SCHMITT Acuity User manual

1
7” Touch Panel Display System
Operator’s Manual
LLL009070 - Revision: 07.14.20
sales@acuitylaser.com
Tel: +1 503.227.7908
Fax: +1 503.223.1258
acuitylaser.com
©2020 Schmitt Industries
2765 NW Nicolai St.
Portland, OR 97210 USA
A product line of
Schmitt-ind.com

2
Acuity Touch Panel Display Instruction Manual
Table of Contents
Setup............................................................................................................................................................. 3
Settings Explanation..................................................................................................................................... 5
Calibration .................................................................................................................................................... 5
Measuring..................................................................................................................................................... 6
Utilities.......................................................................................................................................................... 7
Troubleshooting ........................................................................................................................................... 8

3
1. Setup
1.1. Connect the laser connection kit(s) to the port(s) on the bottom of the TPD labeled “Port A” or
“Port B”with a DB-9 cable. Ensure the laser is powered up. When using a single laser, either
port can be used.
Figure 1: Identification of TPD ports and power button.
1.2. Connect the power cable to the TPD and a power source.
1.3. Power up the TPD by pushing and holding the green button on the bottom of the TPD.
The power indicator on the front of the TPD will turn green.
Upon initial power up, the TPD will boot into the Settings screen after approximately 30
seconds.
Figure 2: Settings screen at initial startup.
1.4. Configure the Units by selecting the desired measurement units from the Units dropdown.
(Available display units are millimeter (mm), centimeter (cm), meter, foot decimal, foot
fraction, inch decimal, inch fraction).

4
1.5. Configure the number of decimal places or smallest fraction to display from the Decimals or
Smallest fraction dropdown. When using foot fraction or inch fraction units, the dropdown
menu displays the smallest fraction to round to (down to 1/64 of an inch). When using foot
decimal or inch decimal the dropdown will display the smallest decimal number to round to (1-
9).
1.6. Configure the number of readings to average in the Average readings field (default 25).
Choose rolling or batch for the averaging function. See Averaging section for an explanation.
1.7. Choose the laser model from the Laser dropdown in the Laser a config or Laser b config panes.
1.8. Check Enabled to enable the laser interface. This will start the baud rate query. When query
has completed, the status of the query will be displayed to the right of the baud rate
dropdown.
1.8.1. “Connected” indicates a successful query.
1.8.2. “Connection failed” indicates a connection problem. See trouble shooting section.
1.8.3. “Error: <specific error message>” indicates a mismatch with the connected laser model
and the value in the Laser dropdown.
1.8.4. If not using Thickness mode, enter the Distance limits (if required).
1.9. Repeat steps 1.7 through 1.8 for second laser (if applicable).
1.10. If measuring thickness with two lasers, check Thickness mode and choose a thickness function
(sum or difference). The sum function is for applications where the subject is placed between
opposing lasers. The thickness is determined by the sum of the measurements. The difference
function is for applications were the subject and the zero-point surface are both in motion.
One laser measures the zero point and the other laser measures the subject. The thickness is
determined by the difference in the two measurements.
1.10.1. Enter Thickness limits (if required).
1.11. Press the Done button to change to the Measure screen.
1.12. Press Save to save the current setup. You can now start measuring by pressing Start.
Figure 3: Measure screen during active measuring.

5
2. Settings Explanation
2.1. Averaging: Averaging will steady the display of measurements. The default setting is a rolling
average of 25 measurements. The rolling average will add the latest reading to the back of the
averaging list and remove the first reading from the front of the averaging list so that the
averaging list always contains the latest 25 measurements. These measurements are then
averaged before display. The batch average will average measurements in groups of 25 before
display. Averaging can delay the display of the latest average value as the averaging list is que’d
and de-que’d. To reduce this effect, lower the average readings value. To turn off averaging, set
the mode to ‘off’.
2.2. Thickness mode: Thickness mode uses two lasers to measure the thickness of a target. It
operates in two modes, sum and difference. In sum mode, the thickness is calculated by
summing the values of two laser measurements pointed at opposite sides of the target. In
difference mode one laser is used to track the zero-point of the target (e.g.: roller surface or
conveyor belt), the other laser is used to track the measurement of the target on top of the
zero-point. The thickness of the target is determined by the difference in the target
measurement and the zero-point measurement.
2.3. Distance and Thickness limits: Distance limits (non-thickness mode) or Thickness limits
(thickness mode) can be specified to change the color of the measurement display in four
scenarios:
2.3.1. Upper limit (Upper lim): When measurement is equal or greater than upper limit, the
measurement color will be red.
2.3.2. Upper limit (Upper warn): When measurement is equal or greater than upper warn, the
measurement color will be orange.
2.3.3. Upper limit (Lower warn): When measurement is equal or less than lower warn, the
measurement color will be orange.
2.3.4. Upper limit (Lower lim): When measurement is equal or less than lower limit, the
measurement color will be red.
3. Calibration
3.1. From the Measure screen, press the Start button to begin measuring.
3.2. Single or dual laser non-thickness mode calibration:
3.2.1. Press the Calibrate button in the appropriate measurement pane.
3.2.2. Point the laser to the zero position and press OK.
3.2.3. Place an object of know measurement value under the laser and press OK, then enter the
known value.
3.2.4. Repeat steps 2.2.1 through 2.2.3 for the other laser (if applicable).
3.3. Dual laser thickness mode calibration:
3.3.1. Sum function mode
3.3.1.1. Press the Calibrate button in the measurement pane.
3.3.1.2. Place a calibration block between the lasers and press OK. Then enter the known
thickness.
3.3.1.3. You will then be asked “Would you like to sample an additional calibration
block?”. Choose Yes to sample another calibration block. Choose No to accept the

6
value of the first calibration block. Typically sampling a single calibration block is
sufficient. If greater accuracy is required, sample two calibration blocks.
3.3.2. Difference function mode
3.3.2.1. Press the Calibrate button in the measurement pane.
3.3.2.2. Point the both lasers at the zero position and press OK.
3.3.2.3. Place an object of known value on top the zero position and under only one laser
and press OK. Then enter the known value.
3.4. Slope and offset can be used to fine tune calibrated or non-calibrate setups. Offset (default 0)
will adjust the measurements up or down in an absolute manner by the number specified.
Slope (default 1) will adjust the rate of change between the raw laser values and measurement
value.
3.5. After calibration, press Stop, then press Save to save the current setup.
4. Measuring
4.1. Press and hold the green button on the bottom of the TPD to power up the unit. It will boot
into the Measure screen.
4.2. Press Start to start measuring.
4.3. Use the Tare button to set the zero point if necessary.
4.4. Use the +/-button to change the sign of the measurement value if necessary.
4.5. Press List to display a rolling list of the last 50 measurements.
4.6. Press Graph to display a rolling graph of the last 50 measurements.

7
5. Utilities
5.1. From the Settings screen, press the Utilities button to enter the Utilities screen.
Figure 4: Utilities screen.
5.2. Reset utility:
5.2.1.To erase all application settings, press Reset. A warning dialog will be displayed. If
accepted, the TPD will shut down. After restarting, it will be ready to re-configure.
Note: internal laser settings will not be reset.
5.3. Update utility:
5.3.1.Connect the TPD unit to a wired network with access to acuitylaser.com.
5.3.2.Press Update to update the Acuity Laser UI software via the internet.
5.3.3.If a newer version is available, instructions will be displayed regarding how to update.
Otherwise a message will be displayed indicating the software is up to date.
5.4. Copy data utility:
5.4.1.Press Copy data to copy all measurement data logs, application logs and the settings file to
a connected USB thumb drive.
5.4.2.Instructions will be displayed regarding when to connect and disconnect the USB thumb
drive. Warning! Do not disconnect the USB thumb drive until instructed to do so.
Otherwise damage could occur to the USB thumb drive.
5.5. Set clock utility:
5.5.1.Press Set clock to set the system clock. This is useful when using measurement data logs to
plot time vs measurement in a spreadsheet application.
5.6. Press Back to return to the Settings screen.

8
6. Troubleshooting
6.1. A range error indicates the subject is beyond the lasers range. Adjust the position of the laser
relative to the subject and zero point. Use the range indicator to ensure that maximum subjects
and the zero point are well within the lasers range.
6.2. Various errors displayed to the right of the Baud rate dropdown indicate a likely mismatch
between the Laser dropdown and the actual laser model. Change the Laser dropdown to match
the actual laser model. If errors persist after ensuring a match, power down the laser unit, wait
5 seconds, then power up the laser unit.
6.3. Application logs (out.log) and settings file (default.json) can be useful in diagnosing other
issues. Use the Copy data utility to copy out.log, out_<n>.log and default.json to a USB thumb
drive then send the files to customer support.
Table of contents