R203M13-001 REV. A ENGLISH
AIR-OXYGEN BLENDER (DISS AND NIST CONNECTIONS)
MicroMax™
LOW FLOW / HIGH FLOW
Operating Manual & Instructions For Use
R203P14 / R203P13

866.4.MAXTEC WWW.MAXTEC.COMii
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
Maxtec TEL: 1(800) 748.5355
2305 South 1070 West FAX: 1(801) 973.6090
Salt Lake City, Utah 84119 USA email: sales@maxtec.com
website: www.maxtec.com
AUTHORIZED REPRESENTATIVE:
RECEIVING / INSPECTION
Remove the MicroMax Air-Oxygen Blender from the packaging and inspect for damage. If there is any
damage, DO NOT USE and contact your Provider.
INTENDED USE
The MicroMax Air-Oxygen Blender dispenses a continuous and precise blend of medical air and USP
oxygen via outlet ports to infant, pediatric and adult patients. The exact Fractional Concentration of
Inspired Oxygen (FIO2) blend of gases corresponds to the dialed in FIO2setting indicated by the con-
trol knob (dial).
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Air-Oxygen Blender. This is provided for
your safety and to prevent damage to the Air-Oxygen Blender. If you do not understand this manual,
DO NOT USE the Air-Oxygen Blender and contact your Provider.
DANGER
This product is not intended as a life-sustaining or life-supporting device.
EC REP QNET BV
Kantstraat 19
NL-5076 NP Haaren
The Netherlands

866.4.MAXTEC WWW.MAXTEC.COMiii
EXPLANATION OF ABBREVIATIONS
FIO2Fractional Concentration of Inspired Oxygen
DISS Diameter Indexed Safety System
NIST Non-Interchangeable Screw Thread
psi Pounds Per Square Inch
l/min Liters Per Minute
SAFETY INFORMATION—WARNINGS AND CAUTIONS
DANGER Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
WARNING Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
CAUTION Used without the safety alert symbol indicates a potentially hazardous situation
which, if not avoided, may result in property damage.
OR CONSULT ACCOMPANYING DOCUMENTS
Symbol for “USE NO OIL”
Symbol indicates the device complies with the requirements of Directive 93/42/
EEC concerning medical devices and all applicable International Standards.
Symbol for DO NOT USE
WARNINGS
Only trained, qualified medical personnel under the direct supervision of a licensed physician should
operate the Air-Oxygen Blender .
Use this Air-Oxygen Blender only for its Intended Use as described in this manual.
Confirm prescribed dose before administering to patient. Monitor on a frequent basis.
The Air-Oxygen Blender shall be serviced by a qualified service technician.
Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters and Oxygen Handling.
The stand-alone gas mixer is not suitable for use with O2 93.
An Oxygen Analyzer/Monitor must be used to verify oxygen concentration.

866.4.MAXTEC WWW.MAXTEC.COMiv
Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15
l/min for the High Flow Blender, and 3 l/min for the Low Flow Blender.
DO NOT obstruct the alarm.
DO NOT use Blender when alarm is sounding.
DO NOT use oil in or around the Blender.
DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender.
DO NOT use near any type of flame or flammable/explosive substances, vapors or atmosphere.
Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over
100% oxygen will damage the Blender.
CAUTION
Turn off gas supplies when Air-Oxygen Blender is not in use.
Store the Air-Oxygen Blender in a clean, dry area when not in use.
The Air-Oxygen Blender contains magnetic, ferrous material that may affect the results of an MRI.
Ensure all connections are tight and leak free.
Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets.
DO NOT steam autoclave.
DO NOT immerse Air-Oxygen Blender into any liquid.
DO NOT gas sterilize with (EtO) Ethylene Oxide.
DO NOT use if dirt or contaminants are present on or around the Blender or connecting devices.
DO NOT smoke in an area where oxygen is being administered.
DO NOT clean with aromatic hydrocarbons.
Inlet pressure of device used in conjunction with Blender must match inlet pressure of Blender.
When using a bottled high pressure gas source, always use a pressure reducing regulator set within
30-75 psi (2.1-5.2 bar).

866.4.MAXTEC WWW.MAXTEC.COMv
SPECIFICATIONS
Model MicroMax High Flow
R203P14
MicroMax Low Flow
R203P13
Primary Outlet Flow Range
15 - 120 l/min 3 - 30 l/min
With both supply pressures at 50 psi (3.4 bar)
with BLEED closed
Auxiliary Outlet Flow Range
2 - 100 l/min 0 - 30 l/min
With both supply pressures at 50 psi (3.4 bar) with BLEED open
Bleed Flow 13 l/min or less
at 50 psi (3.4 bar)
3 l/min or less
at 50 psi (3.4 bar)
Maximum Combined
Flow (All Outlets) ≥120 l/min ≥30 l/min
Bypass Flow (Loss of Air or
Oxygen supply) > 85 l/min > 45 l/min
Bypass Alarm Activation
50 psi (3.45 bar) 60 psi
(4.14 bar)
50 psi
(3.45 bar)
60 psi
(4.14 bar)
13-25 psi
16-24 psi 18-22 psi 16-24 psi
0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar
Alarm Reset: When pressure dierential is 6 psi (0.4 bar) or less.
Alarm Sound Level: ≥to 80 db at 1 ft (0.3 m)
Oxygen Concentration
Adjustment Range: 21 - 100%
Gas Supply Pressure: 30 - 75 psi (2.1 - 5.2 bar)
Air and Oxygen within 10 psi (0.69 bar)
of each other
Mixed Gas Stability: ±1% Oxygen
Connection Types:
DISS Type - Air & Oxygen Inlets & Outlets
and / or NIST Type - Air &
Oxygen Inlets
NOTE: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).
DIMENSIONS: (without fittings)
Depth:
4.9 in (12.5 cm)
Width:
2.3 in (5.7 cm)
Height:
4.1 in (10.4 cm)
Weight: 2.29 lbs (1.04 kg)
Shipping Weight: 2.95 lbs (1.34 kg)
Operating Temperature Range: 59°F to 104°F (15°C to 40°C)

866.4.MAXTEC WWW.MAXTEC.COM1
Transport / Storage Requirements
TEMPERATURE RANGE: -10°F to 140°F (-23°C to 60°C)
HUMIDITY: Max 95% Noncondensing
FIO2ACCURACY:* ± 3% of full scale
PRESSURE DROP:
Low Flow: ≤2 psi (0.14 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
10 l/min flow rate at 60% FIO2.
High Flow: ≤3 psi (0.21 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
30 l/min flow rate at 60% FIO2.
The Air-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 80601-2-55.
DRYNESS AND COMPOSITION FOR INLET GASES:
Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973
commodity specification for Air, type 1 grade D or better.
Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen
Grade N.
Dew Point: Both inlets should remain 10°F (5.55°C) or more below the lowest
temperature to which the air distribution system equipment is
exposed. At a temperature of 25°F (-3.9°C) and a pressure of
90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.
* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 l/min for
the high flow Blender, and 3 l/min for the low flow Blender.
* Specifications are subject to change without prior notice.

866.4.MAXTEC WWW.MAXTEC.COM2
DIAGRAMS
CAUTION
Missing or illegible labels must be replaced, contact Maxtec. Depending on model, your fittings and/
or labels may differ from these diagrams.
R203P13 Model
R203P14 Model
a
h
cd
g
b
e
f

866.4.MAXTEC WWW.MAXTEC.COM3
COMPONENT DESCRIPTION
ITEM DESCRIPTION
A
Oxygen Concentration Dial
A dial used for selecting oxygen concentrations between 21%-100%. The FIO2scale is used for
reference only. This Dial does not rotate 360°. The dial starts at 21% and ends at 100%.
B
Primary Outlet Port
A male DISS oxygen fitting with check valve that delivers flow when engaged to any
controlling device, such as a flowmeter.
C
Auxiliary Outlet Port
A male DISS oxygen fitting with check valve that delivers flow when engaged to any
controlling device, such as a flowmeter. This outlet is equipped with a bleed valve that allows
the user to control if the bleed is ON or OFF. With the bleed in the ON position, this outlet
delivers accurate oxygen concentrations in the following flows:
Model Flow Range
High Flow 2 – 100 l/min
Low Flow 0 – 30 l/min
D
Auxiliary Bleed Collar
The collar is used to engage and disengage the bleed. The bleed is necessary to maintain
accurate FIO2Concentration below 15 l/min for the High Flow and
≤3 l/min for the Low Flow. To activate the bleed, slide the knurled collar back until it
contacts the cover. To deactivate the bleed, pull the collar away from cover until it reaches a
positive stop.
E
Oxygen Inlet Fitting
A female DISS or NIST oxygen fitting with one way valve that is used to connect an oxygen
supply hose.
FAir Inlet Fitting
A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose.
G
Alarm
An audible alarm that sounds due to an excessive pressure drop or deletion of either gas
supply.
H Rear Slide Mount with dove tail.

866.4.MAXTEC WWW.MAXTEC.COM4
PRE-USE TESTING
WARNING
•Read this User Manual before installing or operating the Air-Oxygen Blender.
•Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor.
CAUTION
Inspect the Air-Oxygen Blender for visual damage before use, DO NOT USE if damaged.
NOTE: The tests listed below should be performed prior to placing the Blender in service.
Pre-Use Testing consists of:
•Alarm Test
•Reverse Gas Flow Procedure
1. Secure the Air-Oxygen Blender to a wall or pole bracket in an upright position.
2. It is recommended to install a condensation trap in the air supply line.
3. Connect the air and oxygen supply lines to the appropriate inlet fittings on the bottom of the
Blender.
4. Attach a flowmeter, or other metering device to one of the outlet ports and verify FIO2range
for accuracy with an oxygen analyzer.
Primary Outlets Flow capacity:
•High Flow Blender (R203P14 Model) 15 l/min to 120 l/min
•Low Flow Blender (
R203P13 Model
) 3 l/min to 30 l/min
Auxiliary Outlet:
The auxiliary flow outlet maintains the same flow capacity and FIO2accuracy as the Primary Outlets
with Bleed Valve not engaged. When bleed flow is activated, some of the air/oxygen mixture will vent
to atmosphere to maintain FIO2concentration accuracy at the Low Flow settings.
•High Flow Blender (R203P14 Model) 15 l/min or less
•Low Flow Blender (R203P13 Model) 3 l/min or less
5. Attach a supply line to the outlet port of the flowmeter.
ALARM TEST
1. Connect the Air-Oxygen Blender to air and oxygen sources, pressurize the Blender and turn
“ON” the flowmeter.
2. Set Oxygen Concentration Dial to 60% FIO2.
3. Disconnect or turn “OFF” the air supply to the Air-Oxygen Blender. The Blender should alarm
with a loud whistle noise. The whistle indicates the alarm is operating correctly.
4. Reconnect and activate the air supply line to the Blender, the alarm should stop whistling.

866.4.MAXTEC WWW.MAXTEC.COM5
5. Disconnect or turn “OFF” the oxygen supply line to the Blender. The whistle indicates the
alarm is operating correctly.
6. Reconnect and activate the oxygen supply line to the Blender, the alarm should stop whistling.
7. If alarm fails to function properly, DO NOT USE.
REVERSE GAS FLOW PROCEDURE
1. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the
Blender to ensure that there is no outlet flow.
2. While gradually increasing the air supply pressure from 30-75 psi (2.07-5.17 bar) check for
leakage past the oxygen inlet check valve.
3. Replace the Duckbill Check Valve in the oxygen inlet if leakage is > 100 ml/min. Reference Air-
Oxygen Blender Service Manual (P/N R203M10).
4. Repeat steps 1-3 to check for leakage past the air inlet check valve.
OPERATING INSTRUCTIONS
CAUTION
Inspect the Air-Oxygen Blender for visual damage before use, DO NOT USE if damaged.
1. Secure Blender to wall or pole mount bracket.
2. Connect Air and Oxygen supply lines from Blender to wall outlets.
3. Connect flowmeter to Blender outlet.
4. Adjust the Oxygen Concentration Dial to the prescribed concentration.
NOTE: The Oxygen Concentration Dial does not rotate 360°. DO NOT force dial less than 21% or
over 100% oxygen, this will damage the Blender.
5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor. If necessary
activate bleed flow valve to maintain FIO2accuracy.
7. To activate the bleed, slide the knurled collar back until it contacts the cover.
8. To deactivate the bleed, pull the collar away from the cover until bleed flow valve is closed.
9. Turn “OFF” gas supplies when Air-Oxygen Blender is not in use.

866.4.MAXTEC WWW.MAXTEC.COM6
CLEANING
CAUTION
DO NOT steam autoclave.
DO NOT immerse the Air-Oxygen Blender into any liquid.
DO NOT use any strong solvent or abrasive cleaners.
DO NOT gas sterilize with (EtO) Ethylene Oxide.
DO NOT clean with aromatic hydrocarbons.
DO NOT allow the liquid to penetrate the device.
DO NOT gas or heat sterilize.
1. Disconnect all gas connections and equipment before cleaning.
2. Close Bleed Valve to ensure O-Ring is not exposed.
3. Clean exterior surfaces using Super Sani-Cloth germicidal disposable wipes. Remove all visible
contamination from the external surfaces of the device and its accessories. Be sure to closely
inspect and remove contamination from seams and recesses on the device that may trap
contaminants. Wipe with clean paper towel to remove debris.
4. After all visible contamination is removed; use a second germicidal wipe to thoroughly wet
the surfaces of the device and accessories. Allow to remain wet for 4 minutes. Use additional
wipes if needed to assure surfaces are wetted continuously for 4 minutes.
5. Allow device to air dry completely.
6. Visually inspect the device for visible contamination. Repeat cleaning/disinfection process if
necessary.
MAINTENANCE
The following maintenance on the Air-Oxygen Blender must be performed by a trained service
technician:
•The alarm should be tested prior to being placed into
clinical service and periodically there after.
•Every year conduct the Operational Verification Procedure (OVP).
*A detailed description of the OVP tests can be found in the Blender Service
Manual (P/N R203M10), available on the Internet; www.maxtec.com.
•Every 2 years the Air-Oxygen Blender should be serviced.
•Refer to the Air-Oxygen Blender Service Manual (P/N R203M10) for
complete details regarding further maintenance and testing.

866.4.MAXTEC WWW.MAXTEC.COM7
TECHNICAL DESCRIPTION
For a complete Technical Description of the Air-Oxygen Blender and list of Replacement Parts, refer-
ence the Air-Oxygen Blender Service Manual (P/N R203M10) available on the Internet; www.maxtec.
com.
RETURNS
Returned products require a Returned Material Authorization (RMA) number, contact Maxtec. All
returns must be packaged in sealed containers to prevent damage. Maxtec will not be responsible for
goods damaged in transit. Refer to Maxtec Return Policy available on the Internet; www.maxtec.com.
Manuals available on our website; www.maxtec.com
DISPOSAL INSTRUCTIONS
This device and its packaging contain no hazardous materials. No special precautions need to be
taken when disposing the device and/or its packaging.
PLEASE RECYCLE

866.4.MAXTEC WWW.MAXTEC.COM8
TROUBLESHOOTING
If the Air-Oxygen Blender fails to function, consult the Troubleshooting Guide below. If problem can-
not be solved by using Troubleshooting Guide, refer to the Air-Oxygen Blender Service Manual (P/N
R203M10) available on the Internet; www.maxtec.com or consult your Provider.
Problem Probable Cause Remedy
Oxygen concentration
discrepancy between Blender
setting and Analyzer/Monitor
(greater than 3%)
1. High Flow model, flow
requirement below 15 l/min.
Low Flow model, flow
requirement below 3 l/min.
2. Analyzer/Monitor inaccurate
3. Low flow bleed obstructed
4. Gas supply contaminated
5. Downstream device causing
back flow or restricted flow
1. Use auxiliary outlet &
engage bleed
2. Recalibrate Analyzer/
Monitor or Verify with
second Analyzer/Monitor
3. Remove obstruction
4. Check gas sources with
calibrated Oxygen Analyzer/
Monitor to confirm Oxygen
is 100% and Air is 21%
5. Isolate Blender. Check
oxygen concentration at
Blender Outlets
No flow at
Blender outlets
1. Gas sources turned “OFF”
2. Gas sources not connected
1. Turn gas sources “ON”
2. Connect gas sources
Alarm sounding 1. Dierence between Oxygen
and air inlet pressures
greater than specified
1. Correct pressure dierence
until Air and Oxygen
pressures are within
specification

866.4.MAXTEC WWW.MAXTEC.COM9
LIMITED WARRANTY
AND
LIMITATION OF LIABILITY
Maxtec warrants that the Blender, (the Product), will be free of defects
in workmanship and/or material for the following period:
Two (2) years from shipment
Should any failure to conform to this warranty appear within the
applicable period, Maxtec shall, upon written notification thereof and
substantiation that the goods have been stored, installed, maintained
and operated in accordance with the product instructions and standard
industry practice, and that no modifications, substitutions, or alterations
have been made to the goods, correct such defect by suitable repair or
replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Maxtec or any retailers are not authorized to
make oral warranties about the merchandise described in this contract,
and any such statements shall not be relied upon and are not part of
the contract for sale. Thus, this writing is a final, complete and exclusive
statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY
OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR
OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Maxtec shall not under any circumstances be liable for special, incidental
or consequential damages including but not limited to lost profits, lost
sales, or injury to person or property. Correction of non-conformities
as provided above shall constitute fulfillment of all liablities of Maxtec
whether based on contract, negligence, strict tort or otherwise. Maxtec
reserves the right to discontinue manufacture of any product or change
product materials, designs, or specifications without notice.
Maxtec reserves the right to correct clerical or typographical errors
without penalty.
507377 REV2 02/19 PRINTED IN USA

R203M13-001 REV. A FRENCH
MÉLANGEUR AIR-OXYGÈNE (RACCORDS DISS ET NIST)
MicroMax™
LOW FLOW / HIGH FLOW
Manuel Utilisateur • Conserver Ces Instructions
R203P14 / R203P13 (montré)

866.4.MAXTEC WWW.MAXTEC.COM11
ATTENTION: Conformément à la loi fédérale (États-Unis), ce produit
ne peut être vendu que par un médecin ou sur son ordonnance.
Maxtec TEL: 1(800) 748.5355
2305 South 1070 West FAX: 1(801) 973.6090
Salt Lake City, Utah 84119 USA email: sales@maxtec.com
website: www.maxtec.com
REPRÉSENTANT AUTORISÉ:
RÉCEPTION/INSPECTION
Retirer le mélangeur air-oxygène de MicroMax de son emballage et vérifier l’absence de dommages.
En cas de dommages, NE PAS UTILISER et contacter le fournisseur de matériel.
UTILISATION PRÉVUE
Le mélangeur air-oxygène de MicroMax fournit un mélange précis et en continu d’air de qualité médi-
cale et d’oxygène USP via des orifices de sortie à des patients nouveaux-nés, pédiatriques et adultes.
Le mélange de gaz exact de la concentration partielle d’oxygène inspiré (FIO2) correspond à celui qui
apparaît sur le cadran pour le réglage du FIO2indiqué par le bouton de contrôle (cadran).
LIRE TOUTES LES INSTRUCTIONS AVANT UTILISATION
Ce manuel indique à un professionnel comment installer et faire fonctionner le mélangeur air-oxygène.
Il est fourni pour garantir la sécurité d’emploi et éviter tout dommage au mélangeur. Si les instruc-
tions de ce manuel ne sont pas claires, NE PAS UTILISER l’appareil et contacter le fournisseur de
matériel.
DANGER
Cet appareil n’est pas conçu pour servir d’appareil de survie ou de maintien en vie.
EC REP QNET BV
Kantstraat 19
NL-5076 NP Haaren
The Netherlands

866.4.MAXTEC WWW.MAXTEC.COM12
EXPLICATION DES ABRÉVIATIONS
FIO2Concentration partielle d’oxygène inspiré
DISS Système de sécurité indexé sur le diamètre
NIST Filetage de vis non interchangeable
psi Livres par pouce carré
l/min Litres par minute
INFORMATIONS LIÉES À LA SÉCURITÉ—
AVERTISSEMENTS ET MISES EN GARDE
DANGER Indique une situation de danger imminent qui, si elle n’est pas évitée, peut
entraîner la mort ou des blessures graves.
avertissement Indique une situation de danger potentiel qui, si elle n’est pas évitée, peut
entraîner la mort ou des blessures graves.
attention Indique une situation de danger potentiel qui, si elle n’est pas évitée, peut
entraîner des blessures légères ou modérées.
attention Sans le symbole de danger, indique une situation de danger potentiel qui, si
elle n’est pas évitée, peut entraîner des dommages matériels.
OR CONSULTER LA DOCUMENTATION JOINTE
Symbole signifiant « NE PAS UTILISER D’HUILE »
Symbole signifiant que l’appareil satisfait aux exigences de la directive 93/42/
CEE relative aux appareils médicaux et à toutes les normes internationales
correspondantes.
Symbole signifiant « NE PAS UTILISER »
AVERTISSEMENT
Seul un personnel entraîné et habilité sous la supervision directe d’un médecin doit faire fonctionner
le mélangeur air-oxygène.
N’utiliser ce mélangeur air-oxygène que pour son utilisation prévue décrite dans ce manuel.
Vérifier la dose prescrite avant de l’administrer au patient et la contrôler fréquemment.
L’entretien du mélangeur air-oxygène doit être confié à un technicien d’entretien qualifié.
Toujours respecter les normes ANSI et CGA pour les produits gazeux à usage médical et la manipu-
lation des débitmètres et de l’oxygène.
La concentration d’oxygène doit être contrôlée au moyen d’un analyseur/moniteur d’oxygène.

866.4.MAXTEC WWW.MAXTEC.COM13
La précision de la concentration en oxygène est faussée si la purge n’est pas activée pour des
réglages de débit inférieurs à 15 L/min pour le mélangeur à débit élevé et 3 L/min pour le mélangeur
à faible débit.
NE PAS obstruer l’alarme.
NE PAS utiliser le mélangeur lorsque l’alarme retentit.
NE PAS utiliser d’huile sur ou à proximité du mélangeur.
NE PAS boucher ou obstruer l’orifice de purge au niveau de la sortie auxiliaire du mélangeur.
NE PAS utiliser à proximité d’un quelconque type de flamme ou de substances, vapeurs ou atmo-
sphères inflammables/explosives.
Le cadran de concentration en oxygène n’effectue pas une rotation complète (360 °). La rotation du
cadran à moins de 21 % ou à plus de 100 % endommagera le mélangeur.
ATTENTION
Fermer toutes les alimentations de gaz lorsque le mélangeur air-oxygène n’est pas utilisé.
Ranger le mélangeur air-oxygène dans un endroit propre lorsqu’il n’est pas utilisé.
Le mélangeur air-oxygène contient des matières magnétiques et ferreuses pouvant fausser les résul-
tats d’une IRM.
Vérifier que tous les raccordements sont bien serrés et ne présentent pas de fuites.
Éviter des sautes de pression supérieures à 100 psi (6,9 bar) lors de la mise sous pression des entrées
du mélangeur.
NE PAS stériliser à l’autoclave.
NE PAS immerger le mélangeur air-oxygène dans un liquide.
NE PAS stériliser à l’oxyde d’éthylène (EtO).
NE PAS utiliser en cas de présence d’impuretés ou de contaminants sur ou autour du mélangeur
ou des raccords.
NE PAS fumer dans une zone d’administration d’oxygène.
NE PAS nettoyer avec des hydrocarbures aromatiques.
La pression d’entrée du dispositif utilisé en conjonction avec le mélangeur doit correspondre à celle
de ce dernier.
Lors de l’utilisation d’une source de gaz haute pression en bouteille, toujours utiliser un détendeur
réglé sur 2,1-5,2 bar (30-75 psi).

866.4.MAXTEC WWW.MAXTEC.COM14
CARACTÉRISTIQUES TECHNIQUES
Modèle MicroMax
Débit élevé R203P14
MicroMax
Faible débit R203P13
Sortie principale
Plage de débits
15 à 120 L/min 3 à 30 L/min
Avec les deux pressions d’alimentation à 50 psi (3,4 bar) SANS
PURGE
Sortie auxiliaire
Plage de débits
2 à 100 L/min 0 à 30 L/min
Avec les deux pressions d’alimentation à 50 psi (3,4 bar)
PURGE
Débit de purge 13 L/min ou moins
à 50 psi (3,4 bar)
3 L/min ou moins
à 50 psi (3,4 bar)
Débit combiné
maximal (toutes sorties) ≥120 L/min ≥30 L/min
Débit de dérivation
(perte d’alimentation d’air ou
d’oxygène)
> 85 L/min > 45 L/min
Activation de l’alarme de
dérivation
50 psi (3,45 bar) 60 psi
(4,14 bar)
50 psi
(3,45 bar)
60 psi
(4,14 bar)
13 à 25 psi 16 à 24 psi 18 à 22 psi 16 à 24 psi
0,9 à 1,7 bar 1,1 à 1,65 bar 1,2 à 1,5 bar 1,1 à 1,65 bar
Réinitialisation de l’alarme : Lorsque la diérence de pression est égale ou inférieure à 6 psi
(0,4 bar) ou moins.
Niveau sonore de l’alarme : ≥à 80 db à 30 cm (1 pi)
Plage de réglage de la
concentration d’oxygène : de 21 à 100 %
Pression d’entrée du gaz : de 30 à 75 psi (de 2,1 bar à 5,2 bar) Air et oxygène à 10 psi
(0,69 bar) l’un de l’autre
Stabilité du mélange de gaz : ±1 % oxygène
Types de raccords : Type DISS - Entrées et sorties d’air et d’oxygène et/ou
Type NIST - Entrées d’air et d’oxygène
REMARQUE: Toutes les valeurs de débit sont mesurés à partir d’un débitmètre d’oxygène (non corrigées).
DIMENSIONS: (SANS LES RACCORDS)
Profondeur : 12,5 cm (4,9 po)
Largeur : 5,7 cm (2,3 po)
Hauteur : 10,4 cm (4,1 po)
Poids: 1,04 kg (2,29 lb)
Poids d’embarquement : 1,34 kg (2,95 lb)
Plage de températures de fonctionnement: 15 °C à 40 °C (59 °F à 104 °F)

866.4.MAXTEC WWW.MAXTEC.COM15
Exigences de transport/stockage
Plage de températures: -10°F to 140°F (-23°C to 60°C)
Humidité: Max 95% Noncondensing
Précision FIO2:* ± 3% of full scale
Chute de pression:
Faible débit: ≤2 psi (0.14 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
10 l/min flow rate at 60% FIO2.
Débit élevé: ≤3 psi (0.21 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
30 l/min flow rate at 60% FIO2.
Le mélangeur air-oxygène a été dégraissé pour service d’oxygène avant la livraison.
Le débit de gaz inverse du mélangeur air-oxygène est conforme à l’alinéa 6 de la norme ISO 11195.
L’analyseur d’oxygène doit être conforme à la norme ISO 21647 pour répondre aux impératifs CE.
SÉCHERESSE ET COMPOSITION POUR LES GAZ D’ENTRÉE:
Air: L’alimentation en air de qualité médicale doit satisfaire aux exigences
de la norme ANSI Z86.1-1973 de spécification de produit pour l’air,
type 1 qualité D ou meilleure.
Oxygène: L’alimentation en oxygène doit satisfaire aux exigences de l’oxygène
de qualité médicale USP qualité N.
Point de
rosée:
Les deux entrées doivent rester 5,55 °C (10 °F) ou plus en dessous
de la température la plus basse à laquelle est exposé l’équipement du
système de distribution de l’air. À une température de -3,9 °C (25 °F)
et une pression de 90 psi (6,33 kg/cm2) cela équivaut à 2 000 mg/
m3.
* La précision de la concentration en oxygène est faussée si la purge n’est pas activée pour des réglages de débit
inférieurs à 15 L/min pour le mélangeur à débit élevé et 3 L/min pour le mélangeur à faible débit.
Les caractéristiques techniques sont sujettes à modification sans avertissement préalable.
This manual suits for next models
1
Table of contents
Languages:
Popular Blender manuals by other brands

Blendtec
Blendtec Wildside Plus Designer Series Owner's manual and user's guide

Kalorik
Kalorik USK BL 1 operating instructions

croydon
croydon TI8SL instruction manual

ARENDO
ARENDO 303714/20191218NB084 user manual

TREVIDEA
TREVIDEA Girmi FR02 user manual

Black & Decker
Black & Decker Power Pro BLP14600DG Use and care book