safety instruction
For safety operation pls. follow the safety instruction.
WARNING
set. Calibrate, inspect and fix the the weighing indicator is prohibited by
Non professional staff
WARNING
Pls. make sure the weighing display well earthing
WARNING
The indicator is electrostatic sensitive device, pls.
power off during electrical connections, internal
components touched by hand is prohibited, and please
take the anti-static measure

IDS701 User Manual
Page 2 of 25
LIST
1. Summary--------------------------------------------------------------------------3
1.1 Main function--------------------------------------------------------------------3
1.2 Optional function----------------------------------------------------------------3
1.3 Technical parameter------------------------------------------------------------3
1.4 Outline and installation drawing---------------------------------------------4
1.5 Battery using---------------------------------------------------------------------4
2. Installation------------------------------------------------------------------------5
2.1 Connection between indicator and load cell-----------------------------6
2.2 Connection of interface-------------------------------------------------------6
3. Basic operation-----------------------------------------------------------------6
3.1 Display and key instruction----------------------------------------------------6
3.2 Power on------------------------------------------------------------------------8
3.3 Zero function-------------------------------------------------------------------8
3.4 Tare function-------------------------------------------------------------------8
3.5 Accumulating function-------------------------------------------------------9
3.6 Print function-------------------------------------------------------------------9
3.7 Hold function-------------------------------------------------------------------10
4.Calibration-----------------------------------------------------------------------11
4.1 Enter and exit setting--------------------------------------------------------13
4.2 Calibration steps--------------------------------------------------------------15
4.3 Parameter setting-------------------------------------------------------------17
5. Output data format-----------------------------------------------------------18
6. Maintenance--------------------------------------------------------------------21

IDS701 User Manual
Page 3 of 25
1. Summary
IDS701 is specially designed for platform scale with friendly
interface, simple operation, steady feature. The Basic function
includes Weigh, Peak hold, Print. Communicate, Options are
accumulate, Count and animal weighing.
1.1 Main function
》basic weighing function: zero tare retare
》peak hold
》hold
》low battery remind charge and stop charge controlled
》PC communication
》Automatically power off
1.2 Optional function
》accumulating function
》conversion: kg/lb
》animal weighing
》printing function(with time)
1.3 Technical parameter
》Stimulating voltage:+5VDC
》A/D converting speed:10 times/sec
》load capacity:it can connect 4 pcs 350Ωload cell at most
》weight unit:kg. lb
》Resolution:3000e
》Interval:1/2/5/10/20/50
》Display:6-digits LED,word height: 20.3mm
》key:ON/OFF TOTAL TARE ZERO SET
》Interface:RS232C Baud rate optional 1200/2400/4800/9600

IDS701 User Manual
Page 4 of 25
》Ambient temperature:-10~40℃
》Storage temperature:-20~+60℃
》optional power:6V/4Ah rechargeable battery;9VDC adapter
1.4 Outline and installation picture
1.5 Battery
1. when you use the internal battery first time,you should charge
the battery 10-12 hours, to prevent low voltage resulted from self
leakage of battery.
2. when the red battery light is on and flashes,it means low
battery You should charge battery in time.
3. Charge time: 10-12 hoursAnd it works 45 hours
4. When the battery light turns green, it means fully charged
5. If you don’t use the battery long time, take out the battery to
protect t the indicator from battery leakage
6. In order to keep the battery in best using condition, it is suggest
that you fully discharge the battery every month, the method is that
using the indicator till it is automatically power off.

IDS701 User Manual
Page 5 of 25
2. Installation and Calibration
2.1 connection indicator with loadcell
IDS701 can connect four pcs 350Ωload cell at most, both four and
six wire load cell are ok. To make it simple, we use quick disconnect
Or standard plug. As belows
Quick disconnect connection
2.2Connection of interface
RS232 communication interface use 3 cores quick connector
RS232 communication interface connection

IDS701 User Manual
Page 6 of 25
3. Basic operation
3.1 Key and display
IDS701 Weighing indicator display instruction
LED display
instruction
Weighing data display
Total
accumulating

IDS701 User Manual
Page 7 of 25
Hold
Hold weighing data
Tare
Display tare weight
Net
Display net weight
Gross
Display gross weight
Display data keep still
zero, indicating zero weight
Battery
Using battery
Hi
Over setting weight
OK
Within setting weight
Lo
Below setting weight
。
Decimal point
Key’s function
Key symbol
Key name
Key function
SET
Work together with zero, tare, on/off to
perform all operations.
ZERO
1.Clear weight within zero range
2. Work together with SET to perform
Hold and animal-weighing operation.

IDS701 User Manual
Page 8 of 25
TARE
1.At Gross mode, tare the loaded
weight
2.At Net mode, display gross weight
after deduct tare
TOTAL
Work together with SET perform
accumulating operation
ON/OFF
1.Press it for 2 seconds to power on or
power off
2.Work together with “SET”to enter
calibration and function setting.
3.2 Power on
Power on and indicator perform self-checking and go to weighing
mode.
3.3 Zero setting
Within zero range, press “zero”, indicator weighing is cleared. When
Indicator is not stable, zero is unworkable.
3.4 TARE
Press , take the loaded weight as tare, display net
weight, Net weight is zero. “tare” “ net” status light is on.
At the Tare mode, Press , clear the tare, display gross weight.
3.5 TOTAL
Accumulation operation

IDS701 User Manual
Page 9 of 25
At Zero mode, load weight till stable, Press go to accumulating
Mode,” total” light on, display” n001”, and then display loaded weight;
unload weight , back to zero, load weight again till stable. Press ,
display”n002”
Then display the loaded weight. Repeat it maximum 999 times.
Check the total weight operation:
Press and hold it then press
At the same time, display ”n**”, (accumulating times) then display total
weight.
There are 8 data totally. It shows the first 4 digital. then the last 4 digital
For example, the first 4 digital is”0012”, the last 4 digital is”34,56”
It means the actual weight is “1234.56”
At TOTAL (accumulate)mode, Press display “ clr n”, it means don’t
clear the total Weight, Press exit it; if clear total weight,
Press “clr n” change to “clr y” it means clear total weight
display. Press to clear the the total weight and exit accumulating mode.
3.6 Print function
When the data is stable, connection with printer, it will be printed after
press“set”1 second.

IDS701 User Manual
Page 10 of 25
Note:print the gross weight when at tare mode , if the net weight is zero.
Can not print.
3.7 Hold function
There are two different hold function. Peak hold function and data
Hold function.And the setting is different accordingly.
C11=1 Peak hold C11=2 Data hold C11=0 no hold function
Peak-hold: display the maximum weight. At weighing mode, Press
still, then press , display the “lock” weight. “hold”
Light is on.At that time, you load or unload stuff, the weight keep still
Data hold: display the “lock” weight. Press still, then press
, display the “lock” weight. “hold” Light is on.At that time, you load
or unload weight, the weight keep still. At “hold” mode, Press
hold it and then press to exit hold mode. “hold” light is off.
3.8 10 times high resolution
At weighing mode. Press hold it and then press
There is one more digital display. It is 10 times high resolution display

IDS701 User Manual
Page 11 of 25
4. Calibration and parameter setting
4.1 Enter setting
There have two methods to enter the setting menu:
1. when the “span” is not pressed down,
Press Press still and then press enter C08-C39
setting.
2. Take out the sealing screw on the back of indicator, then press
down the “span” Press still and then press enter
C01-C39 setting.
The key functions in setting:
Enter
Up
Down
Left
Power switch. exit setting

IDS701 User Manual
Page 12 of 25
4.2. Step of calibration operation:
According to the second method which can enter setting menu,
C01-C39
step
Method of operation
display
Remark
1
[C01 ]
After you enter to calibration
mode, it display [C01 ]
2
press
[C1 1]
Weight unit
option:1=kg
2=lb
3
press
press
press or
[C02 ]
[C02 0]
[C02 2]
Set decimal digits
option:0/1/2/3/4
Select decimal digit
example:two decimal point:
[C02 2]
4
press
press
press or
[C03 ]
[C03 1]
[C03 5]
Set graduation
option:1/2/5/10/20/50
Select required graduation
example:graduation 5:[C03 5]
5
press
press
press or /
[C04 ]
[0100.00]
[0100.00]
Max capacity
example:max weighing 100kg:
[0100.00]
6
press
press
press
press
[C05 ]
[C05 0]
[C05 1]
[CAL 9]
。。。。。。
Zero calibration
option :0=non-calibration zero
1=need calibration zero
calibration zero please choose 1 and
ensure scale is empty and “stable”

IDS701 User Manual
Page 13 of 25
[0000.00]
light is on
Ensure zero calibration, countdown.
Till show[0.00](example for two
decimal point)。
7
press
press
press or
press
press or
press
[C06 ]
[C06 0]
[C06 1]
[SPAN ]
[0100.00]
[0080.00]
[CAL 9]
……
[0080.00]
[CALEnd]
Loading calibration
option:
0=Non-load calibration
1= load calibration
Basic on max capacity setting,
add suitable weight on scale.
close to the max capacity,
heavier than 10% max at least.
For example: the weight is 80kg
As bellows:
enter loading calibration, count
down over, indicator shows
loaded weight , loading
calibration finish.
If you want to set application
Function parameter. Press
“PRINT”if you want to exit press
“TOTAL”
8
press
press
press or
[C07 ]
[07 0]
[07 1]
Default parameters setting
option:0=non-restore default
parameters
1=restore default parameters
Note: after the above

IDS701 User Manual
Page 14 of 25
parameters setting finish, please
do not set default parameters
often, avoid the original setting
parameters lost.
4.3 Application function parameters setting chart
Function
Setting
Item
parameters setting and instruction
warning
tone
C08
warning
tone
Options: 0 = close warning tone
1 = open warning tone
Automatic
power off
C09
Automatic
power off
option:0=close auto power off
10= keep still within 10 min. power off
automatically
30= keep still within 30 min. power off
automatically
60= keep still within 60 min. power off
automatically
Power saving
setting
C10
Power saving
setting
option:0= close power saving setting
3= keep still within 3 min. stop display
5= keep still within 5 min. stop display
Hold
function
C11
Hold mode
option:0=close hold function
1=Peak hold /2=Data Hold
instruction:
Peak :it shows the max. data, mainly
application for materials testing, such as
tension and pulling force.
Hold:it shows current weight value. Mainly
application for animal weighing.

IDS701 User Manual
Page 15 of 25
Kg/lb
conversion
C12
Kg/lb converstion
C12=0 kg/lb convert is ok
C12=1 kg/lb convert is unworkable
Upper/lower
limit alarm
C13
Upper limit alarm
value
You can set it within the max.
capacity limit
C14
Lower limit alarm
value
Inner Code
display
C15
Check inner code
At setting function mode, after
directly enter C15,indicator will
show inner code
Date and time
C16
Date
Enter C16, you can direct to set
the current date, from left to right:
year/month/day
C17
Time
Enter C16, you can direct to set
the current date, from left to right:
year/month/day
Communication
setting
C18
Serial interface
data output
method
option:0= Close serial interface
data output
1= Continuous sending, connect
big display
2=print method, connect printer.
3= Command request method ,
connect computer.
4=PC continues to sending
format, connect computer.
C19
option:

IDS701 User Manual
Page 16 of 25
Baud rate
0=1200/1=2400/2=4800/3=9600
Zero range
C20
Manually zero
range
option:0= close manually zero setting
1=±1% max capacity
2=±2% max capacity
4=±4% max capacity
C21
Initical zero range
option:0= no initial zero
1=±1% max capacity
5=±5% max capacity
10=±10% max capacity
20=±20% max capacity
Automatic
zero tracking
C22
Automatic zero
tracking range
option:0.0= close automatic zero
tracking
0.5=±0.5d
1.0=±1.0d
2.0=±2.0d
3.0=±3.0d
4.0=±4.0d
5.0=±5.0d
remark:1.d is the set graduation
2、Automatic zero tracking
range can not exceed manual
zero setting range
C23
Automatic zero
tracking time
option:0= close automatic zero
tracking time
1=1 second
2=2 seconds
3=3 seconds
Overload
C24
option:00= close overload range

IDS701 User Manual
Page 17 of 25
range
Overload range
01d~99d
remark:d is the setting graduation
(division)
Negative
display
C25
Negative display
Option:0=-9d
10=10% max. capacity
20=20% max. capacity
Standstill
setting
C26
Standstill time
option:0= quick
1= medium
2= slow
C27
Standstill range
option:1=1d
2=2d
5=5d
10=10d
Note: d=division
Digital filter
C28
Dynamic filter
Instruction :
Dynamic filter is
collecting the data
filter before loaded
weight stable.
When loaded
weight easily
shaking (for
example animal) ,
you can set this
filter to make
weight display
more stable
option:0= close dynamic filter
1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength
4=4 digital filter strength
5=5 digital filter strength
6=6 digital filter strength
Note:Pls setting dynamic filter
strength carefully, the No. is
bigger, more stable. if the
loaded weight shake not too
much. The setting is less
than 3

IDS701 User Manual
Page 18 of 25
C29
Noise filter
option:0=close noise filter
1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength
C30
Print time and date
C30=0 yy.mm.dd
C30=1 mm.dd.yy
C30=2 dd.mm.yy
C30=3 yy.mm.dd
5.Output data format
5.1Computer continuous sending format
,
,
CR
LF
S 1 S 2 S 3 Data S 4
S1: weight status, ST= standstill, US= not standstill, OL= overload
S2: weight mode, GS=gross mode, NT=net mode
S3: weight of positive and negative, “+” or ” –“
S4: measurement unit, “kg” or “lb”
Data: weight value, including decimal point
CR: carriage return
LF: line feed

IDS701 User Manual
Page 19 of 25
5.2 Big display continuous sending format
Output continuous format
S
T
X
S
W
A
S
W
B
S
W
C
X
X
X
X
X
X
X
X
X
X
X
X
C
R
C
K
S
1
2
3
4
5
6
State A
Bits0,1,2
0
1
2
Decimal point position
1
0
0
XXXXXX0
0
1
0
XXXXXXX
1
1
0
XXXXX.X
0
0
1
XXXX.XX
1
0
1
XXX.XXX
Bits3,4
Division
0
1
X1
1
0
X2
State B
BitsS
function
Bits0
gross=0,net=1
Bits1
symbol:positive =0,negative =1
Bits2
overload(or lower zero)=1
Bits3
dynamic=1
Bits4
unit:lb=0, kg=1
Bits5
Constant 1
Table of contents
Popular Measuring Instrument manuals by other brands

Tenmars
Tenmars TM-730 user manual

IOtech
IOtech 655u user manual

Armada Technologies
Armada Technologies Pro50 instruction manual

King Instrument
King Instrument 7650 Series installation instructions

Electronic Technologies
Electronic Technologies ES2010E user manual

Nikon
Nikon Forestry Pro II instruction manual