Niigata seiki WSC-80 User manual

STEEL COMPARATOR STAND
CERAMIC COMPARATOR STAND
Precision Measuring Instrument Holder WSC
-
80
WCC
-
80
Model No.
User Manual
PRODUCT SPECIFICATIONS
շMeasuring Platform
ձBracket
նMain Post
ղBracket Clamp
մStem Mounting
 Hole
ճFine Adjustment
 Knob
յInstrument
Mounting Clamp  Model No.
WSC
-
80
WCC
-
80
Base Flatness
3μm
Measuring
Platform
Hardened Steel
Ceramic
Stem Mounting
Hole
ȭ8 mm
Weight
Approx. 2.5kg
Approx. 2 kg
Vertical Stroke Length
120 mm ≧1 mm
Fine Adjust Vertical
Stroke Length
●Materials:Base ... Cast Iron /Main Post ... Stainless /Bracket ... Aluminum
Moves vertically with attached
measuring instrument.
Fixes bracket to main post.
Vertical fine adjustment of
instrument clamp.
Measuring instrument mounts here.
Clamp secures measuring
instrument. 
Holds bracket for vertical
positioning.
Work surface.
Platform will wear with repeated
use, so please use block gauge
to check for wear before use.
ձBracketǃǃǃǃǃ
ǃǃǃǃ
ղBracket Clamp
ճ
Fine Adjustment Knob
ǃǃ
մStem Mounting Hole
յInstrument Mounting ǃClamp
նMain Post      
շMeasuring Platform
.........................
ࠉ.............
....

......
ࠉ......
  
  ......................
        .......
■Part Identification and Description
Throughout this manual, this mark: “”
indicates a required condition or procedure.
Thank you for purchasing the Comparator Stand. This product holds 䃥8mm stem measuring tools such as dial gauge and
electronic micrometer for measuring dimensions such as outside diameter, thickness, and step heights.
●For safe and proper use of this product, please read this User manual
before use and follow the procedures described. Please keep manual
where it is accessible to user for future reference.
●Keep this manual with the instrument if transferred or leased to a third party.
●For inquiries about this product, please contact dealer or place of
purchase.
SAFETY NOTICE
Read the user manual and follow the instructions.
ʋUse of product other than as described in the manual may
cause accident.
Use only for measuring.
ʋUse for any purpose other than measuring may damage
or wear the instrument. Improper use may also cause
accident.
Use in an environment which meets the following
conditions.ȕDry location protected from rain, water, and oils.
ȕLocation not subject to wide temperature changes.
ȕLocation free from cutting chips and dust.
ȕ
Horizontal surface which is stable and free from vibration.
ȕLocation protected from use by children and
unauthorized people .
ʋUse in location contrary to the above may cause poor
accuracy, damage to the product, or may result in accident
or injury.
Handle with Care.
ʋDo not drop or subject to shock -- instrument accuracy
may be affected.
ʋDo not allow water or oil to sit on measurement base,
use a soft cloth to clean.
Protect measuring platform from scratches.
ʋTo prevent damage, use care when placing workpieces
onto platform surface, and avoid sharp edges.
When mounting measurement instrument to
bracket clamp, make sure clamp is tightened firmly.
ʋIf clamp is loose, measurement instrument may fall and
cause damage or affect instrument accuracy.
Follow instructions to apply rust preventive
treatment after use.
ʋPlease refer to “CLEANING AND STORAGE AFTER
USE” section below to prevent corrosion of instrument.
Wipe measuring platform with a clean, dry cloth to remove any
contamination. Then wipe with cloth moistened with alcohol.
Apply rust preventive oil to the Measuring Platform.
(WSC-80 only)
Do not use a stiff brush or abrasive pad as it will damage the
surface and cause poor accuracy.
Wipe any contamination from the main post and bracket using
a clean, dry cloth.
Store in area free from dust and with low humidity.
CLEANING AND STORAGE AFTER USE
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL http://www.niigataseiki.co.jp
Ƭ
Stem Mounting Hole
ƭInstrument
 Mounting
 Clamp
⫮
Attach
ƩInsure that the measuring table is free of dirt and is
secure and not loose.
Before use, wipe table top with soft cloth and alcohol to remove
anti-rust oil and dirt. Insure that top is tight and is not scratched. Do
not use with damaged or loose table top as it will affect accuracy.
ƪCheck that bracket moves smoothly.
Loosen the bracket clamp and check that it moves smoothly on
the main post. Otherwise, clean post and remove any
contamination or corrosion.
ƫAttach the measurement instrument to the bracket.
Firmly tighten the bracket clamp.
Loosen the instrument mounting clamp and insert the
measuring instrument stem into the mounting hole. Fix the
instrument by tightening the instrument mounting clamp.
Measuring Instrument
ƫ
Fine Adjustment Knob
Stem
ƪBracket
 Clamp
ƩBracket ④Taking Measurements.
1. Make sure that there is no dirt or dust on the measuring table,
measuring instrument, or workpiece.
2. While supporting the bracket by hand, loosen the bracket clamp
and adjust to roughly the desired height before retightening the
bracket clamp.
3. Use the fine adjust knob to fine-tune the height.Perform the
desired measurements.
CHECKING THE PLATFORM FOR WEAR
Measuring platform will wear with continued use.
Please check for wear before each use.①Install an accurate measuring instrument.
②Measure block gauge placed at several points on measuring
surface and make sure the reading is consistent. If the reading
varies significantly, please discontinue use as the wear on the
measuring surface is to large.
Wear and performance degradation will vary greatly
depending on frequency and type of use. We recommend
periodic inspection conforming you your in-house
standards and requirements.
HOW TO USE
Tighten
⫯
Loosen
r
⤞Before start of measurement, please check the
measuring surface for wear using the following
procedure.
Indicates risk of personal injury or property damage if not followed.
Caution
SAFETY PRECAUTIONS
Always follow the proper operating procedures in order to prevent harm to yourself or others, and to prevent damage to property.
Please Observe
この説明書は、必ずお守り頂きたい強制事項に マークを付けて表示しています。測定器を取り付け、上下に移動します。ブラケットを締め付けて、主柱に固定するクランプです。ステム取付穴に差し込んだ測定器の上下位置を微調整します。測定器のステムを差し込む取付用の穴です。
ステム取付穴に差し込んだ測定器のステムを、締め付けて固定するクランプです。
ブラケットを支える柱です。ワークを置く台です。使用を繰り返すうちに磨耗しますので、ご使用前に毎回、ブロックゲージを使用して磨耗確認を行ってください。①ブラケット…………②
ブラケットクランプ
… ③微動送り……………④ステム取付穴………  ⑤測定器クランプ……⑥主柱…………………⑦測定台………………鋼製コンパレータスタンドセラミック製コンパレータスタンド精密測定器保持具WSC
-
80WCC
-
80品番取扱説明書この度は「コンパレータスタンド」をお買上げいただきありがとうございます。この商品は、ダイヤルゲージ、電気マイクロメータなど䃥8ステムの測定器を保持し、ワークの外径・厚さ・段差などを測定する測定補助具です。●正しく安全にお使いいただくため、ご使用の前に本取扱説明書を必ず読み、記載の手順に従ってご使用ください。お読みになった後は、お使いになる方がいつでも見ることができる場所に、大切に保管してください。●第三者に譲渡・貸与される場合も、この説明書を必ず添付してください。●当商品に関するお問い合わせは、お買い求めの販売店もしくは当社にご連絡ください。安全上のお知らせ■各部の名称と機能製品仕様⑦測定台①ブラケット⑥主柱②ブラケットクランプ②ブラケット クランプ④ステム取付穴③微動送り⑤測定器クランプI408-K2004
安全上のご注意
必ずお守りください。
お使いになる人や他の人への危害、財産への損害を未然に防止するため、必ずお守りいただくことを下記のように説明しています。
誤った使い方をした場合に『傷害を負う、または物的損害が発生するおそれがある内容』です。取扱説明書をよく読み、指示に従う。・取扱説明書に記載された内容以外での使用は、事故の原因となります。測定作業にのみ使用する。・指定用途以外へのご使用は、製品の破損や磨耗、予測できない事故の原因となります。下記の条件を満たす環境で使用する。●雨や水、油などがかからない、乾燥した場所●温度変化の少ない場所●切粉や粉塵が発生しない場所●振動やガタツキのない水平な場所●子供や、使用者以外が近づかない場所・上記に反する場所での使用は、精度不良や製品の破損、事故やけがの原因となります。注意本器は大切に取り扱う。・落下などの衝撃を与えないでください。精度不良や製品の破損の原因となります。・測定台に水や油が付着した場合は長時間放置せず、柔らかい布などで拭き取ってください。測定台上面の傷に注意する。・傷防止のため、ワークは静かに測定台へ載せてください。また、鋭利なものは載せないでください。測定器を取り付けている時は、ブラケットクランプをしっかりと締めて固定する。・ブラケットクランプの締め付けがゆるいと、ブラケットごと測定器が落下し、測定器の故障・破損・精度不良の原因となります。使用後は指示に従って防錆処理を行う。・錆防止のため、使用後は「使用後のお手入れ・保管方法」に沿って手入れを行ってください。
使用方法
①測定台に汚れ、傷、ガタツキなどが無いか確認する使用前に、防錆油や汚れは柔らかい乾いた布か、アルコールを少量含ませた布で拭き取ってください。ガタツキや、測定台上面に傷がある場合は精度が出ないため使用を中止してください。②ブラケットの動きがスムーズであることを確認するブラケットクランプをゆるめ、ブラケットを上下させます。動きがスムーズでない場合は、主柱やブラケットの汚れ、錆などを取り除いてください。③測定器をブラケットに取り付けるブラケットクランプをしっかりと締めます。測定器クランプをゆるめて、ステム取付穴に測定器のステムを差し込み、測定器クランプを再度締めて固定してください。④測定する1.ワークや測定器の測定子にごみ・埃・汚れなどがないことを確認し、ワークを測定台の上に置きます。2.ブラケットを手で支えながら、ブラケットクランプをゆるめ、おおよその高さに合わせてブラケットクランプを締めます。3.微動送りで高さを微調整し、任意の位置に合わせたら測定を行ってください。測定台は使用を続けると磨耗します。毎回のご使用前に磨耗状況の確認を行ってください。①信頼できる測定器を取り付けます。②ブロックゲージを測定台の上に置いて数点測定し、お客様の測定に影響するような数値変化が無いか確認してください。数値変化がお客様の測定に影響する状態になったら、磨耗状況が大きいため使用を中止してください。※測定開始前に、下記手順に従って測定台の磨耗状況の確認 を行ってください。性能の劣化は、使用の頻度・環境によって大きく異なります。お客様の社内規格に準じた周期で、定期点検を行うことをおすすめ致します。測定台の汚れは柔らかい乾いた布か、アルコールを少量含ませた布で拭き取り、鋼製(WSC-80)の場合は防錆油を塗布してください。
硬いブラシなどで汚れを落とすと測定台に傷がつき、精度不良の原因となるのでおやめください。
主柱やブラケットの汚れを落としてから保管する。
主柱やブラケットは柔らかい布などで汚れを拭き取り、湿度の少ない環境で保管してください。
使用後のお手入れ・保管方法測定台の磨耗状況の確認締める3ゆるめる1差し込む2⑤測定器クランプ④ステム取付穴測定器ステム③微動送り①ブラケット 品 番WSC
-
80WCC
-
80ベース平面度3μm測定台材質焼入鋼セラミックステム取付穴䃥8mm質 量約2.5kg約2kg
ヘッド上下有効ストローク
120mm1mm以上
微動送り上下ストローク
●材質:ベース…鋳鉄/主柱…ステンレス/ブラケット…アルミ

This manual suits for next models

1

Other Niigata seiki Rack & Stand manuals

Niigata seiki WGC-150 User manual

Niigata seiki

Niigata seiki WGC-150 User manual

Niigata seiki WGC-150 User manual

Niigata seiki

Niigata seiki WGC-150 User manual

Popular Rack & Stand manuals by other brands

Milwaukee 48-08-0551 Operator's manual

Milwaukee

Milwaukee 48-08-0551 Operator's manual

CHIEF RPA287 instruction manual

CHIEF

CHIEF RPA287 instruction manual

Balt 27542 Assembly instructions

Balt

Balt 27542 Assembly instructions

ZKTeco ZKIWBA01B installation guide

ZKTeco

ZKTeco ZKIWBA01B installation guide

HumanCentric 101-1005-V4 manual

HumanCentric

HumanCentric 101-1005-V4 manual

BGS technic BGS 4204 instruction manual

BGS technic

BGS technic BGS 4204 instruction manual

Trane BAYWMKIT001A Installer's guide

Trane

Trane BAYWMKIT001A Installer's guide

Costway TL35172 user manual

Costway

Costway TL35172 user manual

Panasonic TY-ST85PF12 Service manual

Panasonic

Panasonic TY-ST85PF12 Service manual

Wine Rack Shop Corner Liquor Bottle Rack Assembly instructions

Wine Rack Shop

Wine Rack Shop Corner Liquor Bottle Rack Assembly instructions

CommScope 2RU FMT instruction sheet

CommScope

CommScope 2RU FMT instruction sheet

Wolfcraft SB 60 Translation of the original operating instructions

Wolfcraft

Wolfcraft SB 60 Translation of the original operating instructions

sanus accents SWM manual

sanus accents

sanus accents SWM manual

Keysight Technologies Y1217A Installation notes

Keysight Technologies

Keysight Technologies Y1217A Installation notes

Yamaha L-85 Assembly instructions

Yamaha

Yamaha L-85 Assembly instructions

HI-Q TOOLS 60150700060 instruction manual

HI-Q TOOLS

HI-Q TOOLS 60150700060 instruction manual

Nexo N200 manual

Nexo

Nexo N200 manual

K&M 18816 quick start guide

K&M

K&M 18816 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.