NTL DM503-2A User manual

This electric, two-stage rotary vane vacuum pump is designed
for the experiments in the rough and fine vacuum range in the
natural sciences.
Attention!
All pumps are delivered without oil filling.
For filling, the vacuum pump oil supplied in the box
must be used for filling.
Attention:
- The device is to be operated exclusively by qualified persons or by
persons who have been instructed in the use of the device. Duty of
supervision!
- Make sure that the unit does not fall down when being moved, packed or
unpacked. If this should happen, have the unit repaired immediately by
authorized specialists.
- If unexpected problems occur during installation or operation of the unit,
immediately unplug the power cord and contact your dealer.
- If the device does not work properly, there is an increased risk of danger.
Therefore, always make sure that the device is standing stably on a solid
surface.
- Make sure that no objects are placed on the power cord. A damaged
cable can cause a fire or electric shock.
- There are no user-serviceable components inside the instrument.
- The repair of this device may only be carried out by an authorized
specialist.
- Never push objects of any kind through the ventilation slots or other
openings in the housing and never attempt to touch any components
inside the device for repair, maintenance or other reasons.
- Never place any containers with liquids on the housing of the device.
- Do not use connecting cables > 3 m
- Errors and omissions excepted
Fruhmann GmbH - 7372 Karl, Austria
DM503-2A
Vacuum pump, electric 12,
two stages
Technical data:
Delivery rate: 5.5 m³/h
Maximum negative pressure: 0,5 Pa
Lowest working temperature: 10°C
Oil quantity: approx. 450 ml
Noise level: approx. 63 dB (A)
Motor power: 190 W
Supply voltage: 230 V / 50 ... 60 Hz
dimensions: 310 x 130 x 250 mm
Weight: 10 kg
Commissioning:
1. Check whether the mains switch (7) is switched off.
2. The power cable (8), which is permanently connected to the pump, is
plugged into the power outlet to the AC mains.
3. Before starting up the pump, the oil level (2) must be checked or topped up.
4. Remove the cap of the outlet connection (6).
5. During pump operation, make sure that no water and no above-average
amount of water vapor does not get into the suction line of the pump.
6. Care must also be taken that the oil level during operation remains within
the visible range of the sight glass.
7. The rotary vane vacuum pump is very sensitive to mechanical
contamination. If this should be the case, damage will occur and a
general repair will be necessary. If "wrong" air is sucked in, the seal must
be tightened or replaced.
8. During prolonged operation, a permissible temperature rise of the pump
must be expected. If necessary, wait for the pump to cool down after it
has been switched off before tampering with it.
Operating conditions:
Operating position: horizontal, upright
Operating temperature in the room: + 15 °C ... 35 °C
Start-up of the pump at approx. 15 °C should be without load.
Humidity: max. 80% rel.
Maintenance:
The rotary vane vacuum pump is almost maintenance-free. It is sufficient
to regularly oil level and oil quality. When the pump is at a standstill, the oil
limit is in the middle; when the pump is in operation, the oil limit drops
significantly. A deficiency (shortage or clouding due to moisture and
impurities or slide valve wear) leads to performance losses loss of
performance, in the worst case to seizure of the rotary vane. Therefore, it
is recommended to use only original vacuum pump oil, which has viscosity
and lubricating properties, and which guarantees trouble-free function
guaranteed in the long term.
Warranty:
The warranty is 2 years from the date of delivery. The invoice is valid as
warranty certificate.
The warranty does not apply to the cleaning of the device and
malfunctions caused by improper use. The warranty also expires in the
event of operating errors.
Initial start-up:
Filling is carried out with the motor switched off through the opening
provided for this purpose on the top of the pump (1). Please fill the pump
until the oil level has reached the red mark in the center of the sight glass
(2). Then close the filling opening (1) and operate the pump for approx. 2
minutes, then check the oil level again and top up with oil if necessary. The
optimum filling level is in the middle of the sight glass (2). However, the
pump will not be affected if a little too much oil is accidentally filled in.
Description:
The rotary vane vacuum pump consists of the drive unit and the electric
motor. Both are directly coupled to each other, mounted on the base frame
and are air-cooled. The motor carries the mains connection line (8), the
mains switch (7) and the carrying handle (4). The housing of the rotary
vane pump has all the mechanical elements required for operation. The
housing of the rotary vane pump has all the mechanical elements
necessary for operation. On the upper side there are the oil filler opening
(1) and the suction port with bottle (5a). On the front side the sight glass of
the oil level control (2) and the oil drain plug (3).
Depending on the requirements and the accessories available, a different
suction port can be mounted. To do this, simply unscrew 5a and screw on
5b or 5c.
①
②
③
⑧
⑥
④
⑦
5a
5b
5c

Popular Water Pump manuals by other brands

Dover PSG Wilden PRO-FLO P1500 ENGINEERING OPERATION & MAINTENANCE MANUAL

Dover

Dover PSG Wilden PRO-FLO P1500 ENGINEERING OPERATION & MAINTENANCE MANUAL

Burcam 300910 installation instructions

Burcam

Burcam 300910 installation instructions

Stream HYDROBOX 900 manual

Stream

Stream HYDROBOX 900 manual

Alcatel I series user manual

Alcatel

Alcatel I series user manual

Grundfos SCALA 2 Service instruction

Grundfos

Grundfos SCALA 2 Service instruction

Edwards E2M28 instruction manual

Edwards

Edwards E2M28 instruction manual

AquaScape Signature series 400 Instruction and maintenance manual

AquaScape

AquaScape Signature series 400 Instruction and maintenance manual

CET PFP-23HPVGD-2D Instruction handbook

CET

CET PFP-23HPVGD-2D Instruction handbook

Pfeiffer Vacuum TMH 261 Translation of the original operating instructions

Pfeiffer Vacuum

Pfeiffer Vacuum TMH 261 Translation of the original operating instructions

Hydro-Force V115 manual

Hydro-Force

Hydro-Force V115 manual

STA-RITE DYNA-WAVE MPRA Series owner's manual

STA-RITE

STA-RITE DYNA-WAVE MPRA Series owner's manual

salmson EXPANSON COMFORT Installation and starting instructions

salmson

salmson EXPANSON COMFORT Installation and starting instructions

Danfoss Series 40 M46 Tandem Service manual

Danfoss

Danfoss Series 40 M46 Tandem Service manual

Fabit 43075H instruction manual

Fabit

Fabit 43075H instruction manual

Liberty Pumps LSG202M installation manual

Liberty Pumps

Liberty Pumps LSG202M installation manual

JB INDUSTRIES ELIMINATOR operating manual

JB INDUSTRIES

JB INDUSTRIES ELIMINATOR operating manual

Teryair DP 75 AL N Operation and maintenance guide

Teryair

Teryair DP 75 AL N Operation and maintenance guide

Grundfos APG Series Installation and operating instructions

Grundfos

Grundfos APG Series Installation and operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.