Mustang Survival MD3151 User manual

I
FOR MODEL NUMBERS MD3151 / MD3153 / MD3154
OWNER'S MANUAL
INFLATABLE PFD
DO NOT REMOVE PRIOR TO SALE
HT1109 REV 23MAR2020
MD3151 INFLATABLE PFD
MANUAL INFLATION MODEL
APPROVED BY TRANSPORT CANADA
WARNING: THE MD3151 WILL NOT INFLATE
AUTOMATICALLY UPON IMMERSION.
TAB MUST BE PULLED TO ACTIVATE
INFLATION MECHANISM.
MD3153 INFLATABLE PFD
HYDROSTATIC INFLATION MODEL
MANUAL WITH AUTOMATIC BACKUP
APPROVED BY TRANSPORT CANADA
MD3154 INFLATABLE PFD
HYDROSTATIC INFLATION MODEL
WITH HARNESS. MANUAL WITH
AUTOMATIC BACKUP
APPROVED BY TRANSPORT CANADA

II MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
TABLEOFCONTENTS
INFLATABLE PFDs.........................................................................................1
AIRLINE OPERATOR POLICY ON CARRIAGE OF
INFLATABLE PFDS AND CO2CARTRIDGES...........................................1
APPROVAL CONDITIONS AND CARRIAGE REGULATIONS ..........1
MANDATORY CARRIAGE REQUIREMENTS...........................................2
WHAT IS AN INFLATABLE PFD?................................................................2
INSTRUCTIONS FOR USE............................................................................3
COMPONENT INSPECTION INSTRUCTIONS........................................ 4
READINESS CHECKLIST................................................................................. 4
SIZING AND FIT................................................................................................... 5
PUTTING ON (DONNING) YOUR PFD..................................................... 5
INFLATING YOUR PFD .................................................................................... 6
SPECIAL CONSIDERATIONS FOR INFLATABLE DEVICES ...........7
DEFLATING YOUR PFD................................................................................... 8
USAGE BELOW FREEZING............................................................................ 8
RE-ARMING YOUR PFD.................................................................................. 9
REPACKING YOUR PFD .................................................................................14
SAILING HARNESS .................................................................................... 18
IS YOUR PFD IN GOOD AND SERVICEABLE CONDITION?................ 19
CARE AND MAINTENANCE INSTRUCTIONS ....................................... 19
CLEANING AND STORING YOUR PFD ................................................. 20

III
WHY ARE PFDsREQUIRED SAFETY EQUIPMENT? ........................... 21
HOW AND WHY TO TEST YOUR PFD..................................................... 21
HOW TO TEST YOUR PFD USING
THE AUTOMATIC INFLATOR.....................................................................22
HOW TO TEST YOUR PFD USING
THE MANUAL INFLATOR ............................................................................23
HOW TO TEST YOUR PFD USING
THE ORAL INFLATOR...................................................................................24
WEAR YOUR PFD .......................................................................................25
HYPOTHERMIA............................................................................................25
HOW HYPOTHERMIA AFFECTS MOST ADULTS..............................25
SOME POINTS TO REMEMBER ABOUT
HYPOTHERMIA PROTECTION...................................................................26
EACH OF THESE DEVICES IS INTENDED TO HELP
YOU SAVE YOUR OWN LIFE ....................................................................27
INFLATABLE PFD SAFETY ACCESSORIES ...........................................28
ADDITIONAL INFORMATION ...................................................................28
DO NOT ATTACH PFDsTO YOUR BOAT ................................................28

1 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
INFLATABLE PFDS
Inflatable personal flotation devices (PFDs) are special devices
in that the flotation of the wearer is not derived from inherently
buoyant materials. In order for them to do their job, you, the wearer,
must do yours! This means you have the responsibility to ensure
you are familiar with the operation and maintenance requirements
of your inflatable PFD! You must also be aware of the limitations of
the device and know that it is not approved under the Small Vessel
Regulations for certain boating activities.
READ YOUR OWNER’S MANUAL AND THE INFORMATION LABELS ON YOUR PFD!
IT COULD SAVE YOUR LIFE!
AIRLINE OPERATOR POLICY ON CARRIAGE OF INFLATABLE PFDs
AND CO2CARTRIDGES
Please be aware that regulations may apply to the air transport of this
product. With the approval of the aircraft operator, carbon dioxide
cylinder(s) may be carried by a passenger or crew in checked or
carry-on baggage. Please consult airline operator policy.
APPROVAL CONDITIONS AND CARRIAGE REGULATIONS
This inflatable PFD is approved by Transport Canada for use as a
Type II (MD3153 and MD3154) or Type III (MD3151) personal flotation
device. It is not approved for water skiing or other high impact,
high speed activities. This inflatable PFD was designed to be more
comfortable and less restrictive to wear than inherently buoyant
PFDs. When worn, used, and serviced according to this owner’s
manual, this PFD can greatly increase your chances of survival in
the water. Not recommended for non-swimmers or weak swimmers.
Users of inflatable PFDs must be at least 16 years old.
The MD3154 has a built-in sailing harness that can cause injury if
not properly used. When worn, used, and serviced according to
this owner’s manual, these PFDs can greatly increase your chances
of survival in the water. They should always be fitted with the
appropriate, unpunctured, and fully functional inflation cartridge,
or be worn fully inflated. They are NOT for white water paddling or
personal watercraft use. PFDs fitted with an automatic inflator are not
permitted for sailboarding activities. In order to meet the equipment
carriage requirements of the small vessel regulations, these PFDs
must be worn by an adult, in the case of an open boat or on deck or in

2
the cockpit in any other type of boat, or readily available to any other
adult occupant while below deck.
A separate section titled “Sailing Harness”, p. 18, discusses
requirements and features on the MD3154 only.
Transport Canada approval does not apply to the sailing harness
because some of the risks associated with its use have not been
evaluated.
MANDATORY CARRIAGE REQUIREMENTS
The Small Vessel Regulations require that you carry on board a
Transport Canada approved personal flotation device, which bears a
mark or label indicating the approval and is of an appropriate size for
each person on board. To be considered serviceable, this PFD shall
not exhibit deterioration that could diminish its performance, such as
broken or deformed hardware, detached webbing, rotted structural
components, air leaks, or non-functional oral inflation tube. Unless
worn inflated, this PFD must also be properly armed with a full CO2
cylinder, inflation system status indicator, and an accessible manual
inflation lanyard. A PFD which is “approved only when worn” or
“required to be worn” must be worn under the specified conditions.
WHAT IS AN INFLATABLE PFD?
While traditional PFDs are inherently buoyant, inflatable PFDs
rely entirely on inflation for buoyancy. Uninflated, your PFD is
a comfortable slim collar that can be inflated at any time with a
33-gram carbon dioxide (CO2) gas cylinder. The inflatable PFD is
designed to oer maximum mobility with minimum bulk. The MD3153
and MD3154 inflatable PFDs can be inflated either automatically by
water immersion, manually by jerking a pull-tab, or orally by mouth.
The MD3151 inflatable PFD can be inflated manually by jerking the
pull-tab or orally by mouth. It is recommended that you familiarize
yourself with the procedures outlined in Instructions for Use, p. 3.

3 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
INSTRUCTIONS FOR USE
This manual supplies instructional, maintenance, and safety
information for models MD3153, MD3154, and MD3151. The model
number is listed on the underside label.
Oral Inflator
(behind cell)
Velcro
TM
Tab
Pull-tab
Front
Buckle Adjustment
Point
Zipper
End
Waist Belt
Unpacked
Inator
Window
Panel
Single
Point
Status
Indicator
Water Inlet Valve
Lanyard
Manual Inflation
Pull-tab
Yellow Inflator Cap
Cap/Locking Ring
Tabs
Black Locking Ring
Inflator
Body
Expiry
Date
FIGURE 1
FIGURE 2
Black Locking Ring Yellow Inator Cap
Single Point
Status Indicator
Inator Body
Expiry Date
Lanyard
HAMMAR
FIGURE 2A
UNPACKED INFLATABLE PFD
MD3153 AND MD3154 MD3151

4
COMPONENT INSPECTION INSTRUCTIONS
1. Look through the inflator window panel (Fig. 1). Ensure the
indicator is green. If the single point status indicator (Fig.
2 for MD3153 and MD3154, or Fig. 2a for MD3151) is red, the
mechanism has been fired or is incorrectly fitted (see Re-arming
Your PFD, p. 9).
THE SINGLE POINT STATUS INDICATOR MUST BE GREEN BEFORE PROCEEDING.
2. Undo the zipper on the wearer’s left side, exposing the
oral inflator.
3. Ensure the oral inflation cap is in the stowed position
(Fig. 7 on p. 8).
4. Ensure the single point status indicator is green and the pull-
tab lanyard is hanging on the outside; secure the zippers on the
Inflatable PFD (see Repacking Your PFD, p. 14).
5. Ensure the current date is not past the date on the inflator.
If it is, replace the inflator (see Re-arming Your PFD, p. 9).
READINESS CHECKLIST
Check your PFD and ensure all of the checkpoints listed below are
true before use.
¨Single point status indicator is green (Fig. 2 or 2a).
¨Ensure the current date is not past the date on the inflator. If it is,
replace the inflator (see Re-arming Your PFD, p. 9).
¨Ensure both zipper stops are in place under Velcro™ tab (Fig. 1).
¨All zippers, Velcro™, and waist buckle are securely fastened.
¨Oral inflation cap is in the stowed position (Fig. 7).
¨PFD is not twisted.
¨Ensure the inflation pull-tab is hanging on the outside (Fig. 31).
¨No rips, tears, excessive abrasion, or holes; all seams are securely
sewn; and the cover, straps, and hardware are still strong.
CONTACT YOUR MUSTANG DEALER IF YOU HAVE ANY CONCERNS ABOUT THE
READINESS OF YOUR INFLATABLE PFD.

5 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
FIGURE 3
SIZING AND FIT
This PFD provides a minimum of 150 Newtons (33.7 pounds) of
buoyancy when inflated and is designed for wearers over 16 years of
age, with a chest range of 76-132 cm (30-52 inches) and weighing
over 36 kg (80 pounds).
PUTTING ON (DONNING) YOUR PFD
It is important that the inflatable PFD is properly adjusted to fit
the person wearing it. An incorrect fit or improper fastening of
attachments could impede its effectiveness.
1. All belts and straps are already threaded correctly and only need
to be adjusted for fit. For the non-harness models MD3153 and
MD3151, the belt should fit below your rib cage adjusted to a
tight personal fit. For the harness model MD3154, the bottom of
the belt must always be worn at least 5 cm (2 inches) above the
lowest rib and adjusted to a tight personal fit.
2. Put the PFD on just like a jacket (Fig. 3) and fasten the front
buckle. Adjust the waist belt using the side buckle on the belt and
secure the belt’s loose end in the belt loop.
WARNING: DO NOT WEAR INFLATABLE PFDS UNDER CLOTHING AS THE INFLATION
COULD BE RESTRICTED OR YOU COULD BE INJURED.
DONNING YOUR PFD

6
INFLATING YOUR PFD
Follow the below procedures to inflate your PFD:
1. Manual Inflation
Activate the inflation system by
jerking firmly downward on the pull-
tab (Fig. 4). The PFD should fully
inflate within five seconds.
2. Automatic Backup Inflation
(For models MD3153 and
MD3154 only)
When the inflator is lowered more
than about four inches in the water,
the hydrostatic valve opens and lets
the water meet the water sensitive
element that in turn releases a
stainless steel coil spring. The spring
then drives a needle into the end of
the carbon dioxide (CO2) cylinder,
which is now punctured so that
the gas instantly fills the PFD. Total
buoyancy is normally reached within
four to five seconds.
Once the PFD is inflated, the CO2will slowly escape through the fabric
over time. A slight reduction of pressure will be observed after several
hours of inflation. Topping up by oral inflation may be required in the
event of very prolonged immersion.
SEE SPECIAL CONSIDERATIONS FOR INFLATABLE DEVICES.
MD4031 MD4032
Manual Inflator Automatic Inflator
FIGURE 4 MANUAL INFLATION

7 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
3. Oral Inflation
Properly armed and inspected
inflatable PFDs will inflate when
manually or automatically activated.
However, you can fully inflate the
PFD by mouth (Fig. 5) should
inflation not occur, or if you are in a
non-emergency situation.
• To orally inflate the Inflatable PFD,
first ensure that both of the two
zippers are completely opened.
• Locate the oral inflation tube by
opening the wearer’s left front
panel. Remove the cap from the
end of the oral tube and blow
into the tube until the PFD is
fully inflated.
ORAL INFLATION MAY CAUSE THE INFLATABLE PFD TO BREAK OUT SUDDENLY
(LOUDLY) IF ANY OF THE ZIPPERS IS CLOSED.
SPECIAL CONSIDERATIONS FOR INFLATABLE DEVICES
The inflatable PFD might not be armed when needed, either
knowingly or unknowingly. Inadvertent inflation causes several risks
that you must address to avoid drowning. Much less common but also
important are:
1. Double inflation (CO2cylinder inflation after full oral inflation)
could damage the PFD.
2. Inflation when stored in a tight space could damage the PFD.
3. The PFD could inflate when you are in an awkward place or
position.
4. If this PFD should inflate inadvertently, a re-arm kit is required to
re-arm the device (MA7214 for models MD3153 and MD3154, and
MA3181 for model MD3151).
THE INFLATABLE PFD SHOULD ALWAYS BE RE-ARMED ACCORDING TO THE
INSTRUCTIONS (SEE RE-ARMING YOUR PFD, P. 9).
FIGURE 5 ORAL INFLATION

8
DEFLATING YOUR PFD
1. To deflate the PFD, reverse the oral-inflation tube cap and insert
it into the valve or depress the valve with your fingertip. The cap
will not lock in the deflate position so it is necessary to hold it in
place (Fig. 6).
2. Gently squeeze the PFD until all air or gas has been expelled. To
avoid damage, do not wring or twist the PFD. Put the cap back in
its stowed position on the oral inflation tube (Fig. 7).
ENSURE ALL AIR HAS BEEN REMOVED FROM THE PFD.
3. Examine the single point status indicator (Fig. 2 or 2a on p. 3). If
the indicator is red, the mechanism has been fired and the inflator
requires service (see Re-arming Your PFD, p. 9).
THE SINGLE POINT STATUS INDICATOR MUST BE GREEN BEFORE PROCEEDING.
4. Your device is ready to be repacked (see Repacking Your PFD,
p. 14).
FIGURE 6
Cap in
deflate
position
Cap in
deflate
position
Cap in
stowed
position
FIGURE 7
Cap in
deflate
position
Cap in
stowed
position
USAGE BELOW FREEZING
Do not use inflatable PFDs in below freezing temperatures because a
fully discharged cylinder may not quickly or adequately inflate your
PFD.
DEFLATE POSITION STOWED POSITION

9 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
RE-ARMING YOUR PFD
USE A VALID MUSTANG SURVIVAL RE-ARM KIT FOR THIS PROCEDURE (MA7214 FOR
MODELS MD3153 AND MD3154 AND MA3181 FOR MODEL MD3151). USE OF OTHER
RE-ARM KITS MAY RESULT IN FAILED OR IMPROPER OPERATION AND WILL VOID THE
PRODUCT’S WARRANTY.
1. Place the PFD on a smooth, flat surface and wipe off any water.
Access the inflator by reaching under the cover and pulling it
toward you, turning the cell and cover inside-out to fully expose
the inflator for re-arming. Hold the CO2cylinder through the
fabric, using one hand (Fig. 8 for models MD3153 and MD3154,
and Fig. 8a for model MD3151).
2. Insert the metal key between the black locking ring and the
labeled yellow cap. Turn the key counterclockwise (Fig. 9 for
models MD3153 and MD3154, and Fig. 9a for model MD3151).
Locking Ring Key
FIGURE 8
HAMMAR
JERK TO
INFLATE
FIGURE 8
FIGURE 8A
FIGURE 9
MD3153 AND MD3154 MD3151
MD3153 AND MD3154

10
3. Turn the black locking ring counterclockwise and lift off the cap
(yellow inflator operating head Fig. 10 for models MD3153 and
MD3154 and Fig. 10a for model MD3151). Dispose of the used cap.
FIGURE 9A
FIGURE 10
FIGURE 10A
Cap
HAMMAR
Locking Ring Key
HAMMAR
HAMMAR
HAMMAR
MD3151
MD3153 AND MD3154
MD3151

11 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
4. Squeeze the sealing ring to elongate and remove the inflator body
through the sealing ring (Fig. 11).
FIGURE 11
FIGURE 12 FIGURE 13
FIGURE 14 FIGURE 14A
Red Indicator Inflator Body
Sealing Ring
6. Check that the new inflator body indicator is green. Insert the
new inflator body with CO2cylinder pointing upward inside the
PFD (Fig. 13). Let the sealing ring rest on the adapter around the
four lugs.
Red Indicator
Trash
Green Indicator
Caution:
Do not turn
the center
shaft
Single
Point
Indicator
Water
Inlet
Valve
HAMMAR
JERK TO
INFLATE
Single Point
Indicator
Expiry Date
REMOVE INFLATOR BODY
RED = TRASH GREEN = GO
MD3153 AND MD3154 MD3151
5. Dispose of the used inflator body (Fig. 12).

12
7. Now check the new yellow cap as follows (Fig. 14 for models
MD3153 and MD3154, and Fig. 14a for model MD3151):
• Is the single point status indicator showing green?
• Is the expiry date OK?
If YES is the answer to both questions, then proceed as follows. If
the answer is NO to either question, get a new cap.
8. Hold the CO2cylinder through the fabric of the PFD. Position the
replacement cap with the water inlet valve pointing to the right
and press firmly onto the inflator body and sealing ring (Fig. 15 for
models MD3153 and MD3154, and Fig. 16 for model MD3151).
FIGURE 15
FIGURE 16
HAMMAR
JERK TO
INFLATE
HAMMAR
90˚
MD3153 AND MD3154
MD3151

13 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
9. While pressing FIRMLY onto the inflator body, turn the BLACK
locking ring clockwise into the locked position (Fig. 17 for models
MD3153 and MD3154, and Fig 17a for model MD3151). Pull on the
cap to make sure it has locked onto the inflator body.
10. Check that the single point status indicator on the cap is green;
that the pull-to-inflate lanyard is present; and that the locking
ring is locked. Turn the cell and cover right-side out, returning to
normal state.
11. If your PFD has a black inflator cap cover, position the protective
cap over the inflator body so that the single point status indicator
is visible through the inflator window panel (Fig. 17B).
12. Refold the PFD in accordance with the repacking section, p. 14.
FIGURE 17
HAMMAR
FIGURE 17A
FIGURE 17B
Black
Locking
Ring
Lanyard
Inator Cap
Cover
Single Point
Status Indicator
Inator
Window
Panel
MD3153 AND MD3154 MD3151
BLACK INFLATOR CAP

14
REPACKING YOUR PFD
If your PFD has been used and/or the inflator replaced, always
inflate through the oral tube and check that it stays inflated at least
overnight. With the cap in the top of the oral tube, gently squeeze
the PFD until all air or gas has been expelled (Fig. 6 on p. 8). Do
not wring or twist the PFD. Put the oral tube cap back in its stowed
position on the oral tube (Fig. 7 on p. 8). Let the PFD dry before
packing.
BEFORE FOLLOWING THE REPACKING SEQUENCE, BE SURE TO PROPERLY RE-ARM
YOUR INFLATABLE PFD (SEE RE-ARMING YOUR PFD, P. 9). ENSURE ALL AIR IS
REMOVED BEFORE REPACKING.
1. Lay the PFD on a flat, clean surface (Fig. 18).
THE FOLD LINES, ON THE DIAGRAM, ARE PROVIDED TO ASSIST WITH PACKING.
AB
FIGURE 18
FIGURE 19 FIGURE 20
2. Fold wearer’s right side (A) inward along the full length of the
dotted line (Fig. 19). Fold cover (B) over side (A) (Fig. 20).
LAY FLAT
RIGHT SIDE A RIGHT SIDE B

15 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
FIGURE 21 FIGURE 22
FIGURE 23 FIGURE 24
3. If your PFD has a black inflator cap cover, ensure the inflator body
is still located within its protective cap cover and that the single
point status indicator is visible through the inflator cap.
4. Repeat step 2 on left with side (C), cover (D) (Fig. 21 and 22).
5. Fold over the top side (E) and cover (F) (Fig. 23 and 24).
C
D
E
F
LEFT SIDE C LEFT SIDE D
FOLD TOP FOLD COVER

16
FIGURE 25 FIGURE 26
6. Mustang Survival uses advanced two-step zippers that can be
reset. Reset 1: Pull the slider (G) to the zipper stop (K) and under
the Velcro™ tab (H) (Fig. 25). Ensure both zipper stops (K) are in
place. Close by pulling the slider (G) in the opposite direction to
the end of the zipper (Fig. 26). Ensure that the Inflatable cell does
not get caught by the zipper when closing.
FIGURE 27
Flap Flap
Zipper
Stops
H
RESET 1: UP RESET 1: DOWN
7. Fasten the Velcro™ tab (H) to keep zipper closed (Fig. 27).
CLOSE TAB H

17 MUSTANG SURVIVAL / OWNER’S MANUAL – MD3151 / MD3153 / MD3154 / INFLATABLE PFD
FIGURE 31
9. Use a finger to push the zipper ends into the openings at the
bottom of the packed PFD (Fig. 30).
10. Ensure the pull-tab (J) is on the outside of the folded PFD. Fig. 31
depicts a correctly folded PFD.
J
FIGURE 30
FIGURE 29 RESET 2: DOWN
PUSH IN ENDS CORRECT
FIGURE 28 RESET 2: UP
8. Reset 2: Pull the slider (I) to the Velcro™ tab (H). Close by
pulling slider (I) in the opposite direction to the end of the zipper
(Fig. 29). Ensure that the inflatable cell does not get caught by
the zipper when closing.
This manual suits for next models
2
Table of contents
Languages:
Other Mustang Survival Marine Safety Device manuals

Mustang Survival
Mustang Survival MD2014 User manual

Mustang Survival
Mustang Survival MD3071 User manual

Mustang Survival
Mustang Survival MD318802 User manual

Mustang Survival
Mustang Survival khimera MD7183 User manual

Mustang Survival
Mustang Survival MD3188 User manual

Mustang Survival
Mustang Survival MD2951 User manual