NUVE PN 150 User manual

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş.
PN 150 / PN 300
PLATELET INCUBATORS
USER’S MANUAL
1984
Z14. K25 266 Rev. No: 05 Rev. Tarihi: 01 / 2016

2
Dear Nüve User,
We would like to take this opportunity to thank you for preferring this Nüve product. Please
read the operating instructions carefully and keep them handy for future reference.
Please detain the packing material until you see that the unit is in good condition and it is
operating properly. If an external or internal damage is observed, contact the transportation
company immediately and report the damage. According to ICC regulations, this responsibility
belongs to the customer.
While you are operating the instrument please;
obey all warning labels,
do not remove warning labels,
do not operate damaged instrument,
do not operate instrument with a damaged cable,
do not move instrument during operation.
In case of a problem contact your Nüve agent for an authorized service or maintenance.
The validity of the guarantee is subject to compliance with the instructions and precautions
described in this manual.
Nüve reserves the right to improve or change the design of its products without any obligation
to modify previously manufactured products.
Information contained in this document is the property of Nüve. It may not be duplicated or
distributed without our permission.
NÜVE
SANAYİ MALZEMELERİ
İMALAT VE TİCARET A.Ş.
Saracalar Mah. Saracalar Kümeevleri No: 4/2
Akyurt 06750 Ankara-TURKEY
Tel : (90.312) 399 28 30 (pbx)
Fax : (90.312) 399 21 97

3
WARRANTY CERTIFICATE
1. Nüve warrants that the equipment delivered is free from defects in material and
workmanship. This warranty is given for a period of two years. The warranty period begins
from the delivery date.
2. Warranty does not apply to parts normally consumed during operation or general
maintenance or any adjustments described in the operating instructions provided with the
instrument.
3. Nüve does not accept any liability in case where the goods are not used in accordance with
their proper intent.
4. The warranty may not be claimed for damages incurred during the shipment, for damages
resulting from improper handling or use, abuse, fire, liquid spillage, tampering or
unauthorized repairs by any persons, use of defective or incompatible accessories, exposure
to abnormally corrosive conditions, use of the product in non-standard environmental
conditions, including but not limited to failure to meet requirements of ambient
temperature, lubrication, humidity or magnetic field influences, from the defects in
maintenance, negligence, bad functioning of auxiliary equipment, in the case of force
majeure or accident and incorrect power supply.
5. Any injury, loss or damage caused; due to a failure resulting from negligence of the
instructions given in this manual; is beyond the scope of the warranty conditions.
BEFORE OPERATING THE INSTRUMENT THIS MANUAL SHOULD BE READ CAREFULLY.
THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF THE
INSTRUCTIONS AND PRECAUTIONS DESCRIBED IN THIS MANUAL.
INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NÜVE. IT MAY
NOT BE DUPLICATED OR DISTRIBUTED WITHOUT PERMISSION.
PLEASE REGISTER ONLINE TO VALIDATE YOUR WARRANTY:
To register your warranty online, please visit our webpage www.nuve.com.tr and fill in the
“Warranty Registration Form”.

4
CONTENT
Page
1. INTRODUCTION .................................................................................................................................................... 5
1.1. USE AND FUNCTION ..................................................................................................................................... 5
2. TECHNICAL SPECIFICATIONS................................................................................................................................. 6
2.1. TECHNICAL SPECIFICATIONS TABLE .............................................................................................................. 6
2.2. ACCESSORIES ................................................................................................................................................ 6
2.2.1. FACTORY FITTED ACCESSORIES............................................................................................................ 6
2.2.2. OPTINAL ACCESSORIES......................................................................................................................... 6
3. PRECAUTIONS AND LIMITATIONS ON USE ........................................................................................................... 7
4. SYMBOLS AND LABELS.......................................................................................................................................... 7
5. INSTALLATION....................................................................................................................................................... 8
5.1. ENVIRONMENTAL CONDITIONS ................................................................................................................... 8
5.2. HANDLING AND TRANSPORTATION ............................................................................................................. 8
5.3. UNPACKING .................................................................................................................................................. 8
5.4. MAINS SUPPLY .............................................................................................................................................. 8
5.5. POSITIONING ................................................................................................................................................ 8
5.6. GENERAL PRESENTATION ............................................................................................................................. 9
5.7. CONTROL PANEL......................................................................................................................................... 10
5.8. PRIOR TO OPERATION ................................................................................................................................ 12
6. OPERATING PRINCIPLES...................................................................................................................................... 12
6.1. PROGRAMMING ......................................................................................................................................... 12
6.1.1. OPERATING MODE ............................................................................................................................. 12
6.1.2. SETTINGS............................................................................................................................................ 13
6.1.3. MEMORY............................................................................................................................................ 14
6.1.3.1. INTERNAL MEMORY ...................................................................................................................... 14
6.1.3.2. EXTERNAL MEMORY (USB STICK) .................................................................................................. 16
6.1.4. SMS .................................................................................................................................................... 16
6.1.5. E-MAIL................................................................................................................................................ 17
6.1.6. SET TEMPERATURE ............................................................................................................................ 19
6.1.7. GRAPH WORKING SCREEN ................................................................................................................. 19
7. PERIODIC MAINTENANCE AND CLEANING ......................................................................................................... 19
7.1. PERIODIC MAINTENANCE ........................................................................................................................... 19
7.2. CLEANING ................................................................................................................................................... 19
8. DISPOSAL MANAGEMENT CONCEPT .................................................................................................................. 19
9. TROUBLESHOOTING ........................................................................................................................................... 20
9.1. ERROR CODES ............................................................................................................................................. 20
9.2. FUSE REPLACEMENT ................................................................................................................................... 20
10. OPTIONS.......................................................................................................................................................... 21
10.1. AlerText™ GSM MODULE ....................................................................................................................... 21
10.2. NüveWarn™ REMOTE ALARM SYSTEM .................................................................................................. 22
10.2.1. MOUNTING REMOTE ALARM ................................................................................................................. 22
10.3. CHART RECORDER.............................................................................................................................. 23
11. ELECTRICAL CIRCUIT DIAGRAM....................................................................................................................... 24
11.1. PN 150 / PN 300 ELECTRICAL CIRCUIT DIAGRAM................................................................................... 24

5
1. INTRODUCTION
1.1. USE AND FUNCTION
PN 150 and PN 300 Platelet Incubators are designed for the storage of platelet. Besides this
usage, they can be used for any cooling application in the allowed temperature range. Platelet
Incubators are designed to be used together with PS 54 Platelet Shaker. 1 pc. PS 54 can be used
in PN 150 while 2 pcs. PS 54 can be used in PN 300. PS 54 has 9 shelves and the capacity of each
shelf is 6 platelet bags.
PN series platelet incubators are safely operated via N-SmArt™ microprocessor control system.
Temperature in the chamber is displayed digitally. It is possible to adjust the working
temperature in the range 20.0C to 35.0C with 0.1C sensitivity. PN 150 and PN 300 Platelet
Incubators keep temperature stable within given tolerances. The circulation and homogeneity
of the air which flows over the platelet bags is provided by a fan assembled at the top of the
chamber. Power switch of the device is equipped with a key lock to prevent unauthorized
tampering. The chamber is lit by a fluorescent lamp. Samples can be observed through a triple
heat resistant glass door. The door is equipped with magnetic gasket and with a key lock to
prevent unauthorized access.
PN series have audible and visual alarm system alarm systems for high and low temperature
conditions, power failure, temperature sensor failure and other faults associated with the
microprocessor control system. As an additional safety feature, audible and visible alarm is
activated when the door remains open more than two minutes. PN series store last 50 failures
in the memory and these errors can be followed on the display. Standard Ethernet connection
feature enables email sending to five different e-mail addresses in case of any error. PN series
offers AlerText™GSM module as an option and SMS can be sent to five different telephone
numbers in case of error.
The alarm system is fed by an automatically re-chargeable battery. In case of electricity
interruption, the chamber temperature is measured and displayed as well. Remote alarm outlet
and the outlet for connection to central alarm system are offered as standard.
PN series refrigerators have internal memory which can store temperature values for 10 years.
In addition to internal memory, temperature values can be recorded in USB stick. PN series have
a weekly circular chart recorder as an option. The temperature in the chamber is recorded by
this chart recorder operated by the backup battery. Chart recorder is equipped with a key lock
to prevent unauthorized usage.
PN series Platelet Incubators are manufactured according to the following standards:
EN 60601-1, EN 61010-1, EN 61000-6-3, EN 62366, EN 980, EN 60601-1-4, EN 60601-1-2, EN
60601-1-6, EN ISO 14971, EN 50419.
This device is in compliance with WEEE Regulaiton.
If the warnings mentioned in this manual are not considered, NUVE will not be responsible
from their results.

6
2. TECHNICAL SPECIFICATIONS
2.1. TECHNICAL SPECIFICATIONS TABLE
TECHNICAL SPECIFICATIONS
PN 150
PN 300
Temperature Range
20°C - 35°C
Temperature Sensor
Pt-100
Temperature Set and Reading
Sensitivity
0.1°C
Thermostat Working Sensitivity
± 0.1°C
Temperature Distribution(at 22°C)
± 1°C
Capacity with 250 ml Plasma Bag
54 bags
108 bags
Platelet Agitator
1 pcs.
2 pcs.
Capacity (Liters)
150
306
Refrigerant
R134a
Alarm Back-Up
Automatically re-chargeable battery for 12-hour
Failure Alarm
Audible and Visual
Temperature Alarm Range
± 2°C
Chamber Lighting (Fluorescent)
15 Watt
Door
Insulated Steel Frame with Triple Glass Window
and Key Lock
Control System
N-SmArt™ Programmable Microprocessor Control System
Temperature Display
Colorful LCD
Chart Recorder (Optional)
Circular, Weekly, Battery Charged, Locked
Internal Memory
10-year memory, one record per hour
USB
Standard
Ethernet
Standard
RS 232
Standard
Central Alarm System
Standard
Power Supply
230 V, 50/60Hz.
Power Consumption
600 Watt
Internal Material
Stainless Steel
External Material
Epoxy-polyester coated stainless steel
Internal Dim.s (ExDxY) (mm)
535x545x500
480x545x1050
External Dim.s (ExDxY) (mm)
735x805x1130
610x800x1740
2.2. ACCESSORIES
2.2.1. FACTORY FITTED ACCESSORIES
PN XXX Y Battery operated 7- day chart recorder
2.2.2. OPTINAL ACCESSORIES
A 08 191 AlerText™GSM alarm module
A 08 195 NuveCloser™ Software CD with 3 m. RS 232 cable
K 13 009 NuveWarn™ Remote alarm system with 10 m cable
A 08 070 Pen for chart recorder
A 08 073 Diagram paper for chart recorder (pack of 100 pieces)

7
3. PRECAUTIONS AND LIMITATIONS ON USE
The user shall pay attention to the following:
Do not operate the instrument for purposes other than its main purpose.
The instrument should only be used by authorized and trained staff after the instruction
manual has been read carefully. Only authorized technical staff can handle the product in
case of a failure.
Correctly grounded power supply should be used.
Only original spare parts and original accessories supplied by Nüve should be used.
PN series platelet incubators are designed to work with PS 54 platelet shaker. PS 54 has 9
shelves and each shelf can carry 250 ml 6 units of platelet bags, totally 54 bags can be placed
in PS 54. Do not load the device more than capacities.
PN 150 and PN 300 have sockets in the chamber for the usage of PS 54. These sockets
should be used for only PS 54.
There should not be any material in the chamber that may damage to the device.
The set temperature should be adjusted carefully so that the structures of the samples are
not destroyed.
4. SYMBOLS AND LABELS
Symbol in the operating instructions:
Attention, general hazard area.
This symbol refers to safety relevant warnings and indicates possibly dangerous
situations.
The non-adherence to these warnings can lead to material damage and injury to
personal.
Symbol in the operating instructions:
This symbol refers to important circumstances.
Labels on the product:
PN 150-300 Fuses (2x 10A)

8
5. INSTALLATION
5.1. ENVIRONMENTAL CONDITIONS
The instrument is designed to operate safely under the following conditions:
Indoor use only.
Ambient temperature: 5°C to 40°C.
Maximum relative humidity for temperature up to 31°C: 80%.
Maximum altitude: 2000 m.
Temperature for maximum performance: 15°C / 25°C.
5.2. HANDLING AND TRANSPORTATION
All handling and transportation must be carried out by using proper equipment and experienced
staff. The instrument must be supported underneath, never be turned upside down and be
moved on vertical position.
5.3. UNPACKING
Remove the cardboard box packing and the second nylon wrapping around the instrument.
Check that no damage has occurred during transportation. The below mentioned are provided
with the instrument, please check them;
1 ea. user’s manual and warranty certificate
2 ea. door keys, on/off keys
5.4. MAINS SUPPLY
The instrument requires 230 V, 50/60 Hz.
Please make sure that the supplied mains matches the required power ratings which are written
on the name plate of the instrument located at the back of the instrument.
Always plug-in the instrument to correctly grounded sockets.
A supply fitted with a circuit breaker should be used for protection against indirect
contact in case of isolation fault.
5.5. POSITIONING
Check that the positioning is suitable for the users.
Check that the instrument is stable on its four feet.
Check that the user will be able to follow up the operation even when he deals with
something else.
Check that the positioning of the device prevents interference with other equipment in the
near surrounding.
Leave at least 30 cm free space between the device and the wall/other equipment due to
ventilation of compressor.

9
5.6. GENERAL PRESENTATION
Figure 1
CONTROL
PANEL
SHELF
CHART RECORDER
(OPTIONAL)
DOOR
SWITCH
DOOR
EXTERNAL
SENSOR HOLE
COMMUNICATION
UNIT

10
Figure 2 –Communication Unit
Figure 3
5.7. CONTROL PANEL
Figure 4
The functions of F1, F2, F3 and F4 buttons depend on the meaning of the corresponding symbol
appearing on the display. The following table shows the meaning of these symbols.
FUSES
POWER
CORD
NC-CONTACT
OUTPUT
REMOTE ALARM OUTPUT
USB PORT-1
USB PORT-2
ETHERNET
RS 232
POWER
SWITCH
SAFETY THERMOSTAT

11
This symbol denotes the menu. You can enter the menu by pushing the button
corresponding to this symbol.
This symbol denotes the settings page. You can enter the settings page by pushing
the button corresponding to this symbol.
This symbol denotes the graph screen. The last 24 hours of operation of the
device can be monitored on the graph of the temperature vs. time by pushing the
button corresponding to this symbol. It is recorded every 10 minutes.
This symbol denotes the explanation of the error codes. You can learn detailed
explanation about the error by pushing the button corresponding to this symbol.
It appears in case of failures.
This symbol denotes the alarm mute key. You can mute the audible alarm by
pushing the button corresponding to this symbol when the alarm is activated in
case of error.
This symbol denotes backspace. You can return the previous page or exit from the
page by pushing the button corresponding to this symbol.
This symbol denotes the value increase key. You can increase the value by
pushing the button corresponding to this symbol while adjusting the numerical
values such as temperature or password.
This symbol denotes the value decrease key. You can decrease the value by
pushing the button corresponding to this symbol while adjusting the numerical
values such as temperature or password.
This symbol denotes the left key. It appears on the graph screen. The next graph
appears when the button corresponding to this symbol is pushed.
1) This symbol denotes the right key when the right and left keys appears on the
same screen. The previous graph is passed when the button corresponding to this
symbol is pushed.
2) It denotes the tab button when the left and the right keys do not appear on the
same screen.

12
This symbol denotes enter key. It is used for approval of adjustments.
The key corresponding to this symbol is used to turn the illumination on or off. If
the key is green, it shows that the lamp is off. The lamp goes on when the key is
pushed and the key turns red. Lamp goes off when the red key is pushed.
5.8. PRIOR TO OPERATION
Plug-in the instrument to correctly grounded sockets.
Insert the on / off key to power switch on rear side of the device.
Turn the device on by rotating the key clockwise.
See that the microprocessor control system is activated.
Learn the functions of the control panel (See Section 5.7).
6. OPERATING PRINCIPLES
6.1. PROGRAMMING
At most 70% of the surface area of the shelves should be loaded in order to obtain a
uniform temperature distribution.
6.1.1. OPERATING MODE
When the device is powered on, it starts to
operate automatically according to the last set
values. The screen on the left is operating mode
screen. Set temperature and chamber
temperature values appear in this screen. Current
time and date are shown at the top line. “ ”
symbol flashes on this screen while cooling is
performed. “ ” symbol flashes on this screen
while heating is performed.
Password screen will appear when the MENU
button (F1) is pushed. You can enter the password
by using the value increase (F1) and decrease keys
(F2). The password is user’s password and can be
changed on settings page (See section 6.1.2.). The
password is 0000 for the first usage.

13
Main menu page appears when the password is
approved by pushing enter key (F4). Submenu
can be selected by using the value increase and
decrease keys. The selected menu becomes red
and enter key (F4) is pushed to enter this menu.
6.1.2. SETTINGS
Settings page appears when “settings”
submenu is selected from the menu page by the
value increase (F1) and decrease (F2) keys. The
required submenu is chosen by using the value
increase and decrease keys.
“Date/Time Settings” submenu is selected on
the settings page by the value increase (F1) and
decrease (F2) keys. The page shown on the left
appears when enter key (F4) is pushed. Time is
shown in the format of “hour:minute” and the
cursor is on hour part when “Date/Time
Settings” page comes to the screen. Hour is
adjusted by pushing the value increase (F1) and
decrease (F2) keys. In order to continue to
adjust, push enter key (F4).
If you want to exit this page without any change, push the backspace key (F3).
“Select Language” submenu is selected on the
settings page by using the value increase (F1) and
decrease (F2) keys. The page on the left appears
when enter key (F4) is pushed. The language of
the control panel can be set as Turkish, English,
French, Russian or Spanish. Push the value
increase (F1) and decrease (F2) keys to select the
language and then push the enter key (F4) to save
the selection. If you want to exit this page without
any change, push the backspace key (F3).

14
“Screen Saver Settings” submenu is selected on
the settings page by using the value increase (F1)
and decrease (F2) keys. The page on the left
appears when enter key (F4) is pushed. Screen
saver can be on or off by pushing the value
increase (F1) and decrease (F2). If the option of
“open” is selected, activation time of screen
saver is adjusted on screen saver time in
minutes. If you want to exit this page without
any change, push the backspace key (F3).
It is suggested to use Screen Saver as ON for energy saving.
“Change Password” submenu is selected on the
settings page by the value increase (F1) and
decrease (F2) keys. The page on the left appears
when enter key (F4) is pushed. After old
password value is entered by pushing the value
increase and decrease keys, push the enter key.
New password value can be entered in the “New
Password” and “Retype New” sections. If you
want to exit this page without any change, push
the backspace key (F3).
6.1.3. MEMORY
“Memory” submenu is selected on the menu
page by the value increase (F1) and decrease (F2)
keys. The page on the left appears when enter
key (F4) is pushed.
6.1.3.1. INTERNAL MEMORY
“Failure History” submenu is selected on the
memory page by the value increase (F1) and
decrease (F2) keys. The page on the left appears
when enter key (F4) is pushed. The last 50
failures can be followed on this page. There are
10 pages in total for error history and 5 failures
are listed on each page. Failures are listed from
the current date to the earlier. Push the value
increase (F1) and decrease (F2) keys to pass the
other page. Push the backspace key (F4) in order

15
to exit the page.
“History (TABLE)” submenu is selected on the
memory page by the value increase (F1) and
decrease (F2) keys. The page on the left appears
when enter key (F4) is pushed. Enter a date to
analyze corresponding operation by pushing the
value increase and decrease keys. Push the
enter key (F4) to access the table.
Temperature values corresponding to the
selected date are shown on three pages. Push
the value increase (F1) and decrease (F2) keys to
pass from one page to another. “–“appears in
the temperature column if the refrigerator does
not run on the selected date. Push the
backspace key (F3) to return previous page.
When the temperature sensor is broken, “–“appears in the temperature column since
temperature cannot be read by the sensor.
“History (GRAPH)” submenu is selected on the memory page by the value increase (F1) and
decrease (F2) keys. “Select Record Date” page appears when enter key (F4) is pushed. Enter a
date to analyze corresponding operation and then push the enter key (F4). You can analyze the
data in the graph. The data belonging to the other dates can be followed by using left (F2) and
right (F3) keys. Push the backspace key (F4) to return the previous page.
The page on the left appears when the settings
key (F1) is pushed on the graph page.
Temperature axis (y-axis) can be adjusted on
this page. Working temperature determines the
midpoint of the temperature axis. The working
temperature value should be close to the set
temperature value for more comfortable
observation. Sensitivity value is the difference
between two values on the y-axis. As sensitivity
value getting smaller, a more detailed graph appears.

16
6.1.3.2. EXTERNAL MEMORY (USB STICK)
USB Stick is connected to USP port-1 of the communication unit (Figure 2). “ ” appears on
the operating mode screen when the USB stick is identified by the microprocessor system.
If does not appear on the screen, USB stick may be defective or may not be
connected correctly.
When “USB Stick” submenu is selected on the memory page, there are three options: “Copy
History”, “Instant Record” and “Copy Failure History”.
Copy History: As USB stick is connected to the device, all data in the internal memory is
transferred to USB stick when “Copy History” is selected. When loading bar shows 100%, push
the enter key (F4) to exit the page.
Instant Record: As USB stick is connected to the
refrigerator, the page on the right appears
when “Instant Record” is selected. The period
of the record time can be selected by using the
value increase (F1) and decrease (F2) keys. As
soon as the enter key (F4) is pushed, recording
starts and the display returns the main page. If
you want to exit this page without starting to
record, use backspace key (F3).
You can remove USB stick when you want to finish the record process.
Copy Failure History: As USB stick is connected to the device, all error history in the internal
memory is transferred to USB stick when “Copy Failure History” is selected. When loading is
finished completely, push the enter key (F4) to exit from this page.
6.1.4. SMS
In order to use SMS, optional GSM module is mandatory to have. Refer to Section
10.1 for information regarding GSM module connection.
“SMS” submenu is selected on the menu page by
the value increase (F1) and decrease (F2) keys.
The page on the left appears when enter key (F4)
is pushed. Submenus on SMS page is selected by
using increase (F1) and decrease (F2) keys.
Selected submenu can be adjusted by using enter
key (F4). If you want to exit the page, push the
backspace key (F3).
Activation or deactivation of SMS function can be adjusted on “SMS” submenu. If you want
to activate this function, choose “on” by pushing the enter key (F4). If you want to
deactivate SMS function, choose “Off” by pushing the enter key (F4).

17
“Repeat time” is the frequency of sending SMS. The user is notified again by sending SMS if
the failure still continues. Repeat time can be adjusted to 8 hours, 16 hours and 24 hours by
pushing enter key (F4).
“SMS Numbers” is selected on the SMS page by
using the value increase (F1) and decrease (F2)
keys. The page on the left appears when enter
key (F4) is pushed. The phone numbers can be
entered in this page and SMS notifications are
delivered to these phone numbers in case of
any failure.
SMS can be sent to 5 different mobile phones. According to demand, 1st, 2nd and 3rd
numbers can be users’numbers and 4th and 5th numbers can be technical service staffs’
numbers.
Enter country code before the phone numbers.
A cursor flashes under a digit which means you can set this digit of the number. Each digit of
phone numbers is entered one by one. The first digit of the phone number is entered by
pushing value increase (F1) and decrease keys (F2) and tab key (F3) is pushed to enter the
second digit. The second number is entered pushing value increase (F1) and decrease keys
(F2) and tab key (F3) is pushed to enter the next digit. After all digits of the phone number
are entered completely by this way, push the enter key (F4) to enter the next phone
number.
Tab key (F3) is used for changeover from one digit of phone number to another.
After all phone numbers from “Phone 1” to “Service 2” is entered as mentioned above, push
the enter key (F4) and return to main page.
6.1.5. E-MAIL
“E-Mail” submenu is selected on the menu page
by the value increase (F1) and decrease (F2) keys.
The page shown on the left appears when enter
key (F4) is pushed. The choice of submenus on e-
mail page is changed by pushing increase (F1) and
decrease (F2) keys. After choosing a submenu,
the adjustments for this submenu is made by
pushing enter key (F4). If you want to exit this
page, push the backspace key (F3).
Activation or deactivation of e-mail function can be adjusted from “E-mail” submenu.
choose “on” by pushing the enter key (F4) to activate this function. Choose “off” by pushing
the enter key (F4) to deactivate e-mail function.
Ethernet settings should be adjusted by technical service staff for the first usage.
Otherwise, this function does not work.

18
“E-Mails” is the page which e-mail addresses will be written on.
“Login” is the adjustments for the e-mail addresses.
“E-Mails” is selected on the e-mail page by the
value increase (F1) and decrease (F2) keys. The
page on the left appears when enter key (F4) is
pushed. The e-mail addresses can be typed in this
page and e-mail is sent to these addresses in case
of any failure.
Each character of e-mail addresses is typed one by one. The first character is entered by
pushing value increase (F1) and decrease keys (F2) and tab key (F3) is pushed to enter the
second character. The second character is written by pushing value increase (F1) and
decrease keys (F2) and tab key (F3) is pushed to enter the next character. After all characters
are written completely in this way, push the enter key (F4) to enter the mail server. After
mail server is written in the same way, push the enter key (F4) to write next e-mail address.
After third email address is written, push enter key (F4) to enter other e-mail addresses.
After all e-mail addresses are written as mentioned above, push enter key (F4) and return to
main page.
Tab key (F3) is used for changeover from one character of address to another.
“Login” is selected on the e-mail page by the
value increase (F1) and decrease (F2) keys. The
page on the left appears when enter key (F4) is
pushed. Each character of e-mail addresses is
written one by one. The first character is entered
by pushing value increase (F1) and decrease keys
(F2) and tab key (F3) is pushed to enter the next
character. The second character is entered by
pushing value increase (F1) and decrease keys
(F2) and tab key (F3) is pushed to enter the next character. After all characters of user’s name
are entered by this way, push enter key (F4) to enter the mail server of the e-mail address. After
mail server is entered in the same way, push enter key (F4) to pass the “password” submenu.
Enter the password and push enter key (F4).
The port provided by internet server is entered on the part of “Port”. After port is entered,
enter key (F4) is pushed to return the main menu. Tab key (F3) is used for changes from one
character of address to another.

19
6.1.6. SET TEMPERATURE
When the setting key (F2) is pushed in the
operation mode, password page appears. This
password is the user’s password. The password is
0000 for the first usage. “Working Settings” page
on the left appears. The set temperature can be
adjusted between 20⁰C and 35⁰C by pushing the
value increase (F1) and decrease (F2) keys. After
set temperature value is entered, push the enter
key (F4) to save the value. Push the backspace key
(F3) to exit this page without any change.
In case of active audible alarm, firstly, push the alarm mute key (F2) to access the
working settings page.
6.1.7. GRAPH WORKING SCREEN
When the graph button (F3) is pushed in the operation mode, the graph page comes to the
screen. Daily operation can be followed on this graph. If you want to analyze old operations,
visit the memory page (See Section 6.1.3).
7. PERIODIC MAINTENANCE AND CLEANING
7.1. PERIODIC MAINTENANCE
Clean the condenser surface with a vacuum cleaner every 3 month.
7.2. CLEANING
Clean the platelet incubators at room temperature after disconnecting the power cable.
Remove the PS 54 Platelet Shaker in the incubators while cleaning.
Clean the platelet incubators with a wet cloth to remove dirt and dust.
Liquid detergent can be used to remove tough dirt.
Take precautions handling chemical cleaners.
Check the external condition of the platelet incubators regularly and ensure any rust spots
that may develop are removed.
8. DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the
responsibility of the user to arrange proper disposal of the individual components.
All parts which may comprise potentially infectious materials have to be disinfected by suitable
validated procedures (autoclaving, chemical treatment) prior to disposal. Applicable local
regulations for disposal have to be carefully observed.
The instruments and electronic accessories (without batteries, power packs etc.) must be
disposed off according to the regulations for the disposal of electronic components. Batteries,
power packs and similar power source have to be dismounted from electric/electronic parts and
disposed off in accordance with applicable local regulations.

20
9. TROUBLESHOOTING
If the device fails to operate, please check the followings:
The power switch is on.
The fuse is sound.
The plug is plugged-in properly.
The plug is not defective.
The mains supply is present.
9.1. ERROR CODES
Blue colored screen turns into red in case of any failure and a short explanation of the error
appears at the bottom line on the red screen. Push the key corresponding to question mark (F3)
for detailed explanation of failures.
Failures which may be encountered during operation is listed below:
EoFF: This error code shows power failure.
After power interruption, push the key corresponding to question mark (F3) in order
to see the time when power interruption occurs.
Err1: It appears when temperature sensor is defective.
Err2: It appears when there is a communication error between display and main PCB.
Err3: It appears when there is a failure in cooling system.
Err5: It appears when door is kept open for a while.
Err6: It shows that the temperature in the chamber is not in the temperature alarm range.
Err7: It appears when software version is not compatible. Push the F3 key corresponding to the
question mark to see the explanation. Inform the authorized NÜVE agent regarding the error
explanation.
A list consisting of the codes of errors can be accessed from the memory page (See
section 6.1.3.1). ‘EoFF’, ‘Err 3’, ‘Err5’ and ‘Err6’ do not appear on working screen and
they are only shown in the list of the error history.
If an error occurs, please contact to an authorized Nüve agent to seek technical help.
9.2. FUSE REPLACEMENT
The fuses shall be always be replaced by the authorized personnel.
This manual suits for next models
1