Operating Manual
PT-Node
Network Adapter for Printing and
Data Transfer
Version 1.0

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 2 of 64
The Documentation
This document describes the set-up, operation and general maintenance of the Pharma
Test instrument. It should be used by the operators and the technical support staff
responsible for the installation and set-up of equipment.
All attached equipment and parts have to be used in compliance with the manufacturer’s
manuals and papers supplied.
Copyrights
© Pharma Test Apparatebau AG 2023
We reserve all rights.
This manual should be used by the owner of the instrument only. He is allowed to copy the
manual for his own use. It is forbidden to supply any copy of this document for any other
purpose other than the instrument use without previous approval from Pharma Test
Apparatebau AG.
How to Use the Manual
To understand the different information we use different formatting:
▪< >Use any key (i.e. <1> OR <STOP>)
▪“ “ Display information
▪
Information entries
▪“ “ Select from a menu
▪Note: informs about special use OR possibility

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 3 of 64
Table of Contents
Table of Contents ...................................................................................................................3
List of Figures ........................................................................................................................4
List of Tables ..........................................................................................................................6
Document History ...............................................................................................................7
Section 1 General Information .........................................................................................8
Section 1.1 Technical Specifications ...............................................................................9
Section 1.2 Standard Supply Scope.................................................................................9
Section 1.3 Available Connecting Cables ........................................................................9
Section 2 Instrument Overview ......................................................................................10
Section 2.1 Power Supply..............................................................................................10
Section 2.2 Serial Connections Channel 1 and 2...........................................................10
Section 2.3 Ethernet Connection...................................................................................10
Section 3 Setting-Up the Instrument .............................................................................11
Section 3.1 Initial Set Up of the Network Configuration ...............................................11
Section 3.1.1 Connecting a PT-Node directly to a PC.................................................14
Section 4 Using the Web Interface.................................................................................15
Section 4.1 File Index without Login..............................................................................17
Section 4.2 Display Reports and Printing......................................................................19
Section 4.3 Login as an Operator ..................................................................................20
Section 4.3.1 Downloading Folders and Files.............................................................21
Section 4.3.2 Deleting Folders and Files: ...................................................................23
Section 4.4 Login as Administrator ...............................................................................25
Section 4.4.1 Device Manager.....................................................................................26
Section 4.4.2 Network Settings...................................................................................27
Section 4.4.3 Settings..................................................................................................29
Section 5 Connecting a Pharma Test Instrument to PT-Node ......................................35
Section 5.1 PTB 111E and PTB 111EP...........................................................................35
Section 5.1.1 Capturing Serial Result Data (“FreeWeigh”).........................................35
Section 5.1.2 Capturing PTB 111EP Internal Printer Test Reports ............................38
Section 5.1.3 Capturing PTB 111E Ticket Printer Test Reports .................................38
Section 5.2 PTB 311E and PTB 511E (Firmware Version 1.xx)......................................41
Section 5.2.1 Capturing Serial Result Data (“FreeWeigh”).........................................41
Section 5.2.2 Capturing PTB 311E and PTB 511E Ticket Printer Test Reports..........42
Section 5.3 PTB 311E (Firmware Version 4.xx) .............................................................45

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 4 of 64
Section 5.3.1 Capturing Serial Result Data (“FreeWeigh”).........................................45
Section 5.4 PT-TD300 ....................................................................................................50
Section 5.4.1 Capturing Test Reports .........................................................................50
Section 5.5 PTZ 100, PTZ 300 ........................................................................................52
Section 5.5.1 Capturing Test Reports .........................................................................52
Section 5.6 PTF 100, PTF 200, PTF 300, PTF 600 ..........................................................54
Section 5.6.1 Capturing Test Reports .........................................................................54
Section 5.7 PTWS 120D, PTWS 120S, PTWS 820D, PTWS 620, PTWS 1220, PTWS D620
and PTWS 1420 instruments.............................................................................................56
Section 5.7.1 Capturing Run Time and OQ Reports ....................................................56
Section 6 Connecting to PT-Node Through an FTP Connection to Transfer Date to an
External Software.................................................................................................................59
Section 7 Frequently Asked Questions ..........................................................................61
Section 8 Troubleshooting .............................................................................................63
Section 9 Cleaning and Maintenance.............................................................................64
Section 10 Safety Information..........................................................................................64
List of Figures
Figure 1: PT-Node unit...........................................................................................................8
Figure 2 PT-Node Schematic ...............................................................................................10
Figure 3 Hotspot SSID is displayed on screen .....................................................................11
Figure 4: Connecting to the hotspot.....................................................................................12
Figure 5: Network connection manager - Ethernet.............................................................13
Figure 6: Network connection manager –Wi-Fi..................................................................14
Figure 7: Unique name of PT-Node is displayed on screen.................................................15
Figure 8: Accessing the web interface using PT-Node’s name ...........................................15
Figure 9: Accessing the web interface using PT-Node’s IP address...................................16
Figure 10 IP address is displayed on screen .......................................................................16
Figure 11: File index.............................................................................................................17
Figure 12: Empty folders vs. folders with content ...............................................................18
Figure 13: List of files...........................................................................................................18
Figure 14: Example of a test report displayed using a web browser...................................19
Figure 15: Login as user.......................................................................................................20
Figure 16: Logout button......................................................................................................20
Figure 17: Downloading folders ...........................................................................................21
Figure 18: Marking all list items ..........................................................................................21

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 5 of 64
Figure 19: Downloading files................................................................................................22
Figure 20: Deleting folders...................................................................................................23
Figure 21: Error message: trying to delete a folder containing files...................................24
Figure 22: Deleting files .......................................................................................................24
Figure 23: Login as administrator........................................................................................25
Figure 24: Device manager ..................................................................................................26
Figure 25: Network settings - Ethernet ...............................................................................27
Figure 26: Network settings - Wi-Fi.....................................................................................27
Figure 27: PT-Node network settings for direct connection ...............................................28
Figure 28: Windows network adapter settings for direct connection..................................28
Figure 29: User settings.......................................................................................................29
Figure 30: Date & time settings ...........................................................................................30
Figure 31: Date & time settings - no NTP server.................................................................30
Figure 32: Upload menu.......................................................................................................31
Figure 33: Uploading a new firmware version .....................................................................31
Figure 34: Firmware filename .............................................................................................32
Figure 35: Firmware update in process...............................................................................32
Figure 36: Firmware update completed...............................................................................32
Figure 37: Uploading a new configuration file .....................................................................33
Figure 38: Configuration file filename .................................................................................33
Figure 39: Updating the configuration file in process..........................................................33
Figure 40: Device manager ..................................................................................................34
Figure 41: Communication cable connection to PTB 111EP................................................36
Figure 42: Example of a serial result of a PTB 111EP instrument ......................................37
Figure 43: Communication cable connection to PTB 111E ..................................................39
Figure 44: Example of a PTB 111E test report.....................................................................40
Figure 45: Example of a serial result of a PTB 311E/PTB 511E instrument........................41
Figure 46: Communication cable connection to PTB 311E/PTB 511E.................................42
Figure 47: Communication cable connection to PTB 311E and PTB 511E...........................43
Figure 48: Example of a PTB 311E test report.....................................................................44
Figure 49: Communication cable connection to PTB 311E/PTB 511E.................................46
Figure 50: Example of a serial result of a PTB 311E instrument ........................................47
Figure 51: Communication cable connection to PTB 311E and PTB 511E...........................48
Figure 52: Example of a PTB 311E test report.....................................................................49
Figure 53: Communication cable connection to PT-TD300 .................................................50

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 6 of 64
Figure 54: PT-Node device settings for PT-TD300 ..............................................................50
Figure 55: Example of a PT-TD300 test report ....................................................................51
Figure 56: Communication cable connection to PTZ x00.....................................................52
Figure 57: PT-Node device settings for PTZ x00..................................................................53
Figure 58: Example of a PTZ 300 test report.......................................................................53
Figure 59: Communication cable connection to PTF x00.....................................................54
Figure 60: PT-Node device settings for PTF x00..................................................................55
Figure 61: Example of a PTF 100 test report .......................................................................55
Figure 62: Communication cable connection to PTWS 1420 ...............................................56
Figure 63: Printer port setting for PT-Node ........................................................................57
Figure 64: PT-Node device settings for PTWS x20 ..............................................................57
Figure 65: Example of a run time report of a PTWS 1420 instrument.................................58
Figure 66: Example of an OQ report of a PTWS 1420 instrument........................................58
Figure 67. FTP client - connection .......................................................................................59
Figure 68: FTP client - files..................................................................................................60
List of Tables
Table 1 Document History......................................................................................................7
Table 2: Technical Data ..........................................................................................................9
Table 3: Supply Scope ............................................................................................................9
Table 4: Available connecting cables .....................................................................................9
Table 5: PTB 111E and PTB 111EP serial result data (“FreeWeigh”) details ......................35
Table 6: PT-Node device settings for PTB 111E and PTB 111EP for serial data .................36
Table 7: PTB 111E ticket printer communication details ....................................................38
Table 8: PT-Node device settings for PTB 111E and ticket printer .....................................39
Table 9: PTB 311E and PTB 511E serial result data (“FreeWeigh”) details.........................41
Table 10: PTB 111E ticket printer communication details...................................................42
Table 11: PT-Node device settings for PTB 311E and PTB 511E and ticket printer ............44
Table 12: PTB 311E serial result data (“FreeWeigh”) details ..............................................45
Table 13: PT-Node device settings for PTB 311E.................................................................46
Table 14: PT-Node device settings for PTB 311E and PTB 511E and ticket printer ............48
Table 15: PT-Node FTP connection settings........................................................................59

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 7 of 64
Document History
Version
Valid from
Author
Change
Remark
1.0
29.03.2023
Pharma Test
N
First release
Table 1 Document History
Index Information - Change:
N = New Document
C = Correction
R = Revision

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 8 of 64
Section 1 General Information
Figure 1: PT-Node unit
PT-Node is an adapter that connects up to two Pharma Test instruments simultaneously
to a network using a wired LAN or wireless Wi-Fi connection. This way you can print test
results from the instrument via your web browser on any local or network printer.
Furthermore, it is possible to transfer the test results from the instruments to external
systems in the same network.
If the instrument is used in any other way as described in the manual, the
integrated safety features may be affected and there could be the
possibility of injuries to the operator.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 9 of 64
Section 1.1 Technical Specifications
Parameter
Specification
Display
OLED screen shows network status, IP address, connected
instruments and date & time
Configuration
Web-interface through web browser
Data storage
Internal SD card
Interfaces
2 x RS-232 serial port to connect up to two Pharma Test
instruments, 1 x LAN port or Wi-Fi to connect to network
Printer support
Printing via web browser on any local or network printer
Data access
FTP server connection
Power
100-240 Volt AC, 50/60 Hz
CE / EMC certification
All CE / EMC Certification provided
Validation
All IQ & OQ documents included
Table 2: Technical Data
Section 1.2 Standard Supply Scope
Part-No.
No.
Description
240-0210
1
Wall plug, 12V/DC, 2000mA, 24W, with international adapters
34-01123
1
Patch cable, RJ45, approx. 1.0m
Table 3: Supply Scope
Section 1.3 Available Connecting Cables
Part-No.
Description
34-00311
Connecting cable for PTB 111E, PTB 311E, PTB 511E and PTBA 211E
instruments
34-01203
Connecting cable for PTWS 120D, PTWS 120S, PTWS 820, PTWS 620, PTWS
1220, PTWS D620 and PTWS 1420 instruments
34-01205
Connecting cable for PT-TD300, PTF 100, PTF 200, PTF 300, PTF 600, PTZ
100 and PTZ 300 instruments
Table 4: Available connecting cables

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 10 of 64
Section 2 Instrument Overview
Check the indicated mains voltage at the instruments type plate, it should be
the same as within your laboratory. If not, you are not allowed to use the
instrument or connect it to the local mains supply.
PT-Node is used to collect serial data from Pharma Test instruments and stores
on a SD card inside of PT-Node. The data can be accessed through an FTP server
via ethernet (TCP/IP) or via a HTTP server through a web interface. This way the
supported Pharma Test instruments can be connected to a network with minimal
effort.
Figure 2 PT-Node Schematic
Section 2.1 Power Supply
PT-Node must be used with the supplied 12V/DC power supply. The supply voltage must be
between 9 and 12VDC. The current is at least 2A. The (+) connection is in the middle (pin). The
ground connection GND is on the outside of the plug-in system. The power supply must be
connected to the port on the rear side of the instrument.
Section 2.2 Serial Connections Channel 1 and 2
PT-Node allows the simultaneous connection of two Pharma Test instruments via serial
connection. The instruments must be connected to either Channel 1 or Channel 2 by the
corresponding instrument serial data cable.
Section 2.3 Ethernet Connection
PT-Node is to be connected to a network using the ethernet port on the rear side of the
instrument.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 11 of 64
Section 3 Setting-Up the Instrument
To use PT-Node you must:
▪Connect PT-Node to your network to be able to print reports via a web browser
and/or to export data to an external system (such as a LIMS)
▪Connect PT-Node to one or two Pharma Test instruments using the supplied
connecting cables
▪Configure the corresponding instrument communication settings in PT-Node
Section 3.1 Initial Set Up of the Network Configuration
By default, PT-Node tries to connect through a wired Ethernet connection to a network
using DHCP. This means PT-Node expects to be assigned an IP address automatically from
your network.
In case this fails, PT-Node creates a Wi-Fi hotspot. You can connect to this hotspot to
configure your network and other settings. If your network is using static IP addresses, you
will have to perform this next step:
Figure 3 Hotspot SSID is displayed on screen
Once the hotspot is created the name of the hotspot (“SSID”) is displayed on the screen of
PT-Node. In this case it is “PTNODEC265A3”.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 12 of 64
Use a Wi-Fi enabled device, for example a smartphone and connect to this hotspot by Wi-
Fi. Find the name of the hotspot as displayed on the PT-Node (example below showing an
Android smartphone):
Figure 4: Connecting to the hotspot

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 13 of 64
Then the PT-Node Connection Manager is displayed:
Figure 5: Network connection manager - Ethernet

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 14 of 64
Here you can select to use a wired Ethernet connection (shown above) and a wireless Wi-Fi
connection (shown below):
Figure 6: Network connection manager –Wi-Fi
Consult with your IT department for information on what settings you need for your network.
In case you are working with static IP addresses disable “DHCP” and manually enter the IP
address you want PT-Node to use.
Section 3.1.1 Connecting a PT-Node directly to a PC
In case you do not want to use any existing network, you can also directly connect a PT-
Node to a PC using an Ethernet cable. Refer to Section 4.4.2.1 below for more information
on this.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 15 of 64
Section 4 Using the Web Interface
After setting up the network connection, connect to the web interface of PT-Node using your
web browser from a device (for example a Windows PC) connected to the same network.
We recommend using either Google Chrome or Microsoft Edge.
In the bottom line of the display of PT-Node you will see a unique name being displayed:
Figure 7: Unique name of PT-Node is displayed on screen
In this example case it is “PTNODEC33673”. This name is unique for each PT-Node unit.
The display alternates between this name and the IP address currently assigned to this PT-
Node.
Now open your browser and enter the name as shown on the PT-Node display followed by
“.local” into the address line of the browser:
Figure 8: Accessing the web interface using PT-Node’s name
The web interface of PT-Node will be displayed.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 16 of 64
Alternatively, you can also connect to a PT-Node by entering the displayed IP address, for
example “192.168.15.3”:
Figure 9: Accessing the web interface using PT-Node’s IP address
The current IP address of PT-Node is shown on the bottom of its display, alternating with
its name:
Figure 10 IP address is displayed on screen
Note, that depending on how your company network is configured, it is possible that PT-
Node gets assigned a new IP address when you connect it to a different ethernet port.
Because of this, we recommend connecting to PT-Node by using its name, as this will never
change.
If no IP address is displayed refer to Section 3.1 on how to setup the network configuration.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 17 of 64
Section 4.1 File Index without Login
After entering the IP address, the file index is displayed. This shows all folders currently
located on the internal SD card of PT-Node:
Figure 11: File index
1: The serial number and currently installed firmware version of your PT-Node are
displayed here.
2: Click on “Login” to log in with username and password. See below for more details.
3: The current date and time are displayed here.
4: Use the circular arrow to re-fresh to current view, to show any new files or folder that
have been created.
1
2
3
4

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 18 of 64
A folder containing files is displayed in yellow. An empty folder is displayed in white:
Figure 12: Empty folders vs. folders with content
Click on any of the folders to display its contents:
Figure 13: List of files
For each file the following information is listed:
▪File name: the files names are automatically generated with a time stamp at the
point where they are created. The format for the file name is:
yyyymmdd_hhmmss.txt
▪File size: the file size in bytes
▪Date: the creation date of the file in the format: mm/dd/yyyy, hh:mm:ss AM/PM

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 19 of 64
Section 4.2 Display Reports and Printing
By default, PT-Node will collect the serial data sent by a Pharma Test instrument, convert
it to a text file and store it in the corresponding folder on its internal SD card.
Click on any of the text files in a folder to display the report. The report will be opened in a
new browser tab:
Figure 14: Example of a test report displayed using a web browser
The contents of the text file are displayed in a new tab. You can print the text report using
the standard printing functionality of your web browser on any printer connected to your
device, either directly or through a network. Click on “Print Text” to call up your operating
systems standard printing dialog.
The text reports here are displayed only. They cannot be edited or changed in any way. The
copy & paste function is disabled to prevent any kind manipulation of the results.
You can close this tab when you do not need it any longer. To keep working with PT-Node
simply switch back the tab labeled “PTNODE”.

Pharma Test Apparatebau AG
Operating Manual
B-24-00100_PT-NODE_v1.0e.docx Page 20 of 64
Section 4.3 Login as an Operator
To be able to download or delete any files stored on PT-Node you have to login as an
operator with password. By default these user settings are defined:
Username: user
Password: user
For security reasons this password should be changed when you first receive your PT-Node.
To do this you first have to login as the administrator. See the instructions below on how to
do that.
Enter the username and password and click on “Login” to proceed:
Figure 15: Login as user
If you entered the correct username and password the file index will be displayed again.
The login button on the top right has changed to logout:
Figure 16: Logout button
You can click on “Logout” to return the standard view of the file index (without the option to
download or delete files).
Note that a standard operator, who only wants to view and print reports does not need to
use any username or password. He can still access the reports but will not be able to
download or delete any files.