Now the display shows you the main menu:
You have to wait until the temperature has reached 37.3°C, otherwise “TEMP“is flashing.
4. Test
Prepare your reagent and place it into the heated reagent stations (=upper row). The second row is
for preheating up the cuvettes.
Place one cuvette into the measuring channel . The measuring programme will be activated if you press
the left button. The system will start the measuring programme and “RESULT/ 0.0“ will flash up for
5 sec. During these 5 sec you pipette the sample and start the stored incubation time with the centre
button .
To abort the incubation press the centre button again.
5 sec before the incubation time ends the MC 1 will give you an acoustic signal.
After these 5 sec you have to reactivate the measuring programme with the left button and start the
measurement within the next 5 sec with the right button (if you don’t use an automatic-pipette) or with the
automatic pipette through pipetting the start-reagent.
The system will stop when the clot moves the ball out of his position!
(You can simulate this by lifting up the cuvette)
For returning to the main menu press the left button again.
5. Settings-Menu
When you start the instrument (main switch ) press the centre button and you will get into the settings-
menu:
With the right / left button you can switch between the input for the INR–standard, ISI-value calibration
curve for PT, incubation time, date/time, default parameter, software-version, language and quit.
INR-standard and ISI-value : to get the INR-value
Calibration : setting / storage the calibration curve (percent and time)
Incub time / IVM : setting the incubation time and the interval mode
Date/time : day, month, year, ...
Default-parameters : default calibration curve
Software-Version : only information
Language : English / German / Chinese (each language possible)
Quit : to leave the settings-menu