Presys FCY-15-IS User manual

®
FCY-15-IS
Hart
Field
Communicator
TECHNICAL MANUAL
presys
presys

presys

IMPORTANT INSTRUCTIONS:
•This manual contains instructions for the FCY-15-IS designed for use in hazardous areas. Read the entire
manual before using the calibrator.
•Before using the calibrator, carefully read the section “Special conditions for safe use”.
•Keep the calibrator in a dry environment whenever possible.
•In case of failure or suspected failure, especially in safe operation, send the instrument for repair to the
factory. Always send the instrument to the factory for repair.
•When not in daily use, before starting up, let the calibrator be turned on for at least one hour.
The warranty conditions are available on our website:
www.presys.com.br/warranty
EM0443-00 (09/20/2021) –Translated version of the document EM0422-00
presys
IMPORTANT INSTRUCTIONS
presys
IMPORTANT INSTRUCTIONS
:
presys
:
15
presys
15
-
presys
-
IS
presys
IS
designed for use in hazardous areas. Read
presys
designed for use in hazardous areas. Read
Before using the calibrator, carefully read the section
presys
Before using the calibrator, carefully read the section
“Special conditions for safe use”.
presys
“Special conditions for safe use”.
Keep the calibrator in a dry environment whenever possible
presys
Keep the calibrator in a dry environment whenever possible
.
presys
.
In case of failure or suspected failure, especially in safe operation, send the instrument for repair to the
presys
In case of failure or suspected failure, especially in safe operation, send the instrument for repair to the
lways send the instrument to the
presys
lways send the instrument to the
presys
factory for repair
presys
factory for repair
When not in daily use, before starting up, let the calibrator be turned on for at least one hour.
presys
When not in daily use, before starting up, let the calibrator be turned on for at least one hour.
presys
The warranty conditions are available on our website:
presys
The warranty conditions are available on our website:
presys
www.presys.com.br/
presys
www.presys.com.br/
presys
Translated version of
presys
Translated version of
the document EM0422
presys
the document EM0422

1
Table Of Contents
Marking Details ............................................................................................................................................................2
Special conditions for safe use...................................................................................................................................3
1 - Introduction...............................................................................................................................................................4
1.1. General Description.............................................................................................................................................4
1.2. General Specificaions .........................................................................................................................................5
2 - Operation.................................................................................................................................................................7
2.1. Parts Identification...............................................................................................................................................7
2.2. Battery and charger...........................................................................................................................................10
2.3. Main Menu ........................................................................................................................................................12
2.4. Transmitter Power Supply .................................................................................................................................13
2.5. HART®............................................................................................................................................................... 14
2.5.1. HART®connections ....................................................................................................................................14
2.5.2. Starting HART®Communication .................................................................................................................20
2.5.3. Adjusting the HART®Transmitter Measurement Range (CH option)..........................................................21
2.5.4. Adjustment of the HART®transmitter Measurement Range with reference (CH option) ............................23
2.5.5. HART®transmitter mA output adjustment - loop test / output trim (CH option)........................................... 25
2.5.6. Full HART®Communicator (FH option) ...................................................................................................... 26
2.5.7. Configuration Files (Save / Download) .......................................................................................................33
2.6. Measure (mA) ...................................................................................................................................................38
2.7. Videos ...............................................................................................................................................................41
2.8. Settings .............................................................................................................................................................42
3 - Maintenance..........................................................................................................................................................45
3.1. Inclusion of DD Files (Device Description) ........................................................................................................45
presys
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
Communication
presys
Communication
................................
presys
................................
presys
Transmitter Measurement Range (CH option)
presys
Transmitter Measurement Range (CH option)
2.5.4. Adjustment of the HART
presys
2.5.4. Adjustment of the HART
®
presys
®
transmitter Measurement Range with reference (CH option)
presys
transmitter Measurement Range with reference (CH option)
transmitter mA output adjustment
presys
transmitter mA output adjustment
-
presys
-
loop test / output trim (CH option)
presys
loop test / output trim (CH option)
Communicator (FH option)
presys
Communicator (FH option)
................................
presys
................................
2.5.7. Configuration Files (Save / Download)
presys
2.5.7. Configuration Files (Save / Download)
2.6. Measure (mA)
presys
2.6. Measure (mA)
................................
presys
................................
presys
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
3.1. Inclusion of DD Files (Device Description)
presys
3.1. Inclusion of DD Files (Device Description)

2
FCY-15-IS
Marking Details
Note: The Ex Compliance Certificate is sent with the instrument and its accessories.
presys
presys
presys
The Ex Compliance Certificate is sent with the instrument and its accessories.
presys
The Ex Compliance Certificate is sent with the instrument and its accessories.

3
Special conditions for safe use
•Use the calibrator only as described in this technical manual.
•The battery should only be charged in a safe area, using the charger provided. To avoid explosion or fire, use
only the battery (BT15-IS) and charger (CG15-IS) specified by Presys.
•Never replace the battery in a hazardous area.
•Do not use the serial communication port in an explosive atmosphere.
•The calibrator's aluminum metallic enclosure is protected by a leather case (BC15-IS) that should always house
it when in a hazardous area.
•The intrinsic safety of the instrument is only valid for the connections shown in this manual, respecting the
intrinsically safe input and output parameters.
•To avoid damage to the instrument and invalidate the Ex certification, never apply a voltage greater than 30 V
between the terminals and the metallic enclosure of the instrument.
•Never open the calibrator. Opening the enclosure may void the calibrator Ex certification.
•Do not use tools on the calibrator that may cause sparks; this practice can cause an explosion.
•Never perform maintenance on the calibrator; the components used are specified and cannot be changed.
•Never use the calibrator in an area close to explosive dust.
presys
should only be charged in a safe area, using the charger provided. To avoid explosion or fire, use
presys
should only be charged in a safe area, using the charger provided. To avoid explosion or fire, use
IS) specified by
presys
IS) specified by
Presys
presys
Presys
.
presys
.
in an explosive atmosphere.
presys
in an explosive atmosphere.
The calibrator's aluminum metallic enclosure is protected by a leather case (BC15
presys
The calibrator's aluminum metallic enclosure is protected by a leather case (BC15
The intrinsic safety of the instrument is only valid for the connections shown in this
presys
The intrinsic safety of the instrument is only valid for the connections shown in this
intrinsically safe input and output parameters.
presys
intrinsically safe input and output parameters.
To avoid damage to the instrument and invalidate the Ex certification, never apply a voltage greater than 30 V
presys
To avoid damage to the instrument and invalidate the Ex certification, never apply a voltage greater than 30 V
between the terminals and the metallic enclosure of the instrument.
presys
between the terminals and the metallic enclosure of the instrument.
r open the calibrator. Opening the enclosure may void the calibrator Ex certification.
presys
r open the calibrator. Opening the enclosure may void the calibrator Ex certification.
Do not use tools on the calibrator that may cause sparks; this practice can cause an explosion.
presys
Do not use tools on the calibrator that may cause sparks; this practice can cause an explosion.
Never perform maintenance on the calibrator; the components used are spec
presys
Never perform maintenance on the calibrator; the components used are spec
Never use the calibrator in an area close to explosive dust.
presys
Never use the calibrator in an area close to explosive dust.

4
1 - Introduction
1.1. General Description
The FCY-15-IS configurator allows the reading and configuration of parameters of field devices that have the
HART®protocol. It has a complete and updated DD (Device Description) configuration library registered in
FieldComm Group. It also allows the inclusion of new DD files through the USB port.
Besides HART®configurator, it has internal 15 Vdc TPS source (Transmitter Power Supply), selectable
250 Ω resistor (min.) and input for measurement of current (mAdc), allowing trim and loop test of the field instrument
with HART protocol without the need of a calibrator.
Its construction considers the use in the field, thus includes items of great value such as: bag with shoulder
straps allowing freedom for the hands, 5.7" display with led backlight facilitating visibility in low light and rechargeable
battery.
presys
configurator allows the reading and configuration of parameters of field devices that have the
presys
configurator allows the reading and configuration of parameters of field devices that have the
protocol. It has a complete and updated DD (Device Description) configuration library registered in
presys
protocol. It has a complete and updated DD (Device Description) configuration library registered in
D files through the USB port.
presys
D files through the USB port.
configurator, it has internal
presys
configurator, it has internal
15
presys
15
Vdc TPS source (Transmitter Power Supply), selectable
presys
Vdc TPS source (Transmitter Power Supply), selectable
presys
) and input for measurement of current (mAdc), allowing trim and
presys
) and input for measurement of current (mAdc), allowing trim and
HART protocol without the need of a calibrator.
presys
HART protocol without the need of a calibrator.
the use in the field, thus includes items of great value such as: bag with shoulder
presys
the use in the field, thus includes items of great value such as: bag with shoulder
straps allowing freedom for the hands, 5.7"
presys
straps allowing freedom for the hands, 5.7"
display with led backlight facilitating visibility i
presys
display with led backlight facilitating visibility i

5
1.2. General Specifications
Full HART®Configurator, with the latest complete DD library, registered by FieldComm Group.
Transmitter Power Supply
(TPS)
15 ± 1 V (0 to 24 mA),
with short-circuit protection (30 mA).
Internal Resistor
For mA Measurement: 100 to 120 Ω.
For mA input + HART®:250 to 280 Ω.
Safety condition (startup) / HART®only: open.
Input for current
measurement (mA)
Range from -5 to 24.5 mAdc.
Resolution of 0.0001 mA. Accuracy of ± 0.02% FS (Full Scale).
Special
software functions
Document: saves the entire configuration tree of the HART®field instrument.
Download: loads settings saved to the instrument with HART®protocol.
Quick/Calibration HART®: easy access to basic HART®commands.
Battery Rechargeable Nickel Metal Hydride with 4200 mAh and up to 8hours of continuous
use. Full charge in just 3 (three) hours.
Display 5.7’’ TFT VGA touch screen 640 x 480 pixels.
presys
Configurator, with the latest complete DD library, registered by Fiel
presys
Configurator, with the latest complete DD library, registered by Fiel
d
presys
d
Comm Group.
presys
Comm Group.
presys
protection (30 mA).
presys
protection (30 mA).
presys
For mA Measurement: 100 to 120 Ω.
presys
For mA Measurement: 100 to 120 Ω.
+ HART
presys
+ HART
®
presys
®
:
presys
:
250 to 280 Ω.
presys
250 to 280 Ω.
Safety condition (startup) / HART
presys
Safety condition (startup) / HART
®
presys
®
only: open.
presys
only: open.
presys
Range from
presys
Range from
-
presys
-
5 to 24.5 mAdc.
presys
5 to 24.5 mAdc.
Resolution of 0.0001 mA.
presys
Resolution of 0.0001 mA.
Accuracy of ± 0.02% FS (
presys
Accuracy of ± 0.02% FS (
presys
software functions
presys
software functions
Document
presys
Document
: saves the entire configuration tree of the HART
presys
: saves the entire configuration tree of the HART
Download
presys
Download
: loads settings saved to the instrument with HART
presys
: loads settings saved to the instrument with HART
Quick/Calibration HART
presys
Quick/Calibration HART
presys
presys
Battery
presys
Battery
Rechargeable
presys
Rechargeable
Nickel Metal Hydride
presys
Nickel Metal Hydride
use. Full charge in just 3
presys
use. Full charge in just 3
presys
presys
presys
presys
Display
presys
Display
5
presys
5
.
presys
.
7
presys
7
’’
presys
’’
TFT VGA touch
presys
TFT VGA touch
presys
presys
presys
presys
presys
presys
presys

6
Memory 16 GB available to the user. Memory access via USB cable x micro USB.
Processor Dual Core 1 GHz, with 512 MB of RAM and 1GB of internal Flash.
Operating Environment
Temperature of 0 to 50 °C and Relative Humidity of 90% maximum (without
condensation).
5 (five) minutes warm-up.
Dimensions / Weight 137 mm x 227 mm x 73 mm (HxWxD). 2.4 kg nominal.
Included Accessories
Soft Carrying case for transport and field use.
Battery charger.
USB cable for HART®library update.
Test Leads kit.
mA input Calibration Certificate.
Technical Manual.
Notes: *FCY-15-IS is a Presys registered trademark.
*Changes can be made to the instrument by changing the specifications described in this technical manual.
*HART®is a FieldComm Group trademark.
presys
user. Memory access via USB cable x micro USB.
presys
user. Memory access via USB cable x micro USB.
presys
Dual Core 1 GHz, with 512 MB of RAM and 1GB of internal Flash.
presys
Dual Core 1 GHz, with 512 MB of RAM and 1GB of internal Flash.
presys
C and Relative Humidity of 90% maximum (without
presys
C and Relative Humidity of 90% maximum (without
presys
73 mm (HxWxD). 2.
presys
73 mm (HxWxD). 2.
4
presys
4
kg nominal.
presys
kg nominal.
presys
presys
Soft Carrying case for transport and field
presys
Soft Carrying case for transport and field
use.
presys
use.
charger.
presys
charger.
USB cable for HART
presys
USB cable for HART
®
presys
®
library
presys
library
update.
presys
update.
Test Lead
presys
Test Lead
s
presys
s
kit.
presys
kit.
mA input Calibration
presys
mA input Calibration
Certificate.
presys
Certificate.
Technical Manual.
presys
Technical Manual.
presys
presys
presys
presys
IS is a Presys registered trademark
presys
IS is a Presys registered trademark
.
presys
.
presys
Changes can be made to the instrument by changing the specifications described in this technical manual.
presys
Changes can be made to the instrument by changing the specifications described in this technical manual.
Changes can be made to the instrument by changing the specifications described in this technical manual.
presys
Changes can be made to the instrument by changing the specifications described in this technical manual.
is a FieldComm Group
presys
is a FieldComm Group
trademark.
presys
trademark.

7
2 - Operation
2.1. Parts Identification
Front Panel
Fig. 01 -Front Panel
presys
presys

8
Left side panel Right side panel
Fig. 02 -Side Panels
presys
Right side panel
presys
Right side panel
presys

9
How to use the carrying case
Fig. 03 -Using the carrying case
Accessories:The bag has three compartments; one for accommodate the calibrator and the others to hold various
accessories including test leads, handles for transport in field use and technical manual.
presys
presys
Fig. 03
presys
Fig. 03
-
presys
-
Using the carrying case
presys
Using the carrying case
presys
The bag has three compartments; one for accommodate the calibrator and the others to hold various
presys
The bag has three compartments; one for accommodate the calibrator and the others to hold various
accessories including test leads, handles for transport
presys
accessories including test leads, handles for transport

10
2.2. Battery and charger
The FCY-15-IS is supplied with rechargeable battery which enables up to 12 hours of continuous use. This
autonomy is reduced according to the active functions (for example, using the 15 Vdc TPS source, or increasing
screen brightness). A battery charger that can be plugged into power supply from 100 to 240 Vac is included. The
time for full charge is 3 (three) hours.
The battery level is displayed in the main menu, as shown below.
Fig. 04 - Main Menu
presys
eable battery which enables up to 12 hours of continuous use. This
presys
eable battery which enables up to 12 hours of continuous use. This
autonomy is reduced according to the active functions (for example, using the
presys
autonomy is reduced according to the active functions (for example, using the
15
presys
15
Vdc TPS source, or increasing
presys
Vdc TPS source, or increasing
screen brightness). A battery charger that can be plugged into power supply fr
presys
screen brightness). A battery charger that can be plugged into power supply fr
om 100 to 240 Vac is included. The
presys
om 100 to 240 Vac is included. The
presys
The battery level is displayed in the main menu, as shown below.
presys
The battery level is displayed in the main menu, as shown below.
presys

11
Clicking on the battery icon, the following screen is shown. This screen shows the battery power (in percent)
and voltage (it decreases while the battery is discharging).
Fig. 05-Battery information
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
normally while the battery is being charged.
The batteries used by FCY-15-IS are made of Nickel Metal Hydride (Ni-MH). This new technology for
rechargeable batteries does not have the undesirable characteristics of memory effect as their preceding batteries
made of Nickel Cadmium (Ni-Cd). To prevent explosion or fire, use only the battery charger supplied by Presys.
Do not short circuit or damage the battery.
presys
Clicking on the battery icon, the following screen is shown. This screen shows the
presys
Clicking on the battery icon, the following screen is shown. This screen shows the
presys
Fig. 0
presys
Fig. 0
5
presys
5
-
presys
-
Battery information
presys
Battery information
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
presys
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
the battery is being charged.
presys
the battery is being charged.
The batteries used by FCY
presys
The batteries used by FCY
-
presys
-
presys
15
presys
15
-
presys
-
IS
presys
IS
rechargeable batteries does not have the undesirable characteristics of memory effect as their preceding batteries
presys
rechargeable batteries does not have the undesirable characteristics of memory effect as their preceding batteries
Nickel Cadmium (Ni
presys
Nickel Cadmium (Ni
-
presys
-
Cd).
presys
Cd).
To
presys
To
prevent explosion or fire,
presys
prevent explosion or fire,
presys
Do not short circuit or damage the battery.
presys
Do not short circuit or damage the battery.

12
2.3. Main Menu
When powered on, the FCY-15-IS goes through a self-test routine. In case of failure, it displays a message
to indicate error; if that occurs, contact Presys Technical Support.
After the self-test is completed, the display shows the main menu, as showed in Fig. 04.
The main menu is divided into 04 (four) functions:
HART®–allows communication with devices that have HART®protocol, see section 2.5.
MEASURE (mA) –mA input for current measurement up to 24.5 mAdc with high accuracy, see section 2.6.
HELP DESK –has videos made by PRESYS to assist in the use of the configurator, and can also store videos made
by the user, see section 2.7.
SETTINGS –general settings of the instrument, see section 2.8.
presys
test routine. In case of failure, it dis
presys
test routine. In case of failure, it dis
Technical Support.
presys
Technical Support.
test is completed, the display shows the main menu, as showed in
presys
test is completed, the display shows the main menu, as showed in
functions:
presys
functions:
allows communication with devices that have
presys
allows communication with devices that have
HART
presys
HART
®
presys
®
protocol, see
presys
protocol, see
mA input for current measurement up to 24.5 mAdc with high accuracy, see
presys
mA input for current measurement up to 24.5 mAdc with high accuracy, see
PRESYS
presys
PRESYS
to assist in the use of the configura
presys
to assist in the use of the configura
general settings of the instrument, see
presys
general settings of the instrument, see
section 2.8
presys
section 2.8

13
2.4. Transmitter Power Supply
The FCY-15-IS has a 15 Vdc (15 ± 1 V -0 to 24 mA) Transmitter Power Supply voltage source, with
short-circuit protection (current limited to 30 mA).
Fig. 06 -15 Vdc TPS Power Supply
presys
0 to 24 mA) Transmitter Power Supply voltage source, with
presys
0 to 24 mA) Transmitter Power Supply voltage source, with
presys
Fig.
presys
Fig.

14
2.5. HART®
The FCY-15-IS enables the reading and setting instrument parameters that have HART®communication
protocol. The HART®protocol allows digital communication between the master (in this case, the FCY-15-IS
configurator) and the slave (field instrument) superimposed on the 4-20 mA analog signal. To access this function,
from the main menu, select the HART®option.
The configurator comes with the latest DD (Device Description) library registered in FieldComm Group,
allowing the configuration of instrument-specific parameters.
2.5.1. HART®connections
When HART®is selected from the main menu, the mA INPUT + HART®and ONLY HART®(INCLUDING
NETWORK) options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
option must be chosen according to the type of connection that will be made.
presys
enables the reading and setting instrument parameters that have HART
presys
enables the reading and setting instrument parameters that have HART
protocol allows digital communication between the master (in this case, the FCY
presys
protocol allows digital communication between the master (in this case, the FCY
configurator) and the slave (field instrument) superimposed on th
presys
configurator) and the slave (field instrument) superimposed on th
e 4
presys
e 4
-
presys
-
20 mA analog signal. To access this function,
presys
20 mA analog signal. To access this function,
The configurator comes with the latest DD (
presys
The configurator comes with the latest DD (
Device Description
presys
Device Description
) library registered in
presys
) library registered in
specific parameters.
presys
specific parameters.
is selected from the main menu, the
presys
is selected from the main menu, the
presys
mA INPUT + HART
presys
mA INPUT + HART
options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
presys
options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
option must be chosen according to the type of connection that will be made.
presys
option must be chosen according to the type of connection that will be made.

15
mA Input + HART®(mA input of the FCY-15-IS in series with the HART®instrument)
Fig. 07 -mA input + HART®(Example 1)
2-wire transmitter powered by internal TPS source. Internal resistor enabled.
presys
in series with the HART
presys
in series with the HART
presys
Fig.
presys
Fig.
07
presys
07
2
presys
2
-
presys
-
presys
wire transmitter powered by internal TPS source. Internal resistor enabled.
presys
wire transmitter powered by internal TPS source. Internal resistor enabled.

16
mA Input + HART®(FCY-15-IS mA input in series with the HART®instrument)
Fig. 08 - mA input + HART®(Example 2)
2-wire transmitter, powered by external power supply. Internal resistor enabled.
presys
with the HART
presys
with the HART
presys
Fig.
presys
Fig.
08
presys
08
presys
wire transmitter
presys
wire transmitter

17
mA Input + HART® FCY-15-IS mA input in series with the HART®instrument)
Fig. 09 -mA input + HART®(Example 3)
4-wire transmitter with 115/230 Vac power and 4-20 mA output. Internal resistor enabled.
presys
series with the HART
presys
series with the HART
presys
Fig.
presys
Fig.
09
presys
09
presys
wire transmitter with
presys
wire transmitter with
115/230 Vac power and 4
presys
115/230 Vac power and 4
Table of contents
Other Presys Cell Phone manuals