8.0KPa 200mmHg =26.7KPa
70mmHg =9.3KPa 210mmHg =28.0KPa
80mmHg =10.7KPa 220mmHg =29.3KPa
90mmHg =12.0KPa 230mmHg =30.7KPa
100mmHg =13.3KPa 240mmHg =32.0KPa
110mmHg =14.7KPa 250mmHg =33.3KPa
120mmHg =16.0KPa 260mmHg =34.7KPa
130mmHg =17.3KPa 270mmHg =36.0KPa
140mmHg =18.7KPa 280mmHg =37.3KPa
09

Safety Instructions
Warning
1) Hypertension can not be judged by the measured value through this
product. The value is used for monitoring blood pressure.
2) Self-diagnosis of measurement results and self-treatment are
dangerous. Contact your physician about the measured value.
3) Do not use the device on the injured arm or the arm under medical
treatment.
4) Operate the device only as intended. Do not use the device for any
other purpose.
4) Do not disassemble or attempt to repair the unit or components.
5) If battery fluid should get in your eyes, immediately rinse with plenty
of clean water. Contact a physician immediately.
6) Pls purchase AC adapter with CE certification. Do not plug or unplug
the power cord into the electrical outlet with wet hands.
7) WARNING: No modification of this equipment is allowed.
Caution:
1). This device is intended for use in measuring blood pressure and pulse
rate in the adult population. Do not use this device on infants or
persons who cannot express their intentions.
2) People with severe blood flow problems, or blood disorders, should
consult a physician before using the device, as the arm cuff inflation
can cause bruising.
3) Do not use this product on patients with severe arrhythmia, infants and
perple who can not express their intentions.
4). Dispose of the device, components and optional accessories
accordin to applicable local regulations. Unlawful disposal maycause
environmental pollution.
5). Wireless communications equipment such as wireless home network
devices, mobile phones, cordless telephones and their base stations,
walkie-talkies can affect this equipment and should be kept at least a
distance d = 3,3 m away from the equipment.
10
6) Once you feel uncomfortable with inflation or the inflation becomes
ceaseless, please loosen the cuff and pull out the air tube immediately.

Packing list
A automatic digital blood pressure monitor arm cuff Storage bag
User manual Batteries(optional) AC adapter(optional)
Abnormal phenomenon and Solutions
Abnormal
Phenomenon Potential cause Solutions
Power on
doesn’t work
Low battery or wrong
polarities of batteries aligned
Replace old batteries with new
ones; Align the polarities of the
battery correctly.
Power off
when inflation
Low battery Replace old batteries with new
ones
No inflation Air tube plug is not correctly
connected
Check the connection of the
air tube and connect properly
if necessary.
Abnormal value
Cuff is not fitted on the arm
correctly
Fit the cuff correctly
Talk during the measurement Keep quiet during measurement
Roll-up sleeve pressed
on the arm
Take off the clothes on the arm,
restart measuring
Nervous or excited Keep quiet, take a deep breath
to calm down
Wrong measurement Adjust the posture
ERROR INDICATORS
TROUBLE SHOOTING
7) Do not use the monitor if you find it is damaged or notice anything
unusual.
8) Do not take measurement more than necessary. It may cause bruising
due to blood flow interference.
11
Symbol Cause Solutions
E4
Air tube plug is not
correctly connected
Check the connection of the air
tube and connect it properly
if necessary. Restart measuring.
The monitor or arm is moved
during the measurement
Keep your body still, never move
the monitor. Restart measuring.
The inflation is 30mmHg lower
than expected systolic value
Manually inflate to meet the
requirement. Restart measuring
E5
The inflation is above 300
mmHg
The device will deflate automaticly
E1
Failure inflation Fit the cuff correctly

Maintentance and storage
1) Do not subject the monitor and package to shocks, such us dropping it onto
the floor.
2) Do not submerge the device or any of the components in water. Do not
subject the monitor to extreme temperatures, dust, humidity, corrosive gas
and drect sunlight, do not use it in an oxygen rich environment.
3) Use the cuff carefully. The cuff contains a sensitive gasbag and should not
be forcely bended.
4) Clean the monitor with soft and dry cloth. Don’t use gasoline, thinners or
similar solvents.
5) Carefully remove spots on the cuff with a damp cloth and soapsuds.
It should not be washed with detergent powder and other detergents.
6) Use only approved parts and accessories. Not approved parts and
accessories may damage the unit.
7) Store the device and the components in a clean, safe location.
8) The device doesn’t need calibration within 4 years liftime, If you want to
continue to use, calibrate the device once every two years, by the dealer or
the local qualified measurement institute for calibration, the calibration
method will be supported by manufacture.
9) About batteries: If the monitor is not to be used for a long time, remove the
batteries to prevent battery leakage.
12

Technical Specifications
13
Model
Display
Accuracy
Work Mode
Measuring Method
Power source
Operating Condition
Storage Condition
Weight
Dimension
Cuff
Cuff Circumference
KYX8012(D)-USB
LCD digital display
Pressure: within
±
0.4kPa (3mmHg
)
Pulse:
within
±
5% of the reading
Automatic inflation and measurement
Oscillometric method
DC 6V (4X “AA”batteries )
Temperature: 5
℃~
40
℃
Humidity: 15%RH~85%RH
Air Pressure: 860hPa~1060hPa
300g(without batteries)
120mmx93mmx59mm
Temperature: -20
℃~
70
℃
Humidity: 10%RH~95%RH
Air Pressure: 860hPa~1060hPa
BF applied part
size 23~33cm
size 25~35cm (optional)
size 22~42cm (optional)
Accessories Arm cuff
Measuring Range
Pressure:
0kPa~38.6kPa (0mmHg~290mmHg)
Pulse: 40~180/min

This product needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided, and this unit can
be affected by portable and mobile RF communications equipment.
Guidance and manufacture’s declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the user of the
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The device use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
(Battery operated
device)
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
(Battery operated
device)
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Guidance and Manufacturer’s Declaration
Important information regarding Electro Magnetic Compatibility (EMC)
a) * Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
b) Caution: This unit has been thoroughly tested and inspected to assure proper
performance and operation!
c) * Caution: this machine should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this machine should
be observed to verify normal operation in the configuration in which it will be used.
14
device

Guidance and manufacture’s declaration –
electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
(Battery operated
device)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Not applicable
(Battery operated
device)
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Not applicable
(Battery operated
device)
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the
device
requires continued
operation during power mains
interruptions, it is recommended
that the
device
be powered from
an uninterruptible power supply or
a battery.
Power frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
15

Guidance and manufacture’s declaration – electromagnetic immunity
The
device
is intended for use in the electromagnetic environment specified below. The customer or the
user of the
device
should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not
applicable
(Battery
operated
device)
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment should
be used no closer to any part of the device , including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
Where Pis the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and dis the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
16

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the device is used exceeds the
applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the device .
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the device .
T
he device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Th
e
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
(m)
Rated maximum output
power of transmitter
(W)
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
reflection from structures, objects and people.
and
17

EXPLANATION OF SAFETY SIGNS AND SYMBOLS
Means type BF applied part.
Symbol for “THE OPERATION GUIDE MUST BE READ”
Symbol for “MANUFACTURER”
Symbol for “COMPLIES WITH MDD 93/42/ECC REQUIREMENTS”
DISPOSAL: Do not dispose this product as unsorted municipal waste.
Collection of such waste separately for special treatment is necessary.
The first number 2:Protected against access to hazardous parts with a finger,
and the jointed test finger of 12mm¢,80mm length, shall have adequate
clearance from hazardous parts. And protected against solid foreign objects of
12,5mm¢and greater.The second number 2: Protected against vertically falling
water drops when enclosure titled up to 150. Vertically falling drops shall have
no harmful effects when the enclosure is titled at any angle up to 15”on either
side of the vertical.
IP22:
Version 1.0
This manual suits for next models
1
Table of contents