NSK Primado2 User manual



1
Read these safety precautions thoroughly before use and operate the product properly.
These indicators are to allow you to use the product safely, prevent danger and harm to you and others. These are classified by degree of danger, damage
and seriousness. All indicators concern safety, therefore be sure to follow them.
These Attachments/Handpieces are to be used exclusively with Primado2, our surgical operation control units.
We recommend that prior to use, you carefully read this document regarding instructions for use, handling method, or maintenance check so that you can
carry on using the Attachments/Handpieces in the future. In addition, keep this operation manual in a place where a user can refer to it at any given time.
Cautions for handling and operation
Class
Degree of Risk
Danger
Hazard that could result in personal death or serious injury if the safety instructions are not correctly
followed.
Warning Hazard that could result in serious injury or damage to the device if the safety instructions are not correctly
followed.
Caution
Hazard that could result in light or moderate injury or damage to the device if the safety instructions are not
correctly followed.
Notice General product specification information highlighted to avoid product malfunction and performance
reduction.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 1 2017/10/16 11:10:32

2
Intended to use
●Only qualified persons are allowed to use these Attachments/Handpieces for bone operations (bone removal, cutting, perforation and formation).
Danger
●Do not use this product in a room where there is a danger of explosion or in the vicinity of flammable substances. It is especially important not to use this
product on a patient to whom a flammable anesthetic agent (dinitrogen monoxide) has been administered.
●Never disassemble nor tamper with the mechanism. Personal injury, electric shock or fire may result.
Warning
●When operating the product always consider the safety of the patient.
●This is a medical system ; do not allow anybody to use it other than surgeons who are familiar with the procedures. Use this system in accordance with its
intended use and proper method of use.
●This system may be affected when used in rooms with electromagnetic interference and portable or mobile RF communication equipments. Do not use this
system around any equipment that generates electromagnetic waves such as an ultrasonic generator or electrical scalpel. In case equipment generating
electromagnetic waves are used around this system, turn off the Control Unit.
Caution
●U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
●Read these safety cautions thoroughly before use.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 2 2017/10/16 11:10:32

3
Caution
●The Attachments/Handpieces are not sterilized before delivery. Therefore, be sure to sterilize them prior to use.
●Prior to use, always operate and inspect the device with care for vibration, noise, and heat. If anything unusual is found, do not use. Contact your dealer.
●Should the product function abnormally, cease operation immediately and return the Attachment/Handpiece to your dealer for repair.
●Do not contact the latch or twist collet during rotation.
●Do not allow any impact on to the product. Do not drop the product. Personal injury, abnormal heating and fire may result caused by the internal damage.
●Do not use bent, damaged or deformed burs or drills that are not suitable for the Attachment/Handpiece. The Attachment/Handpiece may not match and it
may break or cause injury/malfunction.
●Always keep bur/drill shanks clean. Dirt or debris in a chuck could cause poor bur/drill concentricity or low chuck retention force.
●Turn off the power when mounting or removing the motor and Attachment/Handpiece. Mounting or removing it with the power on may cause personal
injury due to an unexpected rotation.
●Use new a bur or drill for every surgical operation.
●Do not clean the product with solvents such as thinner or benzine solution. Discoloration or deformation may result.
●The temperature at the tip of the Attachment/Handpiece may reach up to 60°C depending on the environment.
Notice
●The user shall be responsible for maintenance and operation.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 3 2017/10/16 11:10:32

4
Latch
Irrigation Nozzle
Twist Collet
1Parts Names
* The figure shows type PD-CA-16. * The figure shows type PD-2ADM.
<Contra Angle Handpiece> <Drill Handpiece>
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 4 2017/10/16 11:10:33

5
Conduct pre-use checks by the following procedure to confirm that there are no abnormalities.
1) Mount a bur or drill.
2) Operate the Attachment/Handpiece at the rotational speed permissible for bur or drill mounted and confirm that there is no looseness, abnormal vibration,
abnormal heating or abnormal noise. If any abnormality is observed, stop using it and request your dealer for a repair.
3) Ensure there are no abnormalities before use.
2Check before treatment
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 5 2017/10/16 11:10:33

6
Fig. 1
Micromotor
Attachment /
Handpiece
Alignment PinAlignment Hole
1) Push the Attachment/Handpiece into the Micromotor straight without twisting.
2) Turn the Attachment/Handpiece either clockwise or counterclockwise so that the Alignment Pin aligns,
then push it in till it clicks to connect firmly.
* For removing, pull out the Attachment/Handpiece straight without twisting.
●When mounting or removing the Attachment/Handpiece, set the Safety Lock to “OFF” on the Hand Switch of the Micromotor (Micromotor
with Hand Control).
●Refer to the Operation Manual of Micromotor.
●Prior to mounting/removing the Attachment/Handpiece, remove the bur or drill. Otherwise, injury might occur.
●After mounting the Micromotor and Attachment/Handpiece, confirm that they are connected properly.
CAUTION
3Connecting / Disconnecting Motor and Attachment / Handpiece
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 6 2017/10/16 11:10:33

7
●Use only burs and drills specified by NSK. The use of other burs and drills may cause a fault or an accident.
●Use burs and drills that are suitable for the Attachment/Handpiece. The use of unsuitable burs and drills may cause a fault or an
accident.
●After mounting a bur or drill, pull it lightly to confirm that it is securely connected.
●Before use, confirm that the bur or drill does not come off by operating the Attachment/Handpiece.
●Do not operate the Attachment/Handpiece when a bur or a drill is not mounted.
●Remove the bur or drill from the Attachment/Handpiece after the rotation of the Micromotor has completely stopped.
CAUTION
4Mounting and Removing Burs and Drills
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 7 2017/10/16 11:10:33

8
Fig. 2
Fig. 3
OPEN
CLOSE
Contra Angle Handpiece
For the bur, use a commercially available Ø2.35mm contra bur.
<Mounting>
1) Slide the Latch in the direction of “OPEN” (Fig. 2).
2) Match the notch of the bur to the corresponding part of the shaft and push the bur all the way
through (Fig. 3).
3) Slide the Latch in the direction of “CLOSE” to set the bur in place (Fig. 2).
<Removing>
1) Slide the Latch in the direction of “OPEN” to remove the bur.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 8 2017/10/16 11:10:33

9
Fig. 4
Fig. 5
OPENCLOSE
Drill Handpiece
<Mounting>
1) Turn the Twist Collet in the direction of ‘ ’ until it clicks (Fig. 4).
2) Insert the Bur or Drill through the tip of the Attachment/Handpiece all the way, twisting it slightly
with your fingers.
3) Turn the Twist Collet in the direction of ‘ ’ until it clicks (Fig. 4) to connect the Bur/Drill firmly.
After mounting the Bur/Drill, confirm that it is firmly connected by pulling it lightly.
Chuck Closed Chuck Open
●Insert the Bur/Drill all the way through the Attachment/Handpiece for a firm
connection. If it is not connected firmly, the Bur/Drill may come off during a surgical
procedure and interrupt it.
●When the Bur/Drill is firmly mounted, confirm that ‘ ’symbol and ‘ ’symbol are
aligned.
CAUTION
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 9 2017/10/16 11:10:34

10 Fig. 6
Irrigation Nozzle Irrigation Tube
Contra Angle Handpiece
1) Insert the Irrigation Tube onto the Irrigation Nozzle (Fig. 6).
* Irrigation nozzle is included in Contra Angle Handpiece as standard equipment.
<Removing>
1) Turn the Twist Collet in the direction of ‘ ’ till it clicks (Fig. 4).
2) Pull the Bur/Drill through the tip of the Attachment/Handpiece straight without twisting.
●Mount the Irrigation Nozzle to the Attachment/Handpiece after removing the Bur/Drill.
●Do not use irrigation nozzles other than those specified by NSK.
●Also see the Operation Manuals of Primado2 Control Unit.
CAUTION
5Mounting Irrigation Nozzle
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 10 2017/10/16 11:10:34

11
Fig. 7
Fig. 8
Irrigation Nozzle
Irrigation Nozzle Irrigation Tube
Drill Handpiece
1) Insert the Irrigation Nozzle onto the Bur Guard of the Attachment/Handpiece, making sure the correct
end of the Irrigation Nozzle goes in (Fig. 7).
2) Insert the Irrigation Tube onto the Irrigation Nozzle (Fig. 8).
3) For the related connection and the setting of the Control Unit, see the Operation Manual of Primado2.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 11 2017/10/16 11:10:34

12
6-1 Preparation before cleaning
1) Remove the Bur/Drill from the Attachment/Handpiece. Dispose the Bur/Drill as it cannot be reused. For the disposal of the commercially available bur,
follow the instructions of its manufacturer.
2) Remove the Attachment/Handpiece from each Micromotor.
3) Confirm that there are no abnormalities or damage or deterioration in its appearance. Confirm that the connection part of the Micromotor is not damaged
or broken.
6-2 Cleaning
1) Clean the surface of each component product, using a lint-free cloth or brush moistened with a neutral detergent for medical treatment. Do not use a
wire brush.
2) Remove the detergent from the surface using a lint-free cloth moistened with distilled, deionized or demineralized water.
●Perform cleansing and sterilization after every use.
●Clean and lubricate the Attachment/Handpiece before autoclaving. Autoclaving the Attachment/Handpiece stained with blood or debris
could cause damage to the Attachment/Handpiece.
●In order to prevent infection, wear a mask, rubber gloves and glasses.
CAUTION
6Cleaning / Sterilization
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 12 2017/10/16 11:10:34

13
6-3 Washing
The Attachment/Handpiece can be washed via Thermo Disinfector.
Disinfect either by using a Thermo-Disinfector or by soaking.
6-3-1 Auto disinfection (It is recommended to use a thermo disinfector compatible with ISO 15883)
The following conditions are recommended for Thermo Disinfector ;
a. Cleaning : at 45°C for 3 minutes
b. Intermediate rinse : At least 1 minute (using distilled, deionized or demineralized water)
c. Thermal disinfecting : at 93°C for 5 minutes (using distilled, deionized or demineralized water)
d. Drying : at least 10 minutes
●Prior to use of Thermo Disinfector, read carefully the Operation Manual of Thermo Disinfector.
CAUTION
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 13 2017/10/16 11:10:35

14 Fig. 9
Attachment /
Handpiece EZ Spray Grip
Release Button
6-3-2 Manual disinfection
1) Soaking : Soak the Attachment/Handpiece in a neutral enzymatic detergent.
Example : For 20 to 30 minutes at 40 to 50°C. (Follow the time and temperature recommended by the manufacturer of the neutral enzymatic detergent)
2) Rinse : Remove debris and detergent from the Attachment/Handpiece by rinsing with distilled, deionized or demineralized water.
3) Drying : Wipe the Attachment/Handpiece dry with a lint-free cloth.
6-4 Lubrication
Lubricate the Attachment/Handpiece using separately sold EZ Spray (Order Code : PD-EZ-M) /E-Type
Spray Nozzle (Order Code : Z019090) and PANA SPRAY Plus (Order Code : Z182600).
1) When using EZ Spray
Push the EZ Spray Grip into the rear part of the Attachment/Handpiece and turn either clockwise or
counterclockwise. When the Alignment Pin clicks into position, the Spray is securely connected.
(Fig. 9)
When using E-Type Spray Nozzle
Insert the E-type Spray Nozzle into the PANA SPRAY Plus.
Insert the E-type Spray Nozzle into the rear part of the Attachment/Handpiece.
2) Spray until the oil coming out of the tip of the Attachment/Handpiece is clear of any debris or foreign
substance (approximately 3 seconds).
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 14 2017/10/16 11:10:35

15
3) Remove the Attachment/Handpiece by depressing the Release Button on the EZ Spray Grip (when using EZ Spray).
6-5 Sterilization
Autoclavable products : Contra Angle Handpiece, Drill Handpiece, Irrigation Nozzle.
We verify the effectiveness of product sterilization under the following conditions.
1) By placing the autoclavable products in a Primado Sterilization Cassette.
2) Autoclave under the following conditions ;
• Pre-vacuum Autoclave : at 134°C for 3 minutes or longer
• Gravity-displacement Autoclave : at 132°C for 15 minutes or longer
●Continue spraying PANA SPRAY Plus until it comes out clear at the tip of the Attachment/Handpiece. Due to its structure, PANA SPRAY
Plus comes out from the tip of the Attachment/Handpiece and the Twist Collet part. To prevent the dispersal of the lubrication oil in the
surrounding area, it is recommended to use a Spray Mist Absorber (Order Code: Y900084).
●Shake the PANA SPRAY Plus well before use.
●After cleaning and prior to sterilization, check the Attachment/Handpiece in the same method as in Step “2. Check before treatment”.
CAUTION
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 15 2017/10/16 11:10:35

16
6-6 Storage
Store the sterilized Attachment/Handpiece in a place where it can be kept sterile until the next use.
●The sterilization parameters are indicated for sterilization and immediate use only.
●Attachment/Handpieces cannot be sterilized using the EOG sterilization method.
●Sterilize Attachments/Handpieces after each patient.
●Do not use a chlorinated cleaner.
●Do not clean Attachments/Handpieces by an ultrasonic cleaning apparatus.
●Do not clean Attachments/Handpieces by a washer sterilizer.
●For detailed information for washing and sterilizing the Micromotor, refer to the Primado2 Micromotor Operation Manual.
CAUTION
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 16 2017/10/16 11:10:35

17
If you are suspecting any faults, check the following once again prior to asking for repair work. If none of these cases is applicable or if the situation does not
improve despite your efforts or measures, the damage could be serious ; call your dealer to fix the problem.
Symptoms Causes/Points of checking Countermeasures
Cannot be connected to the Micromotor. The Alignment Pin is not in position. Turn further till the Alignment Pin aligns, then push
further to connect to the Micromotor.
Abnormality of the Removal part. Contact your dealer.
The Bur or Drill cannot be firmly mounted. The Bur or Drill has not been inserted deep
enough.
Insert the Bur/Drill all the way through.
The Twist Collet/Latch is in ‘OPEN’. After inserting the bur/drill, fasten the Twist Collet/
Latch.
Abnormality of the Twist Collet/Latch. Contact your dealer.
No rotation when the Bur or Drill has been
securely mounted.
The Twist Collet/Latch is in ‘OPEN’. After inserting the Bur/Drill, fasten the Twist Collet/
Latch.
· The bearing has overheated.
· Foreign materials inside.
· The bearing has worn out.
Contact your dealer.
7Troubleshooting
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 17 2017/10/16 11:10:35

18
Symptoms Causes/Points of checking Countermeasures
Overheating (during operation). Continuous operation for a long period of time. Wait until handpiece cools down.
· The bearing has overheated.
· Foreign materials inside.
· The bearing has worn out.
Contact your dealer.
Abnormal vibration and noise.
Bur run-out.
· The bearing has overheated.
· Foreign materials inside.
· The bearing has worn out.
Contact your dealer.
OM-SH0081E_Primado2 Surgical Drill Attachment_EN_for FDA_140124.indd 18 2017/10/16 11:10:35
Other manuals for Primado2
5
Table of contents
Popular Drill manuals by other brands

Ryobi
Ryobi R12DD-LL13S Original instructions

Power works
Power works XB 20V DDG303 Operator's manual

Craftsman
Craftsman CMCD713 instruction manual

Buffalo Tools
Buffalo Tools PS07216 Assembly & operating instructions

XPOtool
XPOtool 64076 Operation manual

Bosch
Bosch GSH 11 E Professional operating instructions