Rohde & Schwarz R&S RT-Z2T User manual

R&S®RT-Z2T
Probe Interface Adapter
Manual
1326438302
Version 04
(=J[á2)

This manual describes the R&S RT-Z2T probe interface adapter (1338.0007.02).
© 2022 Rohde & Schwarz GmbH & Co. KG
Muehldorfstr. 15, 81671 Muenchen, Germany
Phone: +49 89 41 29 - 0
Email: [email protected]
Internet: www.rohde-schwarz.com
Subject to change – data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
TEKTRONIX, TEK, and TEKPROBE are registered trademarks of Tektronix, Inc.
All other trademarks are the properties of their respective owners.
1326.4383.02 | Version 04 | R&S®RT-Z2T
Throughout this manual, products from Rohde & Schwarz are indicated without the ® symbol, e.g.
R&S®RT-Z2T is indicated as R&S RT-Z2T.

Contents
R&S®RT-Z2T
3Manual 1326.4383.02 ─ 04
Contents
1 Safety information................................................................. 4
2 Product description...............................................................6
2.1 Supported Tektronix probes................................................................ 6
2.2 Deliveries............................................................................................... 6
2.3 Inspecting the contents........................................................................6
2.4 Description of the adapter................................................................... 7
3 Putting into operation........................................................... 9
3.1 Connecting the adapter to the oscilloscope...................................... 9
3.2 Connecting the Tektronix probe to the adapter............................... 10
4 Maintenance and service.................................................... 11
4.1 Cleaning............................................................................................... 11
4.2 Contacting customer support............................................................11
4.3 Returning for servicing...................................................................... 12
4.4 Calibration interval............................................................................. 12
4.5 Storage.................................................................................................13
4.6 Disposal............................................................................................... 13

Safety information
R&S®RT-Z2T
4Manual 1326.4383.02 ─ 04
1 Safety information
The product documentation helps you use the product safely and efficiently. Fol-
low the instructions provided here and in the following chapters.
Intended use
The R&S RT-Z2T probe interface adapter is designed for measurements on cir-
cuits that are only indirectly connected to the mains or not connected at all. It is
not rated for any measurement category.
The product is intended for the development, production and verification of elec-
tronic components and devices in industrial, administrative, and laboratory envi-
ronments. Use the product only for its designated purpose. Observe the operating
conditions and performance limits stated in the data sheet.
The R&S RT-Z2T probe interface adapter is designed to connect active Tektronix
oscilloscope probes with TekProbe BNCTM level II interface to oscilloscopes that
have a Rohde & Schwarz probe interface.
Where do I find safety information?
Safety information is part of the product documentation. It warns you of potential
dangers and gives instructions on how to prevent personal injury or damage
caused by dangerous situations.
●Printed "Safety Instructions for Oscilloscopes and Accessories" in many lan-
guages is delivered with your Rohde & Schwarz oscilloscope and probes.
●The documentation of the oscilloscope and measurement equipment provides
safety instructions when you need to take care during setup or operation.
Operating site
Only use the product indoors, and keep it dry. The product casing is not water-
proof.
You can operate the product up to the altitude specified in the data sheet. The
lowest specified altitude for a product of the measurement setup defines the alti-
tude for the complete setup.
The product is suitable for pollution degree 2 environments where nonconductive
contamination can occur. For more information on environmental conditions such
as ambient temperature and humidity, see the data sheet.

Safety information
R&S®RT-Z2T
5Manual 1326.4383.02 ─ 04
Product integrity
The R&S RT-Z2T can withstand a moderate amount of physical and electrical
stress. To avoid damage, treat the adapter with care:
●Do not exceed the specified voltage limits.
●Connect the R&S RT-Z2T only to an instrument with Rohde & Schwarz probe
interface. Never connect it to a usual BNC jack. Although the 7 mm coaxial
connector looks like a standard BNC connector, it is constructed differently
and does not fit to the standard BNC jack. The interface of the R&S RT-Z2T
can withstand a higher frequency limit.
●Prevent the adapter from receiving mechanical shock.
●Do not spill liquids on the adapter.
●Store the adapter in a shock-resistant case, e.g. in the foam-lined shipping
case.

Product description
R&S®RT-Z2T
6Manual 1326.4383.02 ─ 04
2 Product description
The R&S RT-Z2T probe interface adapter is designed to connect active Tektronix
oscilloscope probes with TekProbe BNCTM level II interface to oscilloscopes that
have a Rohde & Schwarz probe interface.
The adapter provides power to the probe. The Rohde & Schwarz oscilloscope is
adjusted to the correct settings by selecting the probe from a list of predefined
probes.
2.1 Supported Tektronix probes
The following probes are supported by the R&S RT-Z2T adapter:
●Single-ended probes: P6205, P6243, P6245, P6241, P6249
●Current probes: TCP202
●Differential probes: P6246, P6247, P6248, P6250, P6251
●High-voltage differential probes: P5205, P5210
●Electro-optical probes: P6701B, P6703B, P6711, P6713
2.2 Deliveries
The R&S RT-Z2T delivery package contains the following items:
●R&S RT-Z2T probe interface adapter
●Manual
2.3 Inspecting the contents
●Inspect the package for damage.
Keep the package and the cushioning material until the contents have been
checked for completeness and the device has been tested.
Inspecting the contents

Product description
R&S®RT-Z2T
7Manual 1326.4383.02 ─ 04
If the packaging material shows any signs of stress, notify the carrier and your
Rohde & Schwarz service center. Keep the package and cushioning material
for inspection.
●Inspect the adapter.
If there is any damage or defect, or if the R&S RT-Z2T probe interface adapter
does not operate properly, notify your Rohde & Schwarz service center.
●Inspect the accessories.
If the contents are incomplete or damaged, notify your Rohde & Schwarz ser-
vice center.
2.4 Description of the adapter
The R&S RT-Z2T has a Rohde & Schwarz probe interface on one side, and a
TekProbe BNCTM level II interface on the other side. The interfaces provide power
to the probe.
Figure 2-1: R&S probe interface of the R&S
RT-Z2T
1 = Release knob
2 = Rohde & Schwarz probe interface with 7 mm (276 mil) coaxial connector and 6 pogo pins
Description of the adapter

Product description
R&S®RT-Z2T
8Manual 1326.4383.02 ─ 04
Figure 2-2: TekProbe BNCTM level
II interface of the R&S
RT-Z2T
Description of the adapter

Putting into operation
R&S®RT-Z2T
9Manual 1326.4383.02 ─ 04
3 Putting into operation
Read and observe the instructions in Chapter 1, "Safety information", on page 4
before using the adapter.
For your safety, follow the safety instructions delivered with your
Rohde & Schwarz oscilloscope and probe, and the instructions given in the man-
uals of your measurement equipment.
Observe the voltage limits and ratings of the adapter and the measuring instru-
ment. Limits and ratings are marked on the products and listed in the data sheets.
Do not exceed the lesser of the maximum input voltages specified for the adapter
or the oscilloscope. If the oscilloscope is set to 50 Ω, input coupling, the maxi-
mum input voltage is usually determined by the oscilloscope. If the oscilloscope is
set to 1 MΩ input coupling, the maximum input voltage is usually determined by
the adapter.
Preventing electrostatic discharge (ESD)
Electrostatic discharge is most likely to occur when you connect or disconnect a
DUT.
►NOTICE! Electrostatic discharge can damage the electronic components of
the product and the device under test (DUT).
Ground yourself to prevent electrostatic discharge damage:
a) Use a wrist strap and cord to connect yourself to ground.
b) Use a conductive floor mat and heel strap combination.
3.1 Connecting the adapter to the oscilloscope
1. Make sure that the oscilloscope is properly grounded.
2. NOTICE! Risk of damaging the probe.
Connect the probe only to an instrument with Rohde & Schwarz probe inter-
face. Never connect it to a usual BNC jack, because this can damage the
probe interface.
Connecting the adapter to the oscilloscope

Putting into operation
R&S®RT-Z2T
10Manual 1326.4383.02 ─ 04
Connect the adapter (1) to the Rohde & Schwarz probe interface of the oscil-
loscope (2).
The adapter snaps in when connected properly to the port.
Figure 3-1: Connecting the R&S
RT-Z2T to the Rohde
&
Schwarz oscilloscope
► To disconnect the adapter:
a) Press and hold the release button (3).
b) Pull the adapter away from the oscilloscope.
3.2 Connecting the Tektronix probe to the adapter
For a description of the usage of a Tektronix probe, refer to its user manual.
1. Turn the lever of the probe to the left side in the open position.
2. Connect the BNC connector of the probe to the BNC socket of the R&S RT-
Z2T adapter.
3. Lock the probe by pushing the lever to the right into the closed position.
4. On the oscilloscope, choose the correct predefined probe, or enter the values
for external attenuation, input coupling and display unit manually.
Connecting the Tektronix probe to the adapter

Maintenance and service
R&S®RT-Z2T
11Manual 1326.4383.02 ─ 04
4 Maintenance and service
Like all Rohde & Schwarz products, Rohde & Schwarz probes and adapters are
of high quality and require only minimum service and repair. However, if service
or calibration is needed, contact your Rohde & Schwarz service center. Return a
defective product to the Rohde & Schwarz service center for diagnosis and
exchange.
4.1 Cleaning
1. Clean the outside of the product using a soft cloth moistened with either distil-
led water or isopropyl alcohol. Keep in mind that the casing is not waterproof.
Note: Do not use cleaning agents. Solvents (thinners, acetone), acids and
bases can damage the labeling or plastic parts.
2. Dry the product completely before using it.
4.2 Contacting customer support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz product, contact our customer
support center. A team of highly qualified engineers provides support and works
with you to find a solution to your query on any aspect of the operation, program-
ming or applications of Rohde & Schwarz products.
Contact information
Contact our customer support center at www.rohde-schwarz.com/support, or fol-
low this QR code:
Contacting customer support

Maintenance and service
R&S®RT-Z2T
12Manual 1326.4383.02 ─ 04
Figure 4-1: QR code to the Rohde
&
Schwarz support page
4.3 Returning for servicing
Use the original packaging to return your R&S RT-Z2T to your Rohde & Schwarz-
service center. A list of all service centers is available on:
www.services.rohde-schwarz.com
If you cannot use the original packaging, consider the following:
1. Use a sufficiently sized box.
2. Protect the product from damage and moisture (e.g. with bubble wrap).
3. Use some kind of protective material (e.g. crumpled newspaper) to stabilize
the product inside the box.
4. Seal the box with tape.
5. Address the package to your nearest Rohde & Schwarz service center.
4.4 Calibration interval
The recommended calibration interval for R&S RT-Z2T probe interface adapter is
two years. For servicing, send the probe to your nearest Rohde & Schwarz ser-
vice center (see Chapter 4.3, "Returning for servicing", on page 12).
Calibration interval

Maintenance and service
R&S®RT-Z2T
13Manual 1326.4383.02 ─ 04
4.5 Storage
Protect the product against dust. Ensure that the environmental conditions, e.g.
temperature range and climatic load, meet the values specified in the data sheet.
Store the product in a shock-resistant case, e.g. in the shipping case.
4.6 Disposal
Rohde & Schwarz is committed to making careful, ecologically sound use of natu-
ral resources and minimizing the environmental footprint of our products. Help us
by disposing of waste in a way that causes minimum environmental impact.
Disposing electrical and electronic equipment
A product that is labeled as follows cannot be disposed of in normal household
waste after it has come to the end of its service life. Even disposal via the munici-
pal collection points for waste electrical and electronic equipment is not permitted.
Figure 4-2: Labeling in line with EU directive WEEE
Rohde & Schwarz has developed a disposal concept for the eco-friendly disposal
or recycling of waste material. As a manufacturer, Rohde & Schwarz completely
fulfills its obligation to take back and dispose of electrical and electronic waste.
Contact your local service representative to dispose of the product.
Disposal
Table of contents