Important Safety Information
Electric shock hazard:
•Donotapplymorethantheratedvoltagebetweenanytwoinput
terminals, or between any input terminal and earth ground.
•Donotcontactthetestleadtipsoranyuninsulatedportionofthe
accessory.
•The test leads supplied with this product comply with IEC 61010-
031:2008,UL61010-031:2010andCAN/CSA-C22.2NO.61010-
031A-07–Amendment 1:2010. These safety standards limit the
exposed length of the probe tip to 4 mm for measurement category
III. These test leads include a cap that must be in place when used in
measurement category III or IV applications. Refer to Fig. 1.
Failure to observe these warnings could result in severe injury or death.
Electric shock hazard:
•Donotattempttorepairthisunit.Itcontainsnouser-serviceableparts.
•Donotexposetheunittoextremesintemperatureorhighhumidity.
Refer to “Specifications.”
Failure to observe these precautions may result in injury and can damage
the unit.
Cap can be removed for measurement category I or II applications.Cap must be in place for measurement category III or IV applications.
Figure 1
Lifetime Limited Warranty
Greenlee Textron Inc. warrants to the original purchaser of these goods for use
that these products will be free from defects in workmanship and material for
their useful life, excepting normal wear and abuse. This warranty is subject to
the same terms and conditions contained in Greenlee Textron Inc.’s standard
one-year limited warranty.
For all Test Instrument repairs, contact Customer Service at 800-435-0786 and
request a Return Authorization.
Foritemsnotcoveredunderwarranty(suchasitemsdropped,abused,etc.),a
repair cost quote is available upon request.
For technical assistance: 800-435-0786
Specifications
Measurement Category: Per UL 61010-031 and IEC 61010-031
With Caps: CAT IV 1000 V
Without Caps: CAT II 1000 V
Operating Conditions:
Temperature: –10°Cto50°C(14°Fto122°F)
RelativeHumidity(non-condensing): 80%maximumfortemperatures
upto31°C(88°F),decreasinglinearlyto50%maximum
at50°C(122°F)
Altitude: 2000m(6500’)maximum
Indoor use only
StorageConditions: –20°Cto60°C(–4°Fto140°F),
0%to80%relativehumidity(non-condensing)
Pollution Degree: 3
Measurement Categories
These definitions were derived from the international safety standard for
insulation coordination as it applies to measurement, control, and laboratory
equipment. These measurement categories are explained in more detail
by the International Electrotechnical Commission; refer to either of their
publications: IEC 61010-1 or IEC 60664.
Measurement Category I
Signal level. Electronic and telecommunication equipment, or parts thereof.
Some examples include transient-protected electronic circuits inside
photocopiers and modems.
Measurement Category II
Local level. Appliances, portable equipment, and the circuits they are
plugged into. Some examples include light fixtures, televisions, and long
branch circuits.
Measurement Category III
Distribution level. Permanently installed machines and the circuits they are
hard-wired to. Some examples include conveyor systems and the main
circuit breaker panels of a building’s electrical system.
Measurement Category IV
Primary supply level. Overhead lines and other cable systems. Some
examples include cables, meters, transformers, and other exterior
equipment owned by the power utility.
Cleaning
Periodically wipe the leads with a damp cloth and mild detergent; do not use
abrasives or solvents.