Commissioning:
1. Check whether the mains switch (7) is switched off.
2. The power cable (8), which is permanently connected to the pump, is
plugged into the power outlet to the AC mains.
3. Before starting up the pump, the oil level (2) must be checked or topped up.
4. Remove the cap of the outlet connection (6).
5. During pump operation, make sure that no water and no above-average
amount of water vapor does not get into the suction line of the pump.
6. Care must also be taken that the oil level during operation remains within
the visible range of the sight glass.
7. The rotary vane vacuum pump is very sensitive to mechanical
contamination. If this should be the case, damage will occur and a
general repair will be necessary. If "wrong" air is sucked in, the seal must
be tightened or replaced.
8. During prolonged operation, a permissible temperature rise of the pump
must be expected. If necessary, wait for the pump to cool down after it
has been switched off before tampering with it.
Operating conditions:
Operating position: horizontal, upright
Operating temperature in the room: + 15 °C ... 35 °C
Start-up of the pump at approx. 15 °C should be without load.
Humidity: max. 80% rel.
Maintenance:
The rotary vane vacuum pump is almost maintenance-free. It is sufficient
to regularly oil level and oil quality. When the pump is at a standstill, the oil
limit is in the middle; when the pump is in operation, the oil limit drops
significantly. A deficiency (shortage or clouding due to moisture and
impurities or slide valve wear) leads to performance losses loss of
performance, in the worst case to seizure of the rotary vane. Therefore, it
is recommended to use only original vacuum pump oil, which has viscosity
and lubricating properties, and which guarantees trouble-free function
guaranteed in the long term.
Warranty:
The warranty is 2 years from the date of delivery. The invoice is valid as
warranty certificate.
The warranty does not apply to the cleaning of the device and
malfunctions caused by improper use. The warranty also expires in the
event of operating errors.
Initial start-up:
Filling is carried out with the motor switched off through the opening
provided for this purpose on the top of the pump (1). Please fill the pump
until the oil level has reached the red mark in the center of the sight glass
(2). Then close the filling opening (1) and operate the pump for approx. 2
minutes, then check the oil level again and top up with oil if necessary. The
optimum filling level is in the middle of the sight glass (2). However, the
pump will not be affected if a little too much oil is accidentally filled in.
Description:
The rotary vane vacuum pump consists of the drive unit and the electric
motor. Both are directly coupled to each other, mounted on the base frame
and are air-cooled. The motor carries the mains connection line (8), the
mains switch (7) and the carrying handle (4). The housing of the rotary
vane pump has all the mechanical elements required for operation. The
housing of the rotary vane pump has all the mechanical elements
necessary for operation. On the upper side there are the oil filler opening
(1) and the suction port with bottle (5a). On the front side the sight glass of
the oil level control (2) and the oil drain plug (3).
Depending on the requirements and the accessories available, a different
suction port can be mounted. To do this, simply unscrew 5a and screw on
5b or 5c.