Page 10
Ingenio® EZporator® Electroporaon System
Product User Manual | Instrucons for MIR 51000
Set-up Instrucons (Connued)
Post-pulse Display
The EZporator® instrument will display the peak discharge voltage (VP) in volts (V)
and the me constant (t) in milliseconds (ms). It is normal for the measured voltage
to dier slightly from the setpoint. This is due to internal resistance and will uctu-
ate with electrical load variaons (i.e. buer conducvity). Following the delivery of
a pulse, press the Pulse Buon or rotate the Voltage Adjust knob to return to the
ready mode.
NOTE: With exponenal decay generators such as the EZporator®, it is necessary to
idenfy and monitor the me constant (t), as this variable is dependent on the con-
ducvity and osmolarity of the electroporated sample. Signicant changes in the
electroporaon sample (e.g. variaons in cell density, volume of nucleic acid added,
electroporaon soluon used) will alter the me constant.
Voltage Adjust Knob
The EZporator® Voltage Adjust knob controls set voltage AND voltage mode (i.e. LV
and HV modes). There is no separate voltage mode switch. Use the Voltage Adjust
knob to adjust the voltage from 20 to 400 V in LV mode (in 2 V increments) and from
30 to 2,500 V in HV mode (in 10 V increments). If in LV mode, increasing of the volt-
age setpoint past 400 V will result in a switch to HV mode. The voltage then read-
justs to 0 V and connues up to 2,500 V. If in HV mode, decreasing the voltage set-
point past 0 V will result in a switch to LV mode. The voltage then readjusts to 400 V.
NOTE 1: The EZporator® emits a sound when switching between LV and HV modes.
NOTE 2: The Voltage Adjust knob is speed sensive. A quick rotaon increases the
rate of change of the voltage setpoint.
NOTE 3: LV mode is used for mammalian and insect cell types.
Pulse Buon
The Pulse buon is acvated in ‘READY’ mode. Once pressed, the generator will
bleed o the capacitors to the preset voltage, prior to delivering the pulse. The
capacitor bleed me is typically between 1-6 seconds. A pulse sequence may be
aborted by pressing the Pulse buon a second me before delivery of the pulse.
Following the delivery of a pulse, press the Pulse buon or rotate the Voltage Adjust
knob to return to the ‘READY’ mode.
Mirus Bio LLC
www.mirusbio.com
│ [email protected] │ U.S. Toll Free: 844.MIRUSBIO (844.647.8724)│ Direct: +1.608.441.2852