Electronic Technologies ES2010E User manual

-0-

-1-
CONTENT
ISafetyPrecautions and Procedures ………………………………………2
II Introduction …………………………………………………………………3
III.Basic Function Introduction ……………………………………………4
IV ElectricalSymbols …………………………………………………………4
VTechnical Specifications……………………………………………………5
1 Basic WorkingCondition……………………………………………………5
2Rated working conditions……………………………………………………5
3 General Specification………………………………………………………5
4Performanceindicators………………………………………………………7
VI Structure………………………………………………………………………8
VII Operation………………………………………………………………………8
1Switchon/off …………………………………………………………………8
2Data Hold、Cancel、Storage…………………………………………………8
3Data Review、Exit …………………………………………………………9
4Data Delection ………………………………………………………………9
5Test Display Mode Switching………………………………………………9
6Test……………………………………………………………………………10
VIII Chargingbattery…………………………………………………………13
IX OtherExplanations and Considerations………………………… 13
XAccessories……………………………………………………………… 14

-2-
ISafetyPrecautionsAnd Procedures
Thankyouforpurchasingour digitaldoubleclampphase voltammeter.Before
youusetheinstrumentforthefirsttime,inordertoavoidpossibleelectric
shockor personal injury, please be sure to: read and strictlyobserve the safety
rules and precautions listed in this manual.
uThisinstrumentisaliveworkingequipment.Foryoursafety, pleasecomply
withtherelevantnationalsafetyregulationsandstrictlyfollowtheelectric
safety work procedures.
uPayattentiontothetextandsymbolsoftheinstrumentpanelandbackplate.
uBeforeuse,makesurethattheinstrumentandaccessoriesare ingoodcondition
and thattheyare not damaged,bare orbroken.Donotuseitwhenthe instrument
backcoverandbatterycover arenotcoveredproperly.Whenthe instrumentis
inuseandtheenclosure ortest wireisbroken andthemetalisexposed,please
stopusingit.
uCannot be used to test voltages higher than 600V. Do not use under strong
electromagnetic environment to avoidaffectingthe normal operation ofthe
instrument. Do nottest in flammable and hazardousareas.
u
Makesure the connecting plugofthewireis tightlyinserted intotheconnector.
Please pay attentionto the direction whentestingthephase. Thetest line must
beevacuatedfromthemeterbefore beingpulledoutofthemeter.Donot touch
the input jack toavoid electric shock.
uDonotoperate the instrumentwith wet handsor exposingit to rain.Donot
use theinstrumentwhen it is wet.
uDonotplaceandstoretheinstrumentforlongperiodsoftimeunderhigh
temperature and humidity, condensation, and directsunlight.
uKeepthesurfaceoftheproductcleananddry.Donotcleanitwithcorrosives
orcrudematerials.Useasoftcloth(suchaseyeglasscloth),dipalubricant
thatpreventsrustand dehumidification, andgentlywipe the currentclamp.
uDonothitor drop the instrumenttoavoid impact on the instrumentand the
currentclamp, damagethe instrument, especially theJointsurface ofjaw.

-3-
uPay attention to the measuring range and use environment specified by this
instrument.
uIt is forbidden to use, disassemble, calibrate and repair thisinstrument
withoutpermission. It mustbeoperated by authorized personnel.
uThe""inthe manualisasafetywarningsign,andtheusermustoperate
safely in accordance with the contentsofthis manual.
uThe""andotherdangeroussignsinthemanualmustbefollowedbythe
userinstrict accordance withthecontents of thismanual.
II Introduction
Digitaldoubleclamp phase voltammeter isthe latestproductofourcompany's
clamp-typemeterseries.Itisa hand-held, dual-channel input all-digital,
multi-function, high-precision,intelligent multi-parameterfrequency
measurementinstrument.The instrument usesthe latestmicroprocessortechnology
and digital signal processingtechnology toachieve powerfrequencymeasurement
(suchas voltage,currentRMS,activepower, reactivepower,apparentpower,power
frequency,powerfactor,phase relationship, etc.),Determine thetransformer
wiring group,inductive,capacitivecircuit,testthesecondarycircuitand bus
differential protection system, read out thephaserelationship between the
differential protection CT groups, check the power meter wiring is correct or
not,themeteruses theclamp-typecurrenttransformer conversionmethod toinput
the measured current, so there isno needtodisconnectthecircuit undertest.
Provideasafe,accurate,and convenientnew typeof powermeterforelectricity
inspectionpersonnel.
Digitaldouble clamp phase voltammeter is also known asintelligent
double-clampdigital phasevoltmeter, multi-function double-clampdigital phase
volt-ampere meter, etc. It issuitableforelectricpower,petrochemical,

-4-
metallurgy,railway,industrialandminingenterprises,scientific research
institutions,andmeasurementdepartments.Especiallysuitableforenergy
billingsystems and relay protection systems.
III.BasicFunction Introduction
1. Largescreenhighbacklightdisplay,cleardisplay oftheinstrument'sworking
status andtest parameters,the operationisvery convenient.
2. Measure twowire’svoltage, current, phase,frequency, activepower,reactive
power, apparent power, power factor, and total power.
3. Measuregrid frequency
4. Low-current5mAmeasurementphaseis very suitablefornewusersto checkthe
connection statusafterno-load operation, the instrument has highmeasurement
accuracy and widemeasurement range.
5. Automatic identification of transformer windings, capacitive and inductive
loads.
6. Static data storage function, cansaveupto500 sets of data
IV Electrical Symbols
Extremely dangerous! The operator must strictly abideby thesafety
rules, otherwise therewill bedanger ofelectric shock and cause
personal injury orcasualties.
Danger! Theoperator muststrictlyabideby the safety rules,
o
therwise there will be dangerofelectricshockandcausepersonal
injury or casualties.
Caveat!The safetyrules mustbestrictlyobserved,otherwise
personal injury orequipment damagemay result.
AC
DC
Double insulation

-5-
VTechnicalSpecifications
1BasicWorking Conditions
1) Ambienttemperature:(23±5)℃
2) Ambienthumidity:(45~75)%RH
3) Measuredsignalwaveform: Sinewave、β=0.02
4)Measuredsignalfrequency:(50±0.2)Hz
5)The position ofthecurrent-carrying conductorundertest in thejaws:Centered
6)Currentamplitudewhen measuringpower、power factor:1A±0.2A
7)Voltage amplitudewhen measuringpower、power factor:100V±20V
8)External referencefrequencyelectromagneticfieldinterference:shouldbe
avoided
2.Ratedworkingconditions
1) Ambienttemperature:(-15~+45)℃
2) Ambienthumidity:(0~90)%RH
3)Altitude: lessthan1500meters
4) Measuredsignalwaveform: Sinewave、β=0.05
5)Measuredsignalfrequency: 45-65Hz
6)Currentamplitudewhen measuringpower、power factor:5mA~20A
7)Voltage amplitudewhen measuringpower、power factor:10V~600V
8)The positionofthecurrent-carrying conductor under test in thejaws:Anywhere
3.GeneralSpecification
Model No. ES2010E
Function
Measure twoAC voltages, currents, phasebetween voltages,
phasebetween
currents,phasesbetweenvoltagesandcurrents,
frequencies, activepower, reactivepower,apparentpower,

-6-
powerfactor, determinetransformerwiring groups,inductive,
capacitivecircuits,testsecondaryloopsand busbar
protectionsystems,readoutthe
phaserelationship between
the CTsofthe differentialprotectiongroups,check the
wiring of thepower meter, check the lineequipment,etc.
Power DC6VNi-MHrechargeable battery
Power
consumption Turnonthebacklightandconsumeupto50mA.The batter
ycan
workcontinuouslyformore than 12hours.
Displaymode LCD display,62×44mm
Instrument
size L/W/T:192*92.5*36mm
Voltagerange
AC 0.00V~600V
Currentrange
AC 0.0mA~20.0A
Phase range 0.0°~360.0°
Frequency
range 45.00Hz~65.00Hz
Active power
range 0.0W~12kW
Reactivepower
range 0.0W~12kVAR
Apparentpower
range 0.0W~12kVA
Power factor
range -1~+1
Resolution
Voltage:AC 0.01V
Current:AC 0.1mA
Phase:0.1°
Frequency:0.01Hz
Active power:0.1W
Reactive power:0.1VAR
Apparent power:0.1VA
Power factor:0.001
Detection rate
About 2seconds/time
Data hold Press HOLD key tohold data during test, " HOLD"symbol display
Data storage 500 groups
USB interface
USB interface,uploadingdata toacomputer for analysisof
managementdata
Automatic
About15minutes afterpoweron,themeter willautomatically

-7-
shut-down shutdown to reduce batteryconsumption
Backlight
function Suitable for dimplaces andnightuse
Voltage
detection
When the battery voltage islower than5.2V, the low battery
voltagewill display,remind to chargeintime
Instrument
weight
Hose:550g(including battery)
Tip-shapedcurrentclamp:170g×2
Testline:250g
Test line
length 1.5m
Line length of
the current
clamp 2m
Working
temperature
and humidity -10℃~40℃;below80%Rh
Storage
temperature
and humidity -10℃~60℃;below70%Rh
input
impedance Testvoltage inputimpedance is:2MΩ
Pressure
resistance
The withstandvoltageof1000V/50Hz sinewaveACvoltage
betweentheinstrumentcircuit and thehousinglasts 1minute
insulation Betweeninstrumentline andhousing≥100MΩ
c Double insulation
Suitable for
safety
regulations
IEC61010-1CAT Ⅲ600V,IEC61010-031,IEC61326,
Pollution
level 2
4Performance indicators
Class type Range Resolution Intrinsic error
Voltage AC 0.00V~600V 0.01V ±(0.5FS)
Current AC 0.0mA~20.0A 0.1mA ±(0.5FS)
Phase 0.0°~360°0.1°±1°
Active power 0.0W~12kW 0.1W ±(3%rdg+3dgt)
Reactive power 0.0VAR~12kVAR 0.1VAR ±(3%rdg+3dgt)
Apparent power 0.0VA~12kVA 0.1VA ±(3%rdg+3dgt)

-8-
Frequency 45HZ~65HZ 0.01HZ ±0.1Hz
Power factor -1~+1 0.001 ±0.03
VI Structure
1.Twowire’svoltage input interface2.LCD display3.hull
4.Functionkeyarea 5.Twowire’scurrentinput interface
6.Currentclampplug 7.Cuspcurrentclamp 8.Voltageinputtestline
VII Operation
Beforeuse,carefullycheck allpartsoftheinstrumentfordamage,
and use themwithoutanydamage.
Itisforbiddentouse thisinstrumentinhazardousareas
1Switchon/off
PressthePOWER button to turn on, the LCDdisplay. Press the POWER button
to shutdown,themeter will automatically shutdownafterabout15minutes.
2Data Hold、Cancel、Storage
HoldtheHOLDkeyinthetestmodetomaintainthedisplaydata,the"H0LD"
symbolindicates,"MEM"flashonetime,pressHOLDagaintocancelthehold.While

-9-
maintainingthedata, the meter willautomatically number and store the currently
helddata,displaythegroupnumber suchas “NO:002”,andthemeter can store
upto500sets of data.If thestorageis full, the “FULL”symbol is displayed.
3Data Review、Exit
Pressthe READ keyin the test modetoenter the datareference mode, the
"MR"symboldisplays, and start reading fromthe" NO:001"group data.Press to
viewthe increment bystep,and presstoincrease theincrementby10steps.
Pressthe key to decrement the step sizeby1, andpressand holdthekey
forastepsizeof10todecrement.PresstheMODEkeytoswitchtoviewthevoltage,
currentandphase,activepower,reactivepower,apparent power, powerfactor,
frequency,and saveddisplay interface.Press READ toexitdatareviewmode and
return to test mode.
4Data Delection
Press HOLD key to enterthe datadeletionoption in data reviewmode or
testmode, press key toselect “YES”, press key toselect “NO”,
and returntotestmode after selection.After powero
After power on, long press the HOLD button to control the backlight, suitable
for dimplaces.
5Test Display Mode Switching
Afterpower on, the meter automatically enters thevoltage and current test
displaymode(Figure1);pressthe MODE keyor pressthe and keysto
switch the phase, active power, reactive power, apparent power, power factor,
frequency,test display mode(Asshown inFigure 2,3),andfigure (3) two-line
power, power factor is U1I1, U2I2correspondingpower and power factor.

-10-
6Test
Danger:Electricity! Mustoperatedby thequalifiedpersonnelthat
trained andauthorized .The operator must strictly abidebythe
safetyrules,otherwisethereisadangerofelectricshock,
resulting inpersonal injury or equipmentdamage.
D
anger!Cannot beused tomeasure voltagelinesexceeding 600V,
otherwise there is adangerof electric shock, resultingin personal
injury or equipment damage.
Danger! Cannotbe used to measure thelines more than 20A, may damage
the equipment.
Muststrictlyfollowthemanualinstructionstoconnect, I1,I2cannotbe
inserted opposite
Afterthetest iscompleted,thetestlinemustbeevacuated fromtheline
beforeremoved fromthemeter.
The phasetestrelationship ofthisinstrument: U1U2、I1I2、U1I1、U2I2,
the phaseofeachrelation measuredis the phaseof1signal advanced 2
signals.
U1,U2voltage jacks andcorresponding currentclampupward right
arroware thesame.
The directionofthecurrentinputclampin phasetestingisthesameas
the directionofthearrowsign ontheclamp.
Thisinstrumentcantest ACvoltage,current,phasebetween voltages,phase
between currents,phasebetween voltage and current,frequency,activepower,
reactivepower,apparentpower,power factor,inductiveandcapacitivecircuitsof
twolines.
Test wiring isasfollows:
Single-phasetest:The testedvoltagelinesLandNcorrespondtotheU1red
and COMblackjacksoftheaccess instrument, andthecurrent clampI1clampsthe

-11-
Llineundertest totest thesingle-phaselinevoltage, current,phase, frequencyand
power parameters, etc. Seebelow. CanalsouseU2Red,COM, I2channeltest
Whenmeasuringthree-phasethree-wire and three-phasefour-wire,the
inductivecapacitiveload andpolaritycanbe judged accordingtovariousphase
relationships. Ifthe U1I1phaseisdisplayed in the rangeof00 to900, themeasured
load is inductive,and if it is displayed in the range of 2700to3600. The measured
load iscapacitive;ifthe displayphaseiscloseto1200,the polarityisthesame;if
the displayphaseiscloseto1200and3000,thepolarityisreversed (currentclamp
or linewiring maybe reversed).
Three-phasethree-wire test:the testedvoltage linesUAyellow,UBgreen,and
UC red will be connected tothe U1red,COMblack,and U2redjacksofthe
instrument.The currentclampsI1and I2will clampthemeasuredcircuitIAand IC
accordingly,three-phasethree-wire voltage,current,phase,frequency,power,
power factor,etc.can betested.Seewiring figurebelow.

-12-
Three-phasefour-wire test:Thewiring intwostepscan testthree-phase
four-wire voltage,current,phase,frequency,power,power factor,etc.The
correspondingwiring is shownin the following table andfigure.
Step1 Step2
Three-phase
four-wireline
Instrument Three-phase
four-wire line
Instrument
UA yellow U1 red UB green U1 red
UB green U2 red UC red U2 red
Nblack COM black Nblack COM black

-13-
VIII Charging Battery:
The batteryisa6VNiMH batterypack. Whentheinstrumentpower supply
voltage islower than5.2V,thebatterysymbol isdisplayed,indicatingthatthebattery
power is insufficient. Pleasechargeitintime.
IX Other Explanationsand Considerations
1Currentclampspecificity
The twocurrentclampsofthemeterarededicated tothismeter and cannotbe
used for another meter,Atthe sametimethe order ofthetwoclampscannotbe
interchanged,otherwisetheerrorwill increase.Currentclamptopreventthe
impact,thejawplane must be keptclean,completelyclosed test isreliable.
2. Currentclamp maintenance
Afterthecurrentclampisused,thedustintheplaneofthejawsshouldbe
promptly removed.The planeofthejawsshouldnotbecleaned withroughobjects
or corrosion.Itisbesttouseasoftcloth(suchas:WD-40 lubrication) togently
wipe.
3Thisinstrumentisusedforthedetectionofsecondaryloopsandlowvoltage
loops.Itcannotbeusedtomeasurethecurrentinhighvoltage linestoprevent
electric shock.
4Three-phasefour-wire (thephasewhen three-phaseload is balanced):
Phase
relationship
Phase
value
Phase
relationship
Phase
value
Ua-Ub 120 Ia-Ib 120
Ub-Uc 120 Ib-Ic 120
Uc-Ua 120 Ic-Ia 120
5Three-phasethree-wire (the phase whenthree-phase loadis balanced):
Phase
relationship
Phase
value
Phase
relationship
Phase
value

-14-
Uab-Ucb 300°Ia-Ic 240°
Uab-Ia 30°Ucb-Ic 330°
Ifthecurrentclampisreversed orthecurrentlineisreversed,the
displayed phasevalue will be 180°outofphase,whichisan increase
of 180°fromthe abovestandard value.
XI Accessories
Host 1PC
Instrumentbox 1PC
Currentclamp 2PCS
Testline 4(2eachforred andblack)
Battery 6V Ni-MH batterypack(built-in)
charger 6V NiMH BatteryCharger
Manual,certificate 1SET
CD 1PC
Dataline 1PC

-15-
The contentsofthisusermanualcannot be used asareason tousethe product for specialpurposes.
The companyisnot responsiblefor other losses caused byuse.
The companyreserves the righttomodifythe contentsofthe usermanual.Ifthereisanychange,itwill notbe
notified.
Table of contents
Popular Measuring Instrument manuals by other brands

Veeder-Root
Veeder-Root TLS RF Wireless 2 Installation and maintenance guide

Laser Products
Laser Products LT-2D3D user manual

Gammon
Gammon GTP-9330 Series Installation and operation

Philips
Philips VOICETRACER 7790 quick start guide

Apera Instruments
Apera Instruments TN480 instruction manual

ILLKO
ILLKO GIGATESTpro user manual

Keithley
Keithley 6514 instruction manual

Hanna Instruments
Hanna Instruments HI 99171 instruction manual

iKonavs
iKonavs DRS manual

AutomationDirect
AutomationDirect ProSense FG1 Series user guide

Chauvin Arnoux
Chauvin Arnoux ORITEL RW 521 user manual

Endress+Hauser
Endress+Hauser Micropilot S FMR540 quick start guide