
Precautions
Thermo Scientic CW3 |9
CAUTION
1. Turn off the cell washer at the main switch. The mains plug must be freely accessible at all times.
Press STOP to shut the cell washer down.
Pull out the power supply plug or disconnect the power supply in an emergency.
2. Use the cell washer only with correct installed rotor and distributor assembly and correct inserted bowl, splash
guard and drain cover.
3. Do not move or relocate the cell washer while the rotor spins.
4. Do not lean on the cell washer.
5. Do not pour any solution such as water, detergent, or disinfectant directly into the rotor chamber. If you do so,
the bearings of the drive unit might corrode or deteriorate.
6. Connect the saline tank to the pump inlet connector of the cell washer correctly with the hose bands (standard
accessories). Otherwise the liquid might leak from the tube and the liquid might get into the inside of the cell
washer. If you think liquid is inside of the cell washer, contact a Thermo Fisher Scientific authorized sales/service
representative to clean and dry the cell washer.
7. Before you operate the cell washer, remove any dropped objects and tube fragments from inside the rotor
chamber.
8. Always check for corrosion and damage on the rotor before using it. Do not use a corroded or damaged rotor.
9. Be sure to use the specified test tubes with a wall thickness of 1 mm or more. When using thin test tubes such
as disposable test tubes, do not use them repeatedly. If any scratches, crack, inside warp, etc. is found on the
test tubes, do not use them because they cannot bear a centrifugal force.
10. Be sure to load test tubes to all the holders.
11. Use one or two drops (about 50 μl) of 3-5% erythrocyte suspension as sample volume for blood cell washing.
When using precipitated erythrocyte layer, use one or two drops (about 50 μl or less) per test tube. Use the 80%
or less of the test tube capacity as sample volume for centrifugation.
12. If you observe some abnormality in this product, stop using it immediately and contact a Thermo Fisher Scientific
authorized sales/service representative. Notify the service representative of the alarm code if displayed.
13. If the cell washer will not be used for a long time, remove the power cord from the socket.
14. For connection to a different outlet, the power cord might be needed to be replaced.
15. Follow local electrical codes.
16. Depending on the magnitude, an earthquake might damage the cell washer.
If you observe some abnormality, contact a Thermo Fisher Scientific authorized sales/service representative.
NOTICE
Usually the control panel and the surface of the cell washer get warm during operation.