Orbit Merret OMR 700 Series User manual

OMR 700
PAPERLESS RECORDER
USER GUIDE
148.9~C
-263mm
453mV
Outstanding Measurement Value

2| USER GUIDE OMR 700
TECHNICAL DATA
Paperless recorders of the OMR series conform to the Europen regulations //EU and //EU.
They are up to the following European and Czech standards:
EN -, Electrical safety
EN -, Electronic measuring, control and laboratory devices - Requirements for EMC „Industrial use“
Seismic capacity:
IEC : , art.
The recorders are applicable for unlimited use in agricultural and industrial areas
ORBIT MERRET, spol. sr.o.
Vodňanská /
Praha
Czech Republic
Tel: + -
Fax: + -
e-mail: [email protected]
www.orbit.merret.eu
SAFETY INSTRUCTIONS
Please read carefully the enclosed safety instructions and observe them!
Installation, all operational interventions, maintenance and service must be performed by a qualified perso-
nnel and in accordance with the attached information and safety regulations. The manufacturer is not liable
for damage caused by improper installation, configuration, maintenance, and service.
The recorder must be installed according to the respective application. Incorrect installation can cause a
malfunction, which can result in damage or accident.
The recorder uses dangerous voltages that can cause a fatal accident. Before you start solving problems
(e.g. in case of failure or disassembly), the device must be disconnected from the power supply. For safety
information the EN - + A standard must be observed.
When removing or inserting a card, observe the safety instructions and follow the recommended procedure.
During any intervention the recorder must be disconnected from the power supply.
Do not attempt to repair or modify the device. A defective recorder must be sent for repair to the manufacturer.
These devices should be safeguarded by isolated or common fuses (breakers)!
The recorder is not designed for installation in potentially explosive surroundings (Ex). Use it only outside
potentially explosive surroundings.

USER GUIDE OMR 700 | 3
14
8
.
9~C
-263mm
453mV CONTENTS .
. CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. DESCRIPTION OF THE INSTRUMENT. . . . . . . .
. CONNECTION OF THE INSTRUMENT . . . . . . .
Instrument connection. . . . . . . . . . . . . . . . . . .
Card connection . . . . . . . . . . . . . . . . . . . . . . . .
. CONTROL OF THE INSTRUMENT . . . . . . . . . .
. SETTING OF THE INSTRUMENT . . . . . . . . . . .
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . .
Control bar. . . . . . . . . . . . . . . . . . . . . . . . . . . .
First power on . . . . . . . . . . . . . . . . . . . . . . . . .
. TECHNICAL DATA . . . . . . . . . . . . . . . . . . . . . . .
. DIMENSIONS AND ASSEMBLY . . . . . . . . . . . .
. WARRANTY CERTIFICATE . . . . . . . . . . . . . . . .

4| USER GUIDE OMR 700
DESCRIPTION
OF THE INSTRUMENT.
2.1 DESCRIPTION
Paperless recorder OMR 700
This recorder is intended for technologies and workings, where it is needed to display and/or record a number
of electrical and nonelectrical values at one place. Universality, versatility and
in particular good value for money predestine the recorder to fulfil most of your
demanding needs including the IP front panel cover.
Ourpaperlessrecorderhasbeendevelopedwithversatilityandintuitivecontrol
in mind. Thanks to its modularity the user can insert input or output cards into
any of the existing slots. Maximal configuration of the recorder thus allows to
measure and record up to inputs. In order to increase reliability, the recorder
has two systems - primary and backup.
Always on board are digital control iputs and outputs, serial line
RS , Ethernet /, USB connector as well as a MB internal memory
to record the measured data.
Projection
Color 5,7“ TFT display with fine resolution dominates the device. The display is multi-touch and it there-
fore allows an ease of use.
Control
The recorder is controlled by both the touch screen and the push buttons with adjustable functions,
positioned underneath a sliding front door. Two LEDs indicate run/error and the state of data recording.
Setting
All functions and settings can be performed directly on the instrument´s display in a clear graphical menu.
Data recording
The OMR can record measured data from any of its active inputs, nodes and mathematical functions.Data
are stored in the internal MB memory with compression that allows up to four-fold increase in its physical
memory without slowing down. Data can also be stored on an external SD card or USB flash drive. In case of a
limited number of measuring inputs, measurement data can be stored with a period of up to ms.
The records can be either in BIN or „CVS“ format. However, the latter is much more demanding on memory.
Recording speed according to number of channels / memory space
Recording speed 16 inputs 48 inputs 80 inputs 96 inputs
1 ms 2 hours x x x
10 ms 20 hours 7,5 hours x x
1 s 2,5 months 1 months 16 days 13 days
1 min 13 years 5 years 2,5 years 2,2 years
10 min 132 years 52 years 26 years 22 years

USER GUIDE OMR 700 | 5
14
8
.
9~C
-263mm
453mV .
DESCRIPTION
OF THE INSTRUMENT
Modules
The development of the device has been performed with an increased emphasis on technical solutions and
universality. Card design not only allows their use in any position of the recorder, but also their additional
insertion into vacant slots. Thus, if new requirements to increase the number or Typee of inputs and outputs
occur in the course of using the recorder, just order yesther card and insert it into a vacant slot. In this way the
instrument can „grow“ in compliance with your requirements.
All analogue modules are fully isolated from the internal bus, and some cards have galvanic isolation even
between individual channels.
Basic version of the recorder includes power supply module and communication module with Ethernet /,
RS (MODBUS), five digital inputs and two digital outputs.
The files on the DVD
Manuals
Recorder_Manual_2.0.2_en User Manual - EN
Manual_OMR700_2018_2v0_en User Guide - EN
Software
Recorder PC Programme for Remote Viewing of Measurements in Real Time
OMRViewer Recorded Data Viewer for OMR700
and current data sheets

6| USER GUIDE OMR 700
CONNECTION
OF THE INSTRUMENT.
Supply lines of the instrument should not be situated in proximity of the incoming low-voltage signals.
Contactors,motorswithlargerinputpowerandotherpowerelements should not be situated close to the recorder.
Supply lines to the instrument input (measured quantity) should be situated at a sufficient distance from all
power lines and appliances. If this can not be secured, it is necessary to use shielded leads with connection
to the ground (terminal E).
The devices are tested according to the standards for use in industrial area, yet we recommend to abide by
the above principles.
ABCDEFGHIJKLMNOPQ
ABCDEFGHIJKLMNOPQ
ABCDEFGHIJKLMNOPQ
ABCDEFGHIJKLMNOPQ
ABCDEFGHIJKLMNOPQ
ABCDEFGHIJKL
ABCDEFGHIJKL
ABCDEFGHIJKL
ABCDEFGHIJKL
ABC
Power source
Firmly placed card
Communication
Firmly placed card
B5: Card slot
Ethernet
RS 485
Digital
inputs
B4: Card slot
B3: Card slot
B2: Card slot
A3: Card slot
A2: Card slot
A1: Card slot
A4: Card slot
Digital
outputs
A5 - Power supply
ABC
+−
L
N
PE
B1
ABCDEFGHIJKL
+
−
L-
L+ RS 485
GND I/O
GND
GND I/O
IN 4
IN 1
IN 2
IN 3
IN 0
OUT 0
OUT 1
RS 485 - Modbus
2
1
Outputs - 24 VDC
Inputs- 24 VDC
3
Ethernet - Modbus TCP
4
1
2
4
3
Slots A are designated for fast analogue cards.
Slot B5 is designated for cards DO.1/2.
There are no rectrictions for placements of other cards.

USER GUIDE OMR 700 | 7
14
8
.
9~C
-263mm
453mV .
CONNECTION
OF THE INSTRUMENT
IN.2 4x PM input U-I
DC - I: ±5/±20 mA, 0...20/4...20 mA
DC - U: ±2/±5/±10/±40 V, 0...2/5/10/40 V
1
2
IN.02
4 3
ABCDEFGHIJKLMNOPQ
2 1
Shielding
GND 1
GND 2
GND 3
GND 4
++++++++ ----
12121212
IN.1 3x Universal input
IN.01
3
ABCDEFGHIJKLMNOPQ
2 1
Shielding
++++ -
++++ -
++++ -
PM: 0…5/20 mA/4…20 mA
DC: ±60/±150/±300/±1 200 mV
T/C: J/K/T/E/B/S/R/N/L
DU: Lin. potentiometer (> 500 Ω)
OHM: 0…0,1/0,3/1/3/10/30 kΩ
RTD: Pt 50/100/500/1 000
Cu: Cu 50/100
Ni: Ni 1 000/10 000
5
4
2
1
PM: ±2 V/±5 V/±10 V/±40 V
3
E+/E+/ES+
S+/S+
S-
E-/E-/ES-
5
++++
E+/E+/ES+
S+/S+
S-
E-/E-/ES-
5
++++
12
4
3
12
4
3
12
4
3
E+/E+/ES+
S+/S+
S-
E-/E-/ES-
5
IN.4 4x T/C nputIN.3 4x RTD input
IN.03
4
OHM: 0…0,1/0,3/1/3/10/30 kΩ
RTD: Pt 50/100/500/1 000
Cu: Cu 50/100
Ni: Ni 1 000/10 000
E+/ES+
S+
ES-
E+/ES+
S+
ES-
E+/ES+
S+
ES-
E+/ES+
S+
ES-
3
ABCDEFGHIJKLMNOPQ
2 1
Shielding
GND 1
GND 2
GND 3
GND 4
IN.04
4
T/C: J/K/T/E/B/S/R/N/L
3
ABCDEFGHIJKLMNOPQ
2 1
Shielding
GND 1
GND 2
GND 3
GND 4
++++ ----
IN.6 12x DC input, currentIN.5 5x RTD input
IN.06
ABCDEFGHIJKLMNOPQ
123456789101112
++++
-
DC - I: 0...5 mA/0...20 mA/4...20 mA/±5/±20 mA/
++++
-++++
-
IN.05
45
OHM: 0…0,1/0,3/1/3/10/30 kΩ
RTD: Pt 50/100/500/1 000
Cu: Cu 50/100
Ni: Ni 1 000/10 000
3
ABCDEFGHIJKLMNOPQ
2 1
Shielding
E+/ES+
S+
ES-
E+/ES+
S+
ES-
E+/ES+
S+
ES-
E+/ES+
S+
ES-
E+/ES+
S+
ES-

8| USER GUIDE OMR 700
.
CONNECTION
OF THE INSTRUMENT
IN.8 2x input for strain gauges
IN.7 12x DC input, voltage
IN.07
ABCDEFGHIJKLMNOPQ
123456789101112
++++
-
DC - U: 0…2 V/0…5 V/0…10 V/0…40 V/±2/±5/±10/40 V
++++
-++++
-
IN.08
ABCDEFGHIJKLMNOPQ
DMS: 1…16 mV/V
Shielding
1
+++---
GND 2
2
+++---
GND 1
IN.10 2x AC/PWR inputIN.9 3x PM input U-I
1
2
AC - I: 0…60/150/300 mV
0…1/2,5/5 A
AC - U1: 0...10/250 V
1 2
ABCDEFGHIJKL
IN.10
3
1
23
12
3
AC
-
U2: 0...120/450 V
IN.09
ABCDEFGHIJKLMNOPQ
1
Shielding
GND 1
++ -
DC - I: 0…20 mA/4…20 mA/±20 mA
DC - U: 0…5 V/0…10 V/±5 V/±10 V
1
2
12
2
GND 2
++ -
12
3
GND 3
++ -
12
IN.12 12x Pulse input
IN.11 8x Digital input
ABCDEFGHIJKL
Shielding
IN.11
12345678
AC/DC: 12…250 V AC/DC
IN.12
ABCDEFGHIJKLMNOPQ
123456789101112
+++
-- +++++++++ --
Shielding
contact
1
2-wire sensors, NPN NO
2
3-wire sensors, PNP NO
3
3-wire sensors, PNP NO
4
1
23
4
GND
GND
GND
GND
+ External supply
560 R/2W

USER GUIDE OMR 700 | 9
14
8
.
9~C
-263mm
453mV .
CONNECTION
OF THE INSTRUMENT
IN.14 2x input for LVDT sensorsIN.13 2x Fast pulse input
IN.14
ABCDEFGHIJKLMNOPQ
1
GND 1
GND 2
Shielding
2
1
2
1
2
3-wire LVDT sensors
1
2
5-wire LVDT sensors
IN.13
ABCDEFGHIJKLMNOPQ
+++++++++++ -
Shielding
contact
1
2-wire sensors, NPN NO
2
3-wire sensors, PNP NO
3
3-wire sensors, PNP NO
4
1
23
4
Excitation
GND
560 R/2W
IRC sensors, line/NPN/PNP
5
2
5
++
A
B
C
ABCABCABCABC
1

10 | USER GUIDE OMR 700
CONNECTION
OF THE INSTRUMENT.
OUT.2 8x Relay, switch-on contactOUT.1 4x Relay, switch-over contact
ABCDEFGHIJKL
OUT.2
12345678
ABCDEFGHIJKL
OUT.1
1234
OUT.4 16x OC, NPN
OUT.3 8x OC, NPN
OUT.3
ABCDEFGHIJKLMNOPQ
12345678
OUT.4
ABCDEFGHIJKLMNOPQ
12345678910111213141516
++++++++++++++++
-
OUT.6 6x SSR
OUT.5 8x OC, PNP
OUT.5
ABCDEFGHIJKLMNOPQ
+++++++++ --- ---
12345678
10...30 VDC
OUT.6
ABCDEFGHIJKLMNOPQ
123456

USER GUIDE OMR 700 | 11
14
8
.
9~C
-263mm
453mV .
CONNECTION
OF THE INSTRUMENT
AO.1 2x Analogue output
AO.01
ABCDEFGHIJKLMNOPQ
++
-
Analog output - voltage
Analog output - current
1
2
1
1
2
++
-
2
1
2
Shielding
DO.2 1x PROFINET
DO.1 1x PROFIBUS
DO.1
1
9
5
6
B: RxD/TxD-P data reception/transmission, positive
CNTR: signal for repeater control
DGND: reference potential for data and +5 V
VP: +5 V
A: RxD/TxD-N data reception/transmission, negative
Pin asignment
3
4
6
8
5
DO.2
Port 1 Port 2
AO.2 4x Analogue output
AO.02
ABCDEFGHIJKLMNOPQ
++
-
Analog output - voltage
Analog output - current
1
2
1
1
2
++
-
2
1
2
++
-
3
1
2
++
-
4
1
2
Shielding
EXC.1 4x Excitation
EXC.1
ABCDEFGHIJKLMNOPQ
+++++ +++ −
1
5/ 10/ 12/ 24 VDC
5/ 10/ 12/ 24 VDC
5/ 10/ 12/ 24 VDC
5/ 10/ 12/ 24 VDC
234
−− −−−

12 | USER GUIDE OMR 700
CONTROL
OF THE INSTRUMENT.
Control bar
5,7“ TFT display
Stylus
Front panel, IP cover
Opening the lid
Toensurethe I P64 cover oftheinstrunt it is
necessary to arRange for proper panel mounting and
proper click of the front lid. Proper snap of the front lid.
!
The lid can be opened by
pressing the two blue tabs
toward each other and
lowering the lid
with a capacity
touchpanel
USB
USB connector for Flash drive
Com
MicroUSB for connection to PC
SD Card
Slot for SD card
Rst
Reset button
Button „2“
button function can be
set in menu
Button „1“
button function can be
set in menu
Stat
OK
Error
Rec
active data recording

USER GUIDE OMR 700 | 13
14
8
.
9~C
-263mm
453mV
CONTROL
OF THE INSTRUMENT .
Elements under the hinged lid
If necessary, a seal can be fitted to the hinged lid as a mechanical security
against possible accidental opening.
Your SD card or USB Flash drive will remain safely stored.
LED signalling
Signalling during device start up
Stat Rec Display Meaning
off off inactive Device is not powered
bliká
rychle off inactive Normal state
on off inactive Undervoltage, processor not running
bliká
rychle off inactive Undervoltage, processor running
Signalling while device is running
Stat Rec Display Meaning
flashes off active Normal state
flashes off active General error
-- flashes active Recording in progress
flashes off inactive Short term power outage
on off inactive Long term power outage

14 | USER GUIDE OMR 700
SETTING
OF THE INSTRUMENT.
BASIC BUILDING BLOCKS
OMR 700
Functionality of the paperless recorder OMR700 is based on the following parts
• Inputs and outputs
- come from the IO cards (fixed B or expanding A - A, B - B). They themselves contain conversions
• Nodes
- mathematical or other calculations with the goal of providing requested adjustments of the measured
signals or preparation of the output values
• Screens
- graphical representation of the measured or calculated values on the recorder display
• Records
- what, where, how often, in what format, and other parameters of measured or calculated data for recording
• Timers
- provide periodic execution of linked blocks with a given period
• Constants
- spontaneously unchangeable named values for calculations
Block chart of the recorder OMR
Inputs
Constants
Timers
Nodes
Screens
Outputs
Records

USER GUIDE OMR 700 | 15
14
8
.
9~C
-263mm
453mV
SETTING
OF THE INSTRUMENT .
From the block chart it is apparent that the central point of the recorder is created by Nodes. They process the
measured inputs, constants, other nodes, and using the preset calculations they calculate a new value. The
calculation is kept in time by the timer. The calculated values of the nodes can be displayed on the screen and
recorded on the media. The outputs can also be equipped by them. An important feature of the nodes is the
fact that they can have a history (they remember previous values). This is an optional configuration offering the
advantage of displaying the value of the node with its history (running chart).
Timers have an optional setting period. The client can set the period within the Rangeof ms up to days
(by ms, while the dialog limits the settings to the order of ms, s, minutes, hours, and days). There are N timers
in the recorder ( at the moment) so that it is necessary to choose a proper setting for each timer to cover
the needs of the entire recorder. The timers control calculations of the nodes and recording on the media,
while the internal mechanism guarantees that the nodes are calculated first and only then the new values are
recorded on the media.
Inputs and Outputs provide rate, which is different for each Typee of the card and even for each register on the
card. The IO cards are in fact designed as intelligent ones = they conduct their operations in order to relieve the
main processor. They are organized into a set of registers. Some of the registers are configurable (e.g. those of
input Range), some are designated for measured values under different phases of processing. A Typeical input
card provides several values for each input – direct input value of the converter (converter bits), value converted
into electrical value (e.g. mA), and the resulting converted value (e.g. in case of weight, the strain gauge input
card recalculates the voltage of the strain gauge into weight by a preset formula - Range, tare). Similarly, it is
possible to control also the outputs. For example: You set the requested turns of a ventilator and the card will
itself, according to the preset parameters, recalculate the measured value into voltage and set it on its output.
Constants are designated for easy and well-arRanged changes of the settings, e.g. of the required values,
filtration parameters... On one place the value can be changed, used for calculations, displayed and recorded.
Screens are used for displaying the measured values. There are N screens and each of them can occupy up
to M different elements like running chart, bar chart, nummber indicator, finger measuring indicator... You set
the element Typee, size, location, number of displayed values, color, Range... So you can build very diverse
screens. The redraw period is set in such a manner that it enables a smooth and fast enough drawing, which,
at the same time, relieves the main processor. Some elements (as e.g. the running chart) display, besides the
current values, also their previous ones. By these elements it is therefore advantageous if the displayed node
has a history that is used for filling up the chart when switching over to the screen.
Records serve for recording values on the media. To each record you can assign a name, frequency, file format,
number of records in the file, where you want to record, and, of course, what you want to record. All records
can be viewed in the record browser. They can also be downloaded to a PC and displayed there

16 | USER GUIDE OMR 700
SETTING
OF THE INSTRUMENT.
5.2 CONTROL BAR
In the upper part of the display there is a dark blue Control Bar. It shows the main control elements.
Switch screens
Previous screen
Next screen
Logged in users
Nobody logged in
Press icon to log in
User
Press icon to log out
Operator and higher
Press icon to enter menu
Recorder status
OK
Device is OK
Warning
Device is out of pre-defined values
Error
Device is out of pre-defined values
Its functionality may be compromised
Critical error
Device is out of pre-defined values
Its functionality is compromised
Recording status
Recording inactive
Recording in progress
No errors detected
Recording in progress
Medium > 80% full
Recording in progress
Medium > 90% full
Name of current screen
Click item to open menu from defined screen
Datum a čas
Current Date and Time

USER GUIDE OMR 700 | 17
14
8
.
9~C
-263mm
453mV .
SETTING
OF THE INSTRUMENT
State of the memory media
Overwiev of the memory media
When the storage is full at % or higher, the icon colour changes to yellow, when the storage
exceeds % of its capacity, the icon colour changes to red. If the recording is in progress and
there is no error condition, the icon colour is green. If the recording is not turned on, the icon
colour is white. The logged-in user will click on the icon to access the media overview.
State of the recorder
The recorder always operates in one of its four states (further details in the chapter Errors and warnings).
„OK“ - device has not detected any function problem
„Warning“ - device outside the specified values, but without consequences on its functioning.
This state can be caused for instance by an undervoltage greater than %, by filling the
memory in excess of %, and by many other causes. If the device returns within the specified
values, „Warning“ state will disappear and the log record will be carried out.
„Error“ - device outside the specified values, but with possible consequences onitsfunctioning.
This state can be caused for instance by an undervoltage greater than %, by filling the memory
in excess of %, and by many other causes. At the „Error“ state an error window is displayed,
through which you can view the errors. If the device returns within the specified values, „Error“
state will disappear and the logrecord will be carried out.
„Critical error“ - device outside the specified values with consequences on its functioning. This
state can be caused for instance by an undervoltage greater than % or by filling the memory
up to % so that it is not possible to make records. At the „Critical error“ state an error window
is displayed, through which you can view the errors. If the device returns within the specified
values, both „Critical error“ and the error window remain displayed until a confirmation
(acknowledgment) of the „Critical error“ is done. After confirmation the log record will be carried
out.
Date and time of each error is recorded in the log
!

18 | USER GUIDE OMR 700
.
SETTING
OF THE INSTRUMENT
User login or entry into the menu
The appearance of the bar without a logged-in user. Control buttons are disabled and therefore
you can not change the screen, enter the menu or view errors, logs and capacity utilization of
the memory media.
It is not possible to switch screens in the default setting.
User with access rights „User“ has the right of switching over screens, viewing errors, logs and
memory media.
User with access rights „Advanced user“ and beyond has, in addition, access to the menu.
Switching the previous or the next screen
Switching over to yesther screen. If you have defined only one screen, the screen remains
unchanged
Switching over to the previous screen. If you have defined only one screen, the screen remains
unchanged.
Fast screen selection

USER GUIDE OMR 700 | 19
14
8
.
9~C
-263mm
453mV .
SETTING
OF THE INSTRUMENT

20 | USER GUIDE OMR 700
.
SETTING
OF THE INSTRUMENT
5.3 FIRST START
When you turn on the device for the first time, the „Guide“ will help you with the basic settings so that the unit
is ready for further work.
20
| USER
GUIDE
OMR 700
Table of contents