Peraqua iQntrol DOS MODBUS User manual

___________________________________________________________________________________________________________
Subject to technical modifications
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Handelsstraße 8 Tel. +43 (0) 7435 58 488-0 www.peraqua.com
Measuring, control and dosing
technology
for private pools
iQntrol DOS MODBUS
As 01/2022 serial no. 84050
Measuring, control and dosing technology for pH correction and disinfectants

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 2/39
Table of contents
1About this manual......................................................................................................................................................................... 4
1.1 Scope of applicability ............................................................................................................................................................... 4
1.2 Target group............................................................................................................................................................................. 4
1.3 Storage of the manual.............................................................................................................................................................. 4
1.4 Further information ................................................................................................................................................................. 4
1.5 Symbols used ........................................................................................................................................................................... 4
2Safety ............................................................................................................................................................................................ 5
2.1 Intended use ............................................................................................................................................................................ 5
2.2 Safety notices........................................................................................................................................................................... 5
3Important facts about swimming pool water properties.............................................................................................................. 6
3.1 Auxiliary hygiene parameters................................................................................................................................................... 6
3.1.1 pH value .............................................................................................................................................................................. 6
3.1.2 Redox voltage...................................................................................................................................................................... 6
3.1.3 free chlorine........................................................................................................................................................................ 6
3.1.4 combined chlorine............................................................................................................................................................... 6
3.1.5 Cyanuric acide ..................................................................................................................................................................... 6
4Scope of delivery –device description.......................................................................................................................................... 7
4.1 Scope of delivery...................................................................................................................................................................... 7
4.2 Check for transport damage..................................................................................................................................................... 7
4.3 Identification of the device ...................................................................................................................................................... 7
4.4 Device description.................................................................................................................................................................... 7
5Installation .................................................................................................................................................................................... 8
5.1 Select installation location ....................................................................................................................................................... 8
5.2 Mounting the device on the wall ............................................................................................................................................. 8
5.3 Voltage supply.......................................................................................................................................................................... 8
5.4 Assembly notices...................................................................................................................................................................... 8
5.5 Provide external pre-filter........................................................................................................................................................ 8
5.6 Installation suggestion ............................................................................................................................................................. 9
6Commissioning - Notices............................................................................................................................................................. 10
6.1 Control parameters................................................................................................................................................................ 10
6.2 pH electrode........................................................................................................................................................................... 10
6.3 Cleaning beads ....................................................................................................................................................................... 10
6.4 Measuring cell flow................................................................................................................................................................ 10
6.5 Dosing cartridges.................................................................................................................................................................... 10
6.6 Chemicals ............................................................................................................................................................................... 10
6.7 Disinfectants........................................................................................................................................................................... 11
6.8 Swimming pool filter backwashing / vacuuming the floor ..................................................................................................... 11
7Electrical connection................................................................................................................................................................... 12
7.1 Open and close the casing...................................................................................................................................................... 12
7.2 Insert lines.............................................................................................................................................................................. 12
7.3 Overview of the connection diagram..................................................................................................................................... 13
7.3.1 The power pack NT_SCHW-8............................................................................................................................................. 13
7.3.2 The I/O board IO_SCHW-8................................................................................................................................................. 15
7.3.3 The measuring amplifier MV_CPRT_V1................................................................................................................................... 16
7.4 External functions .................................................................................................................................................................. 16
7.4.1 Central control technology off (external OFF)................................................................................................................... 16
7.4.2 Clean water external ......................................................................................................................................................... 16
7.4.3 Temperature control......................................................................................................................................................... 16
8Operating the touch screen ........................................................................................................................................................ 17
8.1 Operating programmes.......................................................................................................................................................... 18
8.1.1 Automatic (auto mode)..................................................................................................................................................... 18
8.1.2 Quick Info .......................................................................................................................................................................... 18
8.1.3 Dosing delay ...................................................................................................................................................................... 18
8.1.4 Start routine ...................................................................................................................................................................... 18

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 3/39
8.1.5 Alarm .................................................................................................................................................................................19
8.2 Main menu..............................................................................................................................................................................19
8.2.1 Main menu Settings ......................................................................................................................................................19
8.2.1.1 Main menu Settings pH .......................................................................................................................................20
8.2.1.2 Main menu Settings Proportional disinfection, 2-point (chlorine, bromine).......................................................20
8.2.1.3 Main menu Settings Disinfection O2 (active oxygen) ..........................................................................................21
8.2.1.4 Main menu Settings Dosing performance ...........................................................................................................23
8.2.1.5 Main menu Settings Temperature.......................................................................................................................24
8.2.1.6 Main menu Settings Dosing delay .......................................................................................................................24
8.2.1.7 Main menu Settings System ................................................................................................................................24
8.2.1.8 Main menu Settings System Password End user –Password level 1..............................................................25
8.2.1.9 Main menu Settings System Password Technician 1 –Password level 2........................................................25
8.2.1.10 Main menu Settings System Network.............................................................................................................25
8.2.2 Main menu Service........................................................................................................................................................26
8.2.2.1 Main menu Service Input test..............................................................................................................................26
8.2.2.2 Main menu Service Output test...........................................................................................................................26
8.2.2.3 Main menu Service Info.......................................................................................................................................26
8.2.2.4 Main menu Service Manual dosing O2................................................................................................................26
8.2.3 Main menu Login ..........................................................................................................................................................27
8.2.4 Main menu Log .............................................................................................................................................................27
8.2.4.1 Main menu Log Event Log ...................................................................................................................................27
8.2.4.2 Main menu Log Event Log Justage.......................................................................................................................28
8.2.4.3 Main menu Log Data Log.....................................................................................................................................28
8.2.4.4 Hauptmenü Log Export........................................................................................................................................28
8.2.5 Main menu Calibration..................................................................................................................................................29
8.2.5.1 Main menu Calibration pH 2-point (pH electrode)..............................................................................................29
8.2.5.2 Main Menu Calibration pH - phenol red (tablet measurement)..........................................................................29
8.2.5.3 Main menu Calibration Redox electrode.............................................................................................................30
9Network connection - Visualisation.............................................................................................................................................31
9.1 Network installation ...............................................................................................................................................................31
9.1.1 Insert the network line into the control housing ...............................................................................................................31
9.1.2 Localise the LAN interface..................................................................................................................................................31
9.2 Establish network access at the device...................................................................................................................................32
9.2.1 Main menu Settings System Network .................................................................................................................32
9.2.2 IP-address ..........................................................................................................................................................................32
9.2.3 Gateway.............................................................................................................................................................................32
9.2.4 Subnetmask .......................................................................................................................................................................32
10 Maintenance and cleaning ..........................................................................................................................................................33
10.1 Fine filter............................................................................................................................................................................33
10.2 pH electrode ......................................................................................................................................................................33
10.3 Dosing valves .....................................................................................................................................................................33
10.4 Flow control valve..............................................................................................................................................................33
10.5 Dosing cartridges ...............................................................................................................................................................34
10.6 Redox electrode.................................................................................................................................................................34
11 Decommissioning - Overwintering - Storage...............................................................................................................................34
12 Technical data..............................................................................................................................................................................34
13 Change history - IQntrol DOS MODBUS.......................................................................................................................................35
14 Commissioning protocol..............................................................................................................................................................36
15 Spare part list...............................................................................................................................................................................38
16 Personal notes.............................................................................................................................................................................39

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 4/39
1About this manual
1.1 Scope of applicability
This manual describes the installation, commissioning and operation of the device.
The manual covers the POOLKLAR Touch XL starting with version -> see footer and cover sheet.
1.2 Target group
Only persons who have received proper instructions regarding the device functions may operate the device. Electrical and water-
side connection work may only be conducted by appropriately trained specialists.
1.3 Storage of the manual
All manuals for the device as well as those for the installed components must be stored in the immediate vicinity of the device and
be accessible to the operating personnel at all times.
1.4 Further information
Further information about special topics, e.g., design of the dosing performance or description of the operating parameters, is
available from your specialist dealer.
1.5 Symbols used
This document uses the following types of safety notices as well as general notices:
DANGER!
"DANGER" denotes a safety notice whose non-observance will directly result in
death or serious injury!
WARNING!
"WARNING" denotes a safety notice whose non-observance may result in death or serious injury!
CAUTION!
"CAUTION" denotes a safety notice whose non-observance may result in minor or moderate injury!
ATTENTION!
"ATTENTION" denotes a safety notice whose non-observance may result in property damages!
Notice
A notice denotes information whose non-observance may result in operational disruptions.
Tip
A tip denotes information that may result in improvements in the operating process.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 5/39
2Safety
2.1 Intended use
The IQntrol DOS MODBUS device is designed for carrying out measuring and control tasks when treating swimming pool water in
private swimming pools.
2.2 Safety notices
The operation manual must be consulted prior to installation, commissioning and maintenance work. Following the
commissioning, the manual must be made available to the operator. In your own interest, please observe the safety notices
contained in this manual.
Always handle chemicals with special care!
Please contact your chemicals supplier for information about the safe handling of chemicals.
WARNING!
There is a risk of chemical burns and poisoning!
Observe the relevant regulations when handling chemicals!
-Never mix different chemicals with each other.
-Only use chemicals that have been approved for the treatment of swimming pool water
-Wear suitable protective clothing when conducting maintenance work.
Further information about the chemicals used is available from your chemicals supplier.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 6/39
3Important facts about swimming pool water properties
3.1 Auxiliary hygiene parameters
The following information is meant to offer a preliminary overview about the most important auxiliary hygiene parameters in the
treatment of swimming pool water. Further information is available from your swimming pool dealer, the chemicals supplier, or
the book trade. The information provided here refers to DIN 19643 for swimming pool water in public pools.
Devices that are installed outside of Germany should comply wit the county-specific regulations and statutes, e.g., ÖNORM, etc.
3.1.1 pH value
The pH value has a decisive effect on the chemicals added to the swimming pool cycle. If the pH value is too high, it negatively
affects the germ-killing speed of disinfectants containing chlorine. The likelihood of lime precipitation and water clouding
increases. If the pH value drops too low for an extended time, there is a risk that the treatment system and the basin may be
damaged.
According to the DIN, a pH value around pH7 should be aimed for.
Values below pH6.5 and above pH7.6 should be avoided.
3.1.2 Redox voltage
The redox voltage indicates the ratio between reducing substances (organic contaminations) and oxidising power (active chlorine
content) in the pool water. The redox voltage serves as a measure for assessing the hygienic basin water quality. The higher the
redox voltage, the faster germs and bacteria will be killed. To achieve a sufficiently fast germicidal effect, the redox voltage in
private pools should also comply with the values specified by the DIN. In this context, the amount of free chlorine measured with
the DPD1 method should be between 0.3 and 0.6 mg/l.
The chlorine's disinfecting power essentially depends on the basin water’s pH value. The higher the pH value, the lower the
disinfecting power; if the pH value decreases, the disinfecting power increases at a constant chlorine level.
If the redox voltage is far below 700 mV at a free chlorine content between 0.3 and 0.6 mg/l, the complete function of the
swimming pool water treatment must be checked.
According to the standard a redox voltage around 750 mV should be aimed for.
Values below 650 and above 800 mV should be avoided.
3.1.3 free chlorine
The chlorine available for disinfection in the pool water is called free chlorine. Free chlorine is detected by means of the DPD1
method.
According to the DIN, the free chlorine content should be kept between 0.3 and 1.5 mg/l.
Values below 0.2 and above 1.2 mg/l should be avoided.
3.1.4 combined chlorine
Combined chlorine is the product of a reaction between organic contamination introduced in the pool and a chlorine-containing
disinfectant. Combined chlorine causes the characteristic indoor pool smell and may lead to skin irritation and red eyes.
Ideally, the organic chlorine compounds will be retained in the swimming pool filter system and flushed out via the regular
backwashing. If the filter system does not work correctly, or if the backwashing is not conducted in a proper fashion, combined
chlorine may accumulate in the swimming pool cycle. If the water starts smelling of chlorine, it is high time to conduct a chlorine
measurement by means of the DPD3 measuring.
The concentration of combined chlorine results from the difference between the overall chlorine content and the amount of free
chlorine in the water. Observe the instructions of the analytic device's manufacturer when determining the concentration.
Combined chlorine can only be reduced by dilution, i.e., with vigorous backwashing or by a shock chlorination, if applicable.
Consult your swimming pool dealer.
According to the DIN, a combined chlorine content below 0.2 mg/l should be aimed for.
“Chlorine odours” can result above a combined chlorine content of approx. 0.3 to 0.4 mg/l. (characteristic indoor pool smell)
3.1.5 Cyanuric acid
Cyanuric acid is found in organic chlorine products (stabilized chlorine). If disinfection is carried out with organic chlorine, cyanuric
acid will accumulate. Cyanuric acid can only be reduced by dilution, i.e. with vigorous backwashing. Ask your swimming pool
retailer.
According to the standard, a maximum concentration of 100 mg/l must be observed.
Tip
Since the pH value is an elementary factor in the swimming pool treatment, the dosing of chlorine-containing
disinfectants is only released once the pH value has entered the control range!

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 7/39
4Scope of delivery –device description
4.1 Scope of delivery
The IQntrol DOS MODBUS is delivered with the following standard accessories.
- Buffer solutions pH7, pH4, redox test solution
- Electrode cleaner, distilled water
- Glass beads, electrolyte solution, replacement gaskets
- 2 pc. ½“ measuring water ball valve with immersion pipe
- 7 m each measuring water pipe 6x1 mm in PE and PTFE, respectively
Customer-specific or order-related modifications are possible.
4.2 Check for transport damage
Please check the device and all accessories immediately upon receipt for transport damage and completeness.
4.3 Identification of the device
For spare part orders and troubleshooting, it is useful to know the device serial number and software version. The device serial
number is located on the identification plate on the right side of the control housing. The programme version can be called up via
the menu item Service
Info.
4.4 Device description
1
2
2
5
7
4
6
8
9
13
3
10
11
12
1. Control IQntrol DOS MODBUS
2. Dosing pumps Standard SR10 (optional Sa)
3. Buffer solutions
4. Redox test cylinder
5. pH electrode
6. Measuring cell illumination
7. Dosing valves
8. Flow control valve
9. Flow monitoring
10. Redox electrode
11. Test water tap
12. Measuring water inlet
13. Measuring water return flow
14. Suction set (not visible in image)
The IQntrol DOS MODBUS is delivered as a ready-assembled unit. All parts are mounted on a plastic plate. This ensures a quick
installation of the device. It also goes toward preventing potential execution errors on the part of the installation personnel, as far
as possible.
For the transport, the factory merely removes the pH electrode from the measuring cell and delivers it in a separate protective
box. In order to avoid deformation of the dosing hoses during extended storage, the two dosing cartridges have been pulled off
the motor shaft.
The IQntrol DOS MODBUS is equipped with a touch-sensitive touch screen. By touching a symbol or a numeric value, this will be
activated for parametrisation. The adjustment menus come with additional text-based instructions.
In order to eliminate the risk of confusing both chemicals as far as possible, the dosing technology is colour-coded throughout. The
colour-coding runs from the suction set over the associated dosing pump to the dosing valve.
The parts used for pH value control are located to the left and coded in red. The components used for disinfection are located on
the right and coded in yellow.
CAUTION!
If the two chemicals are switched, this leads to a malfunction of the addition of chemicals to the pool water! This
may result in significant overdosing!

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 8/39
5Installation
5.1 Select installation location
A freely accessible installation location should be selected to facilitate operation and subsequent maintenance tasks. The
installation location must be protected from frost and the device may not be exposed to direct sunlight.
5.2 Mounting the device on the wall
Select the installation height so that the device is located at eye level. Measure the four drill holes on the mounting plate and mark
their location on the wall. Use the included four black plastic spacers to ensure the required distance from the wall.
5.3 Voltage supply
The devices must be supplied with continuous voltage. I.e., the voltage supply must not be locked with the filter system. For
locking with the filter system, please use the external off input.
The permanent voltage supply is required for the proper execution of the disinfection dosing by means of active oxygen. This also
prevents condensation corrosion.
5.4 Assembly notices
Only "good" basin hydraulics can achieve a satisfactory control of the two auxiliary hygiene parameters, pH value and redox
voltage. The dosed chemicals must reach all areas of the basin within a short time. Even under stress, the concentrations
measured in different areas must be approximately the same. The measuring water should be as close to identical to the basin
water as possible and must reach the measuring cell as quickly as possible. This is the only way to quickly detect and rebalance
changes in the water quality.
If the basin water is conducted through an overflow gutter and a compensation basin (splash water), which is potentially also used
for the freshwater feed-in, only the sampling of the measuring water directly from the basin will lead to satisfactory measuring and
control results.
At very low temperatures <5°C, the touch screen may become impaired due to the stiffness of the frontal film!
5.5 Provide external pre-filter
It is not possible for the pool operator to detect a contaminated pre-filter below the new design cover. For this reason, an external
pre-filter must be integrated in the measuring water inlet. Depending on the local conditions, the pre-filter can be installed directly
at the device or at any other installation site along the measuring water inlet. This has the advantage of making it possible to select
a more convenient access point for the pool operator.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 9/39
5.6 Installation suggestion
The following sketch shows an example of the integration of the IQntrol DOS MODBUS in the swimming pool’s water treatment
cycle. In most cases, a measuring water withdrawal between the filter pump and the filter boiler is sufficient. The water is
recirculated into the clean water line downstream of the heat exchanger. Two ball valves ½“ with immersion pipe for the
connection to the filter piping are provided with the accessories. The immersion pipe of the two ball valves should be shortened so
that it ends approximately in the middle of the pipe.
ATTENTION!
Since both chemicals are mixed in with the measuring water, the measuring water recirculation must occur in the
direction of the flow downstream of the heat exchanger in order to prevent corrosion.
If the pool is being operated with an overflow gutter and a splash water tank, into which freshwater is also allowed
to enter, marked differences in the water quality on the measuring cell and in the basin must be expected. In order
to ensure perfect control quality in this case, it is recommended that the measuring water is withdrawn directly
from the basin, using a separate measuring water pump.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 10/39
6Commissioning - Notices
Before the device can be commissioned, the following measures must be implemented. You can find the position of the parts
listed in the following under Item 4.4 Device description.
6.1 Control parameters
The device is loaded with factory-defined control parameters; see Commissioning protocol on page 25/29. Please adjust the
control parameters for your basin according to the required dosing performance and the desired set points. The factory setting for
the dosing performance is 100%, which should be sufficient for an outdoor pool of approx. 100 m³.
Tip
Parameter changes are saved to the SD card with a time delay. This can take up to two minutes. If the voltage supply
is interrupted prior to the backup on the SD card, the device continues to control using the values prior to the change.
That means, if a voltage interruption occurred shortly after a parameter change, the changed parameters must be
checked again and be readjusted, if necessary!
The newly determined adjustment values are also saved to the SD card with a time delay.
6.2 pH electrode
Remove the pH electrode from the packing box and pull off the protective cap. Remove the electrode cable’s union nut from the
electrode by twisting it to the left and screw the electrode into the measuring cell from above. Subsequently, the electrode plug
must be placed back onto the pH electrode and the union nut must be tightened.
Tip
Following a temperature adjustment and inflow time of approx. one hour, it should be checked whether the pH
electrode needs to be calibrated. For this purpose, the pH electrode must be unscrewed from the measuring cell
once again and be placed in the pH7 buffer solution. If the display value shows a deviation larger than 0.1pH, a
calibration must be conducted.
6.3 Cleaning beads
The enclosed blue glass beads serve for cleaning the redox electrode's platinum surface. Pull the cable from the redox electrode
and unscrew the electrode from the measuring cell. Place approx. 15 of the enclosed blue cleaning beads on the platinum surface
so that this area is covered. Screw the redox electrode back into the measuring cell and plug the cable's black plug back into the
electrode.
When screwing in the redox electrode, ensure that no glass beads are present in the thread.
ATTENTION!
Please ensure that all hose screw connections of the measuring water pipes are firmly tightened. Check all screw
connections at the device and at the two measuring water ball valves ½“. Plastic screw connections may only be
fastened hand-tight!
6.4 Measuring cell flow
Once all electrodes have been screwed hand-tight into the measuring cell and the measuring water pipes have been connected,
the ball valves may be opened. The measuring cell flow volume can be adjusted at the flow control valve. The flow mounting float
must be pushed toward the top in order to release the dosing. In order to keep the redox electrode clean, the cleaning beads must
rotate on the platinum surface, but they must not be allowed to lift off of the platinum surface.
6.5 Dosing cartridges
Push the two dosing cartridges onto the motor shaft.
6.6 Chemicals
Place the two chemical containers under the device. In order to protect the floor, it is recommended to place the containers in a
protective tray. Please observe the proper positioning of the container; oredopH value regulation chemical on the left and
oyellowodisinfection chemical on the right.
ATTENTION!
Please do not use hydrochloric acid for the pH value regulation. Hydrochloric acid diffuses through the dosing hose
and corrodes the dosing motor. This, in turn, causes premature wear and tear of the dosing pump.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 11/39
6.7 Disinfectants
Notice
Please find out which chemical was used to disinfect the basin water prior to the commissioning.
If a different disinfectant is used from now on, the following must be observed.
If other disinfectants (e.g., organic chlorine (dichlorine) or “chlorine-free” disinfection chemicals) were used previously, a reduced
redox voltage is displayed during commissioning. Without changes to the control parameters, this would result in significant
overdosing. The breakdown of the disruptive chemicals can take several weeks.
That means, if other disinfection chemicals were used prior to the installation of the POOLKLAR for the dosing of inorganic
chlorine, we recommend to run the device with removed dosing cartridges for about 0.5-1 hour to be able to see the redox status.
The set point of the redox voltage should then be set at approx. 20-50 mV above the displayed value. The development of the
chlorine concentration and the redox voltage must be observed during the subsequent period, and the redox voltage's set point
must be increased so that the measured chlorine concentration approximates 0.5 mg/l.
In order to avoid this uncertainty, the basin water must be exchanged completely.
6.8 Swimming pool filter backwashing / vacuuming the floor
During a backwashing of the swimming pool filter, the dosing must be switched off. In case of electric locking with the filter pump,
this is done by switching off the filter pump. This is followed by the dosing delay, which prevents the dosing during this time
period. However, if the backwashing procedure takes longer than the set dosing delay, the measuring water intake to the device
must be blocked.
If the basin floor is vacuumed via the filter system, the measuring water intake must also be interrupted during this time, due to
the increased contamination load.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 12/39
7Electrical connection
7.1 Open and close the casing
Picture of slotted screw head
Depending on the type of device, the display lid can be swivelled to
the left or right for installation and maintenance work. The locking
axle must be removed for swivelling. The locking axle is identified
by the plastic slotted screws on both sides. The other side is
equipped with two expanding rivets as pivot bearings.
Tip
It is not necessary to completely remove the display lid for maintenance tasks! Only the locking axle must be removed.
Afterwards, the display can be swivelled to the side.
7.2 Insert lines
The casing comes with several factory-made free screw
connections. Several push-outs for metric cable glands with a jam
nut are available for additional insertions.
The two external screw connections with M25 are intended for the
insertion of a preassembled interface cable with RJ45 plug.
You can run the additional lines introduced by you as well as the mains supply line downward behind the measuring cell.
Alternatively, these lines can also be run laterally below the control housing to the left or right. Push-outs in the protective cover
are provided for this purpose.
Tip
The design cover is equipped with one push-out of approx. 30 x 30 cm each on the left and the right, about 40 cm from
the upper edge. This makes it possible to also run off the mains supply and other lines laterally.
ATTENTION!
Please pay attention to the spatial separation between energy and signal lines when inserting additional lines. The
crossing of energy and signal lines must be avoided!
ATTENTION!
Upon completion of the work, the casing must be properly closed again!

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 13/39
7.3 Overview of the connection diagram
DANGER!
Risk of death due to high voltage.
All electrical work on the device may only be conducted by properly trained specialists under
compliance with the applicable safety regulations!
7.3.1 The power pack NT_SCHW-8
Fuse
Current
Function
F1
315 mA
Without function in this device series
F2
2 A
Primary fuse
F3
1 A
Without function in this device series
F4
4 A
230 VAC outlet relay K2 and K3
F5
1 A
Without function in this device series
Plug
Function
SL1
Power inlet 230 VAC
SL2
Relay - K1 –temperature 230 VAC
SL3
Relay - K2 - disinfection 230 VAC
SL4
Relay - K3 –pH regulation 230 VAC
The picture shows an example of the connection of two external 230 volt dosing pumps (dosing devices).

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 14/39
Fuses
ATTENTION!
Fuses are safety-relevant components!
Only fuses in the specified current range may be used.
Non-compliance poses the risk of fire or destruction of the circuit boards!
Fuse in the control unit housing (accessible from the outside)
Fuse
Current
Fuse type
Function
6.3 A
5 x 20 mm
Primary fuse mains supply
Fuses on the power pack
Fuse
Current
Fuse type
Function
F1
315 mA
Micro fuse
24 VDC sensors (without function in the IQntrol DOS MODBUS)
F2
2 A
Micro fuse
Primary fuse power pack
F3
1 A
Micro fuse
without function in the IQntrol DOS MODBUS
F4
4 A
Micro fuse
230 VDC outlet relay K2 and K3
F5
1 A
Micro fuse
without function in the IQntrol DOS MODBUS
Relay - outlets
The relay outputs K1 to K3 (SL2 to SL4) are non-potential-free 230 VAC. If a potential-free outlet is needed, an external relay must
be used.
ATTENTION!
The maximum load per outlet is a maximum of 450 watts ohmic load. If a higher load is to be switched, this must
be realised via an additional load relay.
The total output of all devices activated simultaneously may not exceed 900 watts (4 ampere)!

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 15/39
7.3.2 The I/O board IO_SCHW-8
Fuses on the I/O board
Fuse
Current
Fuse type
Function
F1
315 mA
Micro fuse
Outlet disinfection 24 VDC
F2
315 mA
Micro fuse
Outlet pH regulation 24 VDC

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 16/39
7.3.3 The measuring amplifier MV_CPRT_V1
7.4 External functions
7.4.1 Central control technology off (external OFF)
The input Central control technology off (external OFF) is used for the controlled deactivation of the IQntrol DOS MODBUS via the
swimming pool filter system’s central control. As long as the contact is open, there is no dosing, no heating of the basin water, no
alarm message given.
7.4.2 Clean water external
A flow monitor with a potential-free contact can be connected to the input Clean water external. As long as the contact is open,
there is no dosing, no heating of the basin water.
7.4.3 Temperature control
The device is equipped with a temperature control. This temperature control can be used to regulate the basin water
temperature. If the basin is heated by means of a flow-through heat exchanger, it must be ensured that the associated heating
circuit pump may only be activated when the filter system is in operation!
Please remember that the regulation of the basin temperature can only be guaranteed if the filter operation times are set to an
appropriate duration.
In order to achieve a more precise measuring result, the sensor must be installed by means of an immersion sleeve in the
swimming pool's piping upstream of the filter.
ATTENTION!
Immersion sleeves that come into contact with the swimming pool water must be corrosion-proof.
ATTENTION!
Please observe the safety notices regarding the topic of temperature control!

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 17/39
8Operating the touch screen
The device is fitted with a touch-sensitive touch screen. Desired parameter changes, adjustments and tests can be done simply by
lightly touching the corresponding symbol or the numeric value.
Quick Info
Red Measuring value
Alarm low
Alarm high
current
measuring values
Start
Bar display
Dos on/ off
IN list
Alarm off
OUT list
Menu
Status line
Date/ time
Symbols used
IN list (inputs)
Chemical container empty
red = pH regulation, yellow = disinfection, green = active oxygen, blue = flocculation
Hose rupture –a chemical leak has occurred at one of the three dosing pumps
Drop colour red = pH regulation, yellow = disinfection, blue = flocculation
No measuring cell flow, or flow too low
No flow in the clean water line to the pool, or flow too low (OPTIONAL if connected)
The dynamic dosing time monitoring has been exceeded.
Colour red = pH regulation, yellow = disinfection
The IQntrol DOS MODBUS has been deactivated by an external command from the swimming pool water treatment
(filter system). There is no dosing, no heating of the basin water, no alarm message given.
OUT list (outputs)
red = pH output active
yellow or green = disinfection output active
blue = flocculation released
Temperature output active
Alarm relay output active
Faulty O2dosing. The last dosing could not be completed correctly. The symbol automatically goes out after the next
successful dosing.
on the right side
DOS on/off for the manual temporary deactivation of the dosing (pH regulation and disinfection)
Manually deactivate the alarm relay (it then remains deactivated until the next alarm)
Cancel the dosing delay
into the main menu

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 18/39
8.1 Operating programmes
Depending on the operation state, the display shows different views. The following describes the main display views.
8.1.1 Automatic (auto mode)
The device is in the status Automatic mode. The auxiliary hygiene parameters and
the temperature (optional) are regulated based on set parameters.
There is no disruption.
The picture below shows the time control variant
In the example, dosing has been selected on four days.
During the dosing of the disinfectant (green pump symbol in the OUT list), the ml
display is reduced by 50 ml per minute.
8.1.2 Quick Info
You can view the set control parameters quickly (without a password prompt) via
the Quick Info function.
Quick Info only serves for the quick viewing of the control parameters e.g., by the
end user; no changes can be made.
8.1.3 Dosing delay
If the device is restarted, the dosing delay is running, during which time no dosing
occurs.
Software alarms are suppressed during this time.
The dosing delay must be set high enough to ensure that after the start of the
filter system actual basin water flows through the measuring cell.
The dosing delay can be cancelled with and the device is set to normal
operation.
8.1.4 Start routine
If one of the two measuring values, pH or mV, falls outside the control range, the
start routine is activated.
During the start routine, the dynamic dosing time monitoring is active. If the
corresponding measuring value does not reach the control range within the
selected time, e.g., due to a malfunction, the dosing stops with an alarm.
The dosing time monitoring alarm is signalled by . This malfunction can only
be reset through a device restart or a manual acknowledgement.

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 19/39
8.1.5 Alarm
An alarm occurs; this is signalled by the symbol in the OUT list. The alarm
relay is activated.
A differentiation is made between alarms (software alarms, e.g., alarm high, alarm
low) and disruptions (switch inputs).
In the event of alarms, the respective measuring value is illustrated in red.
In the event of disruptions, the corresponding symbol appears in the IN list.
Alarms or disruptions must be pending for approx. 5 seconds before an alarm is
triggered.
Tip
An alarm will be automatically deleted when its cause has been remedied. This is the case, e.g., if the empty chemical
container was replaced or if measuring water resumes its flow through the measuring cell.
However, the alarm Switch-off time monitoring must be acknowledged manually!
8.2 Main menu
Automatic
Returns to the automatic operation
Settings
For adjusting the control parameters and system settings
Service
Input and output test, system information
Login
Password entry for authorisation
The password level is displayed in the status line to the left of the date
Log
Event and calibration log, logging of measuring values, data export
Calibration
For calibration of the pH and redox electrode
8.2.1 Main menu Settings
The picture below shows the time control variant
pH
Set pH regulation control parameter
Disinfection
Set disinfection control parameter
Dosing performance
Adjust the dosing performance to the basin size
Temperature
Set basin temperature control parameters
Dosing delay
Set the dosing delay
System
Date/ time, password, display, network, language, device ID
Flocculation
Activate the release for flocculation dosing

iQntrol DOS-MODBUS
Original manual in German Version 1.0.2 2022.10.06 20/39
8.2.1.1 Main menu Settings pH
Alarm low lower alarm value
Set point the device attempts to reach this value
Control range The dosing pumps work proportionally, i.e., the higher the
difference between set point and actual value, the longer the dosing time (max.
50 seconds).
The smaller the selected control range, the faster the measuring value will react,
which can easily lead to overdosing.
Alarm high upper alarm value
Time control if the control range is not reached within three times the set time,
the dosing is blocked.
Important: This disruption must be acknowledged manually!
With 2-point (on/off) control, the P range becomes hysteresis. If the hysteresis is
not reached, the output is activated until the set point is reached. Control
direction pH decrease or pH increase control direction selectable The selected
control direction is displayed in the header.
Tip
To protect against overdosing with disinfectants that contain chlorine, the dosing is only released once the pH value
has entered the control range (pH priority dosing). For this reason, the selected control range must not be too high.
8.2.1.2 Main menu Settings Proportional disinfection, 2-point (chlorine, bromine)
The IQntrol DOS MODBUS control is suitable for the dosing of different disinfectants. The disinfection outlet must be adjusted to the
disinfectant used. According to the selected variant, the menu for setting the associated parameters changes.
Alarm low lower alarm value
Control range The dosing pumps work proportionally, i.e., the higher the
difference between set point and actual value, the longer the dosing time (max.
50 seconds).
The smaller the selected control range, the faster the measuring value will react,
which can easily lead to overdosing.
Set point the device attempts to reach this value
Alarm high upper alarm value
Time control if the control range is not reached within three times the set time,
the dosing is blocked.
Important: This disruption must be acknowledged manually!
In the 2-point control (on/off), the control range becomes hysteresis. If the value
stays below the control range, the output is activated until the set point has been
reached.
Tip
To keep the disinfectant concentration at an even level, the factory-set control range must be reduced when using
the 2-point function. In this case, a value between 10 and 20 mV should be selected.
If the selected hysteresis is too high, this can lead to a significant decrease in disinfecting power. If the hysteresis it
too low, this may lead to a larger number of switching cycles.
Please observe the manufacturer's specifications for the connected dosing device.
This manual suits for next models
1
Table of contents