Table 8
Recommended separation distances between portable and mobile RF
communications equipment and the 240390
The is intended for use in an electromagnetic environment in which radiated 240390
RF disturbances are controlled. The customer or the user of the can help 240390
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the 240390
as recom men ded be low, acco rdi ng to the maxi mum ou tpu t power of th e
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
Rated maximum
output power
150kHz to 80MHz
d =1.2 P
80MHz to 800MHz
d =1.2 P
800MHz to 2.5GHz
d =2.3 P