OM-TZ493E 2016.00.00 00 Y
Coupling
Handpiece connection
Powder case cap
Powder case
Before use, thoroughly dry the coupling, handpiece, hose
joint section of the powder case, inside of the powder case,
and inside of the powder case cap.
AIR AIR
AIR
AIR
AIR
AIR
Attaching the Nozzle Tip ( Refer to the Operation Manual 6-6 )
Securely attach the Nozzle Tip to the handpiece nozzle by referring to the
figure below. After attaching, lightly pinch the Tip and check it is not
removed from the handpiece nozzle. If the Tip is attached incorrectly, the
Tip will come off and accidental ingestion may result.
Removing the Nozzle Tip ( Refer to the Operation Manual 6-6 )
After treatment, remove the Nozzle Tip from the handpiece using the
Nozzle Tip Remover and dispose of it.
Nozzle Tip
Nozzle Tip Remover
Attaching and Removing the Nozzle TipBefore use
AIR
AIR
Discard of the cleaning powder ( Refer to the Operation Manual 7-1~7-2 )
• Cleaning of smartGun PERIO after use and regular maintenance are required.
( Failure to do so may cause powder blockage, water leakage, etc. )
• The Nozzle Tip is a single use product. Do not reuse or resterilize it.
• If powder remains in the Powder Case or Handpiece, it will solidify which may cause clogging.
• Before use, thoroughly dry the coupling, handpiece, hose joint section of the powder case, inside of the powder case, and
nozzle. Also, perform idle running with air alone to expel any internal water.
• Make sure you clean the smartGun PERIO in an ultrasonic bath once a week to prevent the instrument from clogging ( refer to
the Operation Manual 8-1 ).
Caution
12
After removing the
cleaning powder in
the powder case,
perform idle running
with air alone to
discharge water
and powder in the
nozzle.
Blow off the powder
remaining in the
powder case with air.
Fit the supplied
blower nozzle to
the coupling and
attach to the
handpiece. Run idle
with no water for
more than 5 sec. to
clear lines.
Use the blower
nozzle in the same
way to clear the lines
in the smartGun
PERIO nozzle. Blow
from the far end of
the nozzle.
1
3
2
Ultrasonic cleaning ( Refer to the Operation Manual 8-1 )
Be sure to do this once a week or after operating Troubleshooting.
Sterilization ( Refer to the Operation Manual 8-2 )
Dry the O-ring section
of the powder case
and the Adjustment
Ring section, apply a
drop of the supplied
turbine oil and rub it
in with your fingers
( refer to the Operation
Manual 7-4 ).
Carefully clean the
inside of the joint
section of the
handpiece and the
O-ring section of
the powder case
with the supplied
cleaning brush.
If powder adheres
to the O-ring
section, remove the
O-rings, and
carefully clean the
O-rings and
grooves.
Remove the rubber
packing and
disassemble the
powder case cap.
Rinse under
running water.
Perform ultrasonic
cleaning with hot
water ( refer to
the Operation
Manual 8-1).
Disassemble the powder case cap and rinse under
running water.
All the parts are autoclavable. Perform autoclave
sterilization up to 135˚C after cleaning.
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C,
or 3 min. at 134˚C.
※
In case the sterilizer chamber temperature may exceed
135˚C during the drying cycle, skip the drying cycle.
※Blower nozzles for other brands couplings
are also available.
If you fail to lubricate the O-ring section, the powder case and
handpiece may no longer swivel freely and water may leak.
Be sure to remove the cleaning powder and clean it prior to ultrasonic cleaning.
Rubber packing
Care After Use
HandpieceBlower nozzle Blower nozzle
Nozzle
Coupling
When idling, hold the
Nozzle and Handpiece
tight.
1
Cleaning ( Refer to the Operation Manual 7-3~7-4 )
2
3
34
AIR
After removing the solidified powder, perform ultrasonic cleaning and autoclave sterilization.
2
1
Water Leakage ( Refer to the Operation Manual 7-3~7-4 )
Replace the O-ring(s)
at the handpiece joint
section of the powder
case, apply a drop of
the supplied turbine oil
and rub it in with your
fingers.
Troubleshooting
Powder Clogging
Nozzle Powder Case
Cleaning wire (small)
Cleaning wire (large)
Cleaning wire (large)
Internal Nozzles
Handpiece
Cleaning file
Cleaning file
• Basic maintenance methods after use are described in the Quick Maintenance
Manual. Refer to the operation manual for more detailed instruction.
• Keep the Operation Manual and the Quick Maintenance Manual for future
reference.
• Perform ultrasonic cleaning once a week to prevent clogging of powder. (Refer
to the Operation Manual 8-1)
• Thoroughly dry the inside of the Powder Case and the connection part after
each maintenance process.
Caution
( Refer to the Operation Manual 8,10 )
Remove the Nozzle from the Handpiece and clean the inside of the Nozzle and Handpiece with the supplied Cleaning Wires
and Cleaning File. In the same way, clean the Handpiece Joint Section and the Internal Nozzles of the Powder Case.
To avoid bending, do not apply excessive
force to the Cleaning wire or file during use.
OMNIDENT Dental-Handelsgesellschaft mbH
Gutenbergring7-9 D-63110 Rodgau
[email protected] www.omnident.de

Popular Dental Equipment manuals by other brands

BRYANT DENTAL Ignis Gesture A-494322 manual

BRYANT DENTAL

BRYANT DENTAL Ignis Gesture A-494322 manual

BDS C-4250 Service manual

BDS

BDS C-4250 Service manual

DENTSPLY AutoMatrix Snippers+ Instructions for use

DENTSPLY

DENTSPLY AutoMatrix Snippers+ Instructions for use

Ugin PC 30 Instructions for use

Ugin

Ugin PC 30 Instructions for use

Xo FLEX installation instructions

Xo

Xo FLEX installation instructions

3M Elipar DeepCure-S operating instructions

3M

3M Elipar DeepCure-S operating instructions

KaVo ESTETICA E50 Life Instructions for use

KaVo

KaVo ESTETICA E50 Life Instructions for use

Great Lakes Rhea RCU instructions

Great Lakes

Great Lakes Rhea RCU instructions

Durr Dental VistaScan Mini View Installation and operating instructions

Durr Dental

Durr Dental VistaScan Mini View Installation and operating instructions

Aseptico AEU-525S Operation & maintenance manual

Aseptico

Aseptico AEU-525S Operation & maintenance manual

Xo 4 troubleshooting guide

Xo

Xo 4 troubleshooting guide

Renfert Silent compactCAM Translation of the original instructions for use

Renfert

Renfert Silent compactCAM Translation of the original instructions for use

Vaniman 10660 quick start guide

Vaniman

Vaniman 10660 quick start guide

Ivoclar Vivadent Bluephase Style Instructions for use

Ivoclar Vivadent

Ivoclar Vivadent Bluephase Style Instructions for use

SybronEndo Elements Instructions for use

SybronEndo

SybronEndo Elements Instructions for use

Medidenta Freestyle Pro instruction manual

Medidenta

Medidenta Freestyle Pro instruction manual

Aseptico AEU-14CF OPERATION, SERVICE MANUAL & PARTS LIST

Aseptico

Aseptico AEU-14CF OPERATION, SERVICE MANUAL & PARTS LIST

DeguDent SOLARIS Instructions for use

DeguDent

DeguDent SOLARIS Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.