Presys FCY-15 User manual

presys

EM0386-00
IMPORTANT INSTRUCTIONS:
•Keep the FCY-15 in a dry environment whenever possible.
•The fuse which protects the current measurement circuit, code 01.02.0277-21, is a special part.
So, only replace the fuse by another original from factory.
•In case of failure, always contact Presys Technical Support.
presys
IMPORTANT INSTRUCTIONS:
presys
IMPORTANT INSTRUCTIONS:
15 in a dry environment whenever possible.
presys
15 in a dry environment whenever possible.
The fuse which protects the current measurement circuit, code 01.02.0277
presys
The fuse which protects the current measurement circuit, code 01.02.0277
presys
So, only replace the fuse by another original from factory.
presys
So, only replace the fuse by another original from factory.
ase of failure, always contact Presys Technical Support.
presys
ase of failure, always contact Presys Technical Support.

Table of Contents
1 - Introduction...............................................................................................................................................................1
1.1. General Description........................................................................................................................................1
1.2. General Specifications ...................................................................................................................................2
2 - Operation..................................................................................................................................................................4
2.1. Parts Identification..........................................................................................................................................4
2.2. Battery and charger........................................................................................................................................7
2.3. Main Menu....................................................................................................................................................10
2.4. Transmitter Power Supply (TPS)..................................................................................................................11
2.5. HART®..........................................................................................................................................................12
2.5.1. HART®Connections.............................................................................................................................12
2.5.2. Starting HART® Communication.........................................................................................................18
2.5.3. Adjusting the HART® Transmitter Measurement Range (CH option)..................................................19
In this screen you can also edit the primary variable unit and the input filter (damping)................................20
2.5.4. Adjustment of the HART® transmitter Measurement Range with reference (CH option).....................21
2.5.5. HART® transmitter mA output adjustment -loop test / output trim (CH option)...................................23
2.5.6. Full HART®Communicator (FH option) ...............................................................................................24
2.5.7. Configuration Files (Save / Download) ................................................................................................31
2.6. Measure (mA)...............................................................................................................................................36
2.7. Videos..........................................................................................................................................................39
2.8. Settings........................................................................................................................................................40
3 - Maintenance...........................................................................................................................................................43
3.1. Replacing the battery ...................................................................................................................................43
3.2. Replacing the mA Input Fuse.......................................................................................................................45
3.3. Inclusion of DD Files (Device Description) ...................................................................................................46
presys
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
2.5.2. Starting HART® Communication
presys
2.5.2. Starting HART® Communication
................................
presys
................................
................................
presys
................................
2.5.3. Adjusting the HART® Transmitter Measurement Range (CH option)
presys
2.5.3. Adjusting the HART® Transmitter Measurement Range (CH option)
In this screen you can also edit the primary variable unit and the
presys
In this screen you can also edit the primary variable unit and the
2.5.4. Adjustment of the HART® transmitter Measurement Range with reference (CH option)
presys
2.5.4. Adjustment of the HART® transmitter Measurement Range with reference (CH option)
mitter mA output adjustment
presys
mitter mA output adjustment
-
presys
-
loop test / output trim (CH option)
presys
loop test / output trim (CH option)
Communicator (FH option)
presys
Communicator (FH option)
................................
presys
................................
2.5.7. Configuration
presys
2.5.7. Configuration
Files (Save / Download)
presys
Files (Save / Download)
................................
presys
................................
2.6. Measure (mA)
presys
2.6. Measure (mA)
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
................................
presys
................................
3.1. Replacing the battery
presys
3.1. Replacing the battery
................................
presys
................................
3.2. Replacing the mA I
presys
3.2. Replacing the mA I
nput Fuse
presys
nput Fuse
presys
3.3. Inclusion of DD Files (Device Description)
presys
3.3. Inclusion of DD Files (Device Description)

1
1 - Introduction
1.1. General Description
The FCY-15 configurator allows the reading and configuration of parameters of field devices that
have the HART® protocol. It has a complete and updated DD (Device Description) configuration library
registered in FieldComm Group. It also allows the inclusion of new DD files through the USB port.
Besides HART® configurator, it has internal 24 Vdc TPS source (Transmitter Power Supply),
selectable 250 Ω resistor (min) and input for measurement of current (mAdc), allowing trim and loop test
of the field instrument with HART protocol without the need of a calibrator.
Its construction takes into account the use in the field, thus includes items of great value such as:
bag with shoulder straps allowing freedom for the hands, 5.7 "display with led backlight facilitating
visibility in low light and rechargeable battery.
presys
15 configurator allows the reading and configur
presys
15 configurator allows the reading and configur
ation of parameters of field devices that
presys
ation of parameters of field devices that
have the HART® protocol. It has a complete and updated DD (Device Description) configuration library
presys
have the HART® protocol. It has a complete and updated DD (Device Description) configuration library
registered in FieldComm Group. It also allows the inclusion of new DD files through the USB port.
presys
registered in FieldComm Group. It also allows the inclusion of new DD files through the USB port.
configurator, it has internal 24 Vdc TPS source (Transmitter Power Supply),
presys
configurator, it has internal 24 Vdc TPS source (Transmitter Power Supply),
selectable 250 Ω resistor (min) and input for measurement of current (mAdc), allowing trim and loop test
presys
selectable 250 Ω resistor (min) and input for measurement of current (mAdc), allowing trim and loop test
of the field instrument with HART protocol without the need of a calibrator
presys
of the field instrument with HART protocol without the need of a calibrator
Its construction takes into account the use in the field, thus includes items of great value such as:
presys
Its construction takes into account the use in the field, thus includes items of great value such as:
bag with shoulder straps allowing freedom for the hands, 5.7 "display with led backlight facilitating
presys
bag with shoulder straps allowing freedom for the hands, 5.7 "display with led backlight facilitating
visibility in low light and rechargeable battery.
presys
visibility in low light and rechargeable battery.

2
1.2. General Specifications
Full HART® Configurator, with the latest complete DD library, registered by FieldComm Group.
Transmitter Power Supply
(TPS) 24.5 ± 1 V (0 to 24 mA),
with short-circuit protection (30 mA).
Internal Resistor For mA Measurement: 100 to 120 Ω.
For mA input + HART®:250 to 280 Ω.
Safety condition (startup) / HART® only: open.
Input for current
measurement (mA) Range from -5 to 24.5 mAdc.
Resolution of 0.0001 mA. Accuracy of ± 0.02% FS (Full Scale).
Special
software functions
Document:saves the entire configuration tree of the HART® field instrument.
Download:loads settings saved to the instrument with HART® protocol.
Quick/Calibration HART®:easy access to basic HART® commands.
Battery Rechargeable Lithium Polymer with 4200 mAh and up to 12 hours of continuous
use. Full charge in just 3 hours.
Display 5.7’’ TFT VGA touch screen 640 x 480 pixels.
presys
Full HART® Configurator, with the latest complete DD library, registered by Fiel
presys
Full HART® Configurator, with the latest complete DD library, registered by Fiel
d
presys
d
Comm Group.
presys
Comm Group.
presys
circuit protection
presys
circuit protection
(30 mA)
presys
(30 mA)
.
presys
.
presys
ent: 100 to 120 Ω.
presys
ent: 100 to 120 Ω.
+ HART®
presys
+ HART®
:
presys
:
250 to 280 Ω.
presys
250 to 280 Ω.
Safety condition (startup) / HART® only: open
presys
Safety condition (startup) / HART® only: open
presys
Range from
presys
Range from
-
presys
-
5 to 24.5 mAdc.
presys
5 to 24.5 mAdc.
Resolution of 0.0001 mA. Accuracy of ± 0.02% FS (
presys
Resolution of 0.0001 mA. Accuracy of ± 0.02% FS (
presys
Special
presys
Special
ctions
presys
ctions
Document
presys
Document
:
presys
:
saves the entire configuration tree of the HART® field instrument.
presys
saves the entire configuration tree of the HART® field instrument.
Download
presys
Download
:
presys
:
loads settings saved to the instrument with HART® protocol.
presys
loads settings saved to the instrument with HART® protocol.
Quick
presys
Quick
/
presys
/
Quick/Quick
presys
Quick/Quick
Calibration
presys
Calibration
/Calibration/
presys
/Calibration/
HART
presys
HART
presys
presys
Ba
presys
Ba
ttery
presys
ttery
Rechargeable Lithium Polymer
presys
Rechargeable Lithium Polymer
use. Full charge in just 3 hours.
presys
use. Full charge in just 3 hours.
presys
presys
presys
presys
Display
presys
Display
5
presys
5
.
presys
.
7
presys
7
’’
presys
’’
TFT VGA touch
presys
TFT VGA touch
presys
presys
presys
presys
presys
presys
presys

3
Memory 16 GB available to the user. Memory access via USB cable x micro USB.
Processor Dual Core 1 GHz, with 512 MB of RAM and 1GB of internal Flash.
Operating Environment Temperature of 0 to 50 °
C and Relative Humidity of 90% maximum (without
condensation).
5 minutes Warm-up.
Dimensions / Weight 140 mm x 255 mm x 80 mm (HxWxD). 1.5 kg nominal.
Included Accessories
-Technical manual;
-Test Leads;
- Carrying case;
- USB x micro USB cable;
-32 mA reserve fuse;
(mA input)
-Charger 100 to 240 Vac.
Notes: *FCY-15 is a registered trademark of Presys.
*Changes can be made to the instrument by changing the specifications described in this technical manual.
*HART® is a registered trademark of FieldComm Group.
presys
16 GB available to the user. Memory access via USB cable x micro USB.
presys
16 GB available to the user. Memory access via USB cable x micro USB.
presys
of RAM and 1GB of internal Flash.
presys
of RAM and 1GB of internal Flash.
presys
Temperature of 0 to 50 °
presys
Temperature of 0 to 50 °
C and Relative Humidity of 90% maximum (without
presys
C and Relative Humidity of 90% maximum (without
presys
140 mm x 255 mm x 80 mm (
presys
140 mm x 255 mm x 80 mm (
H
presys
H
x
presys
x
W
presys
W
xWx
presys
xWx
x
presys
x
D
presys
D
)
presys
)
. 1
presys
. 1
.
presys
.
5 kg nominal.
presys
5 kg nominal.
presys
presys
Technical manual;
presys
Technical manual;
Test Leads;
presys
Test Leads;
Carrying
presys
Carrying
case
presys
case
;
presys
;
-
presys
-
USB x micro USB
presys
USB x micro USB
cable
presys
cable
;
presys
;
-
presys
-
32 mA reserve fuse;
presys
32 mA reserve fuse;
(mA input)
presys
(mA input)
-
presys
-
Charger 100 to 240 Vac.
presys
Charger 100 to 240 Vac.
presys
presys
presys
presys
presys
presys
15 is a registered trademark of Presys.
presys
15 is a registered trademark of Presys.
Changes can be made to the instrument by changing th
presys
Changes can be made to the instrument by changing th
HART® is a registered trademark of FieldComm Group.
presys
HART® is a registered trademark of FieldComm Group.
presys

4
2 - Operation
2.1. Parts Identification
Front Panel
Fig. 01 -Front Panel
presys
presys

5
Left side panel Right side panel
Fig. 02 -Side Panels
presys
Right side panel
presys
Right side panel
presys

6
How to use the carrying case
Fig. 03 -Using the carrying case
Accessories: The bag has three compartments; one for accommodate the calibrator and the others to hold various
accessories including test leads, spare fuse, handles for transport and field use and technical manual.
presys
presys
Fig. 03
presys
Fig. 03
-
presys
-
Using the carrying case
presys
Using the carrying case
presys
The bag has three compartments; one for accommodate the calibrator and the others to hold various
presys
The bag has three compartments; one for accommodate the calibrator and the others to hold various
accessories including test
presys
accessories including test
leads, spare fuse, handles for transport and field use and technical manual.
presys
leads, spare fuse, handles for transport and field use and technical manual.

7
2.2. Battery and charger
The FCY-15 is supplied with rechargeable battery which enables up to 12 hours of continuous use. This
autonomy is reduced according to the active functions (for example, using the 24 Vdc TPS source, or increasing
screen brightness). A battery charger that can be plugged into power supply from 100 to 240 Vac is included. The
time for full charge is 3 hours.
The battery level is displayed in the main menu, as shown below.
Fig. 04 -Main menu
FCY-15 Informations
(Serial N, Manual, Software Version ...)
Battery Level
Technical Manual
presys
is supplied with rechargeable battery which enables up to 12 hours of continuous use. This
presys
is supplied with rechargeable battery which enables up to 12 hours of continuous use. This
tions (for example, using the 24 Vdc TPS source, or increasing
presys
tions (for example, using the 24 Vdc TPS source, or increasing
screen brightness). A battery charger that can be plugged into power supply from 100 to 240 Vac is included. The
presys
screen brightness). A battery charger that can be plugged into power supply from 100 to 240 Vac is included. The
u, as shown below.
presys
u, as shown below.
presys
presys

8
Fig. 05 -Low battery level, charge required
Fig. 06 - Charged battery
Fig. 07 -Charger connected, battery charging
Clicking on the battery icon, the following screen is shown. This screen shows the battery power (in percent),
current (positive value if the battery is charging, negative if it is not) and the estimated time of instrument autonomy
based on the current consumption.
Fig. 08 -Battery information, charger disconnected
presys
Low battery level, charge required
presys
Low battery level, charge required
Charged battery
presys
Charged battery
Charger connected, battery charging
presys
Charger connected, battery charging
Clicking on the battery icon, the following screen is shown. This s
presys
Clicking on the battery icon, the following screen is shown. This s
creen shows the battery power (in percent),
presys
creen shows the battery power (in percent),
presys
current (positive value if the battery is charging, negative if it is not) and the estimated time of instrument autonomy
presys
current (positive value if the battery is charging, negative if it is not) and the estimated time of instrument autonomy
presys
Fig. 08
presys
Fig. 08

9
Fig. 09 -Battery information, charger connected
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
normally while the battery is being charged.
The batteries used by FCY-15 are made of Lithium Polymer (Li-Po). This new technology for rechargeable
batteries does not have the undesirable characteristics of memory effect as their preceding batteries made of Nickel
Cadmium (Ni-Cd).
To prevent explosion or fire, use only the battery charger supplied by Presys. Do not short circuit or damage
the battery.
presys
presys
Fig. 09
presys
Fig. 09
-
presys
-
Battery information, charger connected
presys
Battery information, charger connected
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
presys
The charger provides the battery charge while it feeds the calibrator, thus permitting the calibrator to be used
normally while the battery is being charged.
presys
normally while the battery is being charged.
The batteries used by FCY
presys
The batteries used by FCY
-
presys
-
presys
15 are made of Lith
presys
15 are made of Lith
batteries does not have the undesirable characteristics of memory effect as their preceding batteries made of Nickel
presys
batteries does not have the undesirable characteristics of memory effect as their preceding batteries made of Nickel
Cd).
presys
Cd).
presys
To prevent explosion or fire, use only the battery charger suppl
presys
To prevent explosion or fire, use only the battery charger suppl

10
2.3. Main Menu
When powered on, the FCY-15 goes through a self-test routine. In case of failure, it displays a message to
indicate error; if that occurs, contact Presys Technical Support.
After the self-test is completed, the display shows the main menu, as showed in figure 04.
The main menu is divided into 04 functions:
HART®–allows communication with devices that have Hart®protocol, see section 2.5.
MEASURE (mA) –mA input for current measurement up to 24.5 mAdc with high accuracy, see section 2.6.
VIDEOS –has videos made by PRESYS to assist in the use of the configurator, and can also store videos made by
the user, see section 2.7.
SETTINGS –general settings of the instrument, see section 2.8.
presys
test routine. In case of failure, it displays a message to
presys
test routine. In case of failure, it displays a message to
indicate error; if that occurs, contact Presys Technical Support.
presys
indicate error; if that occurs, contact Presys Technical Support.
test is completed, the display shows the main menu, as showed in
presys
test is completed, the display shows the main menu, as showed in
The main menu is divided into 04 functions:
presys
The main menu is divided into 04 functions:
allows communication with devices that have Hart
presys
allows communication with devices that have Hart
®
presys
®
protocol, see
presys
protocol, see
mA input for cu
presys
mA input for cu
rrent measurement up to 24.5 mAdc with high accuracy, see
presys
rrent measurement up to 24.5 mAdc with high accuracy, see
has videos made by
presys
has videos made by
PRESYS
presys
PRESYS
to assist in the use of the configurator, and can also store videos made by
presys
to assist in the use of the configurator, and can also store videos made by
section 2.7
presys
section 2.7
.
presys
.
presys
general settings of the instrument
presys
general settings of the instrument

11
2.4. Transmitter Power Supply (TPS)
The FCY-15 has a 24 Vdc (24.5 ± 1 V -0 to 24 mA) Transmitter Power Supply voltage source, with
short-circuit protection (current limited to 30 mA).
Fig. 10 -24 Vdc TPS Power Supply
presys
0 to 24 mA) Transmitter Power Supply voltage source, with
presys
0 to 24 mA) Transmitter Power Supply voltage source, with
presys
presys

12
2.5. HART®
The FCY-15 enables the reading and setting instrument parameters that have HART® communication
protocol. The HART® protocol allows digital communication between the master (in this case, the FCY-15
configurator) and the slave (field instrument) superimposed on the 4-20 mA analog signal. To access this function,
from the main menu, select the HART® option.
The configurator comes with the latest DD (Device Description) library registered in FieldComm Group,
allowing the configuration of instrument-specific parameters.
2.5.1. HART®Connections
When HART® is selected from the main menu, the mA INPUT + HART® and ONLY HART® (INCLUDING
NETWORK) options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
option must be chosen according to the type of connection that will be made.
presys
15 enables the reading and setting instrument parameters that have HART® communication
presys
15 enables the reading and setting instrument parameters that have HART® communication
protocol. The HART® protocol allows digital communication between the master (in this case, the FCY
presys
protocol. The HART® protocol allows digital communication between the master (in this case, the FCY
osed on the 4
presys
osed on the 4
-
presys
-
20 mA analog signal. To access this function,
presys
20 mA analog signal. To access this function,
presys
The configurator comes with the latest DD (Device Description) library registered in FieldComm Group,
presys
The configurator comes with the latest DD (Device Description) library registered in FieldComm Group,
c parameters.
presys
c parameters.
When HART® is selected from the main menu, the
presys
When HART® is selected from the main menu, the
mA INPUT + HART®
presys
mA INPUT + HART®
presys
options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
presys
options will be displayed on the screen. The internal resistor (250 Ω min.) can also be enabled. The
chosen according to the type of connection that will be made.
presys
chosen according to the type of connection that will be made.

13
mA Input + HART® (mA input of the FCY-15 in series with the HART® instrument)
Fig. 11-mA input + HART® -Example 1
2-wire transmitter powered by internal TPS source. Internal resistor enabled.
presys
15 in series with the HART® instrument)
presys
15 in series with the HART® instrument)
presys
Fig. 1
presys
Fig. 1
2
presys
2
-
presys
-
presys
wire transmitter powered by internal TPS source. Internal resistor e
presys
wire transmitter powered by internal TPS source. Internal resistor e

14
mA Input + HART® (mA input of the FCY-15 in series with the HART® instrument)
Fig. 12 -mA input + HART® -Example 2
2-wire transmitter,powered by external power supply. Internal resistor enabled.
presys
15 in series with the HART® instrument)
presys
15 in series with the HART® instrument)
presys
Fig. 12
presys
Fig. 12
wire transmitter
presys
wire transmitter
,
presys
,
powered by external power supply. Internal resistor enabled.
presys
powered by external power supply. Internal resistor enabled.

15
mA Input + HART® (mA input of the FCY-15 in series with the HART® instrument)
Fig. 13 -mA input + HART® -Example 3
4 wire transmitter with 115/230 Vac power and 4-20 mA output. Internal resistor enabled.
presys
series with the HART® instrument)
presys
series with the HART® instrument)
presys
presys
4 wire transmitter with 115/230 Vac power and 4
presys
4 wire transmitter with 115/230 Vac power and 4

16
ONLY HART® (INCLUDING NETWORK) (mA input of FCY-15 is not used)
Fig. 14 -HART® only -Example 1
Transmitter powered by external source, resistor of at least 250 Ω in series.
presys
15 is not used)
presys
15 is not used)
presys
presys
Transmitter powered by external source, resistor of at least 250 Ω in series.
presys
Transmitter powered by external source, resistor of at least 250 Ω in series.

17
mA Input + HART®(mA input of the FCY-15 in series with the HART® instrument)
For the connections shown in figure 11,figure 12 and figure 13, use the mA Input + HART® and
INTERNAL RESISTOR enabled option. In this mode, the HART® resistor of at least 250 Ω is internally activated, in
series with the mA input of the FCY-15. The FCY-15 configurator can measure the transmitter current and also read
and configure parameters via HART®. If the internal resistor is not enabled, an external resistor of at least 150 Ω
should be inserted in series with the mA input of the FCY-15. To power the two-wire transmitter, the FCY-15 24 Vdc
TPS source (Figure 11) or an external source (Figure 12)can be used. In the case of the 4-wire transmitter, connect
the 4-20 mA output of the transmitter to the mA input of the FCY-15, and connect the HART® terminals (Figure 13).
ONLY HART® (INCLUDING NETWORK) (mA input of FCY-15 is not used)
For the connection shown in figure 14, use the ONLY HART® option. In this mode, the internal resistor and
mA input are disabled. The HART® resistor of at least 250 Ω must be inserted externally in series with the
transmitter. In this case, the configurator does not measure the transmitter current, but can read and configure its
parameters via HART®.
presys
15 in series with the HART® instrument)
presys
15 in series with the HART® instrument)
figure 13
presys
figure 13
, use the
presys
, use the
mA Input + HART®
presys
mA Input + HART®
enabled option. In this mode, the HART® resistor of at least 250 Ω is internally activated, in
presys
enabled option. In this mode, the HART® resistor of at least 250 Ω is internally activated, in
15 configurator can measure the transmitter current and also read
presys
15 configurator can measure the transmitter current and also read
T®. If the internal resistor is not enabled, an external resistor of at least 150 Ω
presys
T®. If the internal resistor is not enabled, an external resistor of at least 150 Ω
should be inserted in series with the mA input of the FCY
presys
should be inserted in series with the mA input of the FCY
-
presys
-
15. To power the two
presys
15. To power the two
-
presys
-
wire transmitter, the
presys
wire transmitter, the
Figure 12
presys
Figure 12
)
presys
)
can be used. In the case of the 4
presys
can be used. In the case of the 4
20 mA output of the transmitter to the mA input of the FCY
presys
20 mA output of the transmitter to the mA input of the FCY
-
presys
-
15, and connect the HART® terminals (
presys
15, and connect the HART® terminals (
ONLY HART® (INCLUDING NETWORK) (mA input of FCY
presys
ONLY HART® (INCLUDING NETWORK) (mA input of FCY
-
presys
-
15 is not used)
presys
15 is not used)
presys
tion shown
presys
tion shown
presys
in figure 14
presys
in figure 14
, use the
presys
, use the
ONLY HART®
presys
ONLY HART®
mA input are disabled. The HART® resistor of at least 250
presys
mA input are disabled. The HART® resistor of at least 250
transmitter. In this case, the configurator does not measure the transmitter current, but can read and configure its
presys
transmitter. In this case, the configurator does not measure the transmitter current, but can read and configure its
parameters via HART®.
presys
parameters via HART®.
Table of contents
Other Presys Cell Phone manuals